Message ID: 72     Entry time: Thu May 30 02:03:33 2019
Author: A. Algora 
Subject: Ref detector position changed, AIDA plastik HV changed 

The position of the reference detector in the ancilliary card was changed from channel 10 to channel 9. Remember that this is important

in relation to all files saved until now.

The HV of AIDA plastic was changed from 1300 V back to 1000 V.

A new configuration file provided by Alvaro and Jose, with some small changes will be loaded.

cp 20190529_Conf_ATJLT.xlsx 20190530_Conf.xlsx

The new configuration file has been uploaded. Reference detector appears in channel V2A8C9. It was necessary to change the range of the histogram, and to recalibrate it.

With the new configuration parameters the pulser appears at 1360 keV in the calibrated spectra of the reference detector.

The threshold of the Ref detector has also been changed. From 4400 to 2800.

Attachment 1: 20190530_Conf.xlsx  35 kB  Uploaded Thu May 30 03:07:07 2019