Wed Apr 10 16:45:59 2024, TD, AIDA interlock
|
Honeywell HSS DPS sensor information https://elog.ph.ed.ac.uk/AIDA/932
|
Thu Mar 28 15:43:47 2024, TD, Unable to start MIDAS DAQ with 1x FEE64 ( aida04 )
|
At AIDA test station T9, STFC DL unable to start MIDAS DAQ with 1x FEE64 ( aida04 ) - see attachments
|
Fri Mar 29 08:39:18 2024, TD, PJCS, Unable to start MIDAS DAQ with 1x FEE64 ( aida04 )
|
>
> At AIDA test station T9, STFC DL unable to start MIDAS DAQ with 1x FEE64 ( aida04 ) - see attachments
>
|
Thu Mar 28 11:10:09 2024, TD, Test of 4x FEE64s at AIDA test station, T9, STFC DL 22x
|
Test of 4x FEE64s at AIDA test station, T9, STFC DL
startup.tcl changed from 'aida01 aida03' to 'aida01 aida02 aida03 aida04'
|
Wed Mar 13 14:50:04 2024, NH, [HowTo] Backup and Restore Raspberry Pi SD Cards
|
Micro SD cards used by the raspberry pis are cheap but prone to failing
If the system won't boot and the SD card is read only it may have failed. Best way is to buy a new micro SD card (16 or 32GB) and restore a backup
|
Tue Jan 30 14:18:16 2024, NH, [HowTo] Update MACB Firmware
|
https://elog.ph.ed.ac.uk/CARME/482 |
Thu Dec 14 10:53:27 2023, NH, MP, CC, GA, Snout Removed
|
Snout removed from S4
All detectors removed
|
Mon Dec 11 23:56:36 2023, TD, Tuesday 12 December 18x
|
00.55 Observe high frequency noise in aida02 & aida04 ( n+n FEE64s ) - attachments 1-5
and aida05 ( p+n FEE64 )
frequency c. 5MHz
|
Mon Dec 11 11:05:27 2023, TD, WR timestamp fail aida05 & aida07 11x
|
MACB config
1x level 0
|
Mon Dec 11 09:57:54 2023, TD, Monday 11 December 27x
|
Water flow and temperature ( gauges outside S4 entrance ) OK
BNC PB-5 pulser connected to p+n FEE64s by diasy chain of test + inputs
|
Sun Dec 10 13:06:27 2023, TD, Sunday 10 December 14x
|
14.05 S4 closed
Water flow and temperature ( gauges outside S4 entrance ) OK
|
Fri Dec 8 12:55:38 2023, TD, NH, FEE64 configuration
|
FEE64 configuration for 1x MSL type BB18(DS)-1000 24cm x 8cm DSSSD
|
Fri Dec 8 11:32:44 2023, NH, CC, VP, TD, Friday 8 December 18x
|
Change n+n FEE64 aida04 d8:80:39:41:f6:b7
asic #1 faulty
|
Thu Dec 7 17:30:36 2023, NH, CC, VP, TD, Thursday 7 December 8x
|
Stack re-assembled - upstream to downstream
|
Wed Dec 6 16:32:03 2023, NH, Replacement of faulty Honeywell HSS-DPS sensor
|
Honeywell HSS-DPS sensor is faulty - attachments 1 & 2
|
Mon Dec 4 10:34:48 2023, NH, AIDA Todo for 11.12.23
|
This is a short list of things todo or check before the test beam time on 11.12.23
Updated after 15:00 CET meeting
|
Fri Nov 24 10:36:06 2023, NH, Friday 24 November
|
Pulser LEMOs removed from p+n fees
Pulser connected to n+n fees: aida02 and aida04
BNC PB-5 Polarity --> NEGATIVE
|
Wed Nov 22 13:44:36 2023, NH, Report - Cannot change per-ASIC Settings
|
On ASIC settings page trynig to change the settings for one ASIC (e.g. ASIC1 of aida09) seems to cause all 4 ASIC settings to change
1. Ensure act on all FEEs and act on all ASICs unticked
|
Thu Nov 23 10:00:16 2023, NH, Report - Cannot change per-ASIC Settings
|
> On ASIC settings page trynig to change the settings for one ASIC (e.g. ASIC1 of aida09) seems to cause all 4 ASIC settings to change
>
> 1. Ensure act on all FEEs and act on all ASICs unticked
|
Wed Nov 22 15:12:40 2023, TD, /tmp/log8015_error
|
|
Wed Nov 22 14:52:43 2023, NH, AIDA Pulser Calibration
|
Connect pulser to all p+n FEEs
Turn on pulser, 100 Hz (attachment 3 for settings), rates OK (attachment 1)
|
Tue Nov 21 13:56:59 2023, NH, CC, HA, MP, AM, VP, Tuesday 21 November 8x
|
Connect the BB7 on top of the snout to the former aida05 (FEE above aida01)
Also remove copper busbar that is unused
|