AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  GELINA  ELOG logo
Message ID: 3     Entry time: Fri Mar 6 17:10:43 2015
Author: TD 
Subject: RAL108 & RAL109 
N.B. The analog output of the RAL109 shaping amplifier module will not
drive a 50 Ohm load nor a 50Ohm coaxial cable > 20cm.

Always onnect the analog output of the RAL109 shaping amplifier module
to an oscilloscope input with 1MOhm/dc input coupling via an IDC ribbon
cable to Lemo-00 adaptor and a short coaxial cable (0.5ns, say).
Attachment 1: ral108.pdf  17 kB  | Hide | Hide all | Show all
ral108.pdf
Attachment 2: ral109.pdf  22 kB  | Hide | Hide all | Show all
ral109.pdf ral109.pdf
Attachment 3: ral108_ral109_noise.pdf  7 kB  | Hide | Hide all | Show all
ral108_ral109_noise.pdf
Attachment 4: ral109gain.txt  4 kB  | Show | Hide all | Show all
Attachment 5: technical_report.pdf  2.708 MB  | Hide | Hide all | Show all
technical_report.pdf technical_report.pdf technical_report.pdf technical_report.pdf technical_report.pdf technical_report.pdf technical_report.pdf technical_report.pdf
Attachment 6: user_guide.pdf  16 kB  | Hide | Hide all | Show all
user_guide.pdf user_guide.pdf user_guide.pdf user_guide.pdf