AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC, Page 21 of 36  ELOG logo
  ID Date Authordown Subject Text Attachments
  54   Fri Jun 28 15:05:49 2019 NHAida 28/6/2019 - WATER OFFThe humidity sensor hasn't reported
humdities under 90% yet. In order to try
and help dry the pipes and turn AIDA the
  
  55   Mon Jul 1 09:42:16 2019 NHAIDA - 01/07/2019Water is still off.

Checked dewpoint sensor still flashing
red light (indicating an error). Pipes warmed.
  
  57   Wed Jul 10 10:46:05 2019 NHAIDA/Water 10.07.2019S4 Conditions:

25.2 C, 35.3% RH, Td = 8.8 C

Water temperature: 20 C
  
  58   Mon Aug 5 12:57:48 2019 NHS4 ConditionsTemp: 26 C
Humidity: 50%
Dew Point: ~15 C
  
  59   Fri Aug 9 12:33:14 2019 NHAida Test 9/8/19 - FEES 1-6The water issue has been resolved - merely
incorrect temperature guages (replaced).
Water is 20 C as required.
  
  60   Tue Aug 13 09:17:02 2019 NHAIDA 12/08/2019The other 6 FEEs were powered on one at
a time to check no issues. None found.
All 12 FEEs plugged back into the
 AIDATemps-1208191457.png 
  61   Tue Aug 13 13:55:45 2019 NHNew Merger/MBS Test RunsAIDA Finally working again so can test
VP's added WAVE histogramming mode.

Fig 1. Waveforms for FEE12 ASIC
 6x 
  63   Wed Aug 14 14:40:18 2019 NHHigh Frequency Noise in AIDAFigures of intense high frequency noise
pickup in AIDA.

Will investigate source in S4.
 HF.pngHFZ.png 
  64   Thu Aug 15 14:09:39 2019 NHAIDA WaveformsNoise seemed less strong today, unsure
of cause. I had re-taped snout as tape was
not sticking well.
 10x 
  65   Fri Aug 16 15:09:05 2019 NHAida Waves (All FEEs)Included waves for all 12 FEEs

FEES 1, 5, 9 all show this HF noise
on it, and are all located at the top of
 12x 
  66   Mon Sep 16 12:24:46 2019 NHAIDA Monitoringhttps://docs.google.com/spreadsheets/d/1fZBybHodSvacUdKpDM_VKznqTrh49CKnqKVQVMEDWGs/edit?usp=sharing

A spreadsheet of ambient conditions
and water temperature in S4 for AIDA.
  
  74   Thu Oct 31 14:27:06 2019 NHWR TimestampsAll 12 FEEs have valid WR Timestamps
Had to powercycle aida09 once as before raw
readout was displaying upper 12 bits of WR
 WRTimes.png 
  77   Thu Oct 31 17:13:54 2019 NHAlpha RunBeginning Alpha run

Thresholds set to 100 (LEC/Slow) 255 (LEC/Fast)
  
  79   Fri Nov 1 09:30:17 2019 NHDatabase error stopping run?Strange error about Database options appearing
during run stop.
Does not seem to affect stopping/starting/data?
 ErrorStopDatabase.png 
  85   Mon Nov 4 08:48:18 2019 NHAlpha Status 04.11.2019Alpha run has been running over the weekend.
Stats looking positive with notable exception
that aida10 is not sending much data. Will
 alphas-bias.pngalphas-temp.pngalpha-aida10.pngstats-0511.pngspec-0511.png 
  86   Mon Nov 4 13:17:12 2019 NHReport: aida10 database corruptionOften aida10 database seems to get corrupted
(ASIC values are all 0xad)
Unsure why it's only aida10, might
 db_working.txtdb_corrupt.txt 
  91   Mon Nov 11 10:34:19 2019 NHAlpha Analysis (11.11.19)Analysis of approx 20GB of data since Thursday
night (not continuous - estimate of hours
incominG)
 9x 
  97   Wed Nov 13 15:38:20 2019 NH[HowTo] MBS for AIDA (Nov 19)For MBS integration AIDA forwards its data
to an MBS Foreign Data Receiver (FDR), currently
the PC assigned for this task is x86l-94.
  
  100   Fri Nov 22 08:55:31 2019 NHWaveformsAIDA Waveforms immediately after beam at 21.11.19
8am or so.
 waves_21119_even.png21119_waves_odd.pngwaves_21119_even_z.pngwaves_21119_odd_z.png 
  101   Mon Nov 25 13:25:51 2019 NHWaveforms> AIDA Waveforms immediately after beam at
21.11.19 8am or so.
 wave_odd.pngwave_even1.png 
ELOG V3.1.4-unknown