ID |
Date |
Author |
Subject |
23
|
Fri Jan 25 09:54:12 2019 |
CA, TD, NH, VP | 25th January 2019 | 09.30: AIDA still running, not writing to file
Still running ok
No water leaks
09.45: FEE64 temperatures ok (attachment 1)
Good event statistics ok (attachment 2)
10.00: Results of System Wide Checks
ADC Calibration check (attachment 3)
Clock Status (attachment 4)
Sync error counter (attachment 5)
White Rabbit Decoder status (attachment 7)
Memory Information (attachment 8)
10.30: Clock ReSync performed (attachment 6)
Start writing to file TapeData/NULL/R7_0
11.50: DAQ stopped, TapeServer stopped
TapeServer restarted, DAQ restarted
Writing to file TapeData/NULL/R9_0 with Clock ReSync at start.
Wasn't writing to disk, restarting (DAQ stop, TapeServer stop, TapeServer start TO DISK, DAQ Start)
Writing to file TapeData/NULL/R10_0
12.04: DAQ stopped, TapeServer stopped, merger stopped
FEE64 power relay switched off
Raspberry PIs switched off and disconnected
water cooler stopped
Mains IEC cables removed from relay
|
22
|
Thu Jan 24 17:16:32 2019 |
CA, TD, NH, VP | MACB time switch settings | AIDA@DESPEC MACB time switch settings before and after WR are as follows.
Attachment 1 - MACB 1-4 from left to right
MACB number: 1 2 3 4
100MHZ clock setting (before WR): 0 2 2 2
White Rabbit setting (after WR): 3 3 3 3 |
21
|
Thu Jan 24 14:53:26 2019 |
TD | Analyses GSIJan19/R2_0 - R4_0 |
Attachment 1 - summary analysis R2_0
increased slow comparator & LEC/MEC fast comparators cf. R1_0 to minimise random noise
Attachment 2 - summary analysis R3_0
per R2_0 but BNC PB-5 pulser frequency increased from c. 2Hz to c. 1kHz
pulser to *all* channels 12x FEE64s
Attachment 3 - summary analysis R4_0
open TapeSever file, DAQ GO, Start ASIC Readout (*correct* DAQ start sequence)
Observations
- some RESUME but *zero* PAUSE data items observed
- information types 4 & 5 not equal for R2_0 and almost, but not quite, equal for R3_0 & R4_0
- information types 4 & 5 start data file with 'correct' DAQ start sequence
|
20
|
Thu Jan 24 13:29:40 2019 |
TD | Analysis of GSIJan19/R1_0 |
Analysis of file GSIJan19/R1_0 (white rabbit data)
attachment 1 - overall statistics
attachment 2 - verbose output listing all data around first six identified timewarps
attachment 3 - analyser.f (old version for 48 bit timestamp)
attachment 4 - analyser.f (new version for 64 bit WR timestamp)
Observations
1) File does not commence with information type 4 & 5 data items
DAQ/TapeServer start sequence issue??
2) DAQ RESUMEs observed for all FEE64s - DAQ PAUSEs *not* observed
3) Number of information type 4 & 5 data items for each FEE64 *not* the same
4) reported idle & elapsed times incorrect - presumably analyser.f bug
|
19
|
Thu Jan 24 10:26:09 2019 |
VP, TD, CA, NH | Start up Merger | 1) Click on item "Merger MK2" in top panel
This kills any existing Merger existence and loads a new copy.
2) Start (if needed) browser conection to localhost:8115.
3) Click on "NewMerger Control"
4) In NewMerger Control, click "Reload"
5) If MergerState = IDLE, click "SETUP".
6) Wait until state = STOPed
7) If links shown are correct, click "GO"
8) Wait until popup about "Resume Merger" appears and click OK
9) when writing to disk, click "Forward Output to Data" |
18
|
Thu Jan 24 10:21:26 2019 |
CA, TD, NH, VP | January 24th 2019 | 10.30 AIDA setup and DAQ start complete
writing to file TapeData/NULL/R1
FEE64 Temperatures and Good event statistics ok (attachments 1 and 2)
11.00 results of analyser - shows significant number of timewarps in aida07, and a few in aida01-04
11.24 compared timestamps between mbs and AIDA, agree to 8 digits (attachment 3)
12.49 DAQ and Merger stopped and restarted
Raised slow comparator threshold from 0xa to 0x64
Raised fast comparator threshold (LEC/MEC) from 0xf to 0xff
12.51 Good event statistics: aida08 running at a normal rate
pulser peak (1.8.L) no longer has noise shoulder (attachment 4)
writing to file TapeData/NULL/R2
13.58 stopped writing to file. Pulser frequency increased to 1 kHz
writing to file TapeData/NULL/R3
15.00 file R4 - DAQ startup
17.58 trimmed and re-crimped ground wire on aida08
lowered pulser freq back to 50Hz
reduced slow comparator threshold to 0xa, fast comparator threshold (LEC) to 0xf
good event stats ok, aida08 no longer running hot (attachment 6)
1.8.L and 1.8.W spectra of aida08 (attachment 5)
pulser peak width now at 55.71
FEE64 temperatures ok (attachment 7) |
17
|
Wed Jan 23 17:03:08 2019 |
CA, TD, NH, VP | January 23rd 2019 | Post White Rabbit update
Attachment 1 - 1.8.W pulser waveform spectra for all FEE64
- pulser signals observed in all FEE64
Attachment 2 - good event statistics - aida08 has significant noise
Attachment 3 - FEE64 temperatures - all ok, but firmware version not displayed
Attachment 4 - 1.8.L pulser peak spectra
- widths:
aida01 - 23.66
aida02 - 19.76
aida03 - 20.43
aida04 - 18.82
aida05 - 20.05
aida06 - 41.04
aida07 - 21.24
aida08 - 205.02 (noisy FEE64)
aida09 - 24.81
aida10 - 22.85
aida11 - 13.98
aida12 - 19.90
Attachment 5 - New Merger statistics
Attachment 6 - Raw data stream |
16
|
Wed Jan 23 09:25:55 2019 |
CA, TD, NH, VP | MACB and setup pics 23.01.19 | Pictures of leaking cooling pipe
MACB switch settings
AIDA mount - needs rotated
Attachment 11 - AIDA@DESPEC MACB settings - 100MHz and WR |
15
|
Wed Jan 23 08:12:01 2019 |
CA, TD, NH, VP | January 22nd 2019 | Attachment 1 - 1.8.W waveform spectra with pulser signal
Attachment 2 - 1.8.L pulser peak spectra
widths: aida01 - 16.06
aida04 - 20.68
aida05 - 14.03
aida08 - 73.02
aida09 - 669.29
aida11 - 405.35
aida12 - 397.6
Attachment 3 - good event statistics
fast comparator threshold (LEC) at 0xff
FEEs 8,9,11,12 running hot
Attachment 4 - FEE64 temperatures - all ok
System wide checks ok, except aida02 fails ADC calibration
Leak found in seal on cooling pipe, switched power off |
14
|
Mon Dec 10 11:09:53 2018 |
CA, TD | Scintillator pictures/system checks for 07.12.18 | 11.12 System wide checks ok (attachment 1)
FEE temperatures ok (attachment 2)
bias/leakage currents ok (attachment 3)
11.13 1.8.W spectra for naida 1,9,10,11,12 (attachment 4)
1.8.W spectra expanded in Y direction (attachment 5)
pulser peaks (attachment 6)
11.20 pulser peak widths:
Aida10 -> 98.82
Aida09 -> 109.17
Aida11 -> 71.72
Aida12 -> 92.82
Aida01 -> 24.05
11.23 hit rates (attachment 7)
11.41 pictures of setup with scintillator mounted (attachment 8-14) |
13
|
Sat Dec 8 15:22:52 2018 |
TD | Raspberry Pi startup | nnrpi1 - FEE64 system consoles
nnrpi2 - USB-controlled ac mains relay, CAEN N1419ET
username: pi
password: *******
To startup TclHttpd server for web access via port 8015 execute
/MIDAS/TclHttpd/Linux-arm/TclHttpd-server
Update - 9 March 2020 command line should be
/MIDAS/TclHttpd/linux-arm/TclHttpd-server |
12
|
Thu Dec 6 20:45:47 2018 |
TD | Analysis of R2_0 |
Analysis of data file R2_0
Attachment 1 - summary output of TDR analyser
Attachment 2 - verbose output of TDR analyser |
11
|
Thu Dec 6 15:31:40 2018 |
TD, CA, NH | December 6th 2018 | 15.32 1.8.W spectra for aida01, aida09, aida10, aida11, aida12 (attachments 1, 2, 3)
2 and 3 expanded in Y and X direction respectively
15.35 good event statistics (attachment 4)
FEE temperatures (attachment 5)
leakage currents/bias (attachment 6)
17.20 attachment 7 - loose ground wire
attachment 8,9,10 - pictures of wire connections
attachment 11 - side on view of dsssd mounted in aida snout
17.28 Pulser widths:
aida10 -> 101.24
aida09 -> 136.22
aida11 -> 66.31
aida12 -> 87.17
aida01 -> 15.29 |
10
|
Sat Dec 1 16:43:47 2018 |
TD | Analysis of WRTest28Nov18.dat |
Attachment 1 - analyser.f
Modified version of TDR format analyser program to handle 64 bit WR timestamps
Attachment 2 - analyser program output
12092 blocks = 773888kb
8 timewarps
0 information code 5 data items
data file does not commence with either information codes 5 or 4
times/rates not correctly calculated - program assumes 10ns/tick
Attachment 3 - od -x output |
9
|
Sat Oct 13 17:26:53 2018 |
TD | Analysis of 3 October 2018 T4 tests | |
8
|
Thu Sep 20 09:39:17 2018 |
TD | AIDA downstream window |
The AIDA downstream window ('endcap' of snout) consists of 2x 50um aluminized mylar sheets |
7
|
Thu Sep 20 09:16:31 2018 |
TD | Images of AIDA setup for DESPEC | Attachment 1 - detail of MSL type BB18(DS)-1000 p+n junction side strips, bias ring, multiple guard rings, filed
plates and bond wires
Attachment 2 - MSL type BB18(DS)-1000 with Kapton PCBs attached
Attachment 3 - MSL type BB18(DS)-1000 with Kapton PCBs + 3M 1245 copper foil-screened Samtec FTSH ribbon cables
attached
Attachment 4 - AIDA snout installed (with 1x DSSSD) with cabling to 4x FEE64s in AIDA support assembly
Attachment 5 - AIDA + FATIMA array
Attachment 6 - AIDA FEE64 PSUs, network switch, USB-controlled ac mains relay
Attachment 7 - AIDA MACBs, pulser, detector bias
Attachment 8 - FATIMA DAQ
Attachment 9 - DESPEC 19" racks on detector platform |
6
|
Thu Sep 20 08:57:48 2018 |
OH | Thursday 20th September | 09:57 Alpha run stoppped R2_0 -> R2_2
Thresholds dropped from 0x64 to 0xa
Pulser turned on again
System wide checks ok
FEE Temp ok - attachment 1
Bias ok, leakage at 1.485uA at 160V
Stats in similar place to yesterday - attachment 2
Low Energy pulser walkthrough
1.0V to 0.4V in 0.1V steps
2 minutes per step
Pulser at 50Hz
10:29 Run started R3_0
10:45 Run stopped R3_2
1.8.L spectra for low energy pulser walkthrough - attachment 3
11:00
High Energy pulser walkthrough
1.8V to 0.V in 0.2V
2 minutes per step
Pulser at 50Hz
11:06 Run started R4_0
10:18 Run stopped R4_2
1.8.L spectra for high energy pulser walkthrough - attachment 4 |
5
|
Wed Sep 19 12:40:41 2018 |
OH, TD | Single DSSD Installation and Tests | On the 18th of September we installed DSSD 3208-14
Was found to have problems including exponential leakage behavior
V Ic
-5 15.4uA
-10 53uA
-15 Tripped at over 100uA
Cables were double checked, bond wires were checked and no problems could be observed.
DSSD replaced with 3208-15
Bias at -100V Ic 10.2uA GP/grnd Jumpers removed,
rising upwards, possible short circuit as well.
Jumpers replaced
Bias -100V Ic 1.6uA
Bias -160V Ic 1.62uA
Possibility that mechanical sample 3208-17 and 3208-14 have been swapped. (To check at later date)
FEEs powered up and Adapter boards and all grounding installed with DSSD going to 9,10,11 and 12
Rates shown - attachment 1
Waveforms - attachment 2-4
1.8.l Spectra - attachment 5
FEEs 10 and 11 around 60FWHM and 9 and 12 around 80-90FWHM
Stats page - attachment 6
Aluminium foil used to make sure all ports are light tight shows a stats drop of around half - attachment 8
New peak widths
aida09 - 79.65
aida10 - 67.37
aida11 - 59.99
aida12 - 97.83
Going to begin a systematic study of shaping time
FWHM
Shaping Time aida09 aida10 aida11 aida12
8 79.65 67.37 59.99 97.83
7 97.55 70.84 62.67 115.34
6 100.72 70.04 62.77 120.05
4 147.47 80.64 71.34 167.23
2 217.62 114.98 133.35 232.47
Shaping time 8us 1.8.L Shots
Shaping time 4us Stats appear obviously worse - attachment 7
1.8.L Widths - attachment 8
Shaping time 2us Stats appear worse again - attachment 9
1.8.L Widths - attachment 10
Shaping time 6us Stats are better than 2 and 4 - attachment 11
1.8.L widths - attachment 12
Shaping time 7us Stats are better than 6 - attachment 13
1.8.L widths - attachment 14
8us is the optimum shaping time.
Taken back to 8us and similar shaping times are observed.
Connected the ground of all 4 adapter boards with direct cables. Observed improvements in both FEEs 9 and 12
however the noise in 11 went
through the roof >500kHz - attachment 15
Tested just having 9,10 and 12 connected the rates are comparable to when disconnected - attachment 16
All 4 grounded together with a ground running to the snout as well stats No difference to all 4 together -
attachment 17
1.8.L Widths - attachment 17
aida09 104.76
aida10 84.58
aida11 281.77
aida12 82.58
Tested just having 10 and 12 directly connected. Same side of the DSSD: Noise is the same as when disconnected.
Stats - attachment 19
1.8.L spectra - attachment 20
aida09 107.13
aida10 67.02
aida11 66.78
aida12 131.51
Tested having 11 and 12 connected directly. This is the two adapter boards responsible for the bias with 11
carrying the core and 12 the braid.
Massively reduced noise in 9 and 12, comparable noise in 10 but much more noise in 11. Stats - attachment 21
aida09 63.96
aida10 76.64
aida11 125.96
aida12 68.05
1.8.L spectra - attachment 23
18:39 Alpha run started in /TapeData/Sep18/R2
Writing at 96kb/s
11 Was fast at start bus asic control brought it down
|
4
|
Thu Sep 13 17:21:59 2018 |
TD | AIDA upstream window |
The upstream window of the AIDA detector assembly consists of 2x 50um aluminized mylar |
|