Thu Oct 6 09:34:30 2022, Claudia, Test setup Photos and cabling    
|
Setup 5.10.22
dE 20 um (ID 2837-33) and 494 um double sided (ID 1194-9)
46 channels (2 connectors missing) |
Thu Oct 6 09:28:33 2022, Claudia, Items to be replaced for next run
|
Missing/broken items identified
- long rails for fixing chamber on support
- vacuum valve (ours is broken) |
Thu May 26 09:20:32 2022, Nick / Claudia, Inventory of Items at n_TOF for (n,cp) experiments 
|
Detectors currently in the boxes:
2x MSL type W1(SS)-50, 2902-4 (53um), 2940-6 (55um)
1x MSL type W1(SS)-20 2837-33 (20um) |
Fri Sep 9 10:01:18 2016, Sarah, Leakage currents 09.09.16
|
EFED 0.04uA
DEED 0.05uA
PHDI 0.13uA |
Fri Sep 2 12:03:40 2016, Sarah, SPD card changed
|
14 bit card replaced for DEED 9-12, to prevent issue with pulser arriving before gamma flash. |
Fri Sep 2 12:01:57 2016, Sarah, Leakage currents 02.09.16
|
E 0.03
dE 0.05
PHDI 0.12 |
Tue Aug 23 15:34:38 2016, Claudia, Leakage currents 23.08.16
|
E = 0.03uA
dE = 0.04uA
PHDI = 0.09uA |
Wed Aug 17 15:10:05 2016, Sarah, Leakage currents 12.08.16
|
E = 0.03uA
dE = 0.04uA
PHDI = 0.06uA |
Thu Aug 11 08:19:37 2016, Sarah, Leakage currents 10.8.16
|
At lunchtime (12.45)
E = 0.03 uA
dE = 0.04 uA |
Tue Aug 9 15:57:52 2016, Sarah, Leakage currents 09.08.16
|
Morning - 9am
E= 0.03uA
dE= 0.04uA |
Mon Aug 8 11:02:13 2016, Sarah, Leakage currents 08.08.16
|
E 0.03 uA
dE 0.04 uA
PHDI 0.05 uA |
Wed Aug 3 14:48:09 2016, Sarah, Leakage currents 02.08 - 03.08
|
02.08:
dE = 0.03 uA
E = 0.03 uA |
Wed Aug 3 09:48:31 2016, Sarah, EFED10 alpha calibration checks
|
|
Tue Aug 2 08:14:26 2016, Sarah, Leakage currents spreadsheet (26/07- 01/08)
|
Steady for dE, E. Possibly rising on PHDI- monitor closely. |
Sun Jul 31 15:56:32 2016, Sarah, DEED16 Alpha calibration checks (205169)
|
***note that DEED 16 was on old sampling rate for 205168- so use next run, also DEED 9-12 not plugged in***
|
Sun Jul 31 15:02:57 2016, Sarah, DEED6 Alpha calibration checks (205168)
|
DEED6: (1 bin = 2 ch)
Pulser (DAQ):
- Centroid = 5145.6(6) ch |
Sun Jul 31 14:32:34 2016, Sarah, DEED1 Alpha calibration checks (205168)
|
DEED1: (rebin x2 -> 1 bin = 4 ch)
Pulser (DAQ):
- Centroid = 5000(1) ch |
Sun Jul 31 13:43:56 2016, Sarah, Alpha spectra    
|
Pulser set to 100 pps.
Shaping amp:
Gain: 1x1.0x50 |
Thu Jul 28 16:00:43 2016, Sarah, List of to-dos
|
Process alpha data
- Compare pulser width and alpha width to calibration spectra (shaping amp with 1us and gain 50x)
- Does photodiode data process OK or is it a problem with the oscillations? |
Wed Jul 27 13:20:53 2016, Sarah, Gafchromic images 6x
|
|