AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  CARME  ELOG logo
Message ID: 69     Entry time: Thu Sep 16 15:24:29 2021
Author: JM RSS 
Subject: CARME vented + thermocouple feedthrough 

Chamber has been vented back to atmosphere. 

Motor was able to be pulled back to insert bloacking pin once at atmosphere.

Thermocouple feedthrough constructed with 4 thermocuples attached. Internal thermocouples each ~200mm in length with external thermocouples each ~1m in length. All were tested by plugging into TC08 and using baking code. Feedthrough will be installed under the motors flange tomorrow. 

Survey team come tomorrow morning, afterwhich ion pump and RGA will also be installed. 

Attachment 1: 1.jpg  1.247 MB  Uploaded Thu Sep 16 16:33:29 2021  | Hide | Hide all
1.jpg
Attachment 2: 2.jpg  820 kB  Uploaded Thu Sep 16 16:34:26 2021  | Hide | Hide all
2.jpg
Attachment 3: 3.jpg  223 kB  Uploaded Thu Sep 16 16:35:26 2021  | Hide | Hide all
3.jpg