Test of AIDA FEE64s
ISOL test station STFC DL, T9
Pulser BNC PB-5
Amplitude 1.0V
Attenuation x1
Polarity +
Frequency 100Hz
Tail pulse, decay time 1ms
Delay 250ns (min)
ASIC settings 2019May22-10.19.28
Slow comparator 0x80
Adaptor PCB - PJCS
FEE64 temperatures OK
System wide checks OK
Statistics OK
Hit/HitRate spectra OK
LEC spectra OK
Data observed for all channels of all ASIC #1-4
Pulser peak widths *.8.L
Wave spectra OK
Data observed for all channels of all ASIC #1-4
Date/time FEE64 MAC Pulser peak width Status
(channels FWHM)
6.7.23 17:50 d8:80:39:41:d8:2a 17,17,16,15 OK
6.7.23 11:20 d8:80:39:41:b4:0b 7,7,7,7 OK
6.7.23 10:03 d8:80:39:41:d0:0e 12,12,12,12 OK
5.7.23 15.31 d8:80:39:41:d7:cc 6,6,6,6 OK
5.7.23 14:04 d8:80:39:41:a0:71 12,11,11,12 OK Rate/Hitrate, *.*.L and *.*.W spectra - attachments 1-10
5.7.23 13:18 d8:80:39:42:0d:16 13,15,14,14 OK
5.7.23 12:43 d8:80:39:41:ee:71 17,17,15,16 OK
6.7.23 11:00 d8:80:39:42:0d:0b 17,15,16,15 https://elog.ph.ed.ac.uk/AIDA/870 Mezzanine replaced - ASIC #2 address bit #2 always on? ASIC #2 fail or mezzanine-FEE64
connection
issue?
7.7.23 09:46 d8:80:39:41:f6:ee https://elog.ph.ed.ac.uk/AIDA/869 HDMI replaced - not fully assembled - LEC fast comparator ASIC #1 channel #2 missing. ADC
does not
calibrate. HDMI issue?
OK FEE64s - attachment 11 |