ID |
Date |
Author |
Subject |
Text |
 |
|
175
|
Wed Jan 26 19:16:18 2022 |
JM, OH | Water cooling, FEE MAC address | Water cooling has been connected to the
FEE's. Flow meter installed to outlet
manifold by way of two right angled pieces. |
  |
|
174
|
Sat Jan 15 15:38:25 2022 |
JM, OH, TD | FEE installation, cable management | All 8 FEEs have been installed in their
holders.
The damaged feedthrough does not |
    |
|
173
|
Sat Jan 15 11:28:10 2022 |
JM | Inventory |
Vaccum
DN250CF 1x gaskets
DN200CF 5x gaskets
DN160CF 1x gaskets, |
 |
|
172
|
Fri Jan 14 10:00:18 2022 |
TD, OH, JM | Friday 14 January |
11.00 Per https://elog.ph.ed.ac.uk/CARME/170
and https://elog.ph.ed.ac.uk/CARME/171 issues |
|
|
171
|
Thu Jan 13 10:33:24 2022 |
JM, OH, TD | Adaptor Board misallignment | The ERNI connector is not central on the
adaptor boards causing misallignment with
the FEE. |
  |
|
170
|
Wed Jan 12 20:07:20 2022 |
JM, OH, TD | FEE holder installation | 4 FEE holders installed on CARME. Mounting
procedure devised due to limited working
space to install. |
10x |
|
168
|
Tue Jan 11 12:45:15 2022 |
TD, JM, OH | Coolant manifold installation + detector pin tests | 2x cooling water manifolds with 2x 8 ports
have been installed - see attachments 1 &
2
|
  |
|
167
|
Mon Jan 10 20:52:53 2022 |
JM, TD, OH | Baking tent removed, VI tests | Baking tent, frame and trays fully
removed from CARME chamber. Removal of baking
frame required removing air lines from upper |
  |
|
166
|
Tue Dec 7 09:39:59 2021 |
CB (remotely) | Gate valves closed | Dry run complete.
YR09 section valves closed. Section
isolated. Manual valve between electron target |
|
|
165
|
Tue Nov 30 10:49:43 2021 |
CB (remotely) GV | Gate valve opened | GSI vacuum division IE514 gauge connected.
Gleb reports gate downstream opened.
Vacuum goes up and recovers as expected. |
|
|
164
|
Sun Nov 28 15:57:45 2021 |
CB (remotely) | Nov 2021 temperatures | Attached |
|
|
163
|
Fri Nov 26 10:01:45 2021 |
CB (remotely) ML | Power cut in the CRYRING cave | ML reports power cut in CRYRING cave. Reasons
unknown. No effect on vacuum, but top valve
closed for safety reason. Ramping down temperature. |
|
|
162
|
Wed Nov 24 12:08:08 2021 |
CB + ML OG | Ion pumps on | Found SIP power unit for D2000 #4 not functioning
correctly. Max current regardless of ion
element connected to. |
  |
|
161
|
Tue Nov 23 08:49:33 2021 |
CB + ML | NEG activation completed | Found temperatures and pressure stabilised
overnight. See attached.
12:00 Began activation of 3x UHV1400 |
   |
|
160
|
Mon Nov 22 08:06:01 2021 |
RSS CB + ML | NEG activation starts | 08:20 Temperature of channels 1-3 and 5-6
were reduced to 120 degrees. Immediately
after, the conditioning of the NEGs was stopped |
  |
|
159
|
Fri Nov 19 12:03:51 2021 |
CB RSS | Conditionings NEGs | Increased conditioning temperature in D2000
#3 and 4 (behind detectors) to 150C from
130/140C. No appreciable change in temperature. |
 |
|
158
|
Thu Nov 18 13:50:51 2021 |
CB RSS | Thermocamera images | Used thermocamera provided by GSI beam
diagnostic division to take some pictures
of YR09. See attached. |
14x |
|
157
|
Thu Nov 18 12:29:50 2021 |
CB RSS | Conditioning D2000 NEGs | D2000 thermocouples report temperatures
around 100C (D2000 #1,2) and 120C (3,4).
1,2 are in the intermediate chamber, 3,4 |
 |
|
156
|
Wed Nov 17 07:12:22 2021 |
RSS CB + ML (remotely) | CARME baking with tape continues 17 Nov | CARME baking with heating tape continues
Left overnight at -
Ch1-3: 140 degrees |
7x |
|
155
|
Tue Nov 16 12:05:13 2021 |
CB RSS | CARME baking with tape continues | CARME baking with heating tape continues
Left overnight at 100C. Turned
up this morning at 9:00 to 120C. |
   |