AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
DESPEC |
 |
|
Message ID: 526
Entry time: Sun Mar 3 16:06:11 2024
|
Author: |
TD |
Subject: |
Sunday 3 March |
|
|
17.00 FEE64 41:a0:71 ASIC #2 u/s
Replaced ASIC mezzanine with new ASIC mezzanine
Installed as aida04
Unable to boot nnrpi2 - manual/local control of BNC PB-5 and FEE64 PSU - no interlock
Water temperature & pressure as measured outside S4 area - OK
Manually power FEE64s
BNC PB-5 local control/ON
Amplitude 1.0V
Attenuation x10
Polarity -
tau_d 1ms
Frequency 22Hz
FEE64 temps OK - attachment 1
All system wide checks OK
WR timestamp OK - attachment 2
aida04 ASIC settings - attachment 3
ADC data item stats OK - attachment 4
aida04 Rate spectrum - attachment 5
aida04 *.*.L spectra - attachments 6-9
1.8.L pulser peak width 15 ch FWHM
aida04 1.8.W spectra - 20us FSR - attachment 10-11
MIDAS configuration - attachment 12
DHCP config - attachment 13
18.30 FEE64 41:f6:b7 ASIC #1 u/s
Replaced ASIC mezzanine with new ASIC mezzanine
Installed as aida02
All system wide checks OK
FEE64 temps - attachment 14
aida02 ASIC temperature c. 512 deg C !
PSU/Virtex temps OK
Mezzanine ambient to touch - assume faulty sensor/poor connection?
aida02 ASIC settings - attachment 15
WR timestamp OK - attachment 16
ADC data item stats OK - attachment 4
aida04 Rate spectrum - attachment 18
aida02 1.8.W spectra - 20us FSR - attachment 19-20
aida02 *.*.L spectra - attachments 21-24
1.8.L pulser peak width 16 ch FWHM
19.30 Power OFF
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|