Entry  Tue Oct 16 03:15:03 2018, ATD & JLT, Start of REP run preparation (RIBF148) Screenshot_from_2018-10-16_14-57-33.png
16-Oct-2018

- Start DACQ computer: OK!
Entry  Fri Dec 1 07:29:47 2017, Vi Ho Phong, [Hardware][Configuration] Return the PB-5 Pulse generator (Pulser 2) and the 428F Linear Fan in-out modules and replace with a new Linear Fan in-out module  7x
The PB-5 Pulse generator labeled "Pulser 2" and the 428F Linear Fan in-out labeled "Clover G7" from BRIKEN NIM crate have to be returned to the RIBF electronics
pool because the 
borrowing time is expired.
Entry  Fri Nov 24 00:45:36 2017, AT, ur17b BRIKEN files 
List of files in BRIKEN DAQ PC:

Experiment DA1702 (including the test with YSO during first night)
Entry  Thu Nov 16 02:32:23 2017, BRIKEN collaboration, YSO thresholds dynVSdynGated.png
YSO Tests 
Thresholds & calibrations
Entry  Tue Oct 31 06:57:53 2017, OH, TD, AIDA Oct17 Beam Test Analysis AIDA_Test_Results.pdf
PDF file attached with the preliminary analysis of the October 2017 AIDA in beam test.
Entry  Fri Oct 27 10:04:31 2017, SG, RY, LISE++ file for BRIKEN1 Cu82_Be5mm_Brho7.7000Tm_F1deg8mm_F5deg8mm_F7_D80mod_F11nuclei.png
Here attached the lise file for BRIKEN1
Entry  Thu Oct 26 09:01:53 2017, BCR, Number of Counts vs Expected Efficiency for Proposals ExpectedBRIKEN_Precision.pdf
 

Here is a doc with an estimate of the statistical uncertainty P1n and P2n as a function of the number of implant-beta correlations for future
experiments.
Entry  Wed Oct 25 19:14:09 2017, NB, First pass efficiency for current experiment 
I roughly (~10%) checked the efficiency for 4 of the peaks from  171025_1222_1355_152EuCloverCal.root.

I checked the peaks at 122, 245, 344, and 1408. The indication is an efficiency of 6.9,  9,   9,8, and 2.6 % for these gamma rays
repectively.
Entry  Wed Oct 25 18:12:43 2017, Sunflower collaboration, Gamma sources web page 
http://www.nishina.riken.jp/collaboration/SUNFLOWER/misc/util/ribfsource.html
Entry  Wed Oct 25 17:49:09 2017, JT, AM, NB, DS, DM, AT, pulser in CLOVERS 
G7 pulser weird behavior

Entry  Wed Oct 25 01:02:14 2017, RG, Status of the YSO array 
The presently used detector has 48x48x5 mm3 YSO array with 1x1mm2 segments. They are separated with ESR.  We use H8500 pspmt and 2 mm quartz diffuser
with rough edges. Coupling is done with sylgard.  Anger logic uses a conventional resistive network with 1kohm and we have 4 position signals and
dynode.  The position signals (xa,xb,ya,yb) are split in Low gain branch connected directly to 5V range input in Struck and High gain branch, which
Entry  Tue Oct 24 13:14:47 2017, RY, Calibration data for clovers 171024_1858_152Eu_G7.root171024_2112_152Eu_D4.root
for energy calibration with 152Eu source

171024_1858_152Eu_G7.root
171024_2112_152Eu_D4.root
Entry  Tue Oct 24 03:08:23 2017, RG, RY, MS, JA, RCF, SG, AT, light pulser YSO 171024_1107_YSOPlots.rootScreenshot_from_2017-10-24_11-09-54.png
light pulser measurement
SN's pulse generator 20ns width
Entry  Mon Oct 23 09:43:00 2017, RG, RY, MS, SG, AT, 137Cs YSO 171023_1655_137CsYSO.root171023_1655_1741_137CsYSO_070.rootScreenshot_from_2017-10-23_17-51-37.pngScreenshot_from_2017-10-23_17-54-53.png
measurement of 137Cs with YSO 
after matching the gains of the 
HighGain channels
Entry  Sun Oct 22 11:32:59 2017, JA,ATD, JLT, New Gasific Configuration file 171022_1846_Conf_BrikenFullYSO.xlsx
 
Entry  Sun Oct 22 11:14:35 2017, JA,ATD,JLT,RG,RY,SG,MS,PV,Liu,KR, 252Cf source Wasabi-BrikenYSO 
252Cf source in the middle of Wasabi and YSO in front of Clover D4

Start: 19:15
Entry  Sun Oct 22 10:26:44 2017, JA,ATD,JLT,RG,RY,SG,MS,PV,Liu,KR, First Wasabi-YSO-BRIKEN correlation test 171022_1846_Conf_BrikenFullYSO.xlsx171022_1846_60Co_WasabiBrikenYSO.png
60Co source between YSO and Wasabi in front of D4 Clover
Briken configuration file: 171022_1811_Conf_BrikenFullYSO.xlsx WRONG!!!
Start: 18:25
Entry  Sun Oct 22 08:52:17 2017, KR,JA,ATD,JLT, Clover resolution 
Resolution check using 40K 1460 keV line 
in the background measurement overnight
(File: 171022_0126_1101_bck.root)
Entry  Sun Oct 22 05:15:26 2017, RG,RY,MS,ATD,JA,JLT, Check of YSO 12x
We start a systematic check of YSO in BRIKEN DACQ

WE check and correct a configuration file for YSO alone
Entry  Sat Oct 21 17:30:23 2017, JA, ATD, JLT, Background measurement overnight Screenshot_from_2017-10-22_11-26-49.pngScreenshot_from_2017-10-22_11-26-10.pngScreenshot_from_2017-10-22_11-26-28.png
22/10/2017
Start: 01:26
Run: 037