Entry time: Fri Jan 31 17:26:47 2025
|
Author: |
CC, TD, MP |
Subject: |
Friday 31 January contd. |
|
|
18.18 bPlas and BB7 installs complete
Restart AIDA DAQ, Merger, Tape Server and re-test
per p+n FEE64 1.8.L spectra - attachment 1
aida09 pulser peak width 69 ch FWHM = 42keV FWHM
per FEE64 Rate Spectra - attachment 2
BB18 p+n FEE64s very good/good, n+n FEE64s OK - could be improved
BB7 aida10 p+n asics good, n+n asics 1x good, 1x OK
per FEE64 1.8.W spectra - 20us FSR - attachments 3-5
ADC data item stats - attachment 6
aida01 aida03 30k, all other BB18 p+n FEE64s < 20k, n+n FEE64s 100/270k
BB7 aida10 19k
WR timestamps OK - attachment 7
Merger, TapeSever etc - attachments 8-9
disk directpry /TapeData/FEB25
working OK
DSSSD bias & leakage current OK - attachment 10
System wide checks OK - attachment 11-15
note global clock status 6 errors reported earlier today have now gone
FEE64 temperatures OK - attachment 16
18.38 Transition to safe state
DAQ STOP
disable data transfer 1
detector bias OFF
FEE64 power OFF
Can restart as follows
1) FEE64 power ON
2) DAQ RESET
3) DAQ SETUP
4) Enable histogramming
5) Enable waveforms
6) Detector bias ON
7) Restore ASIC settings
8) ASIC Control
9) FEE64 temperatures
10) System wide checks
sync ASIC clocks
11) FADC control - calibrate ADCS for *all* FEE64s
12) System wide checks contd.
13) DAQ GO
14) Check ADC data item stats
15) Check WR timestamps
If all OK can re-connect to Merger/TapeServer as follows
1) DAQ STOP
2) enable data transfer 1
3) DAQ GO
Can disconnect from Merger/TapeServer as follows
1) DAQ STOP
2) disable data transfer 1
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|