ID Date Authordown Subject
  746   Tue Oct 2 01:55:11 2018 OH, TDTuesday 2 October
09.54 DAQ continues OK - current file Sep18/R2_26

09.55 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3
      good event statistics OK - see attachment 4
      nnaida2 rate due to one hot channel

      Merger OK - see attachment 5
      TapeServer OK - see attachment 6
      no errors or warnings shown by server terminal sessions 

      1.8.L spectra - see attachments 7 & 8
      1.8.W spectra 20us FSR - see attachments 9 & 10
      
10.00 temperature +25.7 deg C, RH 29.6%, d.p. +6.4 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level c. 80%

10.03 ASIC check load (file Sep18/R2_26)
  747   Wed Oct 3 01:48:21 2018 OH, TDWednesday 3 October
09.42 DAQ continues OK - current file Sep18/R2_36

09.42 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3
      good event statistics OK - see attachment 4
      nnaida2 rate due to one hot channel

      Merger OK - see attachment 5
      TapeServer OK - see attachment 6
      no errors or warnings shown by server terminal sessions 

      1.8.L spectra - see attachments 7 & 8
      1.8.W spectra 20us FSR - see attachments 9 & 10

09.46 ASIC check load (file Sep18/R2_36)
      Note for p+n junction strip FEE64s the 1.8.W spectra are only updating for nnaida2, 10 & 17 - these traces
look OK
      Other FEE64s show a noise transient and do not update - ASIC check load did not fix - probably requires
DAQ STOP/GO  

09.57 temperature +25.4 deg C, RH 39.2%, d.p. +10.6 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level c. 80%

13:17 FEE temperatures and system wide checks all ok
      Alpha run R2 stopped
      R2_0 -> R2_37

13:18 Thresholds set to 0xa from 0x64 for all FEEs and ASICs

13:22 R3_0 started
      Thresholds at 0xa and pulser on at 25Hz setting ~ 2Hz

      ASIC settings saved to EXPERIMENTS/AIDA/2018Oct03-13.26.58
      Merger screenshot - attachment 11
      Stats - attachment 12
      Layouts 7 and 8 - attachments 13 & 14 

13:31 R3_0 Stopped

13:34 R4_0 Started
      ASIC thresholds all at 0x9
      Merger screenshot - attachment 15
      Stats - attachment 16
      Layout 1 - attachment 17

13:38 R4_0 Stopped

13:40 R5_0 Started
      ASIC thresholds all at 0x8
      Merger screenshot - attachment 18
      Stats - attachment 19
      Layout 1 - attachment 20

13:45 R5_0 Stopped

13:49 R6_0 Started
      ASIC slow comp at 0x7
      Merger screenshot - attachment 21
      Stats - attachment 22
      Layout 1 - attachment 23

13:52 R6_1 Stopped

      R7_0 Started
      ASIC Slow comp at 0x6
      Merger screenshot - attachment 24
      Stats - attachment 25
      Layout 1 - attachment 26
      Tapeserver window screenshot - attachment 27
      Top window showing 72% CPU usage - attachment 28

15:06 Pulser measurements carried out with thresholds at 0xa
      Typical pulser peak width was ~70-80 with a couple above 100.
      Double peaking that was observed earlier in the week is no longer observed.
      Both in the FEEs and on a scope

      Backside pulser peaks - attachment 29
      Frontside pulser peaks - attachment 30
  748   Wed Oct 3 07:09:03 2018 OH, TDCurrent FEE Configuration Map October 3rd 2018
             FEE Number
DSSD      A    B    C    D
  1       17   20   22   23
  2       18   19   21   24
  3       9    12   14   15
  4       10   11   13   16
  5       1    4    6    7
  6       2    3    5    8


    DSSD 1 3058-4  Top - Upstream
    DSSD 2 3058-3
    DSSD 3 2998-23
    DSSD 4 3058-5
    DSSD 5 3058-6
    DSSD 6 3058-7  Bottom - Downstream
  623   Fri Jun 2 00:01:48 2017 OH, PW, CBFriday 2 June 08:00-16:00
08:05 Briken and BigRIPS stopping daqs
      AIDA on file R1_1099

08:07 BigRIPS starts run 1088
      BRIKEN starts run  77

08:13 System wide checks all ok
      FEE temperatures all ok - attachment 1
      Good events statistics - attachment 2

08:50 Bias and leakage currents ok - attachment 3

09:04 BigRIPS and BRIKEN stop daqs
      AIDA on file R1_1113

09:05 BigRIPS starts run 1089
      BRIKEN starts run 78
      AIDA on file R1_1113

09:22 Beam lost AIDA on file R1_1117

09:26 Beam back AIDA on file R1_1118

10:06 BigRIPS and BRIKEN stop daqs
      AIDA on file R1_1128

10:07 BigRIPS starts run 1090
      BRIKEN starts run 79
      AIDA on file R1_1128

10:12 System wide checks ok

10:24 DSSD rates from R1_1131
 *** DSSSD # 1 count:     88384 old count:     45978 dt:      1.01 s  LEC rate:  42084.25 Hz
 *** DSSSD # 2 count:    119845 old count:     62135 dt:      1.01 s  LEC rate:  57272.13 Hz
 *** DSSSD # 3 count:     40809 old count:     21150 dt:      1.01 s  LEC rate:  19509.84 Hz
 *** DSSSD # 4 count:     37059 old count:     18986 dt:      1.01 s  LEC rate:  17935.87 Hz
 *** DSSSD # 5 count:     14947 old count:      7772 dt:      1.01 s  LEC rate:   7120.56 Hz
 *** DSSSD # 6 count:     29809 old count:     15349 dt:      1.01 s  LEC rate:  14350.29 Hz
 *** DSSSD # 1 count:        35 old count:        16 dt:      1.01 s  HEC rate:     18.86 Hz
 *** DSSSD # 2 count:       108 old count:        56 dt:      1.01 s  HEC rate:     51.61 Hz
 *** DSSSD # 3 count:        77 old count:        38 dt:      1.01 s  HEC rate:     38.70 Hz
 *** DSSSD # 4 count:        76 old count:        35 dt:      1.01 s  HEC rate:     40.69 Hz
 *** DSSSD # 5 count:        27 old count:        15 dt:      1.01 s  HEC rate:     11.91 Hz
 *** DSSSD # 6 count:         4 old count:         2 dt:      1.01 s  HEC rate:      1.98 Hz

11:07 BigRIPS and BRIKEN stop their daqs
      AIDA on file R1_1142

11:08 BigRIPS starts run 1091
      BRIKEN starts run 80
      AIDA on file R1_1142

11:10 Leakages and biases ok - attachment 4

11.36 DAQ continues OK
      Attached HEC x vs y - attach 5
      Zero'd HEC x vs y

11.59 Analysed R1_1153 - see attachment 6
      No timewarps
      Dead time negligible
      All OK

12:06 BigRIPS and BRIKEN stop their daqs
      AIDA on file R1_1156

12:07 BigRIPS starts run1092
      BRIKEN starts run 81
      AIDA on file R1_1156

13:07 BigRIPS and BRIKEN stop daqs
      AIDA on file R1_1170

13:08 BRIKEN starts run 82
      BigRIPS starts run 1093
      AIDA on file R1_1170

13:11 System wide checks ok

13:55 BRIKEN and BigRIPS stop their daqs
      AIDA on file R1_1181
      Beam gone

14:11 Bias and leakage currents ok - attachment 7
      Ambient Temperature 23.9C - Humidity 37.3% - D.P +8.3

15:26 Beam is back
      AIDA on file R1_1201

15:07 BigRIPS starts run 1094
      BRIKEN starts run 83
      AIDA on file R1_1201

15:34 System wide checks ok
      FEE temperatures ok - attachment 8
      Good Events statistics - attachment 9

15:39 No beam - beam back

15:46 Beam gone for couple of minutes

15:50 Beam back
  644   Thu Jun 8 00:54:26 2017 OH, PWThursday 8 June 00:00-16-00
08:54 DAQ still running

09:01 BRIKEN and BigRIPS stop daqs
      AIDA on file R7_293
      Beam stopped

09:25 System wide checks ok
      Bias and leakage ok - attachment 1
      Good event statistics - attachment 2
      FEE temperatures - attachment 3

13:36 Beam about to be back
      Briken on Run 65 (Background)
      AIDA on file 350

13:37 Beam is returned
      BRIKEN starts run 66
      bigRIPS run 3069
      AIDA on file R7_350
      System wide checks ok

13:40 Bias and leakage ok - attachment 4
      Good event - attachment 5
      FEE temperatures - attachment 6

14:00 Beam down originally five minutes

14:08 Beam will be down for another 20 minutes
      BRIKEN and BigRIPS stop daqs
      AIDA on file R7_356

14:22 beam is back

14:23 BigRIPS starts run 3070
      Briken starts run 67
      AIDA on file R7_360

15.23 DAQ stop
      AIDA file RIBF128/R7_373

15:26 BigRIPS starts run 3071
      Briken starts run 68
      AIDA on file R7_374

15.29 beam off - back in 5 mins
      AIDA file RIBF128/R7_374

15.44 beam back
      AIDA file RIBF128/R7_377
          
  634   Mon Jun 5 16:02:50 2017 OH, PV, AETuesday 06 June 00.00-08.00
00:03 BigRIPS starts run 3007
      BRIKEN starts run 007
      AIDA on file R5_146

00:07 Bias and leakage currents ok - attachment 1
      System wide checks all ok
      FEE temperatures ok - attachment 2
      Good evts still seeing noise on 13 and 14 - attachment 3
      No merger errors since GO

00:48 DSSD Rates from R5_160
 *** DSSSD # 1 count:    250904 old count:    163089 dt:      1.02 s  LEC rate:  86326.91 Hz
 *** DSSSD # 2 count:    335185 old count:    222784 dt:      1.02 s  LEC rate: 110496.28 Hz
 *** DSSSD # 3 count:    157445 old count:    103991 dt:      1.02 s  LEC rate:  52548.18 Hz
 *** DSSSD # 4 count:    102079 old count:     67878 dt:      1.02 s  LEC rate:  33621.44 Hz
 *** DSSSD # 5 count:     40500 old count:     26651 dt:      1.02 s  LEC rate:  13614.32 Hz
 *** DSSSD # 6 count:     64652 old count:     42396 dt:      1.02 s  LEC rate:  21878.85 Hz
 *** DSSSD # 1 count:         0 old count:         0 dt:      1.02 s  HEC rate:      0.00 Hz
 *** DSSSD # 2 count:        41 old count:        27 dt:      1.02 s  HEC rate:     13.76 Hz
 *** DSSSD # 3 count:       161 old count:       103 dt:      1.02 s  HEC rate:     57.02 Hz
 *** DSSSD # 4 count:       131 old count:        85 dt:      1.02 s  HEC rate:     45.22 Hz
 *** DSSSD # 5 count:        22 old count:        12 dt:      1.02 s  HEC rate:      9.83 Hz
 *** DSSSD # 6 count:         9 old count:         6 dt:      1.02 s  HEC rate:      2.95 Hz

01:02 Briken and BigRIPS daq stops

01:03 Briken Starts run 8
      BigRIPS starts run 3008

01:40 Phong noticed a significant difference in the noise in the (raw) low energy spectra a few of the FEEs when
      comparing the runs right before and right after the power-cycle  (See Attachment 12) required to recover 
      the DAQ crash last afternoon (June 5th). 
      NNAIDA 22, 21, 14, and 13 had a much higher count rate after the power cycle, and what seems like a lower
      threshold. We can't observe the pulser peak in any of those.
      NNAIDA 13 has the highest count rate, and the waveforms show what looks like noise with a regular
      frequency (attachments 4 - 7)

      For reference, we include the waveform of other positive FEEs 8-11

02:51 Bias and leakages ok - attachment 13
      System wide checks ok
      Good Event statistics - attachment 14
      Fee Temperatures - attachment 15

03:03 BigRIPS and BRIKEN stop daqs
      AIDA on file R5_208

03:04 BigRIPS starts run 3009
      BRIKEN starts run 9
      AIDA on file R5_208

04:03 BigRIPS and BRIKEN stop daqs
      AIDA on file R5_230

04:04 BigRIPS starts run 3010
      BRIKEN starts run 10
      AIDA on file R5_230

04:09 System wide checks all ok
      Bias and leakages ok - attachment 16
      Fee Temperatures - attachment 17
      Good Event Statistics - attachment 18

04:48 DSSD rates from R5_244
 *** DSSSD # 1 count:   1239207 old count:   1155424 dt:      1.04 s  LEC rate:  80379.42 Hz
 *** DSSSD # 2 count:   1684042 old count:   1567549 dt:      1.04 s  LEC rate: 111760.61 Hz
 *** DSSSD # 3 count:    806937 old count:    752064 dt:      1.04 s  LEC rate:  52643.85 Hz
 *** DSSSD # 4 count:    505814 old count:    472642 dt:      1.04 s  LEC rate:  31824.43 Hz
 *** DSSSD # 5 count:    208086 old count:    193231 dt:      1.04 s  LEC rate:  14251.53 Hz
 *** DSSSD # 6 count:    335686 old count:    313190 dt:      1.04 s  LEC rate:  21582.13 Hz
 *** DSSSD # 1 count:         8 old count:         7 dt:      1.04 s  HEC rate:      0.96 Hz
 *** DSSSD # 2 count:       193 old count:       181 dt:      1.04 s  HEC rate:     11.51 Hz
 *** DSSSD # 3 count:       872 old count:       810 dt:      1.04 s  HEC rate:     59.48 Hz
 *** DSSSD # 4 count:       637 old count:       601 dt:      1.04 s  HEC rate:     34.54 Hz
 *** DSSSD # 5 count:        89 old count:        81 dt:      1.04 s  HEC rate:      7.68 Hz
 *** DSSSD # 6 count:        71 old count:        65 dt:      1.04 s  HEC rate:      5.76 Hz

05:01 BRIKEN and BigRIPS stop DAQs
      AIDA on file R5_250

05:02 BigRips starts run 3011
      BRIKEN starts run 11
      AIDA on file R5_250

06:02 BigRIPS and BRIKEN stop their daqs
      AIDA on file R5_272

06:03 BigRIPS starts run 3012
      BRIKEN starts run 12
      AIDA on file R5_272

06:06 System wide checks all ok
      Bias and leakage currents ok - attachment 19
      Good event statistics - attachment 20
      FEE temperatures ok - attachment 21

06:16 Analysis of R5_270 - attachment 22
      Three MBS errors

06:34 130Cd decay curve - attachments 22 and 23
      Despite the large amounts of noise in the detectors it is still possible to observe decay curves. They are
cleaned up a lot
      by applying a multiplicity cut (attachment 22 with attachment 23 without. Further cuts and vetos could be
applied to clean up further.
      The fit in attachment 22 gives a half life of 182ms which is not quite the expected 162ms but a reasonable
first fit.

07.14 Layout - 1.8.W waveforms - see attachments 24-27
      No significant change in intrinsic/extrinsic noise cf. previous screenshots
      nnaida13 ASIC 1-4 waveforms - see attachments 28-31 - range 100*20us

07.24 BRIKEN and BigRIPS stop DAQs
      AIDA on file R5_300

07.25 Stop AIDA DAQ
      Cycle detector biases
      Start AIDA DAQ

07.38 BigRips starts run 3014
      BRIKEN starts run 14
      AIDA on file R6_1

07.45 nnaida13 ASIC 1-4 waveforms - see attachments 32-35 - range 100*20us

08.07 all system wide checks OK
  642   Wed Jun 7 08:02:42 2017 OH, PVWednesday 7 June 16:00 - 24:00
16:02 System wide checks ok
      Good Events statistics (No beam at time!)- attachment 1
      Fee temperatures - attachment 2
      Bias and leakage ok - attachment 3

16:29 BigRIPS starts run 3051
      BRIKEN starts run 49
      AIDA on file R7_80

      Slits brought in to reduce rate to attempt implant-neutron correlation

16:39 DSSD rates from R7_81
 *** DSSSD # 1 count:    182087 old count:    144548 dt:      1.02 s  LEC rate:  36716.57 Hz
 *** DSSSD # 2 count:    295938 old count:    235754 dt:      1.02 s  LEC rate:  58865.45 Hz
 *** DSSSD # 3 count:     80412 old count:     64397 dt:      1.02 s  LEC rate:  15664.13 Hz
 *** DSSSD # 4 count:     94793 old count:     75541 dt:      1.02 s  LEC rate:  18830.22 Hz
 *** DSSSD # 5 count:     23432 old count:     18547 dt:      1.02 s  LEC rate:   4777.98 Hz
 *** DSSSD # 6 count:     66315 old count:     53096 dt:      1.02 s  LEC rate:  12929.39 Hz
 *** DSSSD # 1 count:         0 old count:         0 dt:      1.02 s  HEC rate:      0.00 Hz
 *** DSSSD # 2 count:         8 old count:         4 dt:      1.02 s  HEC rate:      3.91 Hz
 *** DSSSD # 3 count:        23 old count:        20 dt:      1.02 s  HEC rate:      2.93 Hz
 *** DSSSD # 4 count:        38 old count:        32 dt:      1.02 s  HEC rate:      5.87 Hz
 *** DSSSD # 5 count:        13 old count:        10 dt:      1.02 s  HEC rate:      2.93 Hz
 *** DSSSD # 6 count:        11 old count:         8 dt:      1.02 s  HEC rate:      2.93 Hz

17:13 BigRIPS and BRIKEN stop DAQS
      AIDA on file R7_89

17:30 Beam back
      BigRIPS starts run 3052
      BRIKEN starts run 50
      AIDA on file R7_93

17:52 Analysis of R7_93 No timewarps, no MBS timewarps - attachment 4

18:06 System wide checks all ok 
      Bias and leakage ok - attachment 5
      Good events statistics - attachment 6
      FEE temperatures ok - attachment 7

18:27 fRC problem, beam is off
      
18:29 BigRIPS and BRIKEN stop DAQS
      AIDA on file R7_106

18:32 Beam is back
      BigRIPS starts run 3053
      BRIKEN starts run 51
      AIDA on file R7_106

19:33 BigRIPS and BRIKEN stop daqs
      AIDA on file R7_119

19:34 BRIKEN starts run 52
      BigRIPS starts run 3054
      AIDA on file R7_120

20:09 System wide checks ok
      Bias and leakage ok - attachment 8
      Good event statistics - attachment 9
      Fee temperatures - attachment 10

20:28 BigRIPS and BRIKEN stop their DAQs
      AIDA on file R7_131

20:30 BigRIPS starts run 3055
      BRIKEN starts run 53
      AIDA on file R7_131

20:41 Layout 3 - attachment 11
      1.8.L spectrum for each FEE
      Appears to be 4-5 pulser peaks

21:31 BRIKEN and BigRIPS stop their daqs
      AIDA on file R7_144

21:32 BRIKEN starts run 54
      BigRIPS starts run 2056
      AIDA on file R7_144

22:03 System wide checks ok
      Bias and leakage ok - attachment 12
      Good events statistics - attachment 13
      Fee temperatures - attachment 14
      

22:32 Briken and BigRIPS stop their daqs
      AIDA on file R7_156

22:33 BigRIPS starts run 3057
      BRIKEN starts run 55
      AIDA on file R7_156

23:31 BigRIPS and BRIKEN stop their DAQs
      AIDA on file R7_169

23:32 BigRIPS starts run 3058
      BRIKEN starts run 56
      AIDA on file R7_169

23:46 Analysis of R7_168 - attachmemt 15
  544   Wed Mar 29 13:38:38 2017 OH, DK, CGFirst beam runs
For BigRIPS run numbers, they are available online: http://bripscnt01.rarfadv.riken.jp/runsum/index.php

21:30 AIDA and BRIKEN are ready to go -- so we closed the door and asked the BigRIPS team to send the beam to
F11.  It will take about 30 minutes as they want to recalibrate one of the Faraday cups.

22:09: Beam is coming to F11.  AIDA sub-run R13_17 is running at 22:10 so nearly the same time.
     BigRIPS run number 0140
     BRIKEN file name 170329_2213_40Mg Run 0.  It is started at 22:13.  R13_18 for AIDA
     See attachment 1 for biases.


23:12:  We notice that BigRIPS began the run 0141 at 23:08
     AIDA subrun R13_36 began about or near 23:07 or 23:08
     BRIKEN run number 170329_2313_40Mg Run 1
     All system-wide checks passed (excepting nnaida6 clock status which is a known issue)
     Statistics rates look reasonable, except that the nnaida 16 23 24 are even higher than before when the beam
was off (~200k)
 
23:59 - AIDA crashed.  Merger stopped.  When attempting to stop the run, the following error is displayed:

STATE for nnaida1 returned with an error
error: SOAP http transport timed out after 20000 ms
NONE
error: SOAP http transport timed out after 20000 ms
    while executing
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
    ("eval" body line 1)
    invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
    (procedure "DataAcquisitionControlClient__GetState" line 1)
    invoked from within
"DataAcquisitionControlClient__GetState"

We kill off the merge and tape server.

DAQ itself comes back after a reset.  Attempting to open the tape service, however, yields an error

Cannot issue more Access Tokens
 ClaimTapeServer returned:- 
3 Capability table is full.

30.3.2017

0:56 When attempting to bring AIDA back up, eventually nnaida10 was not showing merger data before enabling the
ASIC readout data.  On the run control, nnaida10 consistently showed that its data transfer was not enabled. 
Attempting to reset etc., then eventually issuing Go would drop all the data transfers.

After a power cycle by the standard method, essentially the problem persists, except that it is nnaida 7 8 9 10
(the whole block) which shows the problem where we cannot get the enabled data transfer to stick.
  549   Sat Apr 1 16:16:05 2017 OH, CGSunday 2 April
00:12 BigRips started run 0223 (Started at 00:12)
      AIDA file R20_344
      System wide checks all ok
      BRIKEN start run 67

01.38 Beam stopped to F11. ~1/2 way through R20_364.

01.43 Beam returned to F11. 1GB through R20_366.
      BigRIPS file 0224.
      BRIKEN run 68.
      System-wise checks all ok, except nnaida1 global clocks failed 6.
      Temps ok. Leakage currents ok (see attachment 1).

02.43 BigRIPS started run 0225. Start of R20_381
      
02.48 BRIKEN run 69 started. Start of R20_382.

03.28 System-wide checks same as previous. (R20_391)

03.43 BigRIPS file 0226. ~1/2 way through R20_394

03.48 BRIKEN run 70 started.

04.46 BigRIPS file 0227 started. End of R20_410.

04.48 BRIKEN run 71 started. Start of R20_411.

05.31 36Ne setting is over. End of R20_421.
      BigRIPS team is changing setting for calibration. No beam will go through Zerodegree
      
05.51 R20 stopped (I was getting food).

06.30 20mm Al degrader + 3mm plastic removed from F8. 23mm Al degrader placed at F11 after IC.

06.35 BigRIPS file 0237 started - 38Mg setting.

06.37 R21 started on 38Mg setting.

06.57 Beam stopped to F11 - accelerator problem.
      ~700MB into R21_5.
      Returned shortly after.

07.23 BigRIPS file 0245 started. ~3/4 way through R21_10.

07.26 BigRIPS file 0246 started. ~1/2 way through R21_11.

07.32 BigRIPS file 0247 started. Start of R21_13.

07.34 BigRIPS file 0248 started. ~1/2 way through R21_13.

07.38 BigRIPS file 0249 started. ~1GB into R21_15.

07.40 BigRIPS file 0250 started. Start of R21_16.

07.41 BigRIPS file 0251 started. Little way through R21_16.

08.06 38Mg setting finished. R21 stopped.

08.08 R22 started - got a bit trigger happy with the DAQ. No beam to F11.

08.10 Degrader set to 10mm Al.

08.16 BigRIPS file 0252 started on 39Al setting.

08.17 R23 started on 39Al setting.

08.20 R23 ended.

08.26 40Si setting. Too high rate for AIDA so beam won't be sent down ZDS.

08.40 F11 degrader set to 15mm Al fixed + 4mm Al variable (all in).

08.43 2nd run at 39Al setting.
      BigRIPS file 0254 started.
      R24 started.

09.00 R24_4 closed. May have been issues with degrader - not sure we were implanting in AIDA...
      End of experiment!

09.07 AIDA background run started - R25.
      Pulser is on in the run though.

09.20 R25 ended.
      BigRIPS run summary attached.
  628   Sat Jun 3 08:07:48 2017 OH, CB, AESaturday 3 June 16:00-24:00
16:26 DAQ continues OK
      all system wide checks ok, *except* nnaida1 clock status (6), nnaida1, 21 ADC calibration
      Leakage currents OK - attachment 1
      FEE64 temperatures OK - attachment 2
      Good event statistics OK - attachment 3
      No new merger errors - attachment 4
      Sorted run R1_1540. DSSSD rates as follows
       *** data items:    8958582 ( 115623.74 Hz)
 *** ADC events:     954672 (  12321.45 Hz)
 *** time warps:          0 (      0.00 Hz)
 *** DSSSD # 1 count:    392147 old count:    361510 dt:      8.78 s  LEC rate:  44646.94 Hz
 *** DSSSD # 2 count:    557051 old count:    514085 dt:      8.78 s  LEC rate:  63421.69 Hz
 *** DSSSD # 3 count:    185346 old count:    172449 dt:      8.78 s  LEC rate:  21102.12 Hz
 *** DSSSD # 4 count:    154380 old count:    142180 dt:      8.78 s  LEC rate:  17576.56 Hz
 *** DSSSD # 5 count:     67195 old count:     62386 dt:      8.78 s  LEC rate:   7650.32 Hz
 *** DSSSD # 6 count:    131951 old count:    122163 dt:      8.78 s  LEC rate:  15022.96 Hz
 *** DSSSD # 1 count:       185 old count:       171 dt:      8.78 s  HEC rate:     21.06 Hz
 *** DSSSD # 2 count:       450 old count:       410 dt:      8.78 s  HEC rate:     51.23 Hz
 *** DSSSD # 3 count:       379 old count:       351 dt:      8.78 s  HEC rate:     43.15 Hz
 *** DSSSD # 4 count:       297 old count:       269 dt:      8.78 s  HEC rate:     33.81 Hz
 *** DSSSD # 5 count:       132 old count:       127 dt:      8.78 s  HEC rate:     15.03 Hz
 *** DSSSD # 6 count:        20 old count:        19 dt:      8.78 s  HEC rate:      2.28 Hz
 
     Pulser event still synchronised - see attachment 5

16:34  BigRIPS & BRIKEN stop runs
       AIDA on file R1_1542

16:35  BigRIPS starts run 1120
       BRIKEN starts run 109
       AIDA on file R1_1543
       Zero'd all SortSas

17:33 bigRIPS and BRIKEN stop daqs
      AIDA on file R1_1556

17:34 bigRIPS starts run 1121
      BRIKEN starts run 110
      AIDA on file R1_1556

17:54 System wide checks all ok

19:50 bigRIPS starts run 1123
      BRIKEN starts run 112
      AIDA on file R1_1588

21:00 End of 78Ni run
      last AIDA file R1_1606
      Stopped AIDA DAQ
      Refilled chiller

21:18 Performing pulser walkthrough. 
      Mistakenly started Tape server before starting DAQ. There will be no R2 run.
      Starting R3 - pulser walkthrough.
      
      Pulser frequency 250 Hz
      Pulser amplitudes 90 000, 80 000, 70 000 ... 10 000

      Can't see pulser peak in spectra. Issue unclear

21:40 Restarted high-amplitude pulser walkthrough run. Run RIBF127R1/R4_0
      Pulser frequency 250 Hz
      Pulser amplitudes 90 000, 80 000, 70 000 ... 10 000

      Pulser peak now visible in histograms

      Started low-amplitude pulser walkthrough run. Run RIBF127R1/R5_0
      Pulser frequency 250 Hz
      Pulser amplitudes 12 000, 10 000, ... 2 000

      Set pulser amplitude back to 90 000, frequency to 25 Hz

23:17 Merger and Tape Server stopped
      Links to FEEs disabled
      All Browser tabs closed
      Terminal sessions closed

23:25 Started R1 in /TapeData/RIBF128/

23:43 Start ASIC data readout
      System wide checks ok
      ASIC load check ok
  841   Fri Jun 14 10:33:44 2019 OH, CA, TD, CG, CB, DKFriday 14th June
Friday 14th June 

Authors: TD, DK, CB

08:40

  elog appears to be down.  Relevant screenshots are in /tmp/june14/ so please upload when we can access

  Julabo chiller 75% M-H markers
  Room Temp +24.5 C deg, dew point +8.1 deg C, Room humidity 35.2%

  Pulser, ASIC settings as yesterday (copy/paste)

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

  DAQ continues OK
  Statistics as before.  nnaida19 counter appears to have overflowed and wrapped around
    See stats.png
  
  System-wide checks OK, except nnaida7 fails ADC calibration as before

  Biases and leak currents reasonable -- see biases.png

  FEE64 temperatures okay, see fee-temps.png

  Merger looks okay, ~600k items / sec; see merger.png

  TapeServer as before, around 4.5 Mbytes/sec; see tapeserver.png


13:52

   Pulser, ASIC settings same as this morning

   DAQ continues OK
   Statistics as before. See attach 190614_1354_stats. Counts RESET.

   System-wide checks OK, except nnaida7 fails ADC calibration as before

   Bias and leak currents OK. See attach 190614_1356_bias

   FEE64 temperatures OK, see attach 190614_1357_temps

   Merger keeps running at ~560k items /s; see 190614_1359_merger
   TapeServer as before. See 190614_tape

17:24 AIDA DAQ crashed by DTAS touching the clock function generator
      Function generator set at 50MHz square high +5V low 0V

18:38 problem loading ASIC settings - mismatch in slow comparator threshold and shaping time
      performed save/restore with 2019Jun13-08.56.35 - all looks ok now
      
      System wide checks ok except nnaida7 fails adc calibration

      FEE64 temperatures ok - attachment 12

      good event statistics ok - attachment 13

      bias/leakage currents ok - attachment 14

     
18:38 for reals

      DAQ Stopped
      R30_0 opened writing to disk
      
      Was to test DTAS sync whilst writing to file
18:43 DAQ stopped on file R30_0

18:45 DAQ restarted on file R31 with no writing to disk

21:50 DAQ found crashed (Valencia group believe it crashed around 20:00)

      Trying to reset FEEs resulted in an error message attachment 15 and 16

      Each FEE was rebooted using telnet to avoid a power cycle
      System reset after reboot

      System wide checks ok
      Check ASIC control 

      FEE temps ok

      DAQ Started again
      On startup FEEs 17, 20, 22 and 23 were all about 150k
      All relate to DSSD0
      Pressed on kapton and noise went down.

      Statistics now good

22.45 Review /var/log/messages and observe nnaida15 rebooted at 19.27 - reason unknown
      This probably explains why DAQ stopped.

Jun 14 19:27:32 aidas1 dhcpd: DHCPDISCOVER from 00:04:a3:2a:f3:ac via p2p2
Jun 14 19:27:32 aidas1 dhcpd: DHCPOFFER on 10.1.2.15 to 00:04:a3:2a:f3:ac via p2p2
Jun 14 19:27:32 aidas1 dhcpd: DHCPREQUEST for 10.1.2.15 (10.1.2.250) from 00:04:a3:2a:f3:ac via p2p2
Jun 14 19:27:32 aidas1 dhcpd: DHCPACK on 10.1.2.15 to 00:04:a3:2a:f3:ac via p2p2
Jun 14 19:27:32 aidas1 rpc.mountd[2868]: authenticated mount request from 10.1.2.15:784 for
/data10/XilinxLinux/ppc_4xx/rfs/nnaida15 (/data10/XilinxLinux/ppc_4xx/rfs)
Jun 14 19:27:59 aidas1 rpc.mountd[2868]: authenticated mount request from 10.1.2.15:864 for /MIDAS@aidas
(/MIDAS@aidas)


22.53 ADC data synchronised - attachment 17

23.05 zero DAQ statistics
      zero all histograms
      

23.20 Check ASIC Control x1 
      FEE64 Rate spectra - attachment 18

      1.8.L spectra - see attachments 19 & 20
      nnaida1 81 ch FWHM, nnaida3 77 ch FWHM

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 21
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 22

23.36 DAQ start file May19/R34_0

      DTAS+AIDA correlation test using gamma ray sources

      Correlation scaler reset issued during R34_0

      ASIC settings 2019Jun13-08.56.35

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

23.53 DAQ stop file May19/R34_2

      nnaida15 did not stop - state unknown - reboot nnaida15

      This did not work - unable to reset nnaida15

      Closed all FireFox broweser tabs, stopped all servers and restarted

00.14 DAQ start

      ASIC settings 2019Jun13-08.56.35

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse
  750   Fri Oct 26 03:28:10 2018 OH, CAFriday 26 October
October 25th:
In anticipation of losing access in the evening on arrival at RIKEN the following was done:
- Chiller was refilled
- All power cables reconnected after being removed in October for the RIKEN power down
- All NIM bins, fan trays and RPis were turned on
- 50MHz AIDA clock was set up and connected
- Remote connections to AC Mains relay and RPi for CAEN bias was tested
- Chiller was powered up

October 26th:

11:31 Still have access to the F11 area due to a bigRIPS delay.
      Double checked wiring from yesterday and all still seems good
      Chiller is still running with set point at +20 deg C 

11:36 FEEs powered up. All 24 mounted

11:57 Bias and leakage currents all ok - attachment 1
      FEE temperatures all ok - attachment 2
      System wide checks all ok

13:26 Bias and leakage currents ok - attachment 3
      FEE temperatures all ok - attachment 4
      System wide checks all ok

14:00 Stats are ok - attachment 5
      Appear similar to last time: https://elog.ph.ed.ac.uk/AIDA/747
      However FEE24 and 24 appear much noisier than last time with rates around 100k

19:24 Pulser walkthrough run from 90000 to 10000 in steps of 10000 for R2
      Opened file Oct18/R3, alpha run, threshold at 100kev, pulser on.
      
  863   Mon Apr 19 15:10:55 2021 OH TDLess HEC events than expected in DSSD3
we are noticing we appear to see less implant events in detector 3 than we expect. This was first noticed in the xy spectra of the DSSD in the Go4 online.
As part of testing we have lowered the threshold in aida09 which from the xy appeared to have less events.

Following this we still seemed to see less events.

OH downloaded a selection of recent files to analyse. Looking at the total number of HEC events it appears that aida09, aida10 and aida12 are all seeing less HEC events than I would expect.
The ratio of HEC events in aida09 to aida11 was 0.36.

Looking at the energies of the HEC events there appeared to be less events everywhere and not just a threshold issue. We thought it could then be radiation damage and poor
charge collection. To test this we raised the bias to 120V on DSSD3.

The same test was again performed and the ratio again came out to be 0.36
We did not appear to recover any HEC events

Comparing the scaled energy distributions of aida09 and aida11 we see no evidence of a different energy response other than an increase in lower energy events in aida09 due to the lower threshold

I now think it may be a hardware issue. Will look at the last file before the most recent powercycle.

Using TD's analyser program can confirm the rate between aida09 and aida11 before 08:30 this morning was ~1:1.
The problem seems to have appeared during the power cycles following this.
  782   Tue Nov 13 22:58:49 2018 OH Wednesday 14 November 08:00 - 16:00
08:01 BigRips and BRIKEN stop
      AIDA on file R5_346

08:02 BigRIPS starts run 1038
      BRIKEN starts run 181114_0800_165Pm_053.dlt
      AIDA on file R5_346

      System wide checks ok * except nnaida 7 fails adc*
      FEE Temps ok - attachment 1
      Stats ok - attachment 2 
      Leakage currents ok - attachment 3

09:15 BigRIPS and BRIKEN stop
      AIDA on file R5_358

09:16 BigRIPS starts run 1039
      BRIKEN starts run 181114_0915_165Pm_054.dlt
      AIDA on file R5_358

10:15 BigRIPS and BRIKEN stop
      AIDA on file R5_368

10:15 BigRIPS starts run 1040
      BRIKEN starts run 181114_1015_165Pm_055.dlt
      AIDA on file R5_368

10:36 Analyser output of R5_368 - attachment 4

11:15 BugRIPS and BRIKEN stop
      AIDA on File R5_378

11:16 BigRIPS starts run 1041
      BRIKEN starts run
      AIDA on file R5_378



08.03 DAQ continues - AIDA file RIBF148/R5_384
      165Pm BigRIPS setting + 2.8mm Al degrader

      ASIC settings 2018Oct03-13.26.58
       from c. 09.00 13.11.18
       n+n FEE64s Ibias LF feedback 0xb -> 0xd
       HEC fast comparator threshold 0x20 -> 0xff
      
11.50 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 5 & 6

      FEE64 temperatures OK - see attachment 7
      good event & AIDA discriminator statistics OK - see attachments 8

      Merger OK - see attachment 9
      TapeServer OK - see attachment 10
      no errors or warnings shown by server terminal sessions 

      per FEE64 Rate spectra - attachment 11
      per FEE64 HitRate spectraq - attachment 12
      1.8.L spectra - attachments 13-16
        some evidence of pulser peak shifts, notably nnaida5, 11 & 15
        nnaida1 pulser peak width 78 ch FWHM
      1.8.H spectra - attachments 17 & 18
 
      1.8.W spectra 20us FSR - attachments 19 & 23
      1.8.W spectra 200us FSR - attachments 20 & 24
      1.8.W spectra 2ms FSR - attachments 21 & 25
      1.8.W spectra 20ms FSR - attachments 22 & 26

12.15 BugRIPS and BRIKEN stop
      AIDA on File R5_388

12.16 BigRIPS starts run 1042
      BRIKEN starts run 57
      AIDA on file R5_388

12.22 *all* online and MIDASsort spectra zero'd

13:15 BigRIPS and BRIKEN stop
      AIDA on file R5_398

13:16 BigRIPS starts run 1043
      BRIKEN starts run58
      AIDA on file R5_398

13:28 BigRIPS and BRIKEN stop
      AIDA on file
      beam halted to give access to F11 area

13.44 beam started
      BigRIPS and BRIKEN start
      AIDA file RIBF148/R5_402
      BRIKEN run 59
      BigRIPS run 1043

14.45 BRIKEN and BigRIPS stop
      AIDA file RIBF148/R5_412

14.47 BRIKEN and BigRIPS start
      AIDA file RIBF148/R5_413
      BRIKEN run 60
      BigRIPS run 1044

15.43 BRIKEN and BigRIPS stop
      AIDA file RIBF148/R5_422

      BRIKEN Resets the timestamp

15:44 BRIKEN and BigRIPS start
      AIDA file RIBF148/R5_422
      BRIKEN run 61
      BigRIPS run 1045

16:02 System wide checks all ok *Except nnaida7 fails adc calibration*
      Leakage currents ok - attachment 27
      Stats ok - attachment 28
      FEE Temps ok - attachment 29

16:27 First 200ms of decay correlations in DSSD 3 and 4 for the 3 degrader settings - attachment 30
      DSSD 3 along the top row and DSSD 4 along the bottom.
      Dgrader settings are 3mm, 2.3mm and 2.8mm
      1 hour data in each
  424   Sat Nov 5 09:40:02 2016 OHUpdated Leakage Current Spreadsheet
Attached is the updated leakage current spreadsheet (Attachment 1)  as pf 5/11/2016 it is up to date with the elog.

Also added is the recording of ambient temperature. This is recorded in degrees C and plotted relative to 21 degrees C to better show on the same plot as leakage currents. (Screenshot in attachement 2)

Also attached is the spreadsheet in csv formatting.
  427   Sun Nov 6 05:16:22 2016 OHUpdate
Leakage current spreadsheet has now been uploaded to google docs for ease of access.

It can be accessed and edited from the following link: https://docs.google.com/spreadsheets/d/12hgbrywB10hGsKt5uc2HnLfZymh8_uvM6cEASbbqnBE/edit?usp=sharing
  532   Thu Feb 2 13:44:02 2017 OHT9 Test Bench
T9 test setup at Daresbury was powered up and noise levels were checked using a pulser peak.

A 1V peak was used. For most channels this corresponded to a peak position at around +-11000 from the zero point at 9000.

An example of this is shown in attachment 1 showing the peak in nnaider 11, ASIC 4, ADC 11.

It was also observed that for both nnaider 11 and 12 (FEE modules corresponding to negative signals) that channels 2, 56 and 59 were not contributing to the rate. See attachment 2.

Upon investigating the missing channels ADC spectra the only apparent issue appeared in nnaida 11 ASIC 4, ADC 8. Which is plot 4.7.L. Here we see the peak much higher coming in at around channel 63000. See attachment 3.
The remaining missing channels in the rate plot appear to have normal amplitudes with maybe slightly higher noise. FWHM of ~60 as oppposed to ~18. Discussed more below.

Also done was a noise measurement across the all the ASICs. To do this an ADC was chosen belonging to each of the ASICs and the pulser peak location and FWHM was recorded. The reaults are as shown below.

FEE#	ASIC/ADC	Centroid Location	FWHM
 11       1.8             42951                19.38
 11       2.8             41012                17.48
 11       3.8             41929                20.01
 11       4.8             41302                17.88
 12       1.8             41830                17.38
 12       2.8             41119                16.01
 12       3.8             42669                18.30
 12       4.8             41811                15.21
 13       1.8             24329                15.05
 13       2.8             24034                13.64
 13       3.8             24209                15.79
 13       4.8             24607                13.29
 14       1.8             24650                16.19
 14       2.8             24618                14.12
 14       3.8             23934                15.04
 14       4.8             22584                14.36
 14       4.8             22584                14.36

It was seen that the noise was fairly low which is to be expected as there were no dtectors connected and only a short length of kapton cables connected to the board.
  Draft   Wed Apr 5 06:10:55 2017 OHBigRIPS particle ID
  553   Wed Apr 5 06:11:08 2017 OHBigRIPS particle ID
I've been going through collecting the BigRIPS root files with Phong and making sure they are on AIDAs1.

The files we have for NP1306 are not the standard format so to get particle ID you need to do this command taken
from the RIKEN elog.

Command to construct the PID in the merged files from preliminary data:

Example file a11111_b1181_e_1130_ion.root

tree->Draw("(F11ICADC[0]+F11ICADC[1]+F11ICADC[2]+F11ICADC[3]+F11ICADC[4]+F11ICADC[5])/6:F11Pla_TM-F3Pla_TM+F5X*0.0383261>>h(500,0,25,500,0,16000)","","colz");

Though these root files were not created by the BigRIPS team so can be cleaned up further but should be good
enough for initial analysis.
  615   Tue May 30 16:05:12 2017 OHWednesday 31 May 00:00 - 08:00
00:05 BigRIPS and BRIKEN stop their daqs
      AIDA currently on file R1_312

00:06 BRIKEN start run 23
      BigRIPS start run 1024
      AIDA on file R1_312
     
      System wide checks all ok

01:05 BigRIPS and BRIKEN stop their daqs
      AIDA on file R1_327

01:06 BigRips Start run 1025
      BRIKEN start run 24
      AIDA file R1_327

      System wide checks all ok

02:02 BigRIPS and BRIKEN stop their daqs
      AIDA on file R1_341
      ASIC load check ok

02:03 BigRIPS start run 1026
      BRIKEN start run 25
      AIDA ON FILE R1_341

02:11 System wide checks all ok
      Leakages and bias ok (Leakages rising again though) - attachment 1
      FEE temperatures ok - attachment 2
      Good event statistics - attachment 3
      Merger and tape server still running - attachment 4

02:23 Beam is off. Will recover soon

02:26 Beam is back

02:27 beam stopped again

02:28 beam back

02:37 Since beam has come back we appear to have lost all implant channels.
      only seeing events in DSSD2 

03:30 ASIC load issue resolved. AIDA data should be good from R1_353

      Beam continues to turn off and on

03:52 Since ~2am beam has been passing through the dssd
      Implantation spectrum - attachment 5
      Upon investigation the fixed degrader had fallen on the left and was at a 30 degree angle.
      It has now been put back in position
      AIDA file when fixed R1_360
      BRIKEN reset the timestamp
      Whilst sorting the ASIC issue bigRips moved onto run 1029
      BRIKEN run 1028. Full details in their elog

03:55 briken and bigRIPS stop runs
      bigRIPS start run 1028
      BRIKEN run 27
      AIDA file R1_360

04:29 Implant hit spectra after the degrader was replaced - attachment 6

04:58 BigRIPS and BRIKEN stop daqs
      AIDA on file R1_376

04:59 BigRIPS starts file 1029
      BRIKEN starts run 28
      AIDA at very end of file R1_376

05:20 System wide checks ok
      bias and leakage currents ok - attachment 7

05:26 Beam still dropping out for short moments of time

06:03 BigRips and BRIKEN stop their runs
      AIDA on file R1_393

06:04 BRIKEN start run 29
      BigRIPS start run 1030
      AIDA on file R1_393

06:28 Beam lost

06:31 Beam returned

06:46 System wide checks ok
      FEE temperatures ok - attachment 8

07:01 BigRIPS and BRIKEN stop their runs
      AIDA on file R1_407

07:02 BRIKEN start run 30
      BigRips run 1031
      AIDA on file R1_407

07:31 system wide checks all ok

07:48 Bias and leakage currents ok - attachment 9

08:01 Briken and BIgRips End current runs
      AIDA file R1_422 E silicon detector has saturation problem (see the attachment). We lower the
gain in round factor of 2. 