746
|
Tue Oct 2 01:55:11 2018 |
OH, TD | Tuesday 2 October | 09.54 DAQ continues OK - current file Sep18/R2_26
09.55 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachment 4
nnaida2 rate due to one hot channel
Merger OK - see attachment 5
TapeServer OK - see attachment 6
no errors or warnings shown by server terminal sessions
1.8.L spectra - see attachments 7 & 8
1.8.W spectra 20us FSR - see attachments 9 & 10
10.00 temperature +25.7 deg C, RH 29.6%, d.p. +6.4 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level c. 80%
10.03 ASIC check load (file Sep18/R2_26) |
|
747
|
Wed Oct 3 01:48:21 2018 |
OH, TD | Wednesday 3 October | 09.42 DAQ continues OK - current file Sep18/R2_36
09.42 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachment 4
nnaida2 rate due to one hot channel
Merger OK - see attachment 5
TapeServer OK - see attachment 6
no errors or warnings shown by server terminal sessions
1.8.L spectra - see attachments 7 & 8
1.8.W spectra 20us FSR - see attachments 9 & 10
09.46 ASIC check load (file Sep18/R2_36)
Note for p+n junction strip FEE64s the 1.8.W spectra are only updating for nnaida2, 10 & 17 - these traces
look OK
Other FEE64s show a noise transient and do not update - ASIC check load did not fix - probably requires
DAQ STOP/GO
09.57 temperature +25.4 deg C, RH 39.2%, d.p. +10.6 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level c. 80%
13:17 FEE temperatures and system wide checks all ok
Alpha run R2 stopped
R2_0 -> R2_37
13:18 Thresholds set to 0xa from 0x64 for all FEEs and ASICs
13:22 R3_0 started
Thresholds at 0xa and pulser on at 25Hz setting ~ 2Hz
ASIC settings saved to EXPERIMENTS/AIDA/2018Oct03-13.26.58
Merger screenshot - attachment 11
Stats - attachment 12
Layouts 7 and 8 - attachments 13 & 14
13:31 R3_0 Stopped
13:34 R4_0 Started
ASIC thresholds all at 0x9
Merger screenshot - attachment 15
Stats - attachment 16
Layout 1 - attachment 17
13:38 R4_0 Stopped
13:40 R5_0 Started
ASIC thresholds all at 0x8
Merger screenshot - attachment 18
Stats - attachment 19
Layout 1 - attachment 20
13:45 R5_0 Stopped
13:49 R6_0 Started
ASIC slow comp at 0x7
Merger screenshot - attachment 21
Stats - attachment 22
Layout 1 - attachment 23
13:52 R6_1 Stopped
R7_0 Started
ASIC Slow comp at 0x6
Merger screenshot - attachment 24
Stats - attachment 25
Layout 1 - attachment 26
Tapeserver window screenshot - attachment 27
Top window showing 72% CPU usage - attachment 28
15:06 Pulser measurements carried out with thresholds at 0xa
Typical pulser peak width was ~70-80 with a couple above 100.
Double peaking that was observed earlier in the week is no longer observed.
Both in the FEEs and on a scope
Backside pulser peaks - attachment 29
Frontside pulser peaks - attachment 30 |
|
748
|
Wed Oct 3 07:09:03 2018 |
OH, TD | Current FEE Configuration Map October 3rd 2018 |
FEE Number
DSSD A B C D
1 17 20 22 23
2 18 19 21 24
3 9 12 14 15
4 10 11 13 16
5 1 4 6 7
6 2 3 5 8
DSSD 1 3058-4 Top - Upstream
DSSD 2 3058-3
DSSD 3 2998-23
DSSD 4 3058-5
DSSD 5 3058-6
DSSD 6 3058-7 Bottom - Downstream |
|
623
|
Fri Jun 2 00:01:48 2017 |
OH, PW, CB | Friday 2 June 08:00-16:00 | 08:05 Briken and BigRIPS stopping daqs
AIDA on file R1_1099
08:07 BigRIPS starts run 1088
BRIKEN starts run 77
08:13 System wide checks all ok
FEE temperatures all ok - attachment 1
Good events statistics - attachment 2
08:50 Bias and leakage currents ok - attachment 3
09:04 BigRIPS and BRIKEN stop daqs
AIDA on file R1_1113
09:05 BigRIPS starts run 1089
BRIKEN starts run 78
AIDA on file R1_1113
09:22 Beam lost AIDA on file R1_1117
09:26 Beam back AIDA on file R1_1118
10:06 BigRIPS and BRIKEN stop daqs
AIDA on file R1_1128
10:07 BigRIPS starts run 1090
BRIKEN starts run 79
AIDA on file R1_1128
10:12 System wide checks ok
10:24 DSSD rates from R1_1131
*** DSSSD # 1 count: 88384 old count: 45978 dt: 1.01 s LEC rate: 42084.25 Hz
*** DSSSD # 2 count: 119845 old count: 62135 dt: 1.01 s LEC rate: 57272.13 Hz
*** DSSSD # 3 count: 40809 old count: 21150 dt: 1.01 s LEC rate: 19509.84 Hz
*** DSSSD # 4 count: 37059 old count: 18986 dt: 1.01 s LEC rate: 17935.87 Hz
*** DSSSD # 5 count: 14947 old count: 7772 dt: 1.01 s LEC rate: 7120.56 Hz
*** DSSSD # 6 count: 29809 old count: 15349 dt: 1.01 s LEC rate: 14350.29 Hz
*** DSSSD # 1 count: 35 old count: 16 dt: 1.01 s HEC rate: 18.86 Hz
*** DSSSD # 2 count: 108 old count: 56 dt: 1.01 s HEC rate: 51.61 Hz
*** DSSSD # 3 count: 77 old count: 38 dt: 1.01 s HEC rate: 38.70 Hz
*** DSSSD # 4 count: 76 old count: 35 dt: 1.01 s HEC rate: 40.69 Hz
*** DSSSD # 5 count: 27 old count: 15 dt: 1.01 s HEC rate: 11.91 Hz
*** DSSSD # 6 count: 4 old count: 2 dt: 1.01 s HEC rate: 1.98 Hz
11:07 BigRIPS and BRIKEN stop their daqs
AIDA on file R1_1142
11:08 BigRIPS starts run 1091
BRIKEN starts run 80
AIDA on file R1_1142
11:10 Leakages and biases ok - attachment 4
11.36 DAQ continues OK
Attached HEC x vs y - attach 5
Zero'd HEC x vs y
11.59 Analysed R1_1153 - see attachment 6
No timewarps
Dead time negligible
All OK
12:06 BigRIPS and BRIKEN stop their daqs
AIDA on file R1_1156
12:07 BigRIPS starts run1092
BRIKEN starts run 81
AIDA on file R1_1156
13:07 BigRIPS and BRIKEN stop daqs
AIDA on file R1_1170
13:08 BRIKEN starts run 82
BigRIPS starts run 1093
AIDA on file R1_1170
13:11 System wide checks ok
13:55 BRIKEN and BigRIPS stop their daqs
AIDA on file R1_1181
Beam gone
14:11 Bias and leakage currents ok - attachment 7
Ambient Temperature 23.9C - Humidity 37.3% - D.P +8.3
15:26 Beam is back
AIDA on file R1_1201
15:07 BigRIPS starts run 1094
BRIKEN starts run 83
AIDA on file R1_1201
15:34 System wide checks ok
FEE temperatures ok - attachment 8
Good Events statistics - attachment 9
15:39 No beam - beam back
15:46 Beam gone for couple of minutes
15:50 Beam back |
|
644
|
Thu Jun 8 00:54:26 2017 |
OH, PW | Thursday 8 June 00:00-16-00 | 08:54 DAQ still running
09:01 BRIKEN and BigRIPS stop daqs
AIDA on file R7_293
Beam stopped
09:25 System wide checks ok
Bias and leakage ok - attachment 1
Good event statistics - attachment 2
FEE temperatures - attachment 3
13:36 Beam about to be back
Briken on Run 65 (Background)
AIDA on file 350
13:37 Beam is returned
BRIKEN starts run 66
bigRIPS run 3069
AIDA on file R7_350
System wide checks ok
13:40 Bias and leakage ok - attachment 4
Good event - attachment 5
FEE temperatures - attachment 6
14:00 Beam down originally five minutes
14:08 Beam will be down for another 20 minutes
BRIKEN and BigRIPS stop daqs
AIDA on file R7_356
14:22 beam is back
14:23 BigRIPS starts run 3070
Briken starts run 67
AIDA on file R7_360
15.23 DAQ stop
AIDA file RIBF128/R7_373
15:26 BigRIPS starts run 3071
Briken starts run 68
AIDA on file R7_374
15.29 beam off - back in 5 mins
AIDA file RIBF128/R7_374
15.44 beam back
AIDA file RIBF128/R7_377
|
|
634
|
Mon Jun 5 16:02:50 2017 |
OH, PV, AE | Tuesday 06 June 00.00-08.00 | 00:03 BigRIPS starts run 3007
BRIKEN starts run 007
AIDA on file R5_146
00:07 Bias and leakage currents ok - attachment 1
System wide checks all ok
FEE temperatures ok - attachment 2
Good evts still seeing noise on 13 and 14 - attachment 3
No merger errors since GO
00:48 DSSD Rates from R5_160
*** DSSSD # 1 count: 250904 old count: 163089 dt: 1.02 s LEC rate: 86326.91 Hz
*** DSSSD # 2 count: 335185 old count: 222784 dt: 1.02 s LEC rate: 110496.28 Hz
*** DSSSD # 3 count: 157445 old count: 103991 dt: 1.02 s LEC rate: 52548.18 Hz
*** DSSSD # 4 count: 102079 old count: 67878 dt: 1.02 s LEC rate: 33621.44 Hz
*** DSSSD # 5 count: 40500 old count: 26651 dt: 1.02 s LEC rate: 13614.32 Hz
*** DSSSD # 6 count: 64652 old count: 42396 dt: 1.02 s LEC rate: 21878.85 Hz
*** DSSSD # 1 count: 0 old count: 0 dt: 1.02 s HEC rate: 0.00 Hz
*** DSSSD # 2 count: 41 old count: 27 dt: 1.02 s HEC rate: 13.76 Hz
*** DSSSD # 3 count: 161 old count: 103 dt: 1.02 s HEC rate: 57.02 Hz
*** DSSSD # 4 count: 131 old count: 85 dt: 1.02 s HEC rate: 45.22 Hz
*** DSSSD # 5 count: 22 old count: 12 dt: 1.02 s HEC rate: 9.83 Hz
*** DSSSD # 6 count: 9 old count: 6 dt: 1.02 s HEC rate: 2.95 Hz
01:02 Briken and BigRIPS daq stops
01:03 Briken Starts run 8
BigRIPS starts run 3008
01:40 Phong noticed a significant difference in the noise in the (raw) low energy spectra a few of the FEEs when
comparing the runs right before and right after the power-cycle (See Attachment 12) required to recover
the DAQ crash last afternoon (June 5th).
NNAIDA 22, 21, 14, and 13 had a much higher count rate after the power cycle, and what seems like a lower
threshold. We can't observe the pulser peak in any of those.
NNAIDA 13 has the highest count rate, and the waveforms show what looks like noise with a regular
frequency (attachments 4 - 7)
For reference, we include the waveform of other positive FEEs 8-11
02:51 Bias and leakages ok - attachment 13
System wide checks ok
Good Event statistics - attachment 14
Fee Temperatures - attachment 15
03:03 BigRIPS and BRIKEN stop daqs
AIDA on file R5_208
03:04 BigRIPS starts run 3009
BRIKEN starts run 9
AIDA on file R5_208
04:03 BigRIPS and BRIKEN stop daqs
AIDA on file R5_230
04:04 BigRIPS starts run 3010
BRIKEN starts run 10
AIDA on file R5_230
04:09 System wide checks all ok
Bias and leakages ok - attachment 16
Fee Temperatures - attachment 17
Good Event Statistics - attachment 18
04:48 DSSD rates from R5_244
*** DSSSD # 1 count: 1239207 old count: 1155424 dt: 1.04 s LEC rate: 80379.42 Hz
*** DSSSD # 2 count: 1684042 old count: 1567549 dt: 1.04 s LEC rate: 111760.61 Hz
*** DSSSD # 3 count: 806937 old count: 752064 dt: 1.04 s LEC rate: 52643.85 Hz
*** DSSSD # 4 count: 505814 old count: 472642 dt: 1.04 s LEC rate: 31824.43 Hz
*** DSSSD # 5 count: 208086 old count: 193231 dt: 1.04 s LEC rate: 14251.53 Hz
*** DSSSD # 6 count: 335686 old count: 313190 dt: 1.04 s LEC rate: 21582.13 Hz
*** DSSSD # 1 count: 8 old count: 7 dt: 1.04 s HEC rate: 0.96 Hz
*** DSSSD # 2 count: 193 old count: 181 dt: 1.04 s HEC rate: 11.51 Hz
*** DSSSD # 3 count: 872 old count: 810 dt: 1.04 s HEC rate: 59.48 Hz
*** DSSSD # 4 count: 637 old count: 601 dt: 1.04 s HEC rate: 34.54 Hz
*** DSSSD # 5 count: 89 old count: 81 dt: 1.04 s HEC rate: 7.68 Hz
*** DSSSD # 6 count: 71 old count: 65 dt: 1.04 s HEC rate: 5.76 Hz
05:01 BRIKEN and BigRIPS stop DAQs
AIDA on file R5_250
05:02 BigRips starts run 3011
BRIKEN starts run 11
AIDA on file R5_250
06:02 BigRIPS and BRIKEN stop their daqs
AIDA on file R5_272
06:03 BigRIPS starts run 3012
BRIKEN starts run 12
AIDA on file R5_272
06:06 System wide checks all ok
Bias and leakage currents ok - attachment 19
Good event statistics - attachment 20
FEE temperatures ok - attachment 21
06:16 Analysis of R5_270 - attachment 22
Three MBS errors
06:34 130Cd decay curve - attachments 22 and 23
Despite the large amounts of noise in the detectors it is still possible to observe decay curves. They are
cleaned up a lot
by applying a multiplicity cut (attachment 22 with attachment 23 without. Further cuts and vetos could be
applied to clean up further.
The fit in attachment 22 gives a half life of 182ms which is not quite the expected 162ms but a reasonable
first fit.
07.14 Layout - 1.8.W waveforms - see attachments 24-27
No significant change in intrinsic/extrinsic noise cf. previous screenshots
nnaida13 ASIC 1-4 waveforms - see attachments 28-31 - range 100*20us
07.24 BRIKEN and BigRIPS stop DAQs
AIDA on file R5_300
07.25 Stop AIDA DAQ
Cycle detector biases
Start AIDA DAQ
07.38 BigRips starts run 3014
BRIKEN starts run 14
AIDA on file R6_1
07.45 nnaida13 ASIC 1-4 waveforms - see attachments 32-35 - range 100*20us
08.07 all system wide checks OK |
|
642
|
Wed Jun 7 08:02:42 2017 |
OH, PV | Wednesday 7 June 16:00 - 24:00 | 16:02 System wide checks ok
Good Events statistics (No beam at time!)- attachment 1
Fee temperatures - attachment 2
Bias and leakage ok - attachment 3
16:29 BigRIPS starts run 3051
BRIKEN starts run 49
AIDA on file R7_80
Slits brought in to reduce rate to attempt implant-neutron correlation
16:39 DSSD rates from R7_81
*** DSSSD # 1 count: 182087 old count: 144548 dt: 1.02 s LEC rate: 36716.57 Hz
*** DSSSD # 2 count: 295938 old count: 235754 dt: 1.02 s LEC rate: 58865.45 Hz
*** DSSSD # 3 count: 80412 old count: 64397 dt: 1.02 s LEC rate: 15664.13 Hz
*** DSSSD # 4 count: 94793 old count: 75541 dt: 1.02 s LEC rate: 18830.22 Hz
*** DSSSD # 5 count: 23432 old count: 18547 dt: 1.02 s LEC rate: 4777.98 Hz
*** DSSSD # 6 count: 66315 old count: 53096 dt: 1.02 s LEC rate: 12929.39 Hz
*** DSSSD # 1 count: 0 old count: 0 dt: 1.02 s HEC rate: 0.00 Hz
*** DSSSD # 2 count: 8 old count: 4 dt: 1.02 s HEC rate: 3.91 Hz
*** DSSSD # 3 count: 23 old count: 20 dt: 1.02 s HEC rate: 2.93 Hz
*** DSSSD # 4 count: 38 old count: 32 dt: 1.02 s HEC rate: 5.87 Hz
*** DSSSD # 5 count: 13 old count: 10 dt: 1.02 s HEC rate: 2.93 Hz
*** DSSSD # 6 count: 11 old count: 8 dt: 1.02 s HEC rate: 2.93 Hz
17:13 BigRIPS and BRIKEN stop DAQS
AIDA on file R7_89
17:30 Beam back
BigRIPS starts run 3052
BRIKEN starts run 50
AIDA on file R7_93
17:52 Analysis of R7_93 No timewarps, no MBS timewarps - attachment 4
18:06 System wide checks all ok
Bias and leakage ok - attachment 5
Good events statistics - attachment 6
FEE temperatures ok - attachment 7
18:27 fRC problem, beam is off
18:29 BigRIPS and BRIKEN stop DAQS
AIDA on file R7_106
18:32 Beam is back
BigRIPS starts run 3053
BRIKEN starts run 51
AIDA on file R7_106
19:33 BigRIPS and BRIKEN stop daqs
AIDA on file R7_119
19:34 BRIKEN starts run 52
BigRIPS starts run 3054
AIDA on file R7_120
20:09 System wide checks ok
Bias and leakage ok - attachment 8
Good event statistics - attachment 9
Fee temperatures - attachment 10
20:28 BigRIPS and BRIKEN stop their DAQs
AIDA on file R7_131
20:30 BigRIPS starts run 3055
BRIKEN starts run 53
AIDA on file R7_131
20:41 Layout 3 - attachment 11
1.8.L spectrum for each FEE
Appears to be 4-5 pulser peaks
21:31 BRIKEN and BigRIPS stop their daqs
AIDA on file R7_144
21:32 BRIKEN starts run 54
BigRIPS starts run 2056
AIDA on file R7_144
22:03 System wide checks ok
Bias and leakage ok - attachment 12
Good events statistics - attachment 13
Fee temperatures - attachment 14
22:32 Briken and BigRIPS stop their daqs
AIDA on file R7_156
22:33 BigRIPS starts run 3057
BRIKEN starts run 55
AIDA on file R7_156
23:31 BigRIPS and BRIKEN stop their DAQs
AIDA on file R7_169
23:32 BigRIPS starts run 3058
BRIKEN starts run 56
AIDA on file R7_169
23:46 Analysis of R7_168 - attachmemt 15 |
|
544
|
Wed Mar 29 13:38:38 2017 |
OH, DK, CG | First beam runs | For BigRIPS run numbers, they are available online: http://bripscnt01.rarfadv.riken.jp/runsum/index.php
21:30 AIDA and BRIKEN are ready to go -- so we closed the door and asked the BigRIPS team to send the beam to
F11. It will take about 30 minutes as they want to recalibrate one of the Faraday cups.
22:09: Beam is coming to F11. AIDA sub-run R13_17 is running at 22:10 so nearly the same time.
BigRIPS run number 0140
BRIKEN file name 170329_2213_40Mg Run 0. It is started at 22:13. R13_18 for AIDA
See attachment 1 for biases.
23:12: We notice that BigRIPS began the run 0141 at 23:08
AIDA subrun R13_36 began about or near 23:07 or 23:08
BRIKEN run number 170329_2313_40Mg Run 1
All system-wide checks passed (excepting nnaida6 clock status which is a known issue)
Statistics rates look reasonable, except that the nnaida 16 23 24 are even higher than before when the beam
was off (~200k)
23:59 - AIDA crashed. Merger stopped. When attempting to stop the run, the following error is displayed:
STATE for nnaida1 returned with an error
error: SOAP http transport timed out after 20000 ms
NONE
error: SOAP http transport timed out after 20000 ms
while executing
"$transport $procVarName $url $req"
(procedure "::SOAP::invoke" line 18)
invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
("eval" body line 1)
invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
(procedure "DataAcquisitionControlClient__GetState" line 1)
invoked from within
"DataAcquisitionControlClient__GetState"
We kill off the merge and tape server.
DAQ itself comes back after a reset. Attempting to open the tape service, however, yields an error
Cannot issue more Access Tokens
ClaimTapeServer returned:-
3 Capability table is full.
30.3.2017
0:56 When attempting to bring AIDA back up, eventually nnaida10 was not showing merger data before enabling the
ASIC readout data. On the run control, nnaida10 consistently showed that its data transfer was not enabled.
Attempting to reset etc., then eventually issuing Go would drop all the data transfers.
After a power cycle by the standard method, essentially the problem persists, except that it is nnaida 7 8 9 10
(the whole block) which shows the problem where we cannot get the enabled data transfer to stick. |
|
549
|
Sat Apr 1 16:16:05 2017 |
OH, CG | Sunday 2 April | 00:12 BigRips started run 0223 (Started at 00:12)
AIDA file R20_344
System wide checks all ok
BRIKEN start run 67
01.38 Beam stopped to F11. ~1/2 way through R20_364.
01.43 Beam returned to F11. 1GB through R20_366.
BigRIPS file 0224.
BRIKEN run 68.
System-wise checks all ok, except nnaida1 global clocks failed 6.
Temps ok. Leakage currents ok (see attachment 1).
02.43 BigRIPS started run 0225. Start of R20_381
02.48 BRIKEN run 69 started. Start of R20_382.
03.28 System-wide checks same as previous. (R20_391)
03.43 BigRIPS file 0226. ~1/2 way through R20_394
03.48 BRIKEN run 70 started.
04.46 BigRIPS file 0227 started. End of R20_410.
04.48 BRIKEN run 71 started. Start of R20_411.
05.31 36Ne setting is over. End of R20_421.
BigRIPS team is changing setting for calibration. No beam will go through Zerodegree
05.51 R20 stopped (I was getting food).
06.30 20mm Al degrader + 3mm plastic removed from F8. 23mm Al degrader placed at F11 after IC.
06.35 BigRIPS file 0237 started - 38Mg setting.
06.37 R21 started on 38Mg setting.
06.57 Beam stopped to F11 - accelerator problem.
~700MB into R21_5.
Returned shortly after.
07.23 BigRIPS file 0245 started. ~3/4 way through R21_10.
07.26 BigRIPS file 0246 started. ~1/2 way through R21_11.
07.32 BigRIPS file 0247 started. Start of R21_13.
07.34 BigRIPS file 0248 started. ~1/2 way through R21_13.
07.38 BigRIPS file 0249 started. ~1GB into R21_15.
07.40 BigRIPS file 0250 started. Start of R21_16.
07.41 BigRIPS file 0251 started. Little way through R21_16.
08.06 38Mg setting finished. R21 stopped.
08.08 R22 started - got a bit trigger happy with the DAQ. No beam to F11.
08.10 Degrader set to 10mm Al.
08.16 BigRIPS file 0252 started on 39Al setting.
08.17 R23 started on 39Al setting.
08.20 R23 ended.
08.26 40Si setting. Too high rate for AIDA so beam won't be sent down ZDS.
08.40 F11 degrader set to 15mm Al fixed + 4mm Al variable (all in).
08.43 2nd run at 39Al setting.
BigRIPS file 0254 started.
R24 started.
09.00 R24_4 closed. May have been issues with degrader - not sure we were implanting in AIDA...
End of experiment!
09.07 AIDA background run started - R25.
Pulser is on in the run though.
09.20 R25 ended.
BigRIPS run summary attached. |
|
628
|
Sat Jun 3 08:07:48 2017 |
OH, CB, AE | Saturday 3 June 16:00-24:00 | 16:26 DAQ continues OK
all system wide checks ok, *except* nnaida1 clock status (6), nnaida1, 21 ADC calibration
Leakage currents OK - attachment 1
FEE64 temperatures OK - attachment 2
Good event statistics OK - attachment 3
No new merger errors - attachment 4
Sorted run R1_1540. DSSSD rates as follows
*** data items: 8958582 ( 115623.74 Hz)
*** ADC events: 954672 ( 12321.45 Hz)
*** time warps: 0 ( 0.00 Hz)
*** DSSSD # 1 count: 392147 old count: 361510 dt: 8.78 s LEC rate: 44646.94 Hz
*** DSSSD # 2 count: 557051 old count: 514085 dt: 8.78 s LEC rate: 63421.69 Hz
*** DSSSD # 3 count: 185346 old count: 172449 dt: 8.78 s LEC rate: 21102.12 Hz
*** DSSSD # 4 count: 154380 old count: 142180 dt: 8.78 s LEC rate: 17576.56 Hz
*** DSSSD # 5 count: 67195 old count: 62386 dt: 8.78 s LEC rate: 7650.32 Hz
*** DSSSD # 6 count: 131951 old count: 122163 dt: 8.78 s LEC rate: 15022.96 Hz
*** DSSSD # 1 count: 185 old count: 171 dt: 8.78 s HEC rate: 21.06 Hz
*** DSSSD # 2 count: 450 old count: 410 dt: 8.78 s HEC rate: 51.23 Hz
*** DSSSD # 3 count: 379 old count: 351 dt: 8.78 s HEC rate: 43.15 Hz
*** DSSSD # 4 count: 297 old count: 269 dt: 8.78 s HEC rate: 33.81 Hz
*** DSSSD # 5 count: 132 old count: 127 dt: 8.78 s HEC rate: 15.03 Hz
*** DSSSD # 6 count: 20 old count: 19 dt: 8.78 s HEC rate: 2.28 Hz
Pulser event still synchronised - see attachment 5
16:34 BigRIPS & BRIKEN stop runs
AIDA on file R1_1542
16:35 BigRIPS starts run 1120
BRIKEN starts run 109
AIDA on file R1_1543
Zero'd all SortSas
17:33 bigRIPS and BRIKEN stop daqs
AIDA on file R1_1556
17:34 bigRIPS starts run 1121
BRIKEN starts run 110
AIDA on file R1_1556
17:54 System wide checks all ok
19:50 bigRIPS starts run 1123
BRIKEN starts run 112
AIDA on file R1_1588
21:00 End of 78Ni run
last AIDA file R1_1606
Stopped AIDA DAQ
Refilled chiller
21:18 Performing pulser walkthrough.
Mistakenly started Tape server before starting DAQ. There will be no R2 run.
Starting R3 - pulser walkthrough.
Pulser frequency 250 Hz
Pulser amplitudes 90 000, 80 000, 70 000 ... 10 000
Can't see pulser peak in spectra. Issue unclear
21:40 Restarted high-amplitude pulser walkthrough run. Run RIBF127R1/R4_0
Pulser frequency 250 Hz
Pulser amplitudes 90 000, 80 000, 70 000 ... 10 000
Pulser peak now visible in histograms
Started low-amplitude pulser walkthrough run. Run RIBF127R1/R5_0
Pulser frequency 250 Hz
Pulser amplitudes 12 000, 10 000, ... 2 000
Set pulser amplitude back to 90 000, frequency to 25 Hz
23:17 Merger and Tape Server stopped
Links to FEEs disabled
All Browser tabs closed
Terminal sessions closed
23:25 Started R1 in /TapeData/RIBF128/
23:43 Start ASIC data readout
System wide checks ok
ASIC load check ok |
|
841
|
Fri Jun 14 10:33:44 2019 |
OH, CA, TD, CG, CB, DK | Friday 14th June | Friday 14th June
Authors: TD, DK, CB
08:40
elog appears to be down. Relevant screenshots are in /tmp/june14/ so please upload when we can access
Julabo chiller 75% M-H markers
Room Temp +24.5 C deg, dew point +8.1 deg C, Room humidity 35.2%
Pulser, ASIC settings as yesterday (copy/paste)
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
DAQ continues OK
Statistics as before. nnaida19 counter appears to have overflowed and wrapped around
See stats.png
System-wide checks OK, except nnaida7 fails ADC calibration as before
Biases and leak currents reasonable -- see biases.png
FEE64 temperatures okay, see fee-temps.png
Merger looks okay, ~600k items / sec; see merger.png
TapeServer as before, around 4.5 Mbytes/sec; see tapeserver.png
13:52
Pulser, ASIC settings same as this morning
DAQ continues OK
Statistics as before. See attach 190614_1354_stats. Counts RESET.
System-wide checks OK, except nnaida7 fails ADC calibration as before
Bias and leak currents OK. See attach 190614_1356_bias
FEE64 temperatures OK, see attach 190614_1357_temps
Merger keeps running at ~560k items /s; see 190614_1359_merger
TapeServer as before. See 190614_tape
17:24 AIDA DAQ crashed by DTAS touching the clock function generator
Function generator set at 50MHz square high +5V low 0V
18:38 problem loading ASIC settings - mismatch in slow comparator threshold and shaping time
performed save/restore with 2019Jun13-08.56.35 - all looks ok now
System wide checks ok except nnaida7 fails adc calibration
FEE64 temperatures ok - attachment 12
good event statistics ok - attachment 13
bias/leakage currents ok - attachment 14
18:38 for reals
DAQ Stopped
R30_0 opened writing to disk
Was to test DTAS sync whilst writing to file
18:43 DAQ stopped on file R30_0
18:45 DAQ restarted on file R31 with no writing to disk
21:50 DAQ found crashed (Valencia group believe it crashed around 20:00)
Trying to reset FEEs resulted in an error message attachment 15 and 16
Each FEE was rebooted using telnet to avoid a power cycle
System reset after reboot
System wide checks ok
Check ASIC control
FEE temps ok
DAQ Started again
On startup FEEs 17, 20, 22 and 23 were all about 150k
All relate to DSSD0
Pressed on kapton and noise went down.
Statistics now good
22.45 Review /var/log/messages and observe nnaida15 rebooted at 19.27 - reason unknown
This probably explains why DAQ stopped.
Jun 14 19:27:32 aidas1 dhcpd: DHCPDISCOVER from 00:04:a3:2a:f3:ac via p2p2
Jun 14 19:27:32 aidas1 dhcpd: DHCPOFFER on 10.1.2.15 to 00:04:a3:2a:f3:ac via p2p2
Jun 14 19:27:32 aidas1 dhcpd: DHCPREQUEST for 10.1.2.15 (10.1.2.250) from 00:04:a3:2a:f3:ac via p2p2
Jun 14 19:27:32 aidas1 dhcpd: DHCPACK on 10.1.2.15 to 00:04:a3:2a:f3:ac via p2p2
Jun 14 19:27:32 aidas1 rpc.mountd[2868]: authenticated mount request from 10.1.2.15:784 for
/data10/XilinxLinux/ppc_4xx/rfs/nnaida15 (/data10/XilinxLinux/ppc_4xx/rfs)
Jun 14 19:27:59 aidas1 rpc.mountd[2868]: authenticated mount request from 10.1.2.15:864 for /MIDAS@aidas
(/MIDAS@aidas)
22.53 ADC data synchronised - attachment 17
23.05 zero DAQ statistics
zero all histograms
23.20 Check ASIC Control x1
FEE64 Rate spectra - attachment 18
1.8.L spectra - see attachments 19 & 20
nnaida1 81 ch FWHM, nnaida3 77 ch FWHM
p+n FEE64s 1.8.W spectra 20us FSR - attachment 21
n+n FEE64s 1.8.W spectra 20us FSR - attachment 22
23.36 DAQ start file May19/R34_0
DTAS+AIDA correlation test using gamma ray sources
Correlation scaler reset issued during R34_0
ASIC settings 2019Jun13-08.56.35
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
23.53 DAQ stop file May19/R34_2
nnaida15 did not stop - state unknown - reboot nnaida15
This did not work - unable to reset nnaida15
Closed all FireFox broweser tabs, stopped all servers and restarted
00.14 DAQ start
ASIC settings 2019Jun13-08.56.35
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse |
|
750
|
Fri Oct 26 03:28:10 2018 |
OH, CA | Friday 26 October | October 25th:
In anticipation of losing access in the evening on arrival at RIKEN the following was done:
- Chiller was refilled
- All power cables reconnected after being removed in October for the RIKEN power down
- All NIM bins, fan trays and RPis were turned on
- 50MHz AIDA clock was set up and connected
- Remote connections to AC Mains relay and RPi for CAEN bias was tested
- Chiller was powered up
October 26th:
11:31 Still have access to the F11 area due to a bigRIPS delay.
Double checked wiring from yesterday and all still seems good
Chiller is still running with set point at +20 deg C
11:36 FEEs powered up. All 24 mounted
11:57 Bias and leakage currents all ok - attachment 1
FEE temperatures all ok - attachment 2
System wide checks all ok
13:26 Bias and leakage currents ok - attachment 3
FEE temperatures all ok - attachment 4
System wide checks all ok
14:00 Stats are ok - attachment 5
Appear similar to last time: https://elog.ph.ed.ac.uk/AIDA/747
However FEE24 and 24 appear much noisier than last time with rates around 100k
19:24 Pulser walkthrough run from 90000 to 10000 in steps of 10000 for R2
Opened file Oct18/R3, alpha run, threshold at 100kev, pulser on.
|
|
863
|
Mon Apr 19 15:10:55 2021 |
OH TD | Less HEC events than expected in DSSD3 | we are noticing we appear to see less implant events in detector 3 than we expect. This was first noticed in the xy spectra of the DSSD in the Go4 online.
As part of testing we have lowered the threshold in aida09 which from the xy appeared to have less events.
Following this we still seemed to see less events.
OH downloaded a selection of recent files to analyse. Looking at the total number of HEC events it appears that aida09, aida10 and aida12 are all seeing less HEC events than I would expect.
The ratio of HEC events in aida09 to aida11 was 0.36.
Looking at the energies of the HEC events there appeared to be less events everywhere and not just a threshold issue. We thought it could then be radiation damage and poor
charge collection. To test this we raised the bias to 120V on DSSD3.
The same test was again performed and the ratio again came out to be 0.36
We did not appear to recover any HEC events
Comparing the scaled energy distributions of aida09 and aida11 we see no evidence of a different energy response other than an increase in lower energy events in aida09 due to the lower threshold
I now think it may be a hardware issue. Will look at the last file before the most recent powercycle.
Using TD's analyser program can confirm the rate between aida09 and aida11 before 08:30 this morning was ~1:1.
The problem seems to have appeared during the power cycles following this. |
|
782
|
Tue Nov 13 22:58:49 2018 |
OH | Wednesday 14 November 08:00 - 16:00 | 08:01 BigRips and BRIKEN stop
AIDA on file R5_346
08:02 BigRIPS starts run 1038
BRIKEN starts run 181114_0800_165Pm_053.dlt
AIDA on file R5_346
System wide checks ok * except nnaida 7 fails adc*
FEE Temps ok - attachment 1
Stats ok - attachment 2
Leakage currents ok - attachment 3
09:15 BigRIPS and BRIKEN stop
AIDA on file R5_358
09:16 BigRIPS starts run 1039
BRIKEN starts run 181114_0915_165Pm_054.dlt
AIDA on file R5_358
10:15 BigRIPS and BRIKEN stop
AIDA on file R5_368
10:15 BigRIPS starts run 1040
BRIKEN starts run 181114_1015_165Pm_055.dlt
AIDA on file R5_368
10:36 Analyser output of R5_368 - attachment 4
11:15 BugRIPS and BRIKEN stop
AIDA on File R5_378
11:16 BigRIPS starts run 1041
BRIKEN starts run
AIDA on file R5_378
08.03 DAQ continues - AIDA file RIBF148/R5_384
165Pm BigRIPS setting + 2.8mm Al degrader
ASIC settings 2018Oct03-13.26.58
from c. 09.00 13.11.18
n+n FEE64s Ibias LF feedback 0xb -> 0xd
HEC fast comparator threshold 0x20 -> 0xff
11.50 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 5 & 6
FEE64 temperatures OK - see attachment 7
good event & AIDA discriminator statistics OK - see attachments 8
Merger OK - see attachment 9
TapeServer OK - see attachment 10
no errors or warnings shown by server terminal sessions
per FEE64 Rate spectra - attachment 11
per FEE64 HitRate spectraq - attachment 12
1.8.L spectra - attachments 13-16
some evidence of pulser peak shifts, notably nnaida5, 11 & 15
nnaida1 pulser peak width 78 ch FWHM
1.8.H spectra - attachments 17 & 18
1.8.W spectra 20us FSR - attachments 19 & 23
1.8.W spectra 200us FSR - attachments 20 & 24
1.8.W spectra 2ms FSR - attachments 21 & 25
1.8.W spectra 20ms FSR - attachments 22 & 26
12.15 BugRIPS and BRIKEN stop
AIDA on File R5_388
12.16 BigRIPS starts run 1042
BRIKEN starts run 57
AIDA on file R5_388
12.22 *all* online and MIDASsort spectra zero'd
13:15 BigRIPS and BRIKEN stop
AIDA on file R5_398
13:16 BigRIPS starts run 1043
BRIKEN starts run58
AIDA on file R5_398
13:28 BigRIPS and BRIKEN stop
AIDA on file
beam halted to give access to F11 area
13.44 beam started
BigRIPS and BRIKEN start
AIDA file RIBF148/R5_402
BRIKEN run 59
BigRIPS run 1043
14.45 BRIKEN and BigRIPS stop
AIDA file RIBF148/R5_412
14.47 BRIKEN and BigRIPS start
AIDA file RIBF148/R5_413
BRIKEN run 60
BigRIPS run 1044
15.43 BRIKEN and BigRIPS stop
AIDA file RIBF148/R5_422
BRIKEN Resets the timestamp
15:44 BRIKEN and BigRIPS start
AIDA file RIBF148/R5_422
BRIKEN run 61
BigRIPS run 1045
16:02 System wide checks all ok *Except nnaida7 fails adc calibration*
Leakage currents ok - attachment 27
Stats ok - attachment 28
FEE Temps ok - attachment 29
16:27 First 200ms of decay correlations in DSSD 3 and 4 for the 3 degrader settings - attachment 30
DSSD 3 along the top row and DSSD 4 along the bottom.
Dgrader settings are 3mm, 2.3mm and 2.8mm
1 hour data in each |
|
424
|
Sat Nov 5 09:40:02 2016 |
OH | Updated Leakage Current Spreadsheet | Attached is the updated leakage current spreadsheet (Attachment 1) as pf 5/11/2016 it is up to date with the elog.
Also added is the recording of ambient temperature. This is recorded in degrees C and plotted relative to 21 degrees C to better show on the same plot as leakage currents. (Screenshot in attachement 2)
Also attached is the spreadsheet in csv formatting. |
|
427
|
Sun Nov 6 05:16:22 2016 |
OH | Update | Leakage current spreadsheet has now been uploaded to google docs for ease of access.
It can be accessed and edited from the following link: https://docs.google.com/spreadsheets/d/12hgbrywB10hGsKt5uc2HnLfZymh8_uvM6cEASbbqnBE/edit?usp=sharing |
|
532
|
Thu Feb 2 13:44:02 2017 |
OH | T9 Test Bench | T9 test setup at Daresbury was powered up and noise levels were checked using a pulser peak.
A 1V peak was used. For most channels this corresponded to a peak position at around +-11000 from the zero point at 9000.
An example of this is shown in attachment 1 showing the peak in nnaider 11, ASIC 4, ADC 11.
It was also observed that for both nnaider 11 and 12 (FEE modules corresponding to negative signals) that channels 2, 56 and 59 were not contributing to the rate. See attachment 2.
Upon investigating the missing channels ADC spectra the only apparent issue appeared in nnaida 11 ASIC 4, ADC 8. Which is plot 4.7.L. Here we see the peak much higher coming in at around channel 63000. See attachment 3.
The remaining missing channels in the rate plot appear to have normal amplitudes with maybe slightly higher noise. FWHM of ~60 as oppposed to ~18. Discussed more below.
Also done was a noise measurement across the all the ASICs. To do this an ADC was chosen belonging to each of the ASICs and the pulser peak location and FWHM was recorded. The reaults are as shown below.
FEE# ASIC/ADC Centroid Location FWHM
11 1.8 42951 19.38
11 2.8 41012 17.48
11 3.8 41929 20.01
11 4.8 41302 17.88
12 1.8 41830 17.38
12 2.8 41119 16.01
12 3.8 42669 18.30
12 4.8 41811 15.21
13 1.8 24329 15.05
13 2.8 24034 13.64
13 3.8 24209 15.79
13 4.8 24607 13.29
14 1.8 24650 16.19
14 2.8 24618 14.12
14 3.8 23934 15.04
14 4.8 22584 14.36
14 4.8 22584 14.36
It was seen that the noise was fairly low which is to be expected as there were no dtectors connected and only a short length of kapton cables connected to the board. |
|
Draft
|
Wed Apr 5 06:10:55 2017 |
OH | BigRIPS particle ID | |
|
553
|
Wed Apr 5 06:11:08 2017 |
OH | BigRIPS particle ID | I've been going through collecting the BigRIPS root files with Phong and making sure they are on AIDAs1.
The files we have for NP1306 are not the standard format so to get particle ID you need to do this command taken
from the RIKEN elog.
Command to construct the PID in the merged files from preliminary data:
Example file a11111_b1181_e_1130_ion.root
tree->Draw("(F11ICADC[0]+F11ICADC[1]+F11ICADC[2]+F11ICADC[3]+F11ICADC[4]+F11ICADC[5])/6:F11Pla_TM-F3Pla_TM+F5X*0.0383261>>h(500,0,25,500,0,16000)","","colz");
Though these root files were not created by the BigRIPS team so can be cleaned up further but should be good
enough for initial analysis. |
|
615
|
Tue May 30 16:05:12 2017 |
OH | Wednesday 31 May 00:00 - 08:00 | 00:05 BigRIPS and BRIKEN stop their daqs
AIDA currently on file R1_312
00:06 BRIKEN start run 23
BigRIPS start run 1024
AIDA on file R1_312
System wide checks all ok
01:05 BigRIPS and BRIKEN stop their daqs
AIDA on file R1_327
01:06 BigRips Start run 1025
BRIKEN start run 24
AIDA file R1_327
System wide checks all ok
02:02 BigRIPS and BRIKEN stop their daqs
AIDA on file R1_341
ASIC load check ok
02:03 BigRIPS start run 1026
BRIKEN start run 25
AIDA ON FILE R1_341
02:11 System wide checks all ok
Leakages and bias ok (Leakages rising again though) - attachment 1
FEE temperatures ok - attachment 2
Good event statistics - attachment 3
Merger and tape server still running - attachment 4
02:23 Beam is off. Will recover soon
02:26 Beam is back
02:27 beam stopped again
02:28 beam back
02:37 Since beam has come back we appear to have lost all implant channels.
only seeing events in DSSD2
03:30 ASIC load issue resolved. AIDA data should be good from R1_353
Beam continues to turn off and on
03:52 Since ~2am beam has been passing through the dssd
Implantation spectrum - attachment 5
Upon investigation the fixed degrader had fallen on the left and was at a 30 degree angle.
It has now been put back in position
AIDA file when fixed R1_360
BRIKEN reset the timestamp
Whilst sorting the ASIC issue bigRips moved onto run 1029
BRIKEN run 1028. Full details in their elog
03:55 briken and bigRIPS stop runs
bigRIPS start run 1028
BRIKEN run 27
AIDA file R1_360
04:29 Implant hit spectra after the degrader was replaced - attachment 6
04:58 BigRIPS and BRIKEN stop daqs
AIDA on file R1_376
04:59 BigRIPS starts file 1029
BRIKEN starts run 28
AIDA at very end of file R1_376
05:20 System wide checks ok
bias and leakage currents ok - attachment 7
05:26 Beam still dropping out for short moments of time
06:03 BigRips and BRIKEN stop their runs
AIDA on file R1_393
06:04 BRIKEN start run 29
BigRIPS start run 1030
AIDA on file R1_393
06:28 Beam lost
06:31 Beam returned
06:46 System wide checks ok
FEE temperatures ok - attachment 8
07:01 BigRIPS and BRIKEN stop their runs
AIDA on file R1_407
07:02 BRIKEN start run 30
BigRips run 1031
AIDA on file R1_407
07:31 system wide checks all ok
07:48 Bias and leakage currents ok - attachment 9
08:01 Briken and BIgRips End current runs
AIDA file R1_422 E silicon detector has saturation problem (see the attachment). We lower the
gain in round factor of 2. |
|