Tue Jun 11 01:45:55 2019, Pancho, Ale, background
|
Start: 09:46
|
Tue May 14 09:41:24 2019, TD, Welcome to the DTAS Elog
|
|
Sun Jun 16 02:07:29 2019, Victor, Total sum (black) and anodes 
|
Different cuts at high energy in individual spectra (anodes) and effect in the total sum |
Mon Jun 17 19:12:33 2019, Victor, Total SUM in coincidence with AIDA Plastic 
|
Also TAC DTAS-Plastic AIDA |
Sun Jun 16 07:31:01 2019, Victor, The bump in the mutiplicities 8x
|
We have calculated the spectra of the module multiplicity for each side (careful! the maximum is 8, since it is
not the total multiplicity but the multiplicity of each side). |
Mon Jun 10 08:07:17 2019, Ale, Jose, Test of pulser count rates
|
File: 20190610_1602_background_AIDA_in.root
SyncPulse: 61777
|
Mon Jun 17 19:11:06 2019, Victor, TAC slowest module vs fastest module DTAS
|
Inside events of 320ns window |
Fri May 31 06:48:58 2019, A. Algora, T spectra for AIDA and F11 plastics
|
T spectra for AIDAPL and F11PLL and F11PLR are defined from now on. The ancilliary card was not saving the CFD values for all channels.
It was modified in order to have those spectra. A new configuration file will be saved. See in the attachment |
Wed May 15 04:03:23 2019, JAR, Synchronization with AIDA working
|
Synchronization with AIDA is working, see attached document to proceed. |
Sun Jun 16 22:02:45 2019, Victor, Sum in coincidence with F11  
|
TAC DTAS (slowest module of the sum) and F11
Total sum in coincidence with F11
|
Wed May 15 07:48:08 2019, Kiss Algora Nacher, Saturation and pulser
|
Saturation and pulser position after calibration
EDy1 16801 25000
|
Thu May 23 04:18:57 2019, A. Algora, SY2527 fixed
|
Following the instructions of CAEN, the SY2527 is working again. Some pushing, massaging appear to solve the situation |
Tue May 28 05:30:03 2019, A. Algora, Reference detector threshold changed
|
Again the Reference detector was firing mainly in the noise. Threshold increased to 5000 !!!!
New config file was saved, rate of Ref detector now: aprox. 400 kHz |
Mon May 27 12:54:48 2019, A. Algora, Reference detector
|
The reference detector was very noisy. The threshold has been changed from 2000 to 3200. We should look at this more carefully.
The position of the fiber was also changed. It moved too close to the Cs peak. Probably it has to do with the flexibility of the plastik screw
that fixes it. |
Thu May 30 02:03:33 2019, A. Algora, Ref detector position changed, AIDA plastik HV changed
|
The position of the reference detector in the ancilliary card was changed from channel 10 to channel 9. Remember that this is important
in relation to all files saved until now.
The HV of AIDA plastic was changed from 1300 V back to 1000 V. |
Wed May 29 09:49:25 2019, A. Algora, New configuration file
|
New configuration file after a few small changes in thresholds. |
Mon Jun 3 12:56:06 2019, A. Algora, New configuration file
|
The E range of the AIDA plastic for the construction of the TAIDA plastic spectrum has been changed. Also the range of the TAC_AIDAPL_CryA has been changed.
The estimated range of AIDA was based on the file 201906011624_background_AIDA_in.root, which was with no beam. But I am not sure is the calibration
of |
Mon May 20 06:55:15 2019, A. Algora, New config file after calibration
|
A file with Co and Cs was collected for 10 min
201905201444_137Cs60Co_align.root
after a minimal correction of the voltages (mentioned in the earlier entry). After this a new calibration was performed. |
Fri May 31 14:17:13 2019, A. Algora, New config file 
|
A new config file provided by Jose and Alvaro is loaded
cp 20190531_Conf_JLTAT.xlsx 20190531_Conf_2.xlsx
Comments: I have modified it afterwards to have TAIDAPL, the threshold in E was too high. The TAC_F11R_CryA was also modified. It is shifted |
Mon Jun 10 12:31:52 2019, Ale, Jose, New calibration
|
New calibration with more statistics
File: 20190610_2030_60Co_137Cs_AIDA_in.root
|