ID Date Author Subject Text Attachments
  585   Sun Apr 21 23:20:45 2024 TDMonday 22 April00.15 Data link to MBS was closed by MBS -
reason(s) unknown
      NH restarted MBS receiver and data
 38x 
  584   Sun Apr 21 00:34:33 2024 TDSunday 21 April01.35 DSSSD bias & leakage current OK - Grafana
- attachment 1
 37x 
  583   Sat Apr 20 17:52:27 2024 TD, NHSaturday 20 April contd.
Note 40kBq 60Co source placed on bottom surface
of snout centred on bPlas detectors and DSSSDs
 Screenshot_from_2024-04-20_18-51-53.pngScreenshot_from_2024-04-20_21-52-08.pngScreenshot_from_2024-04-21_10-52-19.png 
  582   Sat Apr 20 11:40:38 2024 TDOffline analysis R12 & R13 ( pulser walkthrough )
S100 ADC offsets using pulser walkthrough
data from data files R12 & R13
 17x 
  581   Sat Apr 20 03:20:23 2024 TDSaturday 20 April
Note 60Co source (current activity c. 43kBq)
was installed yesterday evening - assume
 23x 
  580   Fri Apr 19 13:52:34 2024 JB, CC, TDFriday 19 April contd.
12.00 Replaced AIDA ASIC mezzanine of aida01
to fix issue with asic #3
 23x 
  579   Fri Apr 19 02:35:14 2024 TDFriday 19 April
03.30
 38x 
  578   Thu Apr 18 13:59:55 2024 TD, JBWR timestamp OK, readout time not OK
WR timestamps appear to be OK but readout
times fort aida03, aida04 and aida09 appear
 Screenshot_from_2024-04-18_14-59-38.pngScreenshot_from_2024-04-18_15-01-15.pngScreenshot_from_2024-04-18_15-01-10.pngScreenshot_from_2024-04-18_16-13-55.pngScreenshot_from_2024-04-18_16-13-48.png 
  577   Thu Apr 18 08:57:48 2024 JB, TDThursday 18 April09.52 DAQ continues file S100_alpha/R5
       note merger not functioning correctly
due to aida03, aida04 & aida09 out of WR
 36x 
  576   Tue Apr 16 23:37:58 2024 TDWednesday 17 April
00.40 Observed zero data fron aida03, aida04,
aida09
 44x 
  575   Tue Apr 16 10:49:37 2024 TD, JB, CCTuesday 16 April11.45 DSSSD bias & leakage current OK - attachments
9-10
 24x 
  574   Mon Apr 15 09:23:43 2024 CC, JB, HMA, TDNoise in AIDA when connecting BB79:40 started up AIDA. TEMP OK. Noise condition
same as Saturday. Attachments 1-3.

9:56 BB7 premaps connected but
 13x 
  573   Sat Apr 13 14:04:44 2024 JB13 April Noise checks15:04 bPlast was left powered over night.
Powering up to check noise conditions - also
to check shifting of Germanium baseline (no
 9x 
  572   Thu Apr 11 22:04:50 2024 JB12.04.2024 AIDA-bPlast noise optimisation15:00 Platform in, biasing detector. Temp
OK Attachment 1. test - Pulser OFF.

          
 Screenshot_from_2024-04-12_15-11-18.pngScreenshot_from_2024-04-12_15-22-54.pngScreenshot_from_2024-04-12_15-22-26.pngScreenshot_from_2024-04-12_18-07-35.pngScreenshot_from_2024-04-12_18-07-27.png 
  571   Thu Apr 11 08:27:21 2024 JB, TD11 April noise tests
TO-DO for 11.04.2024

- Try bringing
bPlast drain wire ground back to the PSU
ground for PN 300 and R&SRMP4040.

-
 48x 
  570   Wed Apr 10 08:37:57 2024 JB, CC, TDWednesday 10 April
09.38 CC completed install of Bplast driver
PCBs yesterday evening.
 47x 
  569   Tue Apr 9 13:22:41 2024 TD, JB, CCContinued of 9th of AprilDisconnected test - pulser from aida08.
Noise conditions improved slightly, probably
less pulser events. (Attachments 1 &
 Screenshot_from_2024-04-09_14-22-20.pngScreenshot_from_2024-04-09_14-22-16.pngScreenshot_from_2024-04-09_17-18-45.png 
  568   Tue Apr 9 09:04:07 2024 TD[How To] Remote control of BNC PB-5
BNC PB-5 set to remote control
 Screenshot_from_2024-04-09_10-03-42.png 
  567   Tue Apr 9 08:37:47 2024 TDTuesday 9 April09.36 Cooling water temperature and flow OK

      FEE64 power ON
 16x 
  566   Mon Apr 8 16:39:00 2024 JB, CC, TDMonday 8 April
17.32 Power and detector bias cycle
 11x 