ID |
Date |
Author |
Subject |
|
|
400
|
Mon Oct 29 12:08:12 2018 |
JR, JA, AT | Background |
Background was measurement
Pulser: NO
DLT: No
Online:
Start:21:07
Stop:
Comment: V2A8C9 contains now the 10Hz clock signal coming from
the same section of the clock where the 100Hz sync signal is extracted.
It will be there even if we inhibit the pulser signals to the preamps
SYNC pulser is 200Hz because
the gate generator produces double peaks from the 100Hz clock
v1a6c13-16 had a configuration error (Trigger window to short)
It's corrected at 14:55 (2018/10/30) on file 181030_Conf_BrikenFull_LowThrAdj_masterCLK.xlsx . |
|
|
399
|
Sun Oct 28 07:41:03 2018 |
AT, JA, AA | Background |
Dacq was reinitialized
Background was measurement
Pulser: YES
DLT: No
Online: 181029_1640prevday_1723_Bkg_Pulser.root
Start:16:40
Stop: 17:23
Comment: We started a Ge filling manually, to check if the funny behavior of the
Corr_BR_BR can be reproduced. It remained clean. So we do not yet what caused the
funny behavior.
Funny behavior definition: periodic peaks appeared in the Corr_BR_BR file
The synch software problem will be solved tomorrow.
Tomorrow definition: a mystical land were 99% of all human productivity, motivation
and achievement is stored |
|
|
398
|
Fri Oct 26 15:15:55 2018 |
JA, ATS, ATD, JLT, AA | Background measurements |
No pulser (inhibit in clock)
DLT: 181026_2244_Bkg_NoPulser_114.dlt
Online: 181026_2259_Bkg_NoPulser.root
TTree: 181026_2244_Bkg_NoPulser_114.root
With pulser:
DLT: 181026_2305_Bkg_Pulser_115.dlt (erased)
Online: 181026_2314_Bkg_Pulser.root
TTree: no
VME crate reset
No pulser:
DLT: 181026_2324_Bkg_NoPulser_116.dlt
Online: 181026_2331_Bkg_NoPulser.root
We adjust all the thresholds reducing them
drastically to see the noise.
Overnight measurement:
Online: 181027_1137_BkgOverNight_NoPulser.root
Start measurements with DLT
No pulser:
Start: 11:42
Stop: 12:46
DLT: 181027_1142_Bkg_NoPulser_VeryLowThr_117.dlt
Online: 181027_1246_Bkg_NoPulser_VeryLowThr.root
Bckg rate:
Fit to RateBRIKEN= 0.340 cps
With pulser:
Start: 12:53
Stop: 14:08
DLT: 181027_1253_Bkg_Pulser_VeryLowThr_118.dlt & *_119.dlt
Online: 181027_1408_Bkg_Pulser_VeryLowThr.root
Bckg rate:
t(10Hz pulse)= 4458s, BRIKEN:2547 counts, Rate=0.571 cps
tube bkg (Integral(l830,1550))=885, net=1662counts, Rate=0.373cps
Fit to RateBRIKEN= 0.566 cps
252Cf source in the center
With pulser:
Start: 14:37
Stop: 16:51
DLT: 181027_1437_252Cf_Pulser_VeryLowThr_120.dlt Runs: 120-123
Online: 181027_1437_252Cf_z0cm_Pulser_VeryLowThr_1652.root
252Cf rate:
t(10Hz pulse)=7973.80 s, BRIKEN: 7.66289e+07 counts, Rate= 9610.09 cps
Fit to RateBRIKEN= 9594.69 cps
No pulser:
Start: 16:57
Stop: 18:47
DLT: 181027_1657_252Cf_NoPulser_VeryLowThr_124.dlt
Online:181027_1657_252Cf_z0cm_NoPulser_VeryLowThr_1847.root
Fit to RateBRIKEN=9591.79 cps
Background:
No Pulser:
DLT: No
Online: 181028_1056_BkgOverNight_NoPulser.root
With pulser:181028_1056_1641_Bkg_Pulser.root
DLT: No
Commments: Corr_BR_BR looks funny. It shows periodic peaks. Probably related to the
filling of the Ge detector. We will stop to check if it appears again |
|
|
397
|
Fri Oct 26 03:57:37 2018 |
ATS & JLT | Background |
Background BRIKEN control measurement
No DLT
2018/10/26
Start: 11:57
Stop: 14:00
ROOT file: 181026_1400_Bkg.root
Measurement re-started at 15:58 |
|
|
396
|
Fri Oct 26 03:31:33 2018 |
AT2 & JA | Test pulser effect on baseline |
We have tested possible change of the baseline in the digitizer due to the pulser connection. NO EFFECT has been
found!
Test for detector He086 ORNL:
He086_NoPulser_samples.png -> Without pulser, baseline around 1000
He086_Pulser_samples.png -> With pulser, baseline around 1000
Test for detector He001 UPC:
He001_Pulser_samples.png -> With pulser, baseline around 1200
He001_NoPulser_samples.png -> Without pulser, baseline around 1200
Test for detector He013 RIKEN:
He013_Pulser_samples.png -> With pulser, baseline around 850
He013_NoPulser_samples.png -> Without pulser, baseline around 850 |
| Attachment 1: He086_NoPulser_samples.png
|
|
| Attachment 2: He086_Pulser_samples.png
|
|
| Attachment 3: He001_Pulser_samples.png
|
|
| Attachment 4: He001_NoPulser_samples.png
|
|
| Attachment 5: He013_Pulser_samples.png
|
|
| Attachment 6: He013_NoPulser_samples.png
|
|
|
|
395
|
Fri Oct 26 03:24:25 2018 |
ATS & JLT | Background |
Background BRIKEN control measurement
No DLT
2018/10/26
Start: 10:00 aprox
Stop: 11:20
ROOT file: 181026_1120_Bkg.root |
|
|
394
|
Thu Oct 25 16:08:53 2018 |
ATS & JLT | Lead bricks |
Lead bricks in BRIKEN hole to protect AIDA |
|
|
393
|
Thu Oct 25 16:04:17 2018 |
ATS & JLT | Background |
Background BRIKEN control measurement
No DLT
2018/10/26
Start: 00:03
Stop: 09:46
ROOT file: 181026_0946_Bkg.root |
|
|
392
|
Thu Oct 25 09:32:25 2018 |
AT & AT | Low threshold, 252Cf, z=25cm, Perpendicular 90deg |
252Cf at z=25cm,ie 25 cm upstream the beam, source axis perpendicular at 90deg with respect to typical
orientation.
Low threshold.
18/10/25 START: 17:32
18/10/25 STOP: 23:30
** Configuration file:
181016_Conf_BrikenFull_2153_test_thr.xlsx
ROOT file:
181025_1732_252Cf_z25cm_2330.root
DLT files:
181025_1732_252Cf_z25cm_109.dlt
181025_1732_252Cf_z25cm_110.dlt
181025_1732_252Cf_z25cm_111.dlt
181025_1732_252Cf_z25cm_112.dlt
181025_1732_252Cf_z25cm_113.dlt
Comment:
- Clovers were placed outside the moderator
- AIDA not inside the moderator.
- Without In beam PE shielding.
****3He tubes 12 and 7 (RIKEN) were shifted 3cm aprox from their position upstream (CHECK DATA). |
|
|
391
|
Thu Oct 25 09:18:24 2018 |
ATS & JLT | F11 plastic |
F11 plastic (long plastic after MUSIC chamber)
connected to DACQ.
The connections were re-cabled and verified.
F11 plastic has 0.2mm thickness (check!)
Current HV: 2000V |
|
|
390
|
Thu Oct 25 09:07:00 2018 |
ATS & JLT | BRIKEN Calibration check |
We put the 252Cf source and check the aligment of 3He tubes gains:
File: 181025_1600_252Cf_gaincheck.root
All OK except 133, 35 and 140
He133: peak at 744keV
He135: peak at 783keV
He140: peak at 739keV
No gain correction applied at the moment but should be verified
before experiment |
|
|
389
|
Thu Oct 25 04:14:00 2018 |
AT & AT | Low threshold, 252Cf, z=25cm, Perpendicular 90deg |
252Cf at z=25cm,ie 25 cm upstream the beam, source axis perpendicular at 90deg with respect to typical
orientation.
Low threshold.
18/10/25 START: 12:13
18/10/25 STOP: 15:42
** Configuration file:
181016_Conf_BrikenFull_2153_test_thr.xlsx
ROOT file:
181025_0213_252Cf_z25cm_1542.root
DLT files:
181025_0213_252Cf_z25cm_106.dlt
Comment:
- Clovers were placed outside the moderator
- AIDA not inside the moderator.
- Without In beam PE shielding.
****** CHECK DATA, possible issue with the clock due to AIDA test. |
| Attachment 1: 181025_0213_252Cf_z25cm.png
|
|
|
|
388
|
Thu Oct 25 00:53:43 2018 |
AT & AT | Low threshold, 252Cf, z=-25cm, Perpendicular 90deg |
252Cf at z=-25cm,ie 25 cm downstream the beam, source axis perpendicular at 90deg with respect to typical
orientation.
Low threshold.
18/10/25 START: 08:55
18/10/25 STOP: 12:05
** Configuration file:
181016_Conf_BrikenFull_2153_test_thr.xlsx
ROOT file:
181025_0854_252Cf_zm25cm_1205.root
DLT files:
181025_0854_252Cf_zm25cm_103.dlt
181025_0854_252Cf_zm25cm_104.dlt
181025_0854_252Cf_zm25cm_105.dlt
Comment:
- Clovers were placed outside the moderator
- AIDA not inside the moderator.
- Without In beam PE shielding. |
|
|
387
|
Wed Oct 24 16:11:32 2018 |
AT & AT | Low threshold, 252Cf, z=-25cm |
252Cf at z=-25cm,ie 25 cm downstream the beam, source axis parallel.
Low threshold.
18/10/24 START: 00:12
18/10/25 STOP: 08:42
** Configuration file:
181016_Conf_BrikenFull_2153_test_thr.xlsx
ROOT file:
181025_0012_252Cf_zm25cm_0842.root
DLT files:
181025_0012_252Cf_zm25cm_096.dlt
181025_0012_252Cf_zm25cm_097.dlt
181025_0012_252Cf_zm25cm_098.dlt
181025_0012_252Cf_zm25cm_099.dlt
181025_0012_252Cf_zm25cm_100.dlt
181025_0012_252Cf_zm25cm_101.dlt
181025_0012_252Cf_zm25cm_102.dlt
Comment:
- Clovers were placed outside the moderator
- AIDA not inside the moderator.
- Without In beam PE shielding. |
|
|
386
|
Wed Oct 24 11:28:42 2018 |
AT & AT | Low threshold, 252Cf, z=-25cm |
252Cf at z=-25cm,ie 25 cm downstream, source axis perpendicular, typical orientation.
Low threshold.
18/10/24 START: 19:31
18/10/25 STOP: 00:01
** Configuration file:
181016_Conf_BrikenFull_2153_test_thr.xlsx
ROOT file:
181024_1930_252Cf_zm25cm_0001.root
DLT files:
181024_1930_252Cf_zm25cm_092.dlt
181024_1930_252Cf_zm25cm_093.dlt
181024_1930_252Cf_zm25cm_094.dlt
181024_1930_252Cf_zm25cm_095.dlt
Comment:
- Clovers were placed outside the moderator
- AIDA not inside the moderator.
- Without In beam PE shielding. |
|
|
385
|
Wed Oct 24 11:27:51 2018 |
AT & AT | Low threshold, 252Cf, z=10cm, source axis parallel |
252Cf at z=10cm, source axis parallel upstream to the beam.
Low threshold.
18/10/24 START: 15:44
18/10/24 STOP: 19:00
** Configuration file:
181016_Conf_BrikenFull_2153_test_thr.xlsx
ROOT file:
181024_1544_252Cf_z10cm_1900.root
DLT files:
181024_1544_252Cf_z10cm_089.dlt
181024_1544_252Cf_z10cm_090.dlt
181024_1544_252Cf_z10cm_091.dlt
Comment:
- Clovers were placed outside the moderator
- AIDA not inside the moderator.
- Without In beam PE shielding.
******* AT: By mistake duplicated entry from 385. |
|
|
384
|
Wed Oct 24 07:42:42 2018 |
AT & AT | Low threshold, 252Cf, z=10cm, source axis parallel |
252Cf at z=10cm, source axis parallel upstream to the beam.
Low threshold.
18/10/24 START: 15:44
18/10/24 STOP: 19:00
** Configuration file:
181016_Conf_BrikenFull_2153_test_thr.xlsx
ROOT file:
181024_1544_252Cf_z10cm_1900.root
DLT files:
181024_1544_252Cf_z10cm_089.dlt
181024_1544_252Cf_z10cm_090.dlt
181024_1544_252Cf_z10cm_091.dlt
Comment:
- Clovers were placed outside the moderator
- AIDA not inside the moderator.
- Without In beam PE shielding. |
|
|
383
|
Wed Oct 24 03:49:01 2018 |
AT & AT | Low threshold, 252Cf, z=0cm |
252Cf at z=10cm, source axis perpendicular to beam axis, typical orientation.
Low threshold.
18/10/24 START: 11:50
18/10/24 STOP: 15:37
** Configuration file:
181016_Conf_BrikenFull_2153_test_thr.xlsx
ROOT file:
181024_1150_252Cf_z10cm_1537.root
DLT files:
181024_1150_252Cf_z10cm_085.dlt
181024_1150_252Cf_z10cm_086.dlt
181024_1150_252Cf_z10cm_087.dlt
181024_1150_252Cf_z10cm_088.dlt
Comment:
- Clovers were placed outside the moderator
- AIDA not inside the moderator.
- Without In beam PE shielding. |
|
|
382
|
Mon Oct 22 16:54:10 2018 |
AT & AT | background, low threshold, No Pulser |
18/10/23 START: 00:53
18/10/24 STOP: 10:12
Background measurements. Checking noise!
Pulser is totally disconnected.
** Configuration file:
181016_Conf_BrikenFull.xlsx
ROOT file:
181023_0053_Nopulser_bckgn_1012.root
DLT files:
181023_0053_Nopulser_bckgn_084.dlt
Comment:
- Clovers were placed outside the moderator
- AIDA not inside the moderator.
- Without In beam PE shielding.
***** NO neutron signals are observed without pulser (?!). Online plot provided. |
| Attachment 1: 181023_0053_Nopulser_bckgn_1012.png
|
|
|
|
381
|
Mon Oct 22 11:42:50 2018 |
AT & AT | 252Cf, z=25cm, no clovers |
252Cf source placed using cardboard piece at z=25cm, source axis parallel upstream to the beam axis.
18/10/22 START: 19:44
18/10/22 STOP: 00:04
** Configuration file:
181016_Conf_BrikenFull_2153_test_thr.xlsx
ROOT file:
181022_1944_252Cf_z25cm_SourceParAxis_0004.root
DLT:
181022_1944_252Cf_z25cm_SourceParAxis_081.dlt
181022_1944_252Cf_z25cm_SourceParAxis_082.dlt
181022_1944_252Cf_z25cm_SourceParAxis_083.dlt
Comment:
- Clovers were placed outside the moderator
- AIDA not inside the moderator.
- Without In beam PE shielding. |