AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
DESPEC |
 |
|
Message ID: 527
Entry time: Mon Mar 4 11:35:30 2024
|
Author: |
TD |
Subject: |
Monday 4 March |
|
|
12.30 FEE64 41:d7:cd ASIC mezzanine u/s, no data, ASIC temperature c. 20 deg C ( ambient ) low
Replaced ASIC mezzanine with new ASIC mezzanine
Installed as aida08
Water temperature & pressure as measured outside S4 area - OK
Manually power FEE64s
BNC PB-5 local control/ON
Amplitude 1.0V
Attenuation x10
Polarity +
tau_d 1ms
Frequency 22Hz
FEE64 temps - attachment 1
aida02 with new ASIC mezzanine continues to ramp to c. 512 deg C - initially reads 0, next refresh c. 70 and third refresh c. 512 deg C
aida02 Virtex and PSU temps OK, aida02 cooling plate ambient to touch
aida08 ASIC temp c. 45 deg C as expected, Virtex and PSU temps OK
All system wide checks OK
WR timestamp OK - attachment 2
aida04 ASIC settings - attachment 3
ADC data item stats OK - attachment 4
aida04 Rate spectrum - attachment 5
aida04 *.*.L spectra - attachments 6-9
1.8.L pulser peak width 13 ch FWHM
aida04 1.8.W spectra - 20us FSR - attachment 10-11 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|