Tue May 3 08:37:50 2016, TD, MIDAS updates   
|
MIDAS updated per instructions of VFEP (I append the new and old versions of startup.tcl and aidacommon)
I have generated the AIDA update for V8 as 2 tarballs.
|
Tue May 3 08:46:25 2016, TD, MIDAS updates - problems 8x
|
1) Unable to enable data transfer
From RunControl unable to 'Enable #1 All' or 'Enable #2 All'
|
Tue May 3 15:45:17 2016, TD, AIDA Temperatures Log
|
AIDA temperatures (ASIC, Virtex, PSU) are recorded in the file:
|
Tue May 3 16:11:57 2016, TD, MIDAS updates - problems   
|
With help from Patrick Coleman-Smith & Vic Pucknell most of the issues identified
appear to be related to one (or more) of the following issues
|
Wed May 4 03:01:15 2016, TD, Wednesday 4 May 2016 11x
|
10.30 Power up.
Normal DAQ startup.
|
Wed May 4 06:35:08 2016, TD, [How To] Save AIDA Settings & Options
|
From 'Control' tab select 'Save/Restore Settings'
|
Wed May 4 07:04:12 2016, TD, WAVE statistics 13x
|
AIDA wave statistics
|
Wed May 4 14:33:34 2016, TD, merge/tapeserver issues   
|
Reboot aidas1
Power cycle FEE64s
|
Wed May 4 14:45:09 2016, TD, merge/tapeserver issues
|
TapeServer issue fixed - disk incorrectly mounted following reboot
|
Thu May 5 07:55:32 2016, TD, [How To] Mount SAS and USB disks 
|
*******************************************************************************************************
UPDATE 24 May 2016
|
Thu May 5 08:12:14 2016, TD, Current AIDA settings and options
|
Current AIDA settings DB Key: 2016May03-20.44.33
|
Thu May 5 08:30:00 2016, TD, [How To] Set MACB Trigger Output Mode
|
From the 'AIDA Control' tab select 'Local Controls'
|
Thu May 5 08:47:49 2016, TD, [How To] Mask Fast Discriminators
|
From the 'AIDA Control' tab select 'Discriminator control'
From the 'Discriminator controls' tab
|
Thu May 5 09:52:20 2016, TD, MERGE test OK 7x
|
Slow comparator 0x10 -> 0x20
Fast comparator LEC/MEC 0x10 -> 0xff
|
Mon May 23 07:28:38 2016, TD, 23 May 2016 25x
|
12.45 Rly16 'Sequence ALL On' operational
15.20 Startup remote control of CAEN N1419 4-ch HV modules
|
Tue May 24 00:06:26 2016, TD, 24 May 2016    
|
Overnight. DAQ stopped. Detector bias off. FEE power and recirculating chiller on.
08.00 Detector bias on OK - see attachments 1-2
|
Tue May 24 07:12:30 2016, TD, 3x WDC WD30EFRX ATA 3Tb disks added to aidas1 LVM configuration 9x
|
The 3x WDC WD30EFRX ATA 3Tb disks installed in aidas1 in April 2015 (formerly /data20, /data30 and
/data40) have now been added to the aidas1 logical volume management (LVM) configuration to form a
|
Fri May 27 11:10:37 2016, TD, [How To] Startup AIDA 34x
|
1) Start Julabo FL11006 Recirculating Chiller
See manual (page 19) http://www2.ph.ed.ac.uk/~td/AIDA/Information/Julabo%20FL11006.pdf
|
Fri May 27 11:16:40 2016, TD, [How To] Start and Stop Runs 7x
|
All AIDA operations are controlled from the Dell T410 Workstation aidas1 using web browsers.
AIDA has multiple, independent data acquisition servers connected by an Ethernet network.
|
Sun May 29 10:31:59 2016, TD, nnaida19 'H' histogramming enabled indicator dispappears following DAQ start 
|
| 18.30 Options for nnaida19 and a.n.other (working) FEE (nnaida18) |