Message ID: 776
Entry time: Mon Nov 12 08:59:43 2018
|
Author: |
LS, CA, TD |
Subject: |
Monday 12 November 6pm onwards |
|
|
17:59 DAQ continues OK
Stats OK - attachment 1
FEE temperatures OK - attachment 2
Bias and leakage currents OK - attachment 3
System-wide checks OK *except* nnaida7 fails ADC calibration
19:07 DAQ continues OK
Stats OK - attachment 4
FEE temperatures OK - attachment 5
Bias and leakage currents OK - attachment 6
System-wide checks OK *except* nnaida7 fails ADC calibration
20.19 DAQ starts
AIDA file RIBF148/R5_4
BRIKEN file 16
BigRIPS 1001
165Pm setting + 3mm Al degrader
20.21 all online & MIDASsort histograms zero'd
20.23 FEE64 good event statistics - see attachment 7
Merger - see attachment 8
TapeServer - see attachment 9
HEC x versus y spectra - see attachment 10
SN thinks DSSSD #3 are the ions of interest, #4 are the ions of interest at different charge state
and #5 & #6 are lighter ions which do not stop in DSSSD #1-6
21.21 DAQ stops
DAQ starts
AIDA file RIBF148/R5_14
BRIKEN file 18
BigRIPS 1003
21.26 HEC x vs y multiplicity spectra - attachment 11
21.29 HEC Ex vs Ey spectra - attachment 12
FEE64 rate spectra - attachment 13
21.31 1.8.L front/rear energy spectra (attachments 14/15) LEC
21.33 1.8.H front/rear energy spectra (attachments 16/17) HEC
21.46 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 good event statistics ok (attachment 18)
FEE64 temperatures ok (attachment 19)
Leakage currents/detector bias ok (attachment 20)
21.49 Merger and TapeServer ok, no error messages in terminal
DAQ still running (RIBF148/R5_18)
22.10 attachment 21 - pulser spectra in all channels fired at same time
- FEE64 in sync
22.12 SYNC100 at 9153.8 Hz
FEE64 discriminator at ~15000 Hz - higher than expected
~1 second deadtime in FEE19 - due to nnaida12 running fast
no timewarps observed in FEE64
all of above in attachment 22
22:32 DAQ continues OK
Stats OK - attachment 24
FEE temperatures OK - attachment 23
Bias and leakage currents OK - attachment 25
System-wide checks OK *except* nnaida7 fails ADC calibration
23:30 DAQ continues OK
Stats OK - attachment 26
FEE temperatures OK - attachment 27
Bias and leakage currents OK - attachment 28