Tue Nov 15 02:22:25 2016, TD, Tuesday 15 November 7x
|
11.20 DSSSD bias & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
11.58 per FEE64 1.8.W spectra - see attachments 4 & 5
14.25 Ambient temperature +24.7 deg C, d.p. +9.0 deg C, RH 36.7%
DSSSD bias & leakage currents OK - see attachments 6 & 7 |
Tue Mar 15 11:15:42 2016, TD, Tuesday 15 March    
|
MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=5.845uA, ambient temperature +19.8 deg C
DSSSD - AIDA adaptor PCB cabling
4 off LH Coupler (Kapton PCB, 5cm), 2x 34-way Samtec ribbon cable (45cm), RH coupler(Kapton PCB, 10cm)
+ 3M 1245 1.4mil copper foil screen ribbon cables + RH coupler only (i.e. not LH coupler)
+ drain wires -> gold-plated Lemo-00 test input connectors
nnaida 11 & 12 AIDA adaptor PCB rev B
nnaida 13 & 14 AIDA adaptor PCB rev C (LK1 fitted)
ground links LK3 & LK7 fitted to nnaida11-14 AIDA adaptor PCBs
Heavy duty copper cable connects copper front end frames of FEE modules
Nitto 5011N conductive gasket between FEE module and front end frames
Standard ASIC settings
nnaida 11 & 12 - negative input
nnaida 13 & 14 - positive input
Pulser BNC PB-5
Fall time 1.0ms
Rate 100Hz
Delay 250ns
Ampl 5.00000V
Polarity +
Pulse top Tail
Atten 10x
Clamp ON
- polarity via Cooknell SA1 Summing Amplifier
207Bi source, approx centred on DSSSD, approx 3cm from DSSSD
AIDA FEE firmware version 8
+ sync 500kHz ASIC clock
File R45
shaping time 8uS
slow comparator 10 (dec)
LEC/MEC fast comparator 10 (dec)
start: 12.08 stop: 10.24 16.3.16
spectra saved to disk 11.26 BST (sic)
Pulser OFF
nnaida 11 disc # 55 disabled
12 disc # 2 & 55 disabled
13 & 14 *no* discs disabled
16.3.16 12.10 Bias +200V, I_L +=6.900uA, ambient temperature +19.8 deg C
See attachments 1-5
File R48
shaping time 8uS
slow comparator 10 (dec)
LEC/MEC fast comparator 255 (dec)
start: 13.45 stop:
spectra saved to disk 1BST (sic)
Pulser walkthrough
Pulser BNC PB-5
Fall time 1.0ms
Rate 100Hz
Delay 250ns
Ampl 10.00000-1.00000V
Polarity +
Pulse top Tail
Atten 5x
Clamp ON
- polarity via Cooknell SA1 Summing Amplifier
file R52
shaping time 8uS
slow comparator 10 (dec)
LEC/MEC fast comparator 10 (dec)
start: stop: 15.46
spectra saved to disk 16.49 BST (sic)
Pulser walkthrough
Pulser BNC PB-5
Fall time 1.0ms
Rate 100Hz
Delay 250ns
Ampl 1.00000-0.10000V
Polarity +
Pulse top Tail
Atten 5x
Clamp ON
- polarity via Cooknell SA1 Summing Amplifier |
Tue May 14 02:19:20 2019, CA, OH, TD, Tuesday 14th May 2019 11x
|
10.20 System wide checks ok (except naida07 ADC calibration failure)
FEE temperatures ok (attachment 1)
good event statistics ok (attachment 2)
detector bias and leakage currents ok (attachment 3)
DAQ stopped
10.40 pulser switched off
slow comparator threshold raised to 0x64
DAQ, merger, TapeServer started - all ok (attachments 4 and 5)
writing to file R1 in TapeData/May19 at 70 kb/s
good event statistics (attachment 6)
16:11 Tapeserver and merger still ok
DAQ still running
System wide checks all ok *except nnaida7 fails ADC calibration*
Bias and leakage currents ok - attachment 7
FEE Temperatures ok - attachment 8
Statistics still ok
Spectrum are showing alpha events (Layout 4) - attachment 9
18:25 Tektronix clock accidentally unplugged. DAQ has stalled as a result.
Clock reset to 50MHz 5V TTL signal
18:29 System wide check of clock status shows error message in attachment 10
DAQ Reset
With slow comp at oxa statistics are still good - attachment 11
System wide checks all ok *except nnaida7 fails ADC calibration*
Resync to the ASIC clock
All histograms zeroed
18:37 Slow comp threshold raised to 0x64 for alpha run
Writing to file:
May19/R2_0
|
Tue Nov 14 16:52:54 2023, TD, Tuesday 14 November 
|
15.06 Quick test of AIDA 24cm x 8cm 'triple' DSSSD before dismounting to install 1x bPlast detector downstream of the DSSSD
Water flow, pressure and temperature OK
Power Up OK
All system wide checks OK
DSSSD bias -150V leakage current -5.29uA
ASIC settings 2021Apr29-13-16-00
FEE64s re-configured to aida01 - aida08 by NH
aida02, aida04 negative input, all other FEE64s positive input
Currently unable to specify settings directory for Options so aida06 & auida08 are still defined as negative input FEE64s - need to change ASIC settings manually
slow comparator p+n FEE64s 0x64->0xa, n+n FEE64s 0x64->0xf
per FEE64 Rate spectra - attachment 1
ADC data item stats OK - attachment 2
p+n FEE64 rates dominated by 'hot' channels, other channels < 1Hz
n+n FEE64 rates ~ 10Hz
Known issues with aida04 asic #1 and aida10 ( now aida06 asic # 4 ) remain to addressed/investigated
No chnages in performance cf. https://elog.ph.ed.ac.uk/AIDA/906 & https://elog.ph.ed.ac.uk/AIDA/907
|
Tue Jun 14 01:42:56 2016, TD, Tuesday 14 June 2016 7x
|
09.35 AIDA67 stalled AIDA67_43 (c. 01.27)
DAQ stopped and file closed 09.32
SYNC error counters - see attachments 1-2
All FEE64s pass all tests *except* SYNC counter error
09.37 AIDA68
ASICs checked
09.42 Detector bias & leakage currents, trigger rates - see attachments 3-5
10.03 AIDA68 stopped (loss of FEE64 synchronisation)
Power-cycle & reboot FEE64s
10.22 AIDA69 starts
All FEE64s pass all system-wide tests *except* nnaida1 fails clock status (0x6)
10.35 SyncCheck looks OK
10.42 AIDA69_4 FEE64 synchronisation OK
13.26 AIDA69 stalled AIDA69_63(c. 13.26)
15.22 AIDA70
ASICs checked
15.31 AIDA70 stopped (loss of FEE64 synchronisation)
See attachment 6
Power-cycle & reboot FEE64s
System-wide checks OK *except* FADC calibration nnaida6, 20, 22 & 23
nnaida1 clock check OK
SYNC errors baseline set
15.54 AIDA71 starts
16.04 AIDA71_1 FEE64 synchronisation OK
17.20 AIDA71_28 FEE64 synchronisation OK
SyncCheck OK
19.55 AIDA71 DAQ stalled AIDA71_64 c. 19.02
See attachment 7
Power-cycle & reboot FEE64s
20.14 AIDA72 starts
ASICs checked
SyncCheck OK
22.13 DAQ running AIDA72_33 |
Tue Nov 13 07:35:37 2018, CA, TD, OH, Tuesday 13 November 16:00 - 00:00 20x
|
16.36 DAQ continues ok
AIDA file RIBF148/R5_198
BRIKEN file 37
BigRIPS file 1022
16.38 system wide checks ok, except nnaida7 ADC calibration failure
16.40 FEE64 temperatures ok (attachment 1)
16.42 good event statistics ok (attachment 2)
16.43 detector bias/leakage currents ok (attachment 3)
16.44 Merger and TapeServer ok, no error messages in terminals (attachments 4 and 5)
16.51 BRIKEN stop
AIDA file RIBF148/R5_200
17.00 Al Degrader thickness changed from 2.3mm to 2.8mm
The 2.3mm plate was previously assumed to be 2.5mm, but this has been corrected by measuring individual Al
plates with digital micrometer.
17.02 BRIKEN start
BRIKEN file 38
BigRIPS file 1023
AIDA file RIBF148/R5_202
18.01 BRIKEN stop
AIDA file RIBF148/R5_211
18.03 BRIKEN start
AIDA file RIBF148/R5_212
BRIKEN file 39
BigRIPS file 1024
18.08 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 6)
good event statistics ok (attachment 8)
detector bias/leakage currents ok (attachment 7)
18.11 Merger and TapeServer ok, no error messages in terminal
19.00 BRIKEN stop
AIDA file RIBF148/R5_221
19.02 BRIKEN start
AIDA file RIBF148/R5_221
BRIKEN file 40
BigRIPS file 1025
19.57 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 9)
good event statistics ok (attachment 10)
detector bias/leakage current ok (attachment 11)
20.01 BRIKEN stop
AIDA file RIBF148/R5_231
20.03 BRIKEN start
AIDA file RIBF148/R5_231
BRIKEN file 41
BigRIPS file 1026
21.01 BRIKEN stop
AIDA file RIBF148/R5_240
21.03 BRIKEN start
AIDA file RIBF148/R5_240
BRIKEN file 42
BigRIPS file 1027
21:18 system wide checks ok - except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 12)
good event statistics ok (attachment 13)
detector bias/leakage current ok (attachment 14)
22:01 BRIKEN stop
AIDA file RIBF148/R5_250
22:03 BRIKEN start
AIDA file RIBF148/R5_250
BRIKEN file 43
BigRIPS file 1028
22:10 system wide checks ok - except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 15)
good event statistics ok (attachment 16)
detector bias/leakage current ok (attachment 17)
Merger and TapeServer ok, no error messages in terminal
23:01 BRIKEN stop
AIDA file RIBF148/R5_260
23:03 BRIKEN start
AIDA file RIBF148/R5_260
BRIKEN file 44
BigRIPS file 1029
23.05 system wide checks ok - except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 18)
good event statistics ok (attachment 20)
detector bias/leakage current ok (attachment 19)
Merger and TapeServer ok, no error messages in terminal
|
Mon Nov 12 23:03:53 2018, TD, Tuesday 13 November 08:00 - 16:00 23x
|
08.03 DAQ continues - AIDA file RIBF148/R5_117
165Pm BigRIPS setting + 3mm Al degrader
08.10 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event & AIDA discriminator statistics OK - see attachments 4 & 5
Merger OK - see attachment 6
TapeServer OK - see attachment 7
no errors or warnings shown by server terminal sessions
08.16 analysis of file RIBF148/R5_116 - see attachment 8
max deadtime nnaida19 ~ 1/420 ~ 0.25%, no timewarps
08.20 BRIKEN stop file
AIDA file RIBF148/R5_119
08.22 BRIKEN start file
AIDA file RIBF148/R5_120
BRIKEN file 28
BigRIPS file 1013
08.44 MIDASsort online spectra zero'd
08.56 1.8.W spectra 20ms FSR - see attachments 16 & 17
observe slow transient recovery per 33nF a.c. coupling but 'sawtooth' transients appear to be absent/rare
09.25 BRIKEN stop file
AIDA file RIBF148/R5_129
09.28 BRIKEN start file
AIDA file RIBF148/R5_130
BRIKEN file 29
BigRIPS file 1014
10.09 per FEE64 hit pattern (rate) - see attachment 9
per FEE64 1.8.L spectra - see attachments 10-13
nnaida1 pulser peak width 76 ch FWHM
per FEE64 1.8.H spectra - see attachments 14 & 15
10.25 BRIKEN stop file
AIDA file RIBF148/R5_139
10.27 Beam off - enter F11 to change BRIKEN amp gain
Correlation scaler pulser trigger frequency changed from 100Hz to 10Hz to minimise BigRIPS DAQ deadtime
Note - we will observe both edges of pulser trigger doubling the apparent frequency and the TDR data format
analyser reports each 16-bit word of the 48-bit correlation scaler so weill observe 6x frequency or 600Hz
(previously)
or 60Hz (now).
10.37 BRIKEN start file
AIDA file RIBF148/R5_141
BRIKEN file 30
BigRIPS file 1015
10.25 BRIKEN stop file
AIDA file RIBF148/R5_144
BRIKEN reloading DAQ configuration, will send correlation scaler reset
10.56 BRIKEN start file
AIDA file RIBF148/R5_145
BRIKEN file 31
BigRIPS file 1016
BRIKEN reload DAQ configuration again
11.00 BRIKEN stop file
AIDA file RIBF148/R5_145
11.06 BRIKEN start file
AIDA file RIBF148/R5_145
BRIKEN file 32
BigRIPS file 1017
11.50 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 18)
good event statistics ok (attachment 19)
detector bias/leakage currents ok (attachment 20)
12.05 BRIKEN stop file
AIDA file RIBF148/R5_155
12.07 BRIKEN start file
BRIKEN file 33
BigRIPS file 1018
AIDA file RIBF148/R5_155
13.06 BRIKEN stop file
AIDA file RIBF148/R5_165
13.08 BRIKEN start file
BRIKEN file 34
BigRIPS file 1019
AIDA file RIBF148/R5_165
13.35 BRIKEN stop file
AIDA file RIBF148/R5_169
BighRIPS DAQ reset required
13.38 BRIKEN start file
BRIKEN file 35
BigRIPS file 1020
AIDA file RIBF148/R5_170
14.39 BRIKEN stop file
AIDA file RIBF148/R5_179
14.40 BRIKEN start file
BRIKEN file 36
BigRIPS file 1021
AIDA file RIBF148/R5_179
15.38 BRIKEN stop file
AIDA file RIBF148/R5_189
15.49 System wide checks ok - except nnaida7 ADC calibration failure
FEE64 temperatures ok - attachment 21
good event statistics ok - attachment 22
detector bias/leakage currents ok - attachment 23
15.51 BRIKEN start file
BRIKEN file 37
BigRIPS file 1022
AIDA file RIBF148/R5_191
|
Mon Nov 12 14:51:44 2018, OH, Tuesday 13 November 00:00 - 08:00 12x
|
00:00 AIDA currently on file R5_38
BRIKEN on file 181112_2321_165Pm_019.dlt (AIDA file at start R5_33)
BigRIPS run 1004
00:23 BRIKEN stop file
AIDA Currently on file R5_43
00:26 BRIKEN start 181113_0025_165Pm_020.dlt
BigRIIPS start 1005
AIDA Currently on file R5_43
System wide checks ok
01:13 Bias and leakage currents ok - attchment 1
Stats ok - attachment 2
FEE Temps ok - attachment 3
01:21 BigRIPS and BRIKEN stop
AIDA on file R5_52
01:25 BigRIPS starts run 1006
BRIKEN starts run 181113_0123_165Pm_021.dlt
AIDA on file R5_53
02:21 BigRIPS and BRIKEN stop
AIDA on file R5_62
02:23 BigRIPS starts run 1007
BRIKEN starts run 181113_0222_165Pm_022.dlt
AIDA on file R5_62
ASIC load check ok
03:15 Beam dropped out unsure for how long but is now back
03:21 BigRIPS and BRIKEN stopped
AiDA on file R5_71
03:23 BigRIPS starts run 1008
BRIKEN starts run 181113_0320_165Pm_023.dlt
AIDA on file R5_71
03:45 Output of analyser on R5_70 - attachment 4
03:59 System wide checks ok
FEE Temps ok - attachment 5
Statistics ok - attachment 6
04:21 BigRIPS and BRIKEN stop
AIDA on file R5_81
04:23 BigRIPS starts run 1009
BRIKEN starts run 181113_0421_165Pm_024.dlt
AIdA on file R5_81
05:19 BigRIPS and BRIKEN stop
AIDA on file R5_90
05:20 BigRIPS start run 1010
BRIKEN starts run 181113_0520_165Pm_025.dlt
AIDA on file R5_90
05:49 System wide checks all ok
Stats ok - attachment 7
FEE Temps ok - attchment 8
Bias and leakage currents ok - attachment 9
High energy hit pattern per dssd - attachment 10
Low energy Ex vs Ey plot per dssd - attachment 11
High energy Ex vs Ey plot per dssd - attachment 12
Also using the previously merged files produced from Phong's sort have confirmed that there is no to
little punch through on DSSD3
Did this through the use of a 2d plot of energy DSSD 3 vs the velocity
06:21 BigRIPS and BRIKEN stop
AIDA on file R5_100
06:23 BigRIPS starts run 1011
BRIKEN starts run 181113_0622_165Pm_026.dlt
AIDA on file R5_100
07:19 BigRIPS and BRIKEN stop
AIDA on file R5_110
07:20 BigRIPS starts run 1012
BRIKEN starts run
AIDA on file R5_110 |
Tue Jun 13 00:11:56 2017, TD,ZLIU, Tuesday 13 June 08:00-16:00 17x
|
08.11 DAQ continues
147Sm BigRIPS setting - 0mm Al (at F7) + 0mm variable + 0mm Al fixed (at F11) = 0mm Al
BNC PB-4
Amplitude 90,000
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency 25Hz range 2.6Hz
delay MIN, tail pulse, INT ref
ASIC settings 2017Mar27-16.12.39
08.17 online analysis
per DSSSD x versus y (HEC) - see attachment 1
per DSSSD E_P versus E_n (HEC) - see attachment 2
ADC & disc data synchronised - see attachment 3
09.00 DAQ stop
AIDA file RIBF148/R3_396
*** END OF BRIKEN MAY/JUNE 2017 CAMPAIGN ***
Detector biases & leakage currents OK - see attachments 4 & 5
FEE64 Temperatures OK - see attachment 6
good events statistics stable & OK - see attachment 7
No warnings/errors reported by merger since start of RIBF148/R3 - see attachment 8
*all* system wide checks OK *except* nnaida13 fails ADC calibration
09.07 analysis RIBF148/R3_398 - deadtime <0.5%, zero MBS & ts timestamps - see attachment 9
*all* online and SortSas spectra zero'd
09.15 DAQ start
AIDA file RIBF148/R3_400
BRIKEN file 48
Background run - AIDA + BRIKEN + Clovers
11:28 252Cf source placed at F11
BRIKEN DAQ stopped bgd_048
AIDA on R3_434
11.31 DAQ stop
AIDA file RIBF148/R3_435
11.34 Detector biases & leakage currents OK - see attachments 10 & 11
Ambient temperature +24.6 deg C, RH 35.6%, d.p. +8.4 deg C
11.41 DAQ start
AIDA file RIBF148/R4
Pulser walkthrough
BNC PB-4
Amplitude 90,000 - 10,000 @ 10,000 step
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency 250Hz range
delay MIN, tail pulse, INT ref
see attachment 12
11.52 DAQ stop
AIDA file RIBF148/R4_2
11.55 DAQ start
AIDA file RIBF148/R5
Pulser walkthrough
BNC PB-4
Amplitude 12,000 - 2,000 @ 2,000 step
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency 250Hz range
delay MIN, tail pulse, INT ref
see attachment 13
12.02 DAQ stop
AIDA file RIBF148/R5_2
13.54 DAQ start
AIDA file RIBF148/R6
Background - pulser off
14.05 Detector biases & leakage currents OK - see attachments 12 & 13
14.11 DAQ stop
AIDA file RIBF148/R6_4
14.20 AIDA retracted from BRIKEN
14.21 Detector biases & leakage currents OK - see attachments 14 & 15
check ASIC control x2
14.50 DAQ start
AIDA file RIBF148/R8
alpha background
BNC PB-4
Amplitude 90,000 step
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency 25Hz range 2.6Hz
delay MIN, tail pulse, INT ref
ASIC settings 2017Mar27-16.12.39
*except*
slow comparator 0x64
LEC/MEC fast comparator 0x14
data rate c. 500kb/s c. 65k data items/s
|
Mon Jun 12 15:51:15 2017, OH, Tuesday 13 June 00:00-08:00 14x
|
00:02 System wide checks ok
Good events - attachment 1
FEE temperatures - attachment 2
Bias and leakages ok - attachment 3
00:31 BRIKEN and BigRIPS stop their DAQs
AIDA on file R3_236
00:32 BRIKEN starts run 40
BigRIPS starts run 2512
AIDA on file R3_236
01:30 BRIKEN and BigRIPS stop their DAQs
AIDA on file R3_254
01:31 BRIKEN starts run 41
BigRIPS starts run 2513
AIDA on file R3_255
01:41 DSSD rates from R3_257
*** DSSSD # 1 count: 119356 old count: 62524 dt: 2.02 s LEC rate: 59170.96 Hz
*** DSSSD # 2 count: 168540 old count: 87963 dt: 2.02 s LEC rate: 83554.01 Hz
*** DSSSD # 3 count: 61578 old count: 32845 dt: 2.02 s LEC rate: 30527.41 Hz
*** DSSSD # 4 count: 49914 old count: 25818 dt: 2.02 s LEC rate: 24744.96 Hz
*** DSSSD # 5 count: 25977 old count: 13821 dt: 2.02 s LEC rate: 12878.15 Hz
*** DSSSD # 6 count: 43384 old count: 22072 dt: 2.02 s LEC rate: 21507.70 Hz
*** DSSSD # 1 count: 0 old count: 0 dt: 2.02 s HEC rate: 0.00 Hz
*** DSSSD # 2 count: 6 old count: 3 dt: 2.02 s HEC rate: 2.97 Hz
*** DSSSD # 3 count: 1 old count: 1 dt: 2.02 s HEC rate: 0.50 Hz
*** DSSSD # 4 count: 1 old count: 1 dt: 2.02 s HEC rate: 0.50 Hz
*** DSSSD # 5 count: 14 old count: 10 dt: 2.02 s HEC rate: 6.94 Hz
*** DSSSD # 6 count: 211 old count: 118 dt: 2.02 s HEC rate: 104.60 Hz
01:43 Analysis of R3_257 - attachment 4
No timewarps
02:10 System wide checks ok
Leakage and bias ok - attachment 5
Good events - attachment 6
FEE temperatures - attachment 7
02:30 BRIKEN and BigRIPS stop their DAQs
AIDA on file R3_274
02:32 BRIKEN starts run 42
BigRIPS starts run 2514
AIDA on file R3_274
03:30 BRIKEN and BigRIPS stop their DAQs
AIDA on file R3_292
03:32 BRIKEN starts run 43
BigRIPS starts run 2514
AIDA on file R3_293
04:28 BRIKEN and BigRIPS stop their DAQs
AIDA on file R3_311
04:29 BRIKEN starts run 44
BigRIPS starts run 2515
AIDA on file R3_311
04:54 System wide checks ok
Bias and leakage ok - attachment 8
Good Events - attachment 9
FEE temperatures - attachment 10
05:00 DSSD rates from R3_319
*** DSSSD # 1 count: 255278 old count: 193923 dt: 1.01 s LEC rate: 60935.40 Hz
*** DSSSD # 2 count: 360584 old count: 272321 dt: 1.01 s LEC rate: 87659.38 Hz
*** DSSSD # 3 count: 133027 old count: 100845 dt: 1.01 s LEC rate: 31961.91 Hz
*** DSSSD # 4 count: 108047 old count: 82416 dt: 1.01 s LEC rate: 25455.71 Hz
*** DSSSD # 5 count: 51776 old count: 38962 dt: 1.01 s LEC rate: 12726.37 Hz
*** DSSSD # 6 count: 88202 old count: 66648 dt: 1.01 s LEC rate: 21406.59 Hz
*** DSSSD # 1 count: 1 old count: 1 dt: 1.01 s HEC rate: 0.00 Hz
*** DSSSD # 2 count: 11 old count: 8 dt: 1.01 s HEC rate: 2.98 Hz
*** DSSSD # 3 count: 2 old count: 2 dt: 1.01 s HEC rate: 0.00 Hz
*** DSSSD # 4 count: 3 old count: 2 dt: 1.01 s HEC rate: 0.99 Hz
*** DSSSD # 5 count: 32 old count: 25 dt: 1.01 s HEC rate: 6.95 Hz
*** DSSSD # 6 count: 439 old count: 335 dt: 1.01 s HEC rate: 103.29 Hz
05:30 BRIKEN and BigRIPS stop DAQs
AIDA on file R3_330
05:31 BigRIPS starts run 2516
BRIKEN starts run 45
AIDA on file R3_331
06:31 BigRIPS and BRIKEN stop DAQs
AIDA on file R3_350
06:32 BRIKEN starts run 46
BigRIPS starts run 2517
AIDA on file R3_350
06:50 System wide checks ok
Bias and leakage currents ok - attachment 11
Good events - attachment 12
FEE temperatures - attachment 13
07:09 Merger still running ok - attachment 14
07:31 BigRIPS and BRIKEN stop their DAQs
AIDA on file R3_369
07:32 BRIKEN starts run 47
BigRIPS starts run 2518
AIDA on file R3_369 |
Mon Dec 11 23:56:36 2023, TD, Tuesday 12 December 18x
|
00.55 Observe high frequency noise in aida02 & aida04 ( n+n FEE64s ) - attachments 1-5
and aida05 ( p+n FEE64 )
frequency c. 5MHz
Oscillation unchanged by
HV cycle
ASIC control load
FEE64 power cycle
BNC PB-5 pulser OFF
BNC PB-5 pulser output disconnected from test + cable ( and remains disconnected )
BB7 HV and preamp power OFF - oscillation disappears
Note - BB7 HV and power have been on all day and AIDA was running OK until at least 22.18 11.12.23 - go figure!
01.42 file DEC23/R7
02.00 checks
DSSSD bias -150V leakage current -5.980uA - attachment 5
FEE64 teperatures OK - attachment 6
ADC data item stats OK - attachment 7
Merger & TapeServer screens - attcahments 8 & 9
per FGEE64 Rate spectra - attachment 10
1.8.W spectra - 20us FSR - attachments 11 & 12
02.18 ASIC check load x2
02.20 file DEC23/R8
Final FRS setting - fragments SE 238U, PID uncertain online
02.22 zero all spectra & stats
03.23 degrader changed to 4350mg
file DEC23/R9
03.25 zero all spectra & stats
06.00 beam stopped
file R9 closed
06.18 pulser walkthrough test -
BNC PB-5
amplitude 9.0-1.0 @ 1.0V step
attenuation x10
polarity -
tail pulse
1ms decay time
file DEC23/R11
|
Tue Oct 11 03:34:02 2016, CG & TD, Tuesday 11 October 16x
|
10.40 DAQ found stopped
All FEE64 AIDA SYNC rates zero, AIDA SYNC count ~ 13.8M => DAQ stopped @ 06.35
System wide checks - see attachments 1-4 - all other system-wide checks OK
11.15 Replace cable for Master MACB external clock signal
- change from c. 4m RG174 (Lemo-00 to Lemo-00) to c. 4m RG58 (BNC to BNC)
Power cycle AIDA FEE64s
Set TS_SYNC_PHASE 0
DSSSD bias/leakage current OK - see attacment 5-6
FEE64 temperatures OK - see attachment 7
*All* system-wide checks OK (N.B. several ReSYNCs required)
B2F/F11 temperature + deg C, d.p. + deg C,RH %
ASIC settings 2016Sep02-18.17.01
slow comparator 0x10 -> 0x0a
Master MACB serial #13 (see https://elog.ph.ed.ac.uk/AIDA/367)
Master MACB mode 5
11.58 DAQ starts
15.41 DAQ OK
15.44 DAQ stop requested
15.46 DAQ start
Data transfer to disk enabled to test DAQ correlation real-time monitor
22.50 AIDA SYNCs OK
# Master SYNCs *not equal* to # FEE64 SYNCs - all FEE64 SYNCs # equal
- all other system-wide checks OK - see attachments 8-15
Merger and TapeServer OK
R17_197 ADC data synchonised
aidas1> ./analyser2 /TapeData/Sept2016/R17_197
*** GREAT format 3.2.0 analyser - TD - May 2014
*** ERROR: READ I/O error: 5002
blocks: 32000
ADC data format: 165074646 ( 1182230.4 Hz)
Other data format: 96013354 ( 687627.8 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 113 ( 0.8 Hz)
RESUME: 113 ( 0.8 Hz)
SYNC100: 1275326 ( 9133.6 Hz)
FEE64 disc: 94733738 ( 678463.4 Hz)
MBS info: 4064 ( 29.1 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 191 ( 1.4 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
Timestamp elapsed time: 139.630 s
FEE module #: 1 elapsed dead time 0.000 s
FEE module #: 2 elapsed dead time 0.206 s
FEE module #: 3 elapsed dead time 0.000 s
FEE module #: 4 elapsed dead time 0.000 s
FEE module #: 5 elapsed dead time 1.084 s
FEE module #: 6 elapsed dead time 4.015 s
FEE module #: 7 elapsed dead time 0.000 s
FEE module #: 8 elapsed dead time 0.000 s
FEE module #: 9 elapsed dead time 0.201 s
FEE module #: 10 elapsed dead time 0.000 s
FEE module #: 11 elapsed dead time 0.000 s
FEE module #: 12 elapsed dead time 0.000 s
FEE module #: 13 elapsed dead time 0.003 s
FEE module #: 14 elapsed dead time 0.193 s
FEE module #: 15 elapsed dead time 0.024 s
FEE module #: 16 elapsed dead time 0.000 s
FEE module #: 17 elapsed dead time 1.198 s
FEE module #: 18 elapsed dead time 0.040 s
FEE module #: 19 elapsed dead time 0.124 s
FEE module #: 20 elapsed dead time 0.000 s
FEE module #: 21 elapsed dead time 0.000 s
FEE module #: 22 elapsed dead time 0.187 s
FEE module #: 23 elapsed dead time 0.258 s
FEE module #: 24 elapsed dead time 0.000 s
FEE module #: 25 elapsed dead time 0.000 s
FEE module #: 26 elapsed dead time 0.000 s
FEE module #: 27 elapsed dead time 0.000 s
FEE module #: 28 elapsed dead time 0.000 s
FEE module #: 29 elapsed dead time 0.000 s
FEE module #: 30 elapsed dead time 0.000 s
FEE module #: 31 elapsed dead time 0.000 s
FEE module #: 32 elapsed dead time 0.000 s
*** Program elapsed time: 19.555s ( 1636.436 blocks/s, 102.277 Mb/s) |
Tue Jun 11 01:45:55 2019, CA, CB, TD, CG, OH, DK, Tuesday 11 June 17x
|
09:46 DAQ continues ok
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachment 1
FEE64 temperatures OK - see attachment 2
good event statistics ok - attachment 3
10:00 Merger & TapeServer ok - attachments 4 & 5
15.13 DAQ continues OK
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range, variable min. c. 2.5Hz
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 6 & 7
FEE64 temperatures OK - see attachment 8
good event statistics - DSSSD#1 noisy - see attachments 9
7x <20k 15x < 50k 20 < 100k cf. 8x <20k, 23x <50k, 24x <100k earlier today
Check ASIC Control x2
good event statistics - DSSSD#1 noisy - see attachments 10
8x <20k 5x < 50k 20 < 100k cf. 8x <20k, 23x <50k, 24x <100k earlier today
A, B. C & D Samtec connectors pressed into adaptor PCBs
15.41 good event statistics - OK - see attachments 11
8x <20k 19x < 50k 20x < 100k
16.53 good event statistics - OK - see attachments 11
8x <20k 19x < 50k 20x < 100k
17.04 p+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 14
n+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 15
1.8.L spectra - see attachments 12 & 13
nnaida1 75 ch FWHM, nnaida3 79 ch FWHM - cf. 76 & 83 ch FWHM respectively yesterday
17.05 all spectra zero'd
DAQ stats zero'd
17.07 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 80% M-H markers
Temperature +23.4 deg C dew point +9.6 deg C RH 42.0%
17.10 leakage currents/detector bias ok - attachment 17 |
Tue Oct 10 08:38:37 2023, TD, Tuesday 10 October 40x
|
09.38 Water flow & temp OK
Power ON
All system wide checks OK
DSSSD HV +120V +5.585uA - attachment 1
CAEN N1419ET grounded output ( internal LK fitted ) plus ThorLabs inline filter
standard coaxial cable to aida04, no HV daisy chain aida12, aida03 & aida11
aida03 LK3 fitted
aida11, aida03, aida12 LK1 fitted
test + & test - daisy chains removed
FEE64 temps OK - attachment 2
slow comparator 0x64
ADC data item stats - attachment 3
per FEE64 Rate spectra - attachments 4
1.8.W spectra - 200us FSR - attachments 5-8
test + to aida01 *only*
ADC data item stats - attachment 9
per FEE64 Rate spectra - attachment 10
aida01 1*L spectra - attachment 11
pulser peak width 1.8.L 141 ch FWHM
ThorLabs inline filter removed
ADC data item stats - attachment 12
per FEE64 Rate spectra - attachment 13
pulser peak width 1.8.L 136 ch FWHM
14.15 DSSSD HV +120V +6.610uA
Disconnect aida01 ribbon cable drain wire from adaptor PCB gnd
aida01 1.8.L pulser peak width 188 ch FWHM - amplitude/structure of noise does not change significantly but observe higher frequency components in addition
1.8.W p+n spectra - 20us FSR - attachment 14
15.20 Switch from HV + to HV - N1419ET ch 3 to ch 2
HV coaxial cable chain aida12, aida03, aida11, aida04 LK1 fitted
aida12, aida03 & aida11 LK1 removed
Connect aida01 ribbon cable drain wire to adaptor PCB gnd
DSSSD HV-120V -6.720uA - attachment 15
ADC data item stats - attachment 16
aida01 1.8.L pulser peak width 138 ch FWHM
1.8.W spectra - 20us FSR - attachments 17 & 18
Little observable difference between HV - and +, as expected
16.20 Jumper aida04 adaptor PCB gnd to aida12 adaptor PCB gnd
slow comparator 0x64 -> 0xf
ADC data item stats - attachment 20
per FEE64 Rate spectra - attachment 21 & 24
1.8.W spectra - 20us FSR - attachments 22 & 23
aida01 1.8.L pulser peak width 99 ch FWHM
17.00 Ground cable from aida04 gnd to aida12 gnd
ADC data item stats - attachment 25
per FEE64 Rate spectra - attachment 26 & 27
1.8.W spectra - 20us FSR - attachments 28 & 29
aida01 1.8.L pulser peak width 97 ch FWHM
17.20 Ground cable from aida04 gnd to aida12 gnd to aida02 gnd
slow comparator 0xf ~ 150keV
ADC data item stats - attachment 30
per FEE64 Rate spectra - attachment 31
1.8.W spectra - 20us FSR - attachments 32 & 33
1.8.W spectra - 200us FSR - attachments 32 & 33
aida01 1.8.L pulser peak width 85 ch FWHM ~ 65keV FWHM => 5s threshold 138keV
aida12 1.8.L pulser peak width 58 ch FWHM ~ 44keV FWHM => 5s threshold 93keV
aida02 1.8.L pulser peak width 133 ch FWHM ~ 100keV FWHM => 5s threshold 212keV
Rate spectra are consistent with these estimates, i.e. observe ~zero rate for p+n channels ( except hot channels ) and low ( ~100-200Hz rate per channel ) for n+n channels
p+n noise aida12, aida03, aida11 excellent
for aida10, aida01 & aida09 good - can probably be improved to match aida12 etc
significant improvement observed for n+n side
aida12 *.*.W spectra - attachments 34-37
per p+n FEE64 1.8.W spectra - 200us FSR - expanded scale - attachment 38 |
Tue Nov 1 02:24:05 2016, TD, Tuesday 1 November 8x
|
11.24 DAQ found stalled
Most recent file R95_168 last accessed 06.31
R95_168 ADC & disc data synchronised - see attachment 8
spectrum s1000 - per ASIC data items as function of timestamp (20us/channel)
aidas1> ./analyser /TapeData/Sept2016/R95_168
*** TDR format 3.3.0 analyser - TD - October 2016
*** ERROR: READ I/O error: 5002
blocks: 14568
ADC data format: 108169612 ( 1126585.0 Hz)
Other data format: 10690700 ( 111343.5 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 31 ( 0.3 Hz)
RESUME: 30 ( 0.3 Hz)
SYNC100: 877566 ( 9139.8 Hz)
FEE64 disc: 9810241 ( 102173.5 Hz)
MBS info: 2832 ( 29.5 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 2160 ( 22.5 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
*** Timestamp elapsed time: 96.015 s
FEE elapsed dead time(s) elapsed idle time(s)
1 0.000 0.000
2 0.000 0.000
3 0.000 0.000
4 0.000 0.000
5 1.994 2.016
6 1.408 1.439
7 0.000 0.000
8 0.000 0.000
9 0.000 0.000
10 0.000 0.000
11 0.000 0.000
12 0.000 0.000
13 0.000 0.000
14 0.000 0.000
15 0.015 0.000
16 0.000 0.000
17 0.000 0.000
18 0.000 0.000
19 0.045 0.000
20 0.000 0.000
21 0.000 0.000
22 0.000 0.000
23 0.099 0.000
24 0.000 0.000
25 0.000 0.000
26 0.000 0.000
27 0.000 0.000
28 0.000 0.000
29 0.000 0.000
30 0.000 0.000
31 0.000 0.000
32 0.000 0.000
*** Statistics
FEE ADC Data Other Data Sample Undefined Pause Resume SYNC100 Disc MBS
Other HEC Data
0 0 0 0 0 0 0 0 0 0
0 0
1 350112 43200 0 0 0 0 36623 6577 0
0 48
2 1247240 43778 0 0 0 0 36622 7156 0
0 0
3 2547061 39987 0 0 0 0 36626 3361 0
0 0
4 881007 40879 0 0 0 0 36623 4256 0
0 5
5 23965264 2536358 0 0 13 12 35831 2500502 0
0 38
6 23702370 1860427 0 0 11 11 36086 1821487 2832
0 844
7 1895595 40743 0 0 0 0 36623 4120 0
0 1
8 1603094 41634 0 0 0 0 36626 5008 0
0 325
9 3180182 1849501 0 0 0 0 36625 1812876 0
0 0
10 2576178 97708 0 0 0 0 36622 61086 0
0 0
11 447806 39812 0 0 0 0 36625 3187 0
0 0
12 905803 39001 0 0 0 0 36618 2383 0
0 0
13 2226595 147679 0 0 0 0 36624 111055 0
0 0
14 3021228 2593022 0 0 0 0 36626 2556396 0
0 513
15 4146089 40401 0 0 2 2 36617 3780 0
0 0
16 4264696 39555 0 0 0 0 36619 2936 0
0 0
17 2934508 416504 0 0 0 0 36623 379881 0
0 0
18 3319005 56583 0 0 0 0 36619 19964 0
0 0
19 6346836 41482 0 0 2 2 36604 4874 0
0 1
20 2392653 39249 0 0 0 0 36623 2626 0
0 283
21 2056421 41512 0 0 0 0 36622 4890 0
0 96
22 2824822 522339 0 0 0 0 36624 485715 0
0 5
23 7402741 39352 0 0 3 3 36590 2756 0
0 1
24 3932306 39994 0 0 0 0 36625 3369 0
0 0
25 0 0 0 0 0 0 0 0 0
0 0
26 0 0 0 0 0 0 0 0 0
0 0
27 0 0 0 0 0 0 0 0 0
0 0
28 0 0 0 0 0 0 0 0 0
0 0
29 0 0 0 0 0 0 0 0 0
0 0
30 0 0 0 0 0 0 0 0 0
0 0
31 0 0 0 0 0 0 0 0 0
0 0
32 0 0 0 0 0 0 0 0 0
0 0
*** Timewarps
FEE ADC Pause Resume SYNC100 Disc MBS Undefined Samples
0 0 0 0 0 0 0 0 0
1 0 0 0 0 0 0 0 0
2 0 0 0 0 0 0 0 0
3 0 0 0 0 0 0 0 0
4 0 0 0 0 0 0 0 0
5 0 0 0 0 0 0 0 0
6 0 0 0 0 0 0 0 0
7 0 0 0 0 0 0 0 0
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
10 0 0 0 0 0 0 0 0
11 0 0 0 0 0 0 0 0
12 0 0 0 0 0 0 0 0
13 0 0 0 0 0 0 0 0
14 0 0 0 0 0 0 0 0
15 0 0 0 0 0 0 0 0
16 0 0 0 0 0 0 0 0
17 0 0 0 0 0 0 0 0
18 0 0 0 0 0 0 0 0
19 0 0 0 0 0 0 0 0
20 0 0 0 0 0 0 0 0
21 0 0 0 0 0 0 0 0
22 0 0 0 0 0 0 0 0
23 0 0 0 0 0 0 0 0
24 0 0 0 0 0 0 0 0
25 0 0 0 0 0 0 0 0
26 0 0 0 0 0 0 0 0
27 0 0 0 0 0 0 0 0
28 0 0 0 0 0 0 0 0
29 0 0 0 0 0 0 0 0
30 0 0 0 0 0 0 0 0
31 0 0 0 0 0 0 0 0
32 0 0 0 0 0 0 0 0
*** Program elapsed time: 2.449s ( 5948.019 blocks/s, 371.751 Mb/s)
DSSSD bias & leakage current OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
System wide checks OK *except* SYNC pulses & errors - see attachments 4-7
Restart merger
DAQ STOP
System wide checks
DAQ RESET & SETUP
System-wide checks
Disable H & data transfer #1
Restart merger & tapeserver
Enable H & data transfer #1
11.46 Start DAQ (R96)
16.28 Stop DAQ OK
16.30 Start DAQ (R97)
BNC PB-4 pulser OFF (background run)
17.04 Stop DAQ
17.08 Ambient temperature +25.0 deg C, d.p. +8.4 deg C, RH 34.8%
17.11 Start DAQ (R98)
BNC PB-4 pulser ON
60Co cource (10-7023) placed on top of AIDA snout approximately above DSSSD #3
17.30 DAQ stalled due to excessive diak activity from data copying & analysis tasks
Power cycle required
18.09 DAQ start (R99)
18.38 DAQ stalled
Most recent file R99_7 last accessed 18.38
Power cycle rquired
19.42 DAQ start (R101)
BNC PB-4 pulser ON
60Co cource (10-7023) placed on top of AIDA snout approximately above DSSSD #3
(Phong's correction: 60Co source position was in front of AIDA snout toward upstream)
ASIC settings 2016Nov01-18.02.17 |
Tue Aug 1 08:50:36 2023, TD, Tuesday 1 August 30x
|
09.50 DSSSD bias +100V leakage current +9.255uA - attachment 2
FEE64 temperatures OK - attachment 1
All system wide checks OK
per FEE64 1.8.W spectra 20 & 200us FSR - attachments 3-6
per p+n FEE64 1.8.L spectra - attachments 7-8
aida01 pulser peak width 87 ch FWHM
adc data item stats - slow comparator 0x64, 0x14, 0xa - attachments 15, 14, 11
per FEE64 rate spectra - attachments 9-10, 12-13
from https://elog.ph.ed.ac.uk/AIDA/872
13x FEE64s installed - see attachment 2 ( courtesy NH )
FEE64 # Cable set #
1 1
2 2
3 3
4 4
5
6 6
7
8 8
9 9
10 5
11 11
12 7
15
Change AIDA PSU cabling from
#1 1-3 2-4 5-6 7-8
#2 9-10 11-12 13-14 15-16
to
#1 1-3 2-4 5-7 9-11
#2 6-8 10-12 13-14 15-16
DSSSD bias +100V leakage current +11.765uA
per FEE64 1.8.W spectra 200us FSR - attachments 16-17
per p+n FEE64 1.8.L spectra - attachments 18-19
aida01 pulser peak width 158 ch FWHM ( note double peaking )
adc data item stats - slow comparator 0x64, 0x14 - attachments 23, 21
per FEE64 rate spectra - attachments 20, 22
DSSSD bias +30V leakage current +35.8uA
13.35 DSSSD bias +100V leakage current +11.785uA
aida04 34-way cable #1 connected to ASIC #1 & 2 disconnected from adaptor PCB
per FEE64 1.8.W spectra 2000us FSR - attachments 26-27
adc data item stats - slow comparator 0x64, 0x14 - attachments 29-30
per FEE64 rate spectra - attachment 28
16.20 Install LK2 & LK4 aida12, aida03, aida11 ( front & rear field plate ground )
DSSSD over current ( 200uA @ c. 30V ) - high resistance short?
LK2 & LK4 installed aida12 *only* - OvC
LK2 & LK4 installed aida11 *only* - OvC
LK2 & LK4 installed aida3 *only* - OvC
LK2 installed aida3 *only* - OvC ( LK2 = rear field plate )
All front & rear field plate connections appear to provide a high resistance short to ground for detector bias. This may be related to issues observed biasing DSSSD via aida01 & aida09 - https://elog.ph.ed.ac.uk/AIDA/874
But why does it affect aida12 too? |
Tue Jun 6 15:26:44 2017, TD, ZL, Tuesday 06 June 16.00-24.00 29x
|
15.53 Bias and leakage currents ok - attachments 1 & 2
FEE temperatures ok - attachment 3
statistics good events still seeing noise on 13 and 14 - attachment 4
No merger errors since GO - attachment 5
16:00 all system wide checks OK
15:59 BigRIPS and BRIKEN stop daqs
AIDA on file R6_147
16:02 BigRIPS starts run 3023
BRIKEN starts run 23
AIDA on file R6_148
implant hit pattern statistics until the end of R6_147: attachment6 spectra cleared
17:00 BigRIPS and BRIKEN stop daqs
AIDA on file R6_165
17:02 BigRIPS starts run 3024
BRIKEN starts run 24
AIDA on file R6_166
18:03 BigRIPS and BRIKEN stop daqs
AIDA on file R6_183
18.04 BigRIPS starts run 3025
BRIKEN starts run 25
AIDA on file R6_184
18.18 Detector biases & leakage currents OK - see attachments 8 & 9
FEE64 Temperatures OK - see attachment 10
good events statistics OK - high but relatively stable - see attachment 11
No warnings/errors reported by merger since DAQ GO - see attachment 12
*all* system wide checks OK
18.23 analysis of file RIBF128/R6_188 - dead time < 0.5%, 0x ts timewarps, 1x MBS timewarps - see attachment 13
online analysis per DSSSD HEC x versus y spectra - see attachment 14
18.28 offline analysis R6_196 - ADC & disc data synchronised - see attachment 14
rates ***
DSSSD # 1 count: 579778 old count: 507154 dt: 1.02 s LEC rate: 71359.29 Hz ***
DSSSD # 2 count: 824589 old count: 719407 dt: 1.02 s LEC rate: 103350.31 Hz
DSSSD # 3 count: 344253 old count: 299859 dt: 1.02 s LEC rate: 43620.90 Hz ***
DSSSD # 4 count: 241567 old count: 210328 dt: 1.02 s LEC rate: 30694.99 Hz ***
DSSSD # 5 count: 70435 old count: 61399 dt: 1.02 s LEC rate: 8878.64 Hz ***
DSSSD # 6 count: 154453 old count: 135358 dt: 1.02 s LEC rate: 18762.47 Hz ***
DSSSD # 1 count: 4 old count: 2 dt: 1.02 s HEC rate: 1.97 Hz ***
DSSSD # 2 count: 113 old count: 96 dt: 1.02 s HEC rate: 16.70 Hz ***
DSSSD # 3 count: 437 old count: 386 dt: 1.02 s HEC rate: 50.11 Hz ***
DSSSD # 4 count: 341 old count: 288 dt: 1.02 s HEC rate: 52.08 Hz ***
DSSSD # 5 count: 53 old count: 46 dt: 1.02 s HEC rate: 6.88 Hz ***
DSSSD # 6 count: 35 old count: 31 dt: 1.02 s HEC rate: 3.93 Hz
all SortSas spectra and histograms zero'd
19.01 BigRIPS and BRIKEN stop daqs
AIDA on file R6_201
19.02 BigRIPS starts run 3026
BRIKEN starts run 26
AIDA on file R6_201
20:00 Detector biases & leakage currents OK - see attachments 20 & 21
FEE64 Temperatures OK - see attachment 22
merger data rate -- - see attachment 23
good events statistics OK - high but relatively stable - see attachment 24, 26
implantation profiles - see attachment 25
*all* system wide checks OK
20:02 BigRIPS and BRIKEN stop daqs
AIDA on file R6_219
20:02 BigRIPS starts run 3027
BRIKEN starts run 27
AIDA on file R6_220
21:02 BigRIPS and BRIKEN stop daqs
AIDA on file R6_237
20:03 BigRIPS starts run 3028
BRIKEN starts run 28
AIDA on file R6_237
21.21 beam lost during R6_243
BigRIPS and BRIKEN stop daqs
AIDA on file R6_243
21:24 beam back
21:25 BigRIPS starts run 3029
BRIKEN starts run 29
AIDA on file R6_244
22:26 BigRIPS and BRIKEN stop daqs
AIDA on file R6_262
22:28 BigRIPS starts run 3030
BRIKEN starts run 30
AIDA on file R6_263
22:36 Detector biases & leakage currents OK
FEE64 Temperatures OK
merger data rate good events statistics OK - high but relatively stable implantation profiles
for the details see attachments 27-32
23:05 beam lost
23:08 beam back
23:28 BigRIPS and BRIKEN stop daqs
AIDA on file R6_281
23:29 BigRIPS starts run 3031
BRIKEN starts run 31
AIDA on file R6_281 |
Mon Jun 5 16:02:50 2017, OH, PV, AE, Tuesday 06 June 00.00-08.00 40x
|
00:03 BigRIPS starts run 3007
BRIKEN starts run 007
AIDA on file R5_146
00:07 Bias and leakage currents ok - attachment 1
System wide checks all ok
FEE temperatures ok - attachment 2
Good evts still seeing noise on 13 and 14 - attachment 3
No merger errors since GO
00:48 DSSD Rates from R5_160
*** DSSSD # 1 count: 250904 old count: 163089 dt: 1.02 s LEC rate: 86326.91 Hz
*** DSSSD # 2 count: 335185 old count: 222784 dt: 1.02 s LEC rate: 110496.28 Hz
*** DSSSD # 3 count: 157445 old count: 103991 dt: 1.02 s LEC rate: 52548.18 Hz
*** DSSSD # 4 count: 102079 old count: 67878 dt: 1.02 s LEC rate: 33621.44 Hz
*** DSSSD # 5 count: 40500 old count: 26651 dt: 1.02 s LEC rate: 13614.32 Hz
*** DSSSD # 6 count: 64652 old count: 42396 dt: 1.02 s LEC rate: 21878.85 Hz
*** DSSSD # 1 count: 0 old count: 0 dt: 1.02 s HEC rate: 0.00 Hz
*** DSSSD # 2 count: 41 old count: 27 dt: 1.02 s HEC rate: 13.76 Hz
*** DSSSD # 3 count: 161 old count: 103 dt: 1.02 s HEC rate: 57.02 Hz
*** DSSSD # 4 count: 131 old count: 85 dt: 1.02 s HEC rate: 45.22 Hz
*** DSSSD # 5 count: 22 old count: 12 dt: 1.02 s HEC rate: 9.83 Hz
*** DSSSD # 6 count: 9 old count: 6 dt: 1.02 s HEC rate: 2.95 Hz
01:02 Briken and BigRIPS daq stops
01:03 Briken Starts run 8
BigRIPS starts run 3008
01:40 Phong noticed a significant difference in the noise in the (raw) low energy spectra a few of the FEEs when
comparing the runs right before and right after the power-cycle (See Attachment 12) required to recover
the DAQ crash last afternoon (June 5th).
NNAIDA 22, 21, 14, and 13 had a much higher count rate after the power cycle, and what seems like a lower
threshold. We can't observe the pulser peak in any of those.
NNAIDA 13 has the highest count rate, and the waveforms show what looks like noise with a regular
frequency (attachments 4 - 7)
For reference, we include the waveform of other positive FEEs 8-11
02:51 Bias and leakages ok - attachment 13
System wide checks ok
Good Event statistics - attachment 14
Fee Temperatures - attachment 15
03:03 BigRIPS and BRIKEN stop daqs
AIDA on file R5_208
03:04 BigRIPS starts run 3009
BRIKEN starts run 9
AIDA on file R5_208
04:03 BigRIPS and BRIKEN stop daqs
AIDA on file R5_230
04:04 BigRIPS starts run 3010
BRIKEN starts run 10
AIDA on file R5_230
04:09 System wide checks all ok
Bias and leakages ok - attachment 16
Fee Temperatures - attachment 17
Good Event Statistics - attachment 18
04:48 DSSD rates from R5_244
*** DSSSD # 1 count: 1239207 old count: 1155424 dt: 1.04 s LEC rate: 80379.42 Hz
*** DSSSD # 2 count: 1684042 old count: 1567549 dt: 1.04 s LEC rate: 111760.61 Hz
*** DSSSD # 3 count: 806937 old count: 752064 dt: 1.04 s LEC rate: 52643.85 Hz
*** DSSSD # 4 count: 505814 old count: 472642 dt: 1.04 s LEC rate: 31824.43 Hz
*** DSSSD # 5 count: 208086 old count: 193231 dt: 1.04 s LEC rate: 14251.53 Hz
*** DSSSD # 6 count: 335686 old count: 313190 dt: 1.04 s LEC rate: 21582.13 Hz
*** DSSSD # 1 count: 8 old count: 7 dt: 1.04 s HEC rate: 0.96 Hz
*** DSSSD # 2 count: 193 old count: 181 dt: 1.04 s HEC rate: 11.51 Hz
*** DSSSD # 3 count: 872 old count: 810 dt: 1.04 s HEC rate: 59.48 Hz
*** DSSSD # 4 count: 637 old count: 601 dt: 1.04 s HEC rate: 34.54 Hz
*** DSSSD # 5 count: 89 old count: 81 dt: 1.04 s HEC rate: 7.68 Hz
*** DSSSD # 6 count: 71 old count: 65 dt: 1.04 s HEC rate: 5.76 Hz
05:01 BRIKEN and BigRIPS stop DAQs
AIDA on file R5_250
05:02 BigRips starts run 3011
BRIKEN starts run 11
AIDA on file R5_250
06:02 BigRIPS and BRIKEN stop their daqs
AIDA on file R5_272
06:03 BigRIPS starts run 3012
BRIKEN starts run 12
AIDA on file R5_272
06:06 System wide checks all ok
Bias and leakage currents ok - attachment 19
Good event statistics - attachment 20
FEE temperatures ok - attachment 21
06:16 Analysis of R5_270 - attachment 22
Three MBS errors
06:34 130Cd decay curve - attachments 22 and 23
Despite the large amounts of noise in the detectors it is still possible to observe decay curves. They are
cleaned up a lot
by applying a multiplicity cut (attachment 22 with attachment 23 without. Further cuts and vetos could be
applied to clean up further.
The fit in attachment 22 gives a half life of 182ms which is not quite the expected 162ms but a reasonable
first fit.
07.14 Layout - 1.8.W waveforms - see attachments 24-27
No significant change in intrinsic/extrinsic noise cf. previous screenshots
nnaida13 ASIC 1-4 waveforms - see attachments 28-31 - range 100*20us
07.24 BRIKEN and BigRIPS stop DAQs
AIDA on file R5_300
07.25 Stop AIDA DAQ
Cycle detector biases
Start AIDA DAQ
07.38 BigRips starts run 3014
BRIKEN starts run 14
AIDA on file R6_1
07.45 nnaida13 ASIC 1-4 waveforms - see attachments 32-35 - range 100*20us
08.07 all system wide checks OK |
Mon Jun 6 19:26:38 2016, CG, DK, BM, AE, Tues 7th June 28x
|
02.08 Tried updating AIDA SYNC on stats page and got this message.
See attachment 8
02.16 Runs R1201-1206 are short runs on the 1GeV, each wit the variable degrader thickness changed to alter the depth of implantation (estimated by visual scalar rates per DSSD)
Started 02.16 - 02.47
02.56 Started R1207 (EURICA run_1407, BigRIPS run 1192).
Settings file 2016Jun07-05.50.53. All nnaida back to 20MeV setting.
Temps ok. Leakage currents ok.
3.68mm variable degrader in + 19.3mm fixed degrader.
Operator stopped beam @ 03.41. Run stopped.
Merger outputting some info to terminal re. unexpected header found (attachment 3)
04:05 Beam is back, but some trouble to get AIDA DAQ happy
04:52 Started R1208 (EURICA run_1408, BigRIPS 1193)
Settings file 2016June07-02.50.53
From 5:05 to 5:20, F82a, and F51a (twice) trip. We reduced the F51a from -745 V; -735 V; to -720 V settings.
06:51 Continued with R1208_19 (EURICA has 1410 begin at 6:40, BigRIPS began 1195 at ~7:00)
Biases uploaded as attachments 4 and 5
7:13 data stopped going to disk
Shortly after that (7:15) operator calls to take control of the beam.
07.40 Beam returned. Some trouble restarting DAQ.
07.56 Run R1209 started.
Settings file 2016Jun07-02.50.53
No changes to settings or degraders. Temps ok. Leakage currents ok.
PPACs tripped repeatedly around 08.10.
09.13 Beam setting changed and run stopped.
11.25 Current status
Julabo FL11006 recirculating chiller set point +15.0deg C (B3F +24.6 deg C, dew point +10.3 deg C, RH 40.7%)
System-wide checks - all FEE64s pass all tests *except* nnaida19 which fails ASIC clock timestamp and SYNC errors not counting
i.e. status is unchanged
nnaida19 merge data link (#18) disabled - see attachment 12
nnaida19 ReSYNCs OK but timestamp is adrift of other FEE64s - see attachment 11
Detector biases & leakage currents - OK - see attachments 9-10
R1211 neutron-rich setting *tuning*
R1212 Prior to start of run, AIDA trigger rates (1GeV FSR MEC setting)
#1 670Hz, #2 820Hz, #3 680Hz, #4 540Hz, #5 900Hz, #6 440Hz F11 650Hz, beta trigger 360Hz
start 15.10
stop 15.41 (run stopped to reboot EURICA VME crate)
ASIC settings DB key 2016Jun07-02.50.53
4mm Be target
Degrader (17+1)mm + 1.08mm (17+1)mm fixed #1 out #2 out #3 in (1.08mm) #4 no change, cannot be observed
Trigger = OR ( DSSSD#1 - #6 ) .AND. EURICA .AND. .NOT.veto (plastic downstream AIDA)
R1213
N.B. EURICA offline due to fault with VME crate RIO3 interface card continue with AIDA+BigRIPS
start 15.05
stop 17.34 (beam lost)
ASIC settings DB key 2016Jun07-02.50.53
4mm Be target
Degrader (17+1)mm + 1.08mm (17+1)mm fixed #1 out #2 out #3 in (1.08mm) #4 no change, cannot be observed
Trigger = OR ( DSSSD#1 - #6 ) .AND. EURICA .AND. .NOT.veto (plastic downstream AIDA)
Attachments 20-28 1*H spectra, FEE64 'right' (n+n ohmic strip, 20GeV FSR energy spectra) & stat/rate hit patterns
17.40 Julabo FL11006 set point +15.0deg C B2F/F11 temp 25.3 deg C, dew point 11.0 deg C, RH 41.8% |
Tue Feb 24 03:04:59 2015, CG, TD, Tues 24th Feb progress 9x
|
24/02/15
1200: Update the FEE firmware on nnaida1-16 without problem.
Gone from file FEE_Oct12_27Nov13.bin ----> FEE_Riken_Apr14_24.bin
Included are a screen shot of the temperature window showing the new firmware versions and a stats screen.
12.15 Update MIDAS DAQ software
su
cp -r /MIDAS@aiidas /MIDAS@aidas.240214 ( /MIDAS is a link to /MIDAS@aidas )
exit
cd <update directory>
cp -r Data_Acq /MIDAS
cp -r Merger /MIDAS
cp -r TclHttpd /MIDAS
New versions of merge64.AD and master64 were installed in /MIDAS/Merger but the process crashed at startup
with multiple messages of the type:
MIDAS Data Link (12345) : Received SEGV signal; exiting.
(see attachment 3)
Is this related to the recent changes to the MIDAS TapeServer software to resolve problems with MIDAS Data
libraries libdataspy.so and libxfer.so ?
As a temporary workaround I have copied the backups of merge64.AD and master64 back to the install
directory /MIDAS/Merger. With the old verison of the merger we continue to obtain error messages
(see attachments 4-8)
|
|