AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  BRIKEN  ELOG logo
Message ID: 75     Entry time: Tue Oct 11 15:58:01 2016
Author: JL Tain et al 
Subject: VETO Det 
Check of Plastic and NaI veto detectors and DACQ settings


Plastic: 10x200x450mm**3, PMT: H1161, HV: -1200V, Anode2
Very small and short signal: width: 20ns, rise: 10ns, amplitude up to 10mV
We decided to put an amplifier to have more flexibility
Amp: Ortec 474 TFA, G: 10xmax, INT=200ns, DIF: 100ns
Signal: width: 200ns, height: 100mV
DACQ Trigger & EFIR: P=2, G:2

NaI: 2"x4"x16", PMT: ?, VD: 1410 (ORNL), HV: +800V
Signal: width: 400ns height: 15mV
DACQ trigger and EFIR: P=20, G=20

File: 161011_2330_PlasticNaI_137Cs.root
File: 161011_2335_PlasticNaI_60Co.root

Photo of the arrangement.
137Cs spectrum of both detectors.
Attachment 1: IMG_20161012_124010.jpg  1.632 MB  Uploaded Wed Oct 12 05:54:49 2016  | Hide | Hide all
IMG_20161012_124010.jpg
Attachment 2: Plastic-137Cs.pdf  21 kB  Uploaded Wed Oct 12 06:04:24 2016  | Hide | Hide all
Plastic-137Cs.pdf
Attachment 3: NaI-137Cs.pdf  17 kB  Uploaded Wed Oct 12 06:04:38 2016  | Hide | Hide all
NaI-137Cs.pdf
ELOG V3.1.4-unknown