AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  CARME  ELOG logo
Message ID: 11     Entry time: Mon Sep 14 12:07:40 2020
Author: JM, CB (remotely) 
Subject: Baking status 
Mon 14 Sep 09:50

Jordan moved the TC08 picolog 2 away from the chamber and built a makeshit heatshield. 
Picolog 2 was facing a slit in the cloth thorugh which the scroll pump flexible hose is connected.
Baking restarted in automatic mode from 80 C

12:09
Baking progressed OK - see attached.

~13:00
CJC 2 getting too high again. Baking went off. Jordan caught it and temporarily set in manual mode.
Moved picolog 2 further away. Increased amount of heatshield. Restarted in auto

17:30
Baking continues OK - see attached.
Will leave overnight.

19:30
Baking continues OK, T=124C , see attached.
However rates shown on Baking.x program appear nonsensical. Too high, too low, sometimes they all go to zero.
Clearly a bug in the code. Should not affect baking for now.
Will try to debug.
Attachment 1: 20200914_1207_temps.png  110 kB  Uploaded Mon Sep 14 13:10:04 2020  | Hide | Hide all
20200914_1207_temps.png
Attachment 2: 20200914_1207_ctrl.png  41 kB  Uploaded Mon Sep 14 13:11:34 2020  | Hide | Hide all
20200914_1207_ctrl.png
Attachment 3: 20200914_1727_temps.png  91 kB  Uploaded Mon Sep 14 18:27:34 2020  | Hide | Hide all
20200914_1727_temps.png
Attachment 4: 20200914_2000_temps.png  102 kB  Uploaded Mon Sep 14 21:01:13 2020  | Hide | Hide all
20200914_2000_temps.png
ELOG V3.1.4-unknown