Change of main power line for MPOD and CLOVER preamp
Because the incident yesterday with the NIM crate,
we try to separate the CLOVER from the rest of
the equipment.
So we put the MPOD, which powers the CLOVER but also
the 3He tubes and the VETO detectors, and the preamp
power supply in a different line.
We check if this has an impact on the noise in either
CLOVER or 3He tubes.
We measure with the 137Cs-60Co source:
File: 161014_1241_137Cs60Co.root
After inspecting the spectra we conclude that the impact
is not appreciable.
See figures
662keV resolution:
D4Black: 3.16keV
D4Red: 2.49keV
D4Green: 3.57keV
D4Blue: 3.40keV |