  142   Wed Jan 20 12:00:19 2016 CG, TDWednesday 20 January 2016
11.58 Repeat yesterday's test - no significant change observed

nnaida11 - input -, shaping time 2us - attachment 1
nnaida14 - input +, shaping time 2us - attachment 2


15.20 invert input polarity - no significant change observed

nnaida11 - input +, shaping time 2us - attachment 3
nnaida14 - input -, shaping time 2us - attachment 4

Note - nnaida11 1.12.L pulser peak width ~45 ch FWHM (integrate)

Test conducted yesterday and today suggest that it is the grounding and/or
the AIDA adaptor PCB that is significant - *not* the particular LH coupler
+ 2x 34-way ribbon cable + 3M 1245 Cu foil screen assembly or FEE64 polarity

Current AIDA adaptor PCB configuration
nnaida11 rev C, LK1 fitted
nnaida14 rev B, no links fitted


15.37 Add LK1 to nnaida14 AIDA adaptor PCB - no significant change

nnaida11 - input +, shaping time 2us - attachment 5
nnaida14 - input -, shaping time 2us - attachment 6
Attachment 1: 5.png
5.png
Attachment 2: 6.png
6.png
Attachment 3: 7.png
7.png
Attachment 4: 8.png
8.png
Attachment 5: 9.png
9.png
Attachment 6: 10.png
10.png
  144   Wed Jan 20 17:02:47 2016 CG, TDRibbon cable + coupler tests at DL

 18/01/16

Set up as last week but with 8us shaping time.

With detector connected, still seeing rail-to-rail noise on waveforms in nnaida13 and double peaking in all nnaida.

Measured various resistances around the system to gauge quality of grounding connections (all measured in Ohms).

     - NIM crate to PCB LEMO connector:                                        4.8

     - NIM crate to mezzanine Cu block:                                           5.3    -> Plastic guide pins now installed.

     - Between LEMO connectors on the adapter boards:           0.19

     - nnaida11/12 drain wire to nnaida13/14 drain wire:            0.25

     - RH coupler nnaida13 to RH coupler nnaida11:                  1.4

     - LEMO to ribbon shielding nnaida13:                                     0.2

     - LEMO to RH coupler shielding nnaida13:                            0.4

 

Double peaking not seen in RIKEN or by PCS during his tests -> most likely pickup of external noise.

Connected multiple different cable types to see effect they have on the sinusoidal wave seen in the waveforms and on the double peaking.

       - nnaida11: LH coupler (not shielded) + 45cm ribbon (shielded) + RH coupler (shielded)

       - nnaida12: LH coupler (not shielded) + 45cm ribbon (shielded)

       - nnaida13: old style 90cm kapton (19/02/14 revision)

       - nnaida14: 45cm thin kapton (shielded) with foil over loose end

Setting as last week, with 2us shaping time.

Peak -to-peak differences in all nnaida for several channels:

               |   nnaida11    |                       nnaida12                    |   nnaida13   |   nnaida14    

    1.4.L  |       157          |                           123                           |         160        |   single peak of width ~400ch

    1.6.L  |      119           |                            111                          |         170        |       "                             "

    1.8.L  |        72           |  Flat top single peak, width ~35    |         155        |      "                             "

   1.10.L |       90            |    "                                             " 42    |         177        |       "                             "                 -> Noticed after cable gnd jumper was not connected on nnaida14 and needs to be for these kaptons. Explains v. large peak width. (Attachments 3-5)

 

19/01/16

Swapped cables on nnaida11 and nnaida13. Now have:

 

       - nnaida11: old style 90cm kapton (19/02/14 revision)

       - nnaida12: LH coupler (not shielded) + 45cm ribbon (shielded)

       - nnaida13: LH coupler (not shielded) + 45cm ribbon (shielded) + RH coupler (shielded)

       - nnaida14: 45cm thin kapton (shielded) with foil over loose end

In nnaida13 see ripple on baseline of waveform with period ~30ch = 600ns (@20ns per ch) => frequency of 1.67 MHz.  (Attachment 6+7)
 

Resistance between drain wire end and ribbon shielding on nnaida13 cabele:   ~0.8 Ohms.

        "                                                          "     RH coupler shielding on nnaida13:    ~1.1 Ohms.

 

For shielded ribbon cable on nnaida13, see fairly flat waveforms. High frequency transients pass through, but no sign of sinusoidal wave over time window of 6us.   (Attachment 9)

After swapping cables on nnaida11 and nnaida13 (only change to system), now see single, Gaussian peaks in nnaida12 with widths ~40ch.

For nnaida12:

           Shaping time (us)     |     Avg peak width (ch)        (avg of 1.8/9/10/11.L peak widths)

                      0.5                     |                   35

                      1.0                     |                   39      (1.11 width ~50ch, all others ~37-42)

                      2.0                     |                   39      (   "                                                      "     )

                      4.0                     |                   35      (1.11 width ~45ch, all others ~32-38)

                      8.0                     |                   37

At 8us shaping time: in nnaida11 (old style cable), see single pulser peak of width ~150 ch.

                                           nnaida13, still see double peaking with peak separation of ~90ch and an extra structure, like a small shoulder, on the outside of the peaks.    (Attachments 8+10)

                                           nnaida14, not fully double peaked, more a peak with a shoulder ~40ch apart.

At 2us shaping time: in nnaida11, pusler peak width back to ~300ch.

                                           nnaida13, retains extra structure on outside of peaks, but peak searation increased to ~150ch.

                                           nnaida14, double peaked, with peak separation of ~65ch.

(Attachments 11-15)

Cannot see waveforms on nnaida12, so swapped cables with nnaida11. Also acts as test of whether the good performance seen in nnaida12 follows the cables or is something to do with the FEE.

Now have:

       - nnaida11: LH coupler (not shielded) + 45cm ribbon (shielded)

       - nnaida12:  old style 90cm kapton (19/02/14 revision)

Now see some Gaussian peaks in nnaida11 with width of ~60ch @2us shaping time, however some peaks seem to have a slight shoulder.

      > suggests issue causing double-peaking is still present (and likely was in nnaida12), but is more/less apparent in some channels than others.

Questions:

     ** Is this just a very well shielded cable, or something to do with the intrinsic properties of this specific cable?

     ** What happens if a RH coupler is added?

     ** What happens if the RH coupler is removed from the cable connected to nnaida13?

Added RH coupler to cable on nnaida11.

Measured some resistances of the shielding on the nnaida11 (good performance - single peaks) cable and the nnaida13 (double peaks) cable:

       - drain wire to far end of shielding:      nnaida11 = 0.2 Ohms           nnaida13 = 0.2 Ohms

       - drain wire to RH coupler:                    nnaida11 = 0.4 Ohms            nnaida13 = 0.5 Ohms

       - far end of nnaida11 cable shielding to far end of nnaida13 cable shielding:     0.05 Ohms   (this seems wrong considering pushing the multimeter probes together gets 0.14 Ohms, but was repeated several times and was always 0.05 Ohms)

With addition of RH coupler to nnaida11 (and shaping time still at 2us):       (Attachments 16+17)

     - nnaida11 peaks show slight shoulder and increased width of ~110ch.

     - nnaida12 now sees noise over full-scale range

     - nnaida13 now quad peaked.

     - nnaida14 still double peaked, but peak separation increased to ~150ch.

Removed RH coupler from nnaida11, all peaks go back to how they were previously, before its addition (Something is repeatable - excellent!)

     - some peaks in nnaida11 now non-Gaussian nad with widths ~130ch. Most Gaussian and width ~50ch.

Removed RH coupler from nnaida13:

    - no real change in the spectra of any nnaida    -> what is the difference between the cables on nnaida11/13???

 

Stock take of ribbon cables currently at DL:

     - 90cm                   = 16

     - 45cm                   = 16   (4 with me for testing in Edinburgh this week. Will be returned to DL next week)

     - LH couplers       = 10   (2 with me for testing in Edinburgh this week. Will be returned to DL next week)

     - RH couplers       = 10  (2 with me for testing in Edinburgh this week. Will be returned to DL next week) 

 

Attachment 1: nnaida11_wav_flatTop.png
nnaida11_wav_flatTop.png
Attachment 2: nnaida11_wav_wavyTop.png
nnaida11_wav_wavyTop.png
Attachment 3: nnaida14_spec_90cmOldKapton_noGNDjumper_2us.png
nnaida14_spec_90cmOldKapton_noGNDjumper_2us.png
Attachment 4: nnaida14_spec_90cmOldGnd_2us.png
nnaida14_spec_90cmOldGnd_2us.png
Attachment 5: nnaida14_spec_90cmOldGnd_2us_fit.png
nnaida14_spec_90cmOldGnd_2us_fit.png
Attachment 6: nnaida11_wav_oldKapton.png
nnaida11_wav_oldKapton.png
Attachment 7: nnaida11_wav_oldKapton_1MhzWave.png
nnaida11_wav_oldKapton_1MhzWave.png
Attachment 8: nnaida13_fullShieldedRibbon_gndJumperOn.png
nnaida13_fullShieldedRibbon_gndJumperOn.png
Attachment 9: nnaida13_wav_shieldedRibbon.png
nnaida13_wav_shieldedRibbon.png
Attachment 10: nnaida13_wav_shieldedRibbon_gndJumperOff.png
nnaida13_wav_shieldedRibbon_gndJumperOff.png
Attachment 11: nnaida12_spec_LHshieldedRibbon.png
nnaida12_spec_LHshieldedRibbon.png
Attachment 12: nnaida14_spec_45cmFoiled.png
nnaida14_spec_45cmFoiled.png
Attachment 13: nnaida13_wav_strangeStructure.png
nnaida13_wav_strangeStructure.png
Attachment 14: nnaida12_spec_LHshieldedRibbon_2us.png
nnaida12_spec_LHshieldedRibbon_2us.png
Attachment 15: nnaida11_wav_goodCable.png
nnaida11_wav_goodCable.png
Attachment 16: nnaida12_spec_goodCablePlusRHcoupler_2us.png
nnaida12_spec_goodCablePlusRHcoupler_2us.png
Attachment 17: nnaida13_spec_addedRHcouplerTo11.png
nnaida13_spec_addedRHcouplerTo11.png
  146   Thu Jan 28 13:53:29 2016 CG, TDRibbon cable + coupler tests at DL - 25-26th Jan

 25/01/16

On start up, received error when trying to access spectrum browser (attachment 1)   => required restart of 'Http for AIDA'

At 2us shaping time, and all other setting as last week, the situation remains the same

    - nnaida11 has width ~50ch, nnaida12 has width ~300ch, nnaida13 is quad peaked and nnaida14 is not quite Gaussian but has width of ~80ch (possibly two peaks very close together)

Change shaping time to 8us.

Swap cables to see if the cable is the source. Currently have two LH coupler + shielded ribbon cables connected to nnaida11/13 (11 good performance, 13 quad peaked)

     - swapped the ribbon cables over but left the couplers where they were to see if the good performance is a consequence of something intrinsic to a particular ribbon or coupler.

     - nnaida11 now has width ~40ch                  (attachment 2)

     - nnaida13 now back to double peaked       (attachment 3)

     - inconclusive.

Could old style kapton laid next to nnaida11 be shielding it due to the built in screen? (the kapton is laid against the ribbon cable)

     - removing the kapton makes no difference to widths, other than nnaida12 which comes down to ~15ch (as expected)

Clipping cable ends together makes no difference, and no combination of jumpers could improve the widths.

Bending cables through 90 degrees makes no difference either (if they were acting as aerials).

Swapped complete PCBs with cables still in place over. Test first thing tomorrow.

 

26/01/16

Swapping PCBs produced no difference

     - nnaida11 has widths of ~50-60ch, although some structure visible in some peaks

     - nnaida14 sees double peaks.

Clipping ribbon ends together now makes things worse.

Swapped PSU cables to nnaida11/14

     - nnaida11 now marginally worse @ ~70ch width

     - nnaida14 sees no obvious change

Connected PSU chasis to NIM crate - no change.

Added two more ribbon + HL/RH couplers to system. Now:

      - nnaida11 = LH + ribbon

      - nnaida12 = LH + ribbon + RH

      - nnaida13 = LH + ribbon + RH

      - nnaida14 = LH + ribon

Clipped drain wire to LEMO connectors on PCB instead of pulser BNC cable.

See marked improvement in performance. All nnaida now have single Gaussian peaks with widths now:   (attachments 4-6)

       - nnaiada11 at ~35ch

       - nnaida12 at ~70ch

       - nnaida13 at ~80-120ch

       - nnaida14 at ~45-55ch (3 at ~80-150ch)

Added RH coupler to nnaida11/14, with thin braid taped over joint in shielding to reduced resistance

       - nnaida11 widths increase slightly to ~40-60ch

       - nnaida14 widths increase slightly to ~60-80ch, few round ~100-120

This is consistent with the RH coupler increasing widths by ~25ch, as seen in previous tests.

See occasional ripple through waveforms in nnaida11/14    (attachments 7-11).

      - expanded view shows period of 34 samples  ==> 680ns   ==> frequency ~1.5MHz

Variation with shaping time (avg peak width over a several channels):

t_shaping (us) nnaida11 nnaida12 nnaida13 nnaida14
0.5 35 48 60 55
1.0 46 60 80 65*
2.0 50 70 100 70
4.0 60 60 85 85
8.0 55 54 60 65

* = some non-Gaussian peaks

Connected detector, biased to +200V (+ve core to nnaida12, -ve braid to nnaida14) with leakage current = 9uA.

At 8us shaping time, and with Lk1 fitted on nnaida13/14 RevC adapter, peaks across all nnaida have widths ~70ch and are all single Gaussian peaks.   (attachments 12-15)

 

 

Attachment 1: spectrumBrowser_InternalError.png
spectrumBrowser_InternalError.png
Attachment 2: nnaida11_spec_BribbonAcoupler.png
nnaida11_spec_BribbonAcoupler.png
Attachment 3: nnaida13_spec_AribbonBcoupler.png
nnaida13_spec_AribbonBcoupler.png
Attachment 4: nnaida11_spec_LHribbon_screenToLEMO.png
nnaida11_spec_LHribbon_screenToLEMO.png
Attachment 5: nnaida13_spec_LHribbonRH_screenToLEMO.png
nnaida13_spec_LHribbonRH_screenToLEMO.png
Attachment 6: nnaida14_spec_LHribbon_screenToLEMO.png
nnaida14_spec_LHribbon_screenToLEMO.png
Attachment 7: nnaida11_wav_fullRibbon_toLEMO.png
nnaida11_wav_fullRibbon_toLEMO.png
Attachment 8: nniada11_wav_ripple.png
nniada11_wav_ripple.png
Attachment 9: nnaida13_wav_fullRibbon_toLEMO.png
nnaida13_wav_fullRibbon_toLEMO.png
Attachment 10: nnaida14_wav_fullRibbon_toLEMO.png
nnaida14_wav_fullRibbon_toLEMO.png
Attachment 11: nnaida14_wav_ripple.png
nnaida14_wav_ripple.png
Attachment 12: nnaida11_spec_screenToLEMO_det.png
nnaida11_spec_screenToLEMO_det.png
Attachment 13: nnaida13_spec_screenToLEMO_det.png
nnaida13_spec_screenToLEMO_det.png
Attachment 14: nnaida11_wav_det.png
nnaida11_wav_det.png
Attachment 15: nnaida13_wav_det.png
nnaida13_wav_det.png
  147   Tue Feb 2 14:08:44 2016 CG, TDReport: Low - No SYNCs from FEEs

Currently running with a source, and everything seems to be running fine in terms of histograms doing what we expect, no error messages etc.

Tried to start the merger and tape server, but the merger reports as waiting for the first SYNC.

Checked the stats page and sure enough, no SYCs are being produced.

Performed a re-sync from the MasterTS page but nothing changed.

Attachment 1: noSYNCs.png
noSYNCs.png
  148   Tue Feb 2 17:32:46 2016 CG, TDTuesday 2 February 2016
Bias +200V, I_L +6.955uA, ambient temp +20.2 deg C

MSL type BB18(DS)-1000 2998-20

~42kBq 207Bi source ~3cm from DSSSD centre

Standard ASIC settings
 shaping time 8us

BNC PB-5
 frequency 100Hz
 amplitude 5.0000V
 x10 attenuation

Slow comparator 16 (dec) - saved to disk c. 18.58 BST (sic) 2.2.16

Slow comparator 10 (dec) - saved to disk c. 10.49 BST (sic) 3.2.16

10.54 3.2.16 I_L +6.305uA, ambient temp +20.1 deg C

Slow comparator 22 (dec) - saved to disk c. c. 12.31 BST (sic) 3.2.16 - see attachment 1

Slow comparator 16 (dec) - saved to disk c. c. 13.04 BST (sic) 3.2.16 - see attachment 2

Slow comparator 12 (dec) - saved to disk c. c. 14.15 BST (sic) 3.2.16 - see attachment 3

Slow comparator 10 (dec) - data to disk file R1 - see attachment 4
start:  stop: 14.20
AIDA adaptor PCB ground links not fitted


Slow comparator 10 (dec) - data to disk file R2 - see attachment 5
start: 14.20 stop: 14.53
4x ground links fitted
14.22 3.2.16 I_L +6.625uA ambient temp +20.2 deg C


Slow comparator 10 (dec) - data to disk file R3 - see attachment 6
                           spectra to disk at 16.19 BST (sic) 3.2.16
start: 14.55 stop: 15.14
4x ground links fitted
BNC PB-5
 amplitudes 9.00000 - 1.00000 @ 1.00000V step 


Slow comparator 9 (dec) - data to disk file R4 - see attachment 7-9
start: 15.22 stop: 16.55
BNC PB-5
 amplitude 5.00000V

From attachments 7-8 we can observe that rates (pulser @ 100Hz + 207Bi) are
generally sensible with slow comparator threshold = 9 (dec). Overall, merge
data rates ~110k data items / s.

Quick calculation suggests threshold (keV) ~ slow comparator setting * 10 
Attachment 1: 20.png
20.png
Attachment 2: 21.png
21.png
Attachment 3: 22.png
22.png
Attachment 4: 23.png
23.png
Attachment 5: 24.png
24.png
Attachment 6: 25.png
25.png
Attachment 7: 26.png
26.png
Attachment 8: 27.png
27.png
Attachment 9: 28.png
28.png
  149   Wed Feb 3 15:34:30 2016 CG, TDRibbon cable + coupler tests at DL 1st-2nd Feb

01/02/16

Started up system. Biased det to +200V 

  150   Tue Feb 9 17:27:12 2016 CG, TDRibbon cable + coupler tests at DL 8-9th Feb

08/02/16

Spoke to PJC-S -> Ethernet cables from FEEs to switch are all 4m Cat6.

System powered on with bias to detector of +200V and leakage current of 6.4uA at room temp 19.0oC.

Double peaking present again in nnaida11/12 -> individual peaks @ ~100ch FWHM with separation ~150ch @ 2us shaping time.

Double peaking also present in nnaida13/14 -> indiv peaks ~100ch FWHM and separation ~90ch.

Changing grounding had no effect.

Made ribbon cable with 2 separately shielded halves plus connecting Cu tape strip at FEE and + drain wire in contact with both halves - single peaks - widths of ~110-160ch.

 

09/02/16

Added conductive gasket to Cu mount blocks

      - resistance between Cu and gasket ~0.2 Ohms (probe-to-probe = 0.16 Ohms)

      - LEMO connector to Cu mount block ~0.35 Ohms

      - LEMO to LEMO (between adapter PCBs) ~0.55 Ohms (connected by heavy duty Cu braid)

Powered system on. Bias of +200V and leakage current = 5.5uA @ room temp 19.5oC.

With gasket added and heavy duty Cu braid between mount blocks, pulser FWHM in nnaida11 ~ 62ch and in nnaida13 ~72ch at shaping time of 8us.

Investigated effect of shaping time on peak FWHM (avg width found from average of 1.(6,8,10,12).L for each FEE.     (see attachments)

    Shaping time (us)    |     nnaida11    |     nnaida12    |     nnaida13     |     nnaida14    

                0.5                   |          95*          |           108*       |          126*        |          99*

                1.0                   |          82            |           116**     |          130^        |         110**  

                2.0                   |           68           |            83^        |           110^       |           92^

                4.0                   |           60           |            77          |             85'        |           90

                8.0                   |            56          |             63         |              77        |            64

     * = all peaks single and Gaussian looking

     ** = not all peaks Gaussian, some flat or sloped top 

     ^ = double peaks, widths estimated from tallest peak

     ' = flat topped

Attachment 1: allnnaida_inSafe_cablesIsolated_0.5us.png
allnnaida_inSafe_cablesIsolated_0.5us.png
Attachment 2: allnnaida_inSafe_cablesIsolated_1us.png
allnnaida_inSafe_cablesIsolated_1us.png
Attachment 3: allnnaida_inSafe_cablesIsolated_2us.png
allnnaida_inSafe_cablesIsolated_2us.png
Attachment 4: allnnaida_inSafe_cablesIsolated_4us.png
allnnaida_inSafe_cablesIsolated_4us.png
Attachment 5: allnnaida_inSafe_cablesIsolated_8us.png
allnnaida_inSafe_cablesIsolated_8us.png
  151   Wed Feb 10 10:56:02 2016 CG, TD9-10 February 2016
MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=6.540uA, ambient temperature +19.9 deg C
 
DSSSD - AIDA adaptor PCB cabling
 
4 off LH Coupler (Kapton PCB, 5cm), 2x 34-way Samtec ribbon cable (45cm), RH coupler(Kapton PCB, 10cm)
 + 3M 1245 1.4mil copper foil screen ribbon cables + RH coupler only (i.e. not LH coupler)
 + drain wires -> gold-plated Lemo-00 test input connectors 
 
nnaida 11 & 12 AIDA adaptor PCB rev B
nnaida 13 & 14 AIDA adaptor PCB rev C (LK1 fitted) 
ground links LK3 & LK7 fitted to nnaida11-14 AIDA adaptor PCBs

Heavy duty copper cable connects copper front end frames of FEE modules
Nitto 5011N conductive gasket between FEE module and front end frames
 
Standard ASIC settings
 nnaida 11 & 12 - negative input
 nnaida 13 & 14 - positive input
 
Pulser BNC PB-5 
Fall time 1.0ms
Rate 100Hz
Delay 250ns
Ampl 5.00000V
Polarity +
Pulse top Tail
Atten 10x
Clamp ON
- polarity via Cooknell SA1 Summing Amplifier
 
207Bi source, approx centred on DSSSD, approx 2cm from DSSSD
 
File R5
shaping time 0.5us
slow comparator 16 (dec)
 
File R6
shaping time 8us
slow comparator 16 (dec)
start: stop: 18.01
 
File R7
shaping time 8uS
slow comparator 12 (dec)
start: 18.02 stop: 09.29 10.2.16
spectra saved to disk 11.37 BST (sic)
 
10.2.16
-------
 
10.50 Bias +200V, I_L +=6.010uA, ambient temperature +20.0 deg C
 
File R8
shaping time 8uS
slow comparator 10 (dec)
start: 10.52 stop: 13.12 
spectra saved to disk 14.13 BST (sic)
 
attachments 1 & 2 - hit patterns during R8

DAQ stalled at start of R9
Run control (updated) showed all FEEs going
No good events from nnaida12 & 14, nnaida11 & 13 OK
nnaida would not stop, would not reset
Power cycle to restart

Ignore R9

File R10
shaping time 8uS
slow comparator 10 (dec)
start: 13.27  stop: 14.09
spectra saved to disk 15.10 BST (sic)
Pulser walkthrough 9.00000V-1.00000V @ 1.0000V step

see attachment 3

File R11
shaping time 8uS
slow comparator 10 (dec)
start:  14.12 stop: 14.38
spectra saved to disk 15.39 BST (sic)
Pulser walkthrough 1.00000V-0.200000 @ 0.1000V step

See attachment 4 & 5

Ignore R12 - same issue as before R9

File R13
shaping time 8uS
slow comparator 8 (dec)
start:  14.50 stop: 15.11
Pulser amplitude 5.00000V

File R14
shaping time 8uS
slow comparator 8 (dec)
start:  15.12 stop: 15.20
Pulser amplitude 5.00000V & 0.20000V

File R15
shaping time 8uS
slow comparator 8 (dec)
start:  15.20 stop: 15.28
spectra saved to disk 16.33 BST (sic)
Pulser amplitude 0.10000V & 1.00000V
see attachments 6 & 7

For slow comaparator setting 8 (dec) and with the above BNC PB-5 pulser settings above
the slow comparator rate decreases from the expected 100Hz (plus source) to < 50Hz
between pulser amplitude 0.20000 and 0.10000V. 


15.32 Bias +200V, I_L +=6.550uA, ambient temperature +19.6 deg C
Attachment 1: 30.png
30.png
Attachment 2: 31.png
31.png
Attachment 3: 32.png
32.png
Attachment 4: 33.png
33.png
Attachment 5: 34.png
34.png
Attachment 6: 35.png
35.png
Attachment 7: 36.png
36.png
  152   Tue Feb 16 17:32:21 2016 CG, TD16 February 2016
MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=6.150uA, ambient temperature +20.1 deg C
 
DSSSD - AIDA adaptor PCB cabling
 
4 off LH Coupler (Kapton PCB, 5cm), 2x 34-way Samtec ribbon cable (45cm), RH coupler(Kapton PCB, 10cm)
 + 3M 1245 1.4mil copper foil screen ribbon cables + RH coupler only (i.e. not LH coupler)
 + drain wires -> gold-plated Lemo-00 test input connectors 
 
nnaida 11 & 12 AIDA adaptor PCB rev B
nnaida 13 & 14 AIDA adaptor PCB rev C (LK1 fitted) 
ground links LK3 & LK7 fitted to nnaida11-14 AIDA adaptor PCBs

Heavy duty copper cable connects copper front end frames of FEE modules
Nitto 5011N conductive gasket between FEE module and front end frames
 
Standard ASIC settings
 nnaida 11 & 12 - negative input
 nnaida 13 & 14 - positive input
 
Pulser BNC PB-5 
Fall time 1.0ms
Rate 100Hz
Delay 250ns
Ampl 5.00000V
Polarity +
Pulse top Tail
Atten 10x
Clamp ON
- polarity via Cooknell SA1 Summing Amplifier
 
207Bi source, approx centred on DSSSD, approx 3cm from DSSSD
  
File R16
shaping time 8uS
slow comparator 8 (dec)
start: 16.2.16 13.15 stop: 09.30 17.2.16
spectra saved to disk 10.30 BST (sic)
Pulser OFF

16.2.16

09.35  Bias +200V, I_L +=5.940uA, ambient temperature +20.0 deg C
  153   Wed Feb 17 11:54:50 2016 CG, TD17 February 2016
File R17 -> pulser walkthrough.
     - Started ~10.00
     - Pulser peak amplitude taken from 9.0V to 1.0 V in 1.0V steps
     - Shaping time 8us, slow disc threshold 0x8
     - File stopped 10.32. Spectra (.msf) saved 10.34
 
File R18 -> pulser walkthrough.
      - Started 10.41.
      - Stepped the pulser amp by 0.1V - too small.
      - Not useful
      - Stopped 10.50

File R19 - pulser walkthrough.
       - Started 10.51
       - empty

File R20 - empty

File R21 -> pulser Walkthrough
     - started 11.18
     - 1.0 - 0.2V in 0.2V steps.
     - Stopped 11.50. Saved histograms 11.51.

File R22 -> source spectrum.
    - slow comp threshold @ 0x7
    - forgot to turn off pulser. => no good.

file R23 -> source spectrum.
    - slow comp threshold @ 0x7
    - started 16.09

18.05 DAQ found stalled, nnaida11 non-responsive
      FORCE reset nnaida11 OK
      but SETUP and FORCE setup failed with timeouts

      Power-cycled all FEE modules to restart

eventual restart following workstation restart

file R29
start 18.50 stop: 09.55 18.2.16
spectra saved to disk 10.55 BST (sic) 
see attachments 1-6

09.56 bias +200V I_L +6.105uA ambient temperature 20.1 deg C
Attachment 1: 40.png
40.png
Attachment 2: 41.png
41.png
Attachment 3: 42.png
42.png
Attachment 4: 43.png
43.png
Attachment 5: 44.png
44.png
Attachment 6: 45.png
45.png
  256   Mon May 30 06:03:51 2016 CG, TDMACB settings

14.00 MACB settings (from left -> right)

MACB# 1 2 3 4 5 6 7 8 9 10 11
Setting 0 2 2 3 3 3 3 3 3 3 3

           MACB#1 is grandparent in tree

           #2 and #3 are parents

           Rest are children.

14.05 MACBs 1 and 3 (master) replaced with spares. No change, clocks remain not locked.

 

  369   Sun Oct 2 15:03:11 2016 CG, TDProblems compiling AIDAsort program on aidas1
I have been trying to compile the sort program I have been working on in Edinburgh so I can work on it locally
while at RIKEN.

When compiling i receive the error shown in make.log (attached) saying that there are undefined reference in
DataSource.cpp.
Also attached is the make output when compiled in Edinburgh - basically just 'unused variable' warnings.

This is exactly the same DataSource.cpp I am using in Edinburgh. I have carried out 'diff' on the DataSpyLib and
DataXferLib directories which are included
in the Makefile and no differences are present. I have also done this for DataSource directory and again, there
are no differences between what I am running here and in Edinburgh.

The only difference I have been able to observe so far is that in Edinburgh I am using g++ version:
 Using built-in specs.
 COLLECT_GCC=g++
 COLLECT_LTO_WRAPPER=/usr/libexec/gcc/x86_64-redhat-linux/4.8.5/lto-wrapper
 Target: x86_64-redhat-linux
 Configured with: ../configure --prefix=/usr --mandir=/usr/share/man --infodir=/usr/share/info
--with-bugurl=http://bugzilla.redhat.com/bugzilla --enable-bootstrap --enable-shared --enable-threads=posix
--enable-checking=release --with-system-zlib --enable-__cxa_atexit --disable-libunwind-exceptions
--enable-gnu-unique-object --enable-linker-build-id --with-linker-hash-style=gnu
--enable-languages=c,c++,objc,obj-c++,java,fortran,ada,go,lto --enable-plugin --enable-initfini-array
--disable-libgcj --with-isl=/builddir/build/BUILD /gcc-4.8.5-20150702/obj-x86_64-redhat-linux/isl-install
--with-cloog=/builddir/build/BUILD/gcc-4.8.5-20150702/obj-x86_64-redhat-linux/cloog-install
--enable-gnu-indirect-function --with-tune=generic --with-arch_32=x86-64 --build=x86_64-redhat-linux
 Thread model: posix
 gcc version 4.8.5 20150623 (Red Hat 4.8.5-4) (GCC)

whereas on aidas1:
 Using built-in specs.
 Target: x86_64-redhat-linux
 Configured with: ../configure --prefix=/usr --mandir=/usr/share/man --infodir=/usr/share/info
--with-bugurl=http://bugzilla.redhat.com/bugzilla --enable-bootstrap --enable-shared --enable-threads=posix
--enable-checking=release --with-system-zlib --enable-__cxa_atexit --disable-libunwind-exceptions
--enable-gnu-unique-object --enable-languages=c,c++,objc,obj-c++,java,fortran,ada --enable-java-awt=gtk
--disable-dssi --with-java-home=/usr/lib/jvm/java-1.5.0-gcj-1.5.0.0/jre --enable-libgcj-multifile
--enable-java-maintainer-mode --with-ecj-jar=/usr/share/java/eclipse-ecj.jar --disable-libjava-multilib
--with-ppl --with-cloog --with-tune=generic --with-arch_32=i686 --build=x86_64-redhat-linux
 Thread model: posix
 gcc version 4.4.7 20120313 (Red Hat 4.4.7-17) (GCC)

The libraries used are all found in LD_LIBRARY_PATH.
On aidas1 this is: /MIDAS/Linux/lib64:/homes/npg/root/lib:/homes/npg/root/lib:/MIDAS/Linux/lib64:/homes/npg/root/lib
and in Edinburgh:
/Disk/ds-sopa-group/np/RIKEN/AIDAsort/MIDAS/Linux/lib64:/Disk/ds-sopa-group/np/thehubdata/thehub6/jlab_software/Linux_Scientific6-x86_64-gcc4.4.6/root/5.34.23/lib:


*********UPDATE 03/10/16 *************

The methods are defined in /MIDAS/DataPackage/DataXferLib/V4_TCP/transfer.c

This is in the expected INCLUDES path of: INCLUDES=  -I/MIDAS/DataPackage/DataXferLib/V4_TCP -I/MIDAS/DataPackage/DataSpyLib
Attachment 1: make.log
aidas1> make
g++ -c -Wall -Wextra -DUNIX -DLINUX -DLINUX64 -DPOSIX -pthread -m64 -I/homes/npg/root/include -I/tmp/DataXferLib_edi/V4_TCP -I/tmp/DataSpyLib_edi  -Iinclude/ main.cpp
g++ -c -Wall -Wextra -DUNIX -DLINUX -DLINUX64 -DPOSIX -pthread -m64 -I/homes/npg/root/include -I/tmp/DataXferLib_edi/V4_TCP -I/tmp/DataSpyLib_edi  -Iinclude/ ./include/DataSource.h ./src/DataSource.cpp
In file included from ./src/DataSource.cpp:12:
include/c_functions.h:10: warning: unused parameter ‘id’
include/c_functions.h:10: warning: unused parameter ‘xclass’
include/c_functions.h:10: warning: unused parameter ‘level’
include/c_functions.h:10: warning: unused parameter ‘source’
include/c_functions.h:10: warning: unused parameter ‘body’
./src/DataSource.cpp: In member function ‘void DataSource::TransferBuffer(double)’:
./src/DataSource.cpp:345: warning: suggest parentheses around comparison in operand of ‘|’
g++ -c -Wall -Wextra -DUNIX -DLINUX -DLINUX64 -DPOSIX -pthread -m64 -I/homes/npg/root/include -I/tmp/DataXferLib_edi/V4_TCP -I/tmp/DataSpyLib_edi  -Iinclude/ ./include/Unpacker.h ./src/Unpacker.cpp
./src/Unpacker.cpp:684:3: warning: "/*" within comment
./src/Unpacker.cpp: In member function ‘void Unpacker::Process(DataSource&)’:
./src/Unpacker.cpp:37: warning: unused variable ‘my_int’
./src/Unpacker.cpp: In member function ‘void Unpacker::LoadParameters(char*)’:
./src/Unpacker.cpp:425: warning: unused variable ‘line’
./src/Unpacker.cpp:427: warning: unused variable ‘ch_first’
./src/Unpacker.cpp:429: warning: unused variable ‘channel’
g++ -c -Wall -Wextra -DUNIX -DLINUX -DLINUX64 -DPOSIX -pthread -m64 -I/homes/npg/root/include -I/tmp/DataXferLib_edi/V4_TCP -I/tmp/DataSpyLib_edi  -Iinclude/ ./include/Calibrator.h ./src/Calibrator.cpp
./src/Calibrator.cpp: In member function ‘void Calibrator::LoadParameters(char*)’:
./src/Calibrator.cpp:299: warning: unused variable ‘line’
./src/Calibrator.cpp:301: warning: unused variable ‘ch_first’
g++ -c -Wall -Wextra -DUNIX -DLINUX -DLINUX64 -DPOSIX -pthread -m64 -I/homes/npg/root/include -I/tmp/DataXferLib_edi/V4_TCP -I/tmp/DataSpyLib_edi  -Iinclude/ ./include/Analysis.h ./src/Analysis.cpp
./src/Analysis.cpp:915:35: warning: "/*" within comment
./src/Analysis.cpp: In member function ‘void Analysis::CloseEvent()’:
./src/Analysis.cpp:146: warning: unused variable ‘b_beam’
./src/Analysis.cpp: In member function ‘void Analysis::WriteOutBuffer(DataSource&)’:
./src/Analysis.cpp:381: warning: comparison between signed and unsigned integer expressions
./src/Analysis.cpp: In member function ‘void Analysis::LoadParameters(char*)’:
./src/Analysis.cpp:1484: warning: unused variable ‘line’
./src/Analysis.cpp:1486: warning: unused variable ‘ch_first’
./src/Analysis.cpp:1488: warning: unused variable ‘channel’
g++ -lrt -m64 -L/homes/npg/root/lib -lGui -lCore -lCint -lRIO -lNet -lHist -lGraf -lGraf3d -lGpad -lTree -lRint -lPostscript -lMatrix -lPhysics -lMathCore -lThread -pthread -lm -ldl -rdynamic  -L/usr/ucblib -lrt -lpthread -L/tmp/MIDAS/Linux/lib64  -ldataspy  -o AIDAsort.v2.exe main.o DataSource.o Unpacker.o Calibrator.o Analysis.o
DataSource.o: In function `DataSource::InitDataSource(int, int, std::basic_string<char, std::char_traits<char>, std::allocator<char> >&, bool)':
DataSource.cpp:(.text+0x1159): undefined reference to `transferMultiBlockSize'
DataSource.cpp:(.text+0x1176): undefined reference to `transferMultiMode'
DataSource.cpp:(.text+0x1193): undefined reference to `transferMultiPort'
DataSource.cpp:(.text+0x11b2): undefined reference to `transferMultiInit'
DataSource.o: In function `DataSource::TransferBuffer(double)':
DataSource.cpp:(.text+0x133c): undefined reference to `transferMultiTxData'
collect2: ld returned 1 exit status
make: *** [AIDAsort.v2.exe] Error 1
Attachment 2: make.logEdi
[s0804967@escrow order_sort_cg]$ make
g++ -c -Wall -Wextra -DUNIX -DLINUX -DLINUX64 -DPOSIX -pthread -std=c++11 -Wno-deprecated-declarations -m64 -I/usr/include/root -I/Disk/ds-sopa-group/np/RIKEN/AIDAsort/DataXferLib/V4_TCP -I/Disk/ds-sopa-group/np/RIKEN/AIDAsort/DataSpyLib -Iinclude/ main.cpp
g++ -c -Wall -Wextra -DUNIX -DLINUX -DLINUX64 -DPOSIX -pthread -std=c++11 -Wno-deprecated-declarations -m64 -I/usr/include/root -I/Disk/ds-sopa-group/np/RIKEN/AIDAsort/DataXferLib/V4_TCP -I/Disk/ds-sopa-group/np/RIKEN/AIDAsort/DataSpyLib -Iinclude/ ./include/DataSource.h ./src/DataSource.cpp
In file included from ./src/DataSource.cpp:12:0:
include/c_functions.h:10:5: warning: unused parameter ‘id’ [-Wunused-parameter]
 int msgDefineMessage (u_int id, u_int xclass, u_int level, char *source, char *body) {
     ^
include/c_functions.h:10:5: warning: unused parameter ‘xclass’ [-Wunused-parameter]
include/c_functions.h:10:5: warning: unused parameter ‘level’ [-Wunused-parameter]
include/c_functions.h:10:5: warning: unused parameter ‘source’ [-Wunused-parameter]
include/c_functions.h:10:5: warning: unused parameter ‘body’ [-Wunused-parameter]
./src/DataSource.cpp: In member function ‘void DataSource::Close()’:
./src/DataSource.cpp:28:9: warning: variable ‘i’ set but not used [-Wunused-but-set-variable]
     int i;
         ^
./src/DataSource.cpp: In member function ‘bool DataSource::ReadBuffer()’:
./src/DataSource.cpp:149:9: warning: variable ‘i’ set but not used [-Wunused-but-set-variable]
     int i;
         ^
./src/DataSource.cpp: In member function ‘void DataSource::InitDataSource(int, int, std::string&, bool)’:
./src/DataSource.cpp:285:9: warning: variable ‘i’ set but not used [-Wunused-but-set-variable]
     int i;
         ^
./src/DataSource.cpp: In member function ‘void DataSource::TransferBuffer(double)’:
./src/DataSource.cpp:345:22: warning: suggest parentheses around comparison in operand of ‘|’ [-Wparentheses]
   if(GetBuffOffset() > MAX_LENGTH_DATA | ts > (ts_0 + DELTA_TS )){
                      ^
g++ -c -Wall -Wextra -DUNIX -DLINUX -DLINUX64 -DPOSIX -pthread -std=c++11 -Wno-deprecated-declarations -m64 -I/usr/include/root -I/Disk/ds-sopa-group/np/RIKEN/AIDAsort/DataXferLib/V4_TCP -I/Disk/ds-sopa-group/np/RIKEN/AIDAsort/DataSpyLib -Iinclude/ ./include/Unpacker.h ./src/Unpacker.cpp
./src/Unpacker.cpp:684:3: warning: "/*" within comment [-Wcomment]
   /***************/
 ^
./src/Unpacker.cpp: In member function ‘void Unpacker::Process(DataSource&)’:
./src/Unpacker.cpp:37:16: warning: unused variable ‘my_int’ [-Wunused-variable]
   unsigned int my_int;
                ^
./src/Unpacker.cpp: In member function ‘void Unpacker::LoadParameters(char*)’:
./src/Unpacker.cpp:425:8: warning: unused variable ‘line’ [-Wunused-variable]
   char line[1024]; // string or char array?
        ^
./src/Unpacker.cpp:427:8: warning: unused variable ‘ch_first’ [-Wunused-variable]
   char ch_first;
        ^
./src/Unpacker.cpp:429:15: warning: unused variable ‘channel’ [-Wunused-variable]
   int module, channel;
               ^
./src/Unpacker.cpp:430:10: warning: variable ‘data’ set but not used [-Wunused-but-set-variable]
   double data;
          ^
g++ -c -Wall -Wextra -DUNIX -DLINUX -DLINUX64 -DPOSIX -pthread -std=c++11 -Wno-deprecated-declarations -m64 -I/usr/include/root -I/Disk/ds-sopa-group/np/RIKEN/AIDAsort/DataXferLib/V4_TCP -I/Disk/ds-sopa-group/np/RIKEN/AIDAsort/DataSpyLib -Iinclude/ ./include/Calibrator.h ./src/Calibrator.cpp
./src/Calibrator.cpp: In member function ‘void Calibrator::LoadParameters(char*)’:
./src/Calibrator.cpp:299:8: warning: unused variable ‘line’ [-Wunused-variable]
   char line[1024];        //
        ^
./src/Calibrator.cpp:301:8: warning: unused variable ‘ch_first’ [-Wunused-variable]
   char ch_first;          //
        ^
g++ -c -Wall -Wextra -DUNIX -DLINUX -DLINUX64 -DPOSIX -pthread -std=c++11 -Wno-deprecated-declarations -m64 -I/usr/include/root -I/Disk/ds-sopa-group/np/RIKEN/AIDAsort/DataXferLib/V4_TCP -I/Disk/ds-sopa-group/np/RIKEN/AIDAsort/DataSpyLib -Iinclude/ ./include/Analysis.h ./src/Analysis.cpp
./src/Analysis.cpp:915:35: warning: "/*" within comment [-Wcomment]
   /*int m_side[32]= {1,1,0,0,     /* 1:4 */
 ^
./src/Analysis.cpp: In member function ‘void Analysis::CloseEvent()’:
./src/Analysis.cpp:146:8: warning: unused variable ‘b_beam’ [-Wunused-variable]
   bool b_beam= false;
        ^
./src/Analysis.cpp: In member function ‘void Analysis::WriteOutBuffer(DataSource&)’:
./src/Analysis.cpp:381:65: warning: comparison between signed and unsigned integer expressions [-Wsign-compare]
     for(int i= offset; i< offset+sizeof(evt_data)+sizeof(int32_t); i++){
                                                                 ^
./src/Analysis.cpp: In member function ‘void Analysis::LoadParameters(char*)’:
./src/Analysis.cpp:1484:8: warning: unused variable ‘line’ [-Wunused-variable]
   char line[1024];        // string or char array?
        ^
./src/Analysis.cpp:1486:8: warning: unused variable ‘ch_first’ [-Wunused-variable]
   char ch_first;
        ^
./src/Analysis.cpp:1488:15: warning: unused variable ‘channel’ [-Wunused-variable]
   int module, channel;
               ^
g++ -lrt -m64 -L/usr/lib64/root -lGui -lCore -lRIO -lNet -lHist -lGraf -lGraf3d -lGpad -lTree -lRint -lPostscript -lMatrix -lPhysics -lMathCore -lThread -lMultiProc -pthread -lm -ldl -rdynamic  -L/usr/ucblib -lrt -lpthread -L/Disk/ds-sopa-group/np/RIKEN/AIDAsort/MIDAS/Linux/lib64 -lxfer -ldataspy -o AIDAsort.v2.exe main.o DataSource.o Unpacker.o Calibrator.o Analysis.o
Attachment 3: transfer.h
#ifndef INCtransferh
#define INCtransferh

#define MAX_STREAM 4
#define MAXPORTS 8

typedef struct block_header {              /* format of data block  */
        unsigned short hdr_flags;          /* see below */
        unsigned short hdr_stream;         /* =0 for forced ack request or 1=>MAX_STREAM  */
        unsigned short hdr_endian;         /*   data byte order */
        unsigned short hdr_ID;
        unsigned int hdr_sequence;        /* for this stream */
        unsigned int hdr_blocklength;     /*  total length of this block including the header */
        unsigned int hdr_datalength;      /*  length of user data in the block  */
        unsigned int hdr_offset;          /*  very large blocks may be fragmented  */
        unsigned int hdr_id1;             /*  for spy to locate header  =0x19062002 */
        unsigned int hdr_id2;             /*  for spy to locate header  =0x09592400 */
} HEADER;

#define HDR_ID1 0x19062002
#define HDR_ID2 0x09592400

typedef struct mbs_block_header {
        unsigned int mbs_endian;
        unsigned int mbs_blocksize;
} MBS_BLKHEADER;


typedef struct ack {
        unsigned short acq_flags;          /* bit 0 = 1 for ack  */
        unsigned short acq_code;           /* =0 for good ack otherwise error reason */
        unsigned short acq_ts_state;       /* tape server state - see below */
        unsigned short acq_stream;         /* =0 for forced ack ack or 1=>MAX_STREAM  */
        unsigned short acq_endian;         /* =1 transmitted in host byte order */
        unsigned short acq_ID;             /* value of ID variable  0 => */
        unsigned int acq_sequence;        /* for stream != 0 is sequence being acked */
        unsigned short acq_stream_state1;           /* see below  */
        unsigned short acq_stream_state2;           /*   */
        unsigned short acq_stream_state3;           /*   */
        unsigned short acq_stream_state4;           /*   */
        unsigned short acq_stream_window1;          /*   */
        unsigned short acq_stream_window2;          /*   */
        unsigned short acq_stream_window3;          /*   */
        unsigned short acq_stream_window4;          /*   */
} ACK;

        
int transferTxData (char *, int, int);
int transferTxRestart ();
int transferBlockSize(int);
int transferMode(int);
int transferBlockMode(int);
int transferEndian(int);
int transferPort(int);
int transferInit (char *);
void transferClose ();
int transferStatus ();
int transferSetVerbose(int);
int transferSetUser(int);
int transferNice(int);
int transferUseOverlap(int);
int transferTxWait();
int transferState();
int transferSndBufSize(int);
int transferRcvBufSize(int);

/*     All these have the thread ID as the first argument  */
int transferMultiTxData (int, char *, int, int);
int transferMultiTxRestart (int);
int transferMultiBlockSize(int, int);
int transferMultiMode(int, int);
int transferMultiBlockMode(int, int);
int transferMultiEndian(int, int);
int transferMultiPort(int, int);
int transferMultiInit (int, char *);
void transferMultiClose (int);
int transferMultiStatus (int);
int transferMultiNice(int, int);
int transferMultiUseOverlap(int, int);
int transferMultiTxWait(int);
int transferMultiState(int);
int transferMultiSndBufSize(int, int);
int transferMultiRcvBufSize(int, int);

#ifndef OK
#define OK 0
#endif

#ifndef ERROR
#define ERROR -1
#endif


#endif
  380   Sun Oct 9 11:47:34 2016 CG, TDSunday 9th October
12.30 Following scheduled power outages of October 8

      power up AIDA and Julabo FL11006 recirculating chiller

      Julabo FL11006 set point +20 deg C
      Temperature +25.3 deg C, d.p. +10.6 deg C, RH 39.3%
     
      boot aidas1 
      aidas1 filesystem checks (fsck) run from c. 12.30 to 15.58

      System wide checks OK *except* nnaida21 failed ADC calibration

16.55 DAQ starts (no data transfer)

      ASIC settings 2016Sep02-18.17.01
      slow comparator threshold 0x10 -> 0x0a
      TS_SYNC_PHASE 0

      DSSSD bias & leakage current OK - see attachments 8 & 9
      FEE64 temperatures OK - see attachment 7

17.30 DAQ stops

      DSSSD bias off
      AIDA withdrawn from (CH2)n moderator 
      DSSSD bias on

17.40 DAQ starts

19.15 Noticed DAQ stopped producing SYNC pulses at some point.
      A reload of the run control tab sees them all as going still, but only 20% are reporting any SYNC output in the stats page.
      All FEEs have failed the SYNC errors since baseline check and the values reported by the SYNC pulses received tests vary
      by +- 3 of a central value.
      See attachments 1-6.
      The merger and tape server were not set up before this run and are not currently running.

      TS_SYNC phase changed from 0 -> 1 and ReSYNC

22.20 DAQ OK

      AIDA SYNCs OK
      All system-wide checks OK *except* nnaida21 fails ADC calibration (no change)
Attachment 1: daqStop1.png
daqStop1.png
Attachment 2: daqStop2.png
daqStop2.png
Attachment 3: daqStop3.png
daqStop3.png
Attachment 4: daqStop4.png
daqStop4.png
Attachment 5: daqStop5.png
daqStop5.png
Attachment 6: daqStop6.png
daqStop6.png
Attachment 7: 3.png
3.png
Attachment 8: 2.png
2.png
Attachment 9: 1.png
1.png
  188   Tue Apr 26 10:06:59 2016 CG, PV & TDTuesday 26 April
2 (of 3) WD 3Tb ATA disk drives are not currently mounted on aidas1.
Attempted to mount them but found device names changed at each boot.
See https://elog.ph.ed.ac.uk/AIDA/186

Mounted Seagate Backup Plus 8Tb USB disk drive - currently /dev/sda2 mounted
as /ntfs
See https://elog.ph.ed.ac.uk/AIDA/187

Started Raspberry Pi nnrpi1 - USB controlled ac mains relay OK

Startup AIDA and test with pulser and no external load.
Typical pulser peak widths ~12-14 ch FWHM as expected. 
  514   Thu Dec 1 16:01:10 2016 CG, PVFriday 2 December
00.07 BigRIPS data file 1135

01.07 BigRIPS data file 1136.

02.09 BigRIPS data file 1137.

02.13 Beam off. ~1/2 way through R44_137

02.16 Beam back. ~600MB into R44_139
      BigRIPS data file 1138.

03.17 BigRIPS data file 1139.

03.37 Beam stopped. End of R44_167.

03.43 Beam returned to F11. ~700MB through R44_170.
      BigRIPS data file 1140.

04.00 Beam off for stripper foil change. End of R44_175.

04.06 Beam returned to F11. 1.2GB into R44_178.
 
04.08 BigRIPS data file 1141.

05.09 BigRIPS data file 1142.

06.02 BigRIPS data fill 1143.

06.26 R44 stopped @ R44_227 ahead of LN2 fill.

07.02 BigRIPS data file 1144.

07.43 Run R45 started.

08.02 BigRIPS data file 1145.

08.05 Local Elog and BigRIPS log uploaded (attachments 3-6).

09.00 Machine time ended, the last data file is R45_18

13.40 Biases are shown as attachments 7 and 8

15.04 DAQ starts (RIBF03R1/R46)
      Pulser walkthrough
      Amplitude 90,000 - 10,000 @ 10,000 step
      x5 attenuator IN
      t_r 50ns tau_d 50us
      frequency 250Hz range
      polarity -, tail pulse, delay MIN

15.19 Walk through finishes.  See attachments 9 and 10 for histograms

15.26 DAQ starts (RIBF03R1/R47)
      Pulser walkthrough
      Amplitude 12,000 - 2,000 @ 2,000 step

15.35 R47_2 finishes.
      See attachments 11 and 12 for histograms

15.41 DAQ starts (RIBF03R1/R48)
      Background - BNC PB-4 pulser OFF

16.35 DAQ stops OK (RIBF03R1/R48_13)

16.55 DSSSD bias & leakage currents Ok - see attachment 13 & 14
      Ambient temperature +25.2 deg C, -4.2 deg C, RH 14.2%
Attachment 1: bias1.png
bias1.png
Attachment 2: bias2.png
bias2.png
Attachment 3: BigRIPSlog.pdf
BigRIPSlog.pdf BigRIPSlog.pdf BigRIPSlog.pdf BigRIPSlog.pdf BigRIPSlog.pdf BigRIPSlog.pdf BigRIPSlog.pdf BigRIPSlog.pdf
Attachment 4: RIBFlog1.pdf
RIBFlog1.pdf RIBFlog1.pdf RIBFlog1.pdf RIBFlog1.pdf RIBFlog1.pdf RIBFlog1.pdf RIBFlog1.pdf RIBFlog1.pdf
Attachment 5: RIBFlog2.pdf
RIBFlog2.pdf RIBFlog2.pdf RIBFlog2.pdf RIBFlog2.pdf RIBFlog2.pdf RIBFlog2.pdf RIBFlog2.pdf RIBFlog2.pdf
Attachment 6: RIBFlog3.pdf
RIBFlog3.pdf RIBFlog3.pdf RIBFlog3.pdf RIBFlog3.pdf RIBFlog3.pdf RIBFlog3.pdf RIBFlog3.pdf RIBFlog3.pdf
Attachment 7: end-exp-bias2.png
end-exp-bias2.png
Attachment 8: end-exp-bias1.png
end-exp-bias1.png
Attachment 9: r46_walkthrough_4.png
r46_walkthrough_4.png
Attachment 10: r46_walkthrough_3.png
r46_walkthrough_3.png
Attachment 11: r47_walkthrough_3.png
r47_walkthrough_3.png
Attachment 12: r47_walkthrough_4.png
r47_walkthrough_4.png
Attachment 13: 40.png
40.png
Attachment 14: 41.png
41.png
  139   Thu Jan 7 16:04:44 2016 CG, PJCSPCS Reply to RIKEN progress - 15/12/15 (late)

Quote:

 Tues 15/12/15

Swapped previously tripped detector (3131-11) on to kaptons already in place for (working) 2977-15 and attempted to bias

   - i.e. made no changes (FEEs, or jumpers etc) to the known working setup of 2977-15, just replaced the detector with 3131-11

   > still trips at ~6-7V

   > conclusion: He's dead Jim.

 

Took some measurements of rough max positions of DSSSDs from the FEEs with the current 90cm kaptons (pictures in attached PowerPoint and PDF)

   - Vertical EURIKA support to centre of EURIKA ~ 60cm

   - AIDA Cu cooling pipe to end of snout ~ 75cm

   - Base of snout to end =  57cm

   - Min distance from end of snout to inner most FEE (horizontal FEEs) ~ 92cm

           - additional 6cm per extra adapter

   - Further DSSSD can go from base of snout ~ 54cm (60cm from Cu pipe)

           - limiting factor is kaptons to nnaida5+6 (bottom stack). Still a few more cm slack in the other kaptons.

   - For kaptons in 4th horizontal connector (furthest slot in 2nd adapter), detector 48cm from base of snout.

   - For kaptons in 8th horizontal connector (furthest slot in 4th adapter), detector 36cm from base of snout.

 

Note (PCS): In disconnecting things to check ground connections, and when putting AIDA away before I left, I noticed ~20% of the network cables are of the shielded variety. I know this didn't seem to make any difference in the setup at DL, but I don't know how concerned you are about this and if you would suggest changing them in future.

 

 I would recommend changing the network cables for unshielded versions to remove a potential ground connection.

  428   Mon Nov 7 02:13:05 2016 CG, OH, TD, DKMonday 7 November
11.10 DAQ found stopped. No storage mode so can't tell when, but presumably at 06.30 during LN2 fill cycle.
      Unable to regain control of (see errors after merger restart in attachment 1) system so power cycling FEEs.

11.15 Bias voltages @ ambient temp 25.5oC before DSSD noise investigation (attachment 2+3).

11.40 NB+DK attempting shielding of LN2 solenoid.
      Manual fill initiated. DAQ still stopped at 11.41.
      Unable to regain control of system. Power cycle FEEs while DK+NB install more shielding.

2:55 Made attempts to solve the noise issue that was plaguing DSSD1.
     Attempts at checking the cabling, making sure the detectors were light sealed and switching the bias cables
         around were unsuccessful.
     A blue cable was found to be laying across the snout of AIDA that runs as a temperature probe for one of the
         clover detectors.
     Moving this cable appears to have solved this issue (Has been locked in position). See pulser spectrum in
         attachments 4 and 5

23.15 DSSSD bias & leakage current Ok - see attachments 6-7
      FEE64 temperatures OK - see attachment 8

00.05 DAQ start (R137)
      Pulser Walkthrough
      BNC PB_4 amplitudes 90,000-10,000 @ 10,000 step
      x5 attenuator IN
      t_r 50ns tau 50us
      Frequency 250Hz range
      Polarity -, tail pulse, delay MIN

      ASIC settings 2016Nov07-23.03.06

00.21 DAQ start (R138)
      Amplitude 14,000-2,000 @ 2,000 step

00.42 Background run R139 started. Pulser set to external trigger.

01.01 Run R139 closed successfully.
      
01.04 Started R140 on no storage mode to run overnight.
      Leakage currents @ 01.06 shown in 9+10.
      R140 still running ok @ 01.40.
Attachment 1: merger_error-DAQstop.png
merger_error-DAQstop.png
Attachment 2: bias1.png
bias1.png
Attachment 3: bias2.png
bias2.png
Attachment 4: layout3_711.png
layout3_711.png
Attachment 5: layout4_711.png
layout4_711.png
Attachment 6: 20.png
20.png
Attachment 7: 21.png
21.png
Attachment 8: 22.png
22.png
Attachment 9: bias1.png
bias1.png
Attachment 10: bias2.png
bias2.png
  548   Fri Mar 31 17:21:56 2017 CG, OH, DKSaturday 1 April - Run status
00.46 34Ne stopped for 36Ne setting (R19_37).
      DAQ kept running as I was upstairs getting some food.
      Stopped at R19_45.

00.57 BigRIPS change to 36Ne setting. Start file 0196. Still tuning.
      Temps ok.
      Leakage currents and bias ok (see attachment 1).

01.32 R20 started on 36Ne setting.
      BigRIPS file 0197 (started @ 01.30).
      BRIKEN file 

02.23 Beam stopped to F11. End of R20_12.

02.28 Beam returned to F11. Start of R20_14.
      BigRIPS file 0198.

02.48 BRIKEN file 48 started.
      Start of R20_19.
      All system checks passed except nnaida1 fails global clocks 6.
      Temps and leakage currents ok.

03:28 BigRips file 0199 (started at 03:27)
      AIDA at end file R20_28
      BRIKEN file 47 started
      All system wide checks ok.
      Temps and leakage currents ok

04:36 BigRIPS file 0200
      Coincides with end of AIDA file R20_46
      BRIKEN file 48
      All system wide checks ok
      Temps and leakage currents ok
      ASIC load check all ok 

04:59 BigRips run 0201 (Started at 4:45)
      Corresponding AIDA file for start of BigRips run is R20_49
      BRIKEN file 49
      Corresponding AIDA file for start of BRIKEn is R20_52

05:33 System wide checks all ok
      Temperatures, biases and leakage currents ok (seet attachment 2)

05:44 BigRips start run 0202
      Start of AIDA file R20_64
      BRIKEN file 50
      System wide checks passed
00.46 34Ne stopped for 36Ne setting (R19_37).
      DAQ kept running as I was upstairs getting some food.
      Stopped at R19_45.

00.57 BigRIPS change to 36Ne setting. Start file 0196. Still tuning.
      Temps ok.
      Leakage currents and bias ok (see attachment 1).

01.32 R20 started on 36Ne setting.
      BigRIPS file 0197 (started @ 01.30).
      BRIKEN file 

02.23 Beam stopped to F11. End of R20_12.

02.28 Beam returned to F11. Start of R20_14.
      BigRIPS file 0198.

02.48 BRIKEN file 48 started.
      Start of R20_19.
      All system checks passed except nnaida1 fails global clocks 6.
      Temps and leakage currents ok.

03:28 BigRips file 0199 (started at 03:27)
      AIDA at end file R20_28
      BRIKEN file 47 started
      All system wide checks ok.
      Temps and leakage currents ok

04:36 BigRIPS file 0200
      Coincides with end of AIDA file R20_46
      BRIKEN file 48
      All system wide checks ok
      Temps and leakage currents ok
      ASIC load check all ok 

04:59 BigRips run 0201 (Started at 4:45)
      Corresponding AIDA file for start of BigRips run is R20_49
      BRIKEN file 49
      Corresponding AIDA file for start of BRIKEn is R20_52

05:33 System wide checks all ok
      Temperatures, biases and leakage currents ok (seet attachment 2)

05:44 BigRips start run 0202
      Start of AIDA file R20_64
      BRIKEN file 50
      System wide checks passed

6:15 Operators call...R20_72 1.5 GB
     The BigRIPS system display PC looks like it crashed

6:22 2017-04-01 06:21:19 BigRIPS run 203 starts
      AIDA run 20_74 is about 25% started
      BRIKEN run 51

6:40 Phone rings, AIDA run R20_79 just started
      Stripper foil seems to have been changed

6:43 2017-04-01 06:42:03 BigRIPS starts a new run     
     R20_79 is 75% the way through.
     BRIKEN run 52
     System checks passed.

7:09 PPACs are discharging very frequently.

7:39 counted 8 discharges.  It's about once in four minutes.  An acceptable discharge rate is several times in
one shift.
     Firstly, each discharge damages the PPAC.  Even 10 to 20 discharges can easily cause dead regions of cm size.
     Secondly, with the frequency of discharge, we are now missing like as much as 25% of all events (at least
at a given PPAC)
     As the detector becomes more damaged, it can discharge more frequently.
     Moving the detector's position slightly can help to alleviate the problem by putting the beam on a fresh region
     Anyway, as for a parasitic run they can do whatever they want, but it seems they are mis-operating the PPACs
     Basically the voltage is too high or the detector is too old and needs to be replaced.

7:40 2017-04-01 07:40:12 BigRIPS run 205 begins
     AIDA run R20_94 is at 721 MB at 7:40
     BRIKEN run 53
     System-wide checks are okay

8:08 Bias and Temp attachments 3 & 4 (bias_1_.png and temp_1_a.png)

9:01 BigRIPS run 206 2017-04-01 09:01:06
     AIDA run R20_115 is at 600 MB at 9:01
     BRIKEN run is 54
     System-wide checks

9:39 Now that the students went home and some more experienced beam physicists are on shift, the PPAC discharge rate
goes down to ~2x per hour.

10:02 New BigRIPS run at 2017-04-01 10:01:40 number 0207
     AIDA run R20_130 was at 1.9 GB at 10:02..
     BRIKEN run 55
     System-wide checks passed

(I went upstairs for a break...it seems that the beam got turned off at some point)

10:55 Phone call, and I hear the chime for the beam coming back.

10:56 BigRIPS starts run 0208 2017-04-01 10:56:06
      AIDA run R20_145 has just started at 10:56
      BRIKEN run was not restarted and continued as run 55
      System-wide checks are okay

11:02 We checked that the synchronization looks okay.

11:58 Noticed that BigRIPS started a new run at 2017-04-01 11:56:42, run 0209
     AIDA run R20_160 should overlap about 50% with this run.
     BRIKEN run is 56
     System wide checks performed.
     Attached are Biases and Temperatures as 5 & 6 ("bias_1_b" ; "temp_1_b")

12:58 BigRIPS starts run 0210 at 2017-04-01 12:57:21
     AIDA run R20_176 was 422 MB at 12:57
     BRIKEN 57
     Pass system-wide checks

14:30 beam current has been fluctuating for awhile

14:50 BigRIPS started a new run 0212 2017-04-01 14:44:18
     R20_203 was at 1.2 GB at 14:44
     BRIKEN run 58
     System checks okay

15:18 Phone rings.  RF is down so the beam stopped.
     R20_212 is at 500 MB at 15:19

15:57 Phone rings again.

16:37 BigRIPS start run 0214
      Start of AUDA file R20_232
      BRIKEN run 59
      System wide check all ok

16:53 Biases and leakage currents ok (See attachment 7)

17:36 BigRips start run 0215
      Strt of AIDA file R20_247
      BRIKEN start run 60
      System wide checks all ok
      Temperatures all ok

18:04 BigRips stop to change gas stripper foil
      AIDA file R20_254

18:09 BigRips start run 0216
      AIDA file R20_255
      BRIKEN start file 61
      System wide checks all ok
6:15 Operators call...R20_72 1.5 GB
     The BigRIPS system display PC looks like it crashed

6:22 2017-04-01 06:21:19 BigRIPS run 203 starts
      AIDA run 20_74 is about 25% started
      BRIKEN run 51

6:40 Phone rings, AIDA run R20_79 just started
      Stripper foil seems to have been changed

6:43 2017-04-01 06:42:03 BigRIPS starts a new run     
     R20_79 is 75% the way through.
     BRIKEN run 52
     System checks passed.

7:09 PPACs are discharging very frequently.

7:39 counted 8 discharges.  It's about once in four minutes.  An acceptable discharge rate is several times in
one shift.
     Firstly, each discharge damages the PPAC.  Even 10 to 20 discharges can easily cause dead regions of cm size.
     Secondly, with the frequency of discharge, we are now missing like as much as 25% of all events (at least
at a given PPAC)
     As the detector becomes more damaged, it can discharge more frequently.
     Moving the detector's position slightly can help to alleviate the problem by putting the beam on a fresh region
     Anyway, as for a parasitic run they can do whatever they want, but it seems they are mis-operating the PPACs
     Basically the voltage is too high or the detector is too old and needs to be replaced.

7:40 2017-04-01 07:40:12 BigRIPS run 205 begins
     AIDA run R20_94 is at 721 MB at 7:40
     BRIKEN run 53
     System-wide checks are okay

8:08 Bias and Temp attachments 3 & 4 (bias_1_.png and temp_1_a.png)

9:01 BigRIPS run 206 2017-04-01 09:01:06
     AIDA run R20_115 is at 600 MB at 9:01
     BRIKEN run is 54
     System-wide checks

9:39 Now that the students went home and some more experienced beam physicists are on shift, the PPAC discharge rate
goes down to ~2x per hour.

10:02 New BigRIPS run at 2017-04-01 10:01:40 number 0207
     AIDA run R20_130 was at 1.9 GB at 10:02..
     BRIKEN run 55
     System-wide checks passed

(I went upstairs for a break...it seems that the beam got turned off at some point)

10:55 Phone call, and I hear the chime for the beam coming back.

10:56 BigRIPS starts run 0208 2017-04-01 10:56:06
      AIDA run R20_145 has just started at 10:56
      BRIKEN run was not restarted and continued as run 55
      System-wide checks are okay

11:02 We checked that the synchronization looks okay.

11:58 Noticed that BigRIPS started a new run at 2017-04-01 11:56:42, run 0209
     AIDA run R20_160 should overlap about 50% with this run.
     BRIKEN run is 56
     System wide checks performed.
     Attached are Biases and Temperatures as 5 & 6 ("bias_1_b" ; "temp_1_b")

12:58 BigRIPS starts run 0210 at 2017-04-01 12:57:21
     AIDA run R20_176 was 422 MB at 12:57
     BRIKEN 57
     Pass system-wide checks

14:30 beam current has been fluctuating for awhile

14:50 BigRIPS started a new run 0212 2017-04-01 14:44:18
     R20_203 was at 1.2 GB at 14:44
     BRIKEN run 58
     System checks okay

15:18 Phone rings.  RF is down so the beam stopped.
     R20_212 is at 500 MB at 15:19

15:57 Phone rings again.

16:37 BigRIPS start run 0214
      Start of AUDA file R20_232
      BRIKEN run 59
      System wide check all ok

16:53 Biases and leakage currents ok (See attachment 7)

17:36 BigRips start run 0215
      Strt of AIDA file R20_247
      BRIKEN start run 60
      System wide checks all ok
      Temperatures all ok

18:04 BigRips stop to change gas stripper foil
      AIDA file R20_254

18:09 BigRips start run 0216
      AIDA file R20_255
      System wide checks all ok

18:25 BRIKEN start file 61
      AIDA file R20_259
      BRIKEN had to restart the computer so a resync needed to be done

19:07 BigRips start file 0217
      AIDA on file R2-_270 (End of file)
      BigRips reducing beam intensity to calibrate the beam current

19:12 Back to normal BigRips condition
      Al degrader 1mm is removed
      BigRips run 0218 (started at 19:11)
      AIDA file R20_271
      BRIKEN start file 62
      System wide checks ok

20:06 Bias and leakage currents ok
      Temperature ok

20:12 BigRips start run 0219
      Start of AIDA file R20_287
      BRIKEN starts file 64
      System wide checks all ok
      ASIC load check ok

21:12 BigRips start run 0220
      AIDA file R20_301
      System wide checks ok
      BRIKEN starts file 65

22:12 BigRips stopped to change stripper foils
      End of AIDA file R20_315

22:20 BigRips start run 0220
      End of AIDA file R20_317
      System wide checks and temperatures ok
      BRIKEN start file 66

23:08 Biases and leakages ok

23:10 BigRips start run 0222
      End of AIDA file R20_329
      BRIKEN start file 67
      System wide checks all ok
Attachment 1: bias_010417_0123.png
bias_010417_0123.png
Attachment 2: 010417_bias1.png
010417_bias1.png
Attachment 3: bias_1_a.png
bias_1_a.png
Attachment 4: temp_1_a.png
temp_1_a.png
Attachment 5: bias_1_b.png
bias_1_b.png
Attachment 6: temp_1_b.png
temp_1_b.png
Attachment 7: 010417_bias4.png
010417_bias4.png
Attachment 8: 010417_bias5.png
010417_bias5.png
Attachment 9: 010417_bias6.png
010417_bias6.png
  533   Tue Mar 21 08:04:43 2017 CG, OHTuesday 21 March - AIDA checks
10.00 Powered system on and biased.
      Bias voltages and leakage currents ok (see att. 1). Ambient temp @23.1oC.
      
14.00 Shunji wanted to do some checks with the LN2 fill.
      Seems to perhaps be a problem with unshielded cables to the valve switches.
      With control box, outlet valve and cables to all valves behind LN2 tank, AIDA survives LN2 fill.
      Parts ordered to facilitate moving of components behind the LN2 tank, away from AIDA.
      Will carry out further tests once these have been fitted.

17.00 FWHM seem reasonable and similar to Dec 2016.
      n-side FEEs see average pulser FWHM of ~70-80ch. nnaida19 noticeably worse @ ~150ch. (att. 2+3)
      p-side FEEs worse with average pulser FWHM of ~200ch. nnaida13 worst of these, showing non-Gaussian peaks
with widths @ ~300ch. (att. 4+5)
      Will not attempt to solve this now as they must all be disconnected to sort problem with nnaida1.
Attachment 1: bias1.png
bias1.png
Attachment 2: layout4.png
layout4.png
Attachment 3: nnaida19.png
nnaida19.png
Attachment 4: layout3.png
layout3.png
Attachment 5: nnaida13.png
nnaida13.png
  534   Thu Mar 23 10:56:29 2017 CG, OHThursday 23rd March - DSSD replacement + move
10.00 Investigating DSSD2 (MSL BB18-1000 3058-9) missing channels, found broken bond wires on rear side (~1/4).
      Replaced with 3058-3.
      Stack order now:
            |   3058-4
      beam  |   3058-3
            |   2998-23
            |   3058-5
            |   3058-6
            V   3058-7

12.00 All detectors, including the new replacement bias correctly (see att. 1).

14.00 Shunji has mentioned moving AIDA+BRIKEN forward by ~1m due to concerns about focusing the accurately at such a rearward position.
      The afternoon was spent removing the EURICA rails to enable the push forward of AIDA+BRIKEN.

18.00 The DAQ was set running with the new detector included. All system checks passed apart from nnaida19 ADC calibration.
      Pulser peak widths are mostly ~70-90ch for all FEEs. This is similar to Tuesday for the n-side FEEs (att. 2-4) and an improvement for the p-side FEEs (att. 5-7).
      Having said that, some spectra show peaks with shoulders or tails and a few feature multiple discreet peaks.
      We will look at this over the next couple of days, depending on when the final move/alignment of AIDA+BRIKEN is completed.

      Complete table of pulser FWHM

       nnaida# | Pulser FWHM (ch) |  Comments
          1    |       64         |
          2    |       168        |  Peak has shoulder
          3    |       77         |
          4    |       74         |
          5    |       79         |  Shoulder
          6    |       86         |  Shoulder
          7    |       80         |
          8    |       78         |
          9    |       68         |
          10   |       70         |
          11   |       80         |
          12   |       85         |
          13   |       70         |
          14   |       65         |
          15   |       92         |  Wide secondary peak in ASICs 1&4 of FWHM ~900ch @ 45k. Only few counts high (~1% height of pulser peak). Noise 10x height of pulser.
          16   |       99         |  Noise peak ~10x height of pulser peak.
          17   |       70         |  Low E tail.
          18   |       75         |
          19   |       96         |
          20   |       83         |
          21   |       68         |  Additional peaks either side of pulser: approx. relative heights > (1/30) (1) (1/100)
          22   |       69         |
          23   |       108        |  Noise peak ~10x height of pulser peak.
          24   |       110        |  Noise peak ~10x height of pulser peak.
Attachment 1: bias_230317.png
bias_230317.png
Attachment 2: nside_wide_230317.png
nside_wide_230317.png
Attachment 3: nside_mid_230317.png
nside_mid_230317.png
Attachment 4: nside_zoom_230317.png
nside_zoom_230317.png
Attachment 5: pside_wide_230317.png
pside_wide_230317.png
Attachment 6: pside_mid_230317.png
pside_mid_230317.png
Attachment 7: pside_zoom_230317.png
pside_zoom_230317.png
ELOG V3.1.3-7933898