AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 889
Entry time: Mon Aug 21 13:17:54 2023
|
Author: |
TD |
Subject: |
Silena 7710 HV test - JCMB 21.8.23 |
|
|
Silena 7710 - see attachment 1
Bias -60V
Without ac mains filter - attachments 2-6
DSO ch#1 AC/1M - y: 10mV/div x: 100ns, 200ns, 1us, 2us & 10us/div
With ac mains filter - attachments 7-11
( see https://elog.ph.ed.ac.uk/nToF/67 attachments 2 & 3 )
DSO ch#1 AC/1M - y: 10mV/div x: 100ns, 200ns, 1us, 2us & 10us/div
Conclusion - Claud Lyons Ltd STF Series Surge & Transient Power Filter makes little observable difference to HF noise transients observed |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|