Message ID: 541
Entry time: Thu Apr 18 13:34:06 2024
|
Author: |
TD, JG |
Subject: |
Thursday 18 April |
|
|
14.30
per FEE64 1.8.L spectra
aida11 pulser peak width 52 ch FWHM
14.45 EG&G Ortec 448 Pulser OFF
all stats & histograms zero'd
alpha background
22.00 DSSSD bias and leakage current OK - attachment 10
FEE64 temperatures OK - attachment 11
ADC data item stats - attachment 12
per FEE64 Stat spectra - attachment 13
Comparison of attachment 9 (per FEE64 Rate spectra with pulser) indicates significant issues (downstream of adaptor PCB) for DSSSD #1