Mon Mar 27 03:29:05 2017, OH, TD, DK, CG, Monday 27 March 15x
|
11:28 Biases and leakages okm (see attachment 1)
14:00 Removed jumpers on cooling rack D. Noise was around 50 channels worse (FWHM of pulser peak) for all
|
Mon Mar 27 03:28:53 2017, DK, TD, JA, Clock settings at the waveform generator stage
|
Clock settings for this experiment are slightly different than in the previous case.
Below are strictly the settings for the waveform generator itself.
|
Sun Mar 26 03:53:43 2017, CG, OH, Sunday 26th March 
|
10.30 Left DAQ running overnight to look for alphas in DSSDs.
DSSD bias still ok @ 23.3oC (see att. 1).
FEE temps ok.
|
Thu Mar 23 10:56:29 2017, CG, OH, Thursday 23rd March - DSSD replacement + move 7x
|
10.00 Investigating DSSD2 (MSL BB18-1000 3058-9) missing channels, found broken bond wires on rear side (~1/4).
Replaced with 3058-3.
|
Tue Mar 21 08:04:43 2017, CG, OH, Tuesday 21 March - AIDA checks    
|
10.00 Powered system on and biased.
Bias voltages and leakage currents ok (see att. 1). Ambient temp @23.1oC.
|
Thu Feb 2 13:44:02 2017, OH, T9 Test Bench  
|
T9 test setup at Daresbury was powered up and noise levels were checked using a pulser peak.
A 1V peak was used. For most channels this corresponded to a peak position at around +-11000 from the zero point at 9000.
|
Wed Feb 1 16:54:40 2017, CG, NP1306 run summary
|
|
Wed Jan 18 14:16:47 2017, Patrick Coleman-Smith, Repaired Modules returning to RIKEN
|
Five modules are being returned to RIKEN after repair.
The MAC addresses are :-
00:04:a3:2a:ED:8f |
Tue Dec 6 06:38:46 2016, TD, Updated Patrick, CAEN N1419 HV output common mode noise 9x
|
There appears to be significant common mode noise on the output of the
CAEN N1419 4 Ch Reversible 500 V/200 ľA NIM HV Power Supply Module (USB)
|
Tue Dec 6 02:30:35 2016, TD, Report - low - significant variation in FEE64 boot times
|
Not something we were not aware of already but /var/log/messages illustrate the issue.
|
Tue Dec 6 02:20:29 2016, TD, Tuesday 6 December   
|
11.20 DSSSD bias & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
Ambient temperature +24.8 deg C, d.p. +3.6 deg C, RH 25.4%
|
Mon Dec 5 11:53:26 2016, TD, MACB test 42x
|
MACB configuration can be found at https://elog.ph.ed.ac.uk/AIDA/357
|
Mon Dec 5 10:51:08 2016, TD, Monday 5 December    
|
19.50 DSSSD bias & leakage currents OK - see attachments 1 & 2
FEE64 temperatures HIGH - see attachment 3
FL11006 E01 low fluid level (actual halfway between minimum and middle visual markers)
|
Sun Dec 4 05:46:42 2016, TD, SyncCheck RIBF03R1 & RIBF123R1 commissioning tests archived
|
The SyncCheck etc software used for the RIBF123R1 and RIBF03R1 commissioning
tests on d05 (user aida) has been archived as
|
Sun Dec 4 04:05:24 2016, TD, Report - low - DAQ SETUP sometimes produces Javascript error
|
DAQ SETUP sometimes produces Javascript error instead of a pop-up displaying
the status of the ReSync - see attachment 1
|
Sun Dec 4 03:44:12 2016, TD, Sunday 4 December 16x
|
12.45 DSSSD bias & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
Ambient temperature +25.1 deg C, d.p. -3.0 deg C, RH 15.6%
|
Sat Dec 3 10:49:36 2016, TD, Saturday 3 December  
|
19.48 DSSSD bias & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
Ambient temperature +25.1 deg C, d.p. -3.0 deg C, RH 15.6% |
Fri Dec 2 09:58:54 2016, DK, Simple PID with LISE++  
|
I wrote this LISE++ file based on the Takechi .lpp modified by AE for 77Ni, 78Ni, 76Ni.
Basically, I changed the primary beam to 48Ca, reduced the production target thickness to 1 mm (I think it may
|
Fri Dec 2 09:47:58 2016, DK, F8 targets
|
F8 target measurements, for posterity.
Plastic: ~39 to 40 mm
|
Fri Dec 2 09:47:45 2016, DK, F8 targets
|
F8 targets, for posterity.
Plastic: 39~40 mm
|