Message ID: 86     Entry time: Sat Jul 12 10:30:06 2025
Author: Francisco García Infantes 
Subject: Report Experimental SetUp Ca41 

I'm attaching a report for the experimental setup so far.

There is a main problem which haven't been solved yet and I don't know if that can cause problem in the future analysis.

Channels #23, #24, #31 and #32 are not working well. The signal os these channels look attenuated from the pulser, make impossible to see then in the Boron run.

As these channels are in the back side of the detector, it could be possible to run like? Unless we have more ideas about to solve the problem.

The front channels work well all of them, and the rest of the back channels work well as well.

I attach the current for the +15 and -15 volt applied to the PreAmp (low voltage) and the leakage current from the detector is 0.1 uA.

Attachment 1: Experimental_Setup_Ca41.pdf  2.302 MB  Uploaded Sat Jul 12 11:30:23 2025  | Hide | Hide all