Mon Apr 25 10:11:40 2016, CG, KM, PV & TD, Monday 25 April
|
AIDA stand, Julabo FL11006 recirculating chiller (with 110VAC/50Hz - 240VAC/50Hz 3ph transformer)
moved from F11 downstream wall to ~2m downstream of EURICA
|
Mon Apr 25 08:58:25 2016, TD, Report - high? - aidas1 - Binding NIS service FAILED
|
During a reboot of aidas1 the message
Binding NIS service FAILED
|
Thu Apr 14 13:56:17 2016, PJCS, Updating the FEE64 firmware.
|
To update the FEE64 firmware:-
Power-up the FEE64 modules that you wish to update.
Identify the file to be used ( FEE_GF_Feb16_15.bin at 14/4/2016
)
This file should be in the Server ( e.g. aidas1 ) at /MIDAS/Aida.
Edit the file /MIDAS/Aida/FlashPgm.csh and change the file |
Tue Mar 22 10:59:52 2016, TD : PJCS, Tuesday 22 March 7x
|
MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=5.515uA, ambient temperature +19.8 deg C
|
Mon Mar 21 14:37:41 2016, PJCS, Vernier Timing document for comment
|
Here is a document about the FEE64 waveform Vernier Timing.
Please comment/criticise/offer advice.
|
Thu Mar 17 10:46:30 2016, TD, Analysis of R45 8x
|
See https://elog.ph.ed.ac.uk/AIDA/179
207Bi source, no pulser
|
Tue Mar 15 11:15:42 2016, TD, Tuesday 15 March
|
MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=5.845uA, ambient temperature +19.8 deg C
|
Mon Mar 14 11:27:47 2016, PJCS, ASIC_clock synchronisation across the system
|
During 'SETUP' the ASIC readout clock ( 500KHz ) is synchronised to the SYNC pulse in V8.10 and later versions of the VHDL.
Attached is a short run for verification.
|
Fri Mar 11 15:31:53 2016, TD, Analysis of R34_0
|
lancre> ./analyser R34_0
*** GREAT format 3.2.0 analyser - TD - May 2014
*** ERROR: READ I/O error: 5002
|
Tue Mar 8 16:29:07 2016, TD, Report - high - AIDA timestamps
|
Version 8 firmware
Analysing R37_* data files (see https://elog.ph.ed.ac.uk/AIDA/160)
|
Thu Mar 10 16:35:39 2016, TD : PJCS, Report - high - AIDA timestamps
|
> Version 8 firmware
>
> Analysing R37_* data files (see https://elog.ph.ed.ac.uk/AIDA/160)
|
Tue Mar 8 15:09:49 2016, TD, Report - high - LEC/MEC fast comparator issues 6x
|
Fast comparator rate spectra and statistics do not appear to working as expected. See enclosed screenshots.
Have the hit/rate spectra and statistics been updated to reflect the firmware version 8 data format? |
Wed Mar 9 17:31:58 2016, TD, Report - high - LEC/MEC fast comparator issues
|
>
> Fast comparator rate spectra and statistics do not appear to working as expected. See enclosed screenshots.
> Have the hit/rate spectra and statistics been updated to reflect the firmware version 8 data format?
|
Wed Mar 9 16:18:01 2016, TD, nndhcp021 AIDA server Httpd/TapeServer/Merger startup
|
Starting Httpd for the TapeServer/Merger, TapeServer and Merger for AIDA produces
three terminal windows which will display the following text messages
|
Wed Mar 9 12:16:01 2016, TD, Wednesday 9 March
|
MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=6.505uA, ambient temperature +19.8 deg C
|
Tue Mar 8 15:40:57 2016, TD, Tuesday 8 March
|
MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=6.705uA, ambient temperature +19.8 deg C
|
Tue Mar 8 15:32:53 2016, TD, Report - low - run control frame issue with multiple streams
|
The run control frame does not appear to update the data transfer 'X' tag
for all FEE modules after selecting 'Enable #1 All' |
Tue Mar 8 06:00:45 2016, CG, AE, RIKEN maintenance progress
|
07/03/16
The new snout doesn't fit due to the screw holes being misaligned. I am going to try modifying the old one (that was in conflict with some material
in the base) over the next couple of days. It is too long by 11mm, so if I can remove that it should then fit.. |
Tue Mar 8 12:11:56 2016, CG, AE, RIKEN maintenance progress
|
Quote:
07/03/16 |
Wed Feb 24 14:53:41 2016, PJCS, First data with two streams merged from one FEE64
|
nnaida2 running Version 8 Firmware and data acquisition software outputting both and ASIC data stream and a Waveform data stream to the Merge process.
Merged data is written to file ... attached.
The waveform is running at 100Hz and ASIC data is running at 10Hz. |
Wed Feb 24 16:13:24 2016, PJCS, First data with two streams merged from one FEE64
|
Quote:
nnaida2 running Version 8 Firmware and data acquisition software outputting both and ASIC data stream and a Waveform data stream to the Merge |
Thu Feb 25 12:47:43 2016, PJCS, First data with two streams merged from one FEE64
|
My GREAT data analyser program reports that the 'fail' bit is set ?
|
Wed Mar 2 11:59:48 2016, PJCS, First data with two streams merged from one FEE64
|
The Fail bit is used to indicate The Timing Vernier word.
I'll post a seperate Elog entry to explain the format.
|
Wed Feb 24 15:46:11 2016, PJCS, OR64 - OR of all the ASIC discriminator channels after application of a mask
|
Firmware version8 number 8 has an OR64 added which can be routed, using the Local control, to the Trigger output and hence seen at the front panel Fast
NIM LEMO00 connector.
The OR64 is the OR of all the ASIC Discriminators. Each rising edge of a discriminator signal will give a 10ns pulse on the Fast NIM signal. If |
Wed Feb 24 13:54:53 2016, TD, Report - low - FEE temperature plot displays incorrect dates
|
See Attachment 1
|
Wed Feb 24 10:25:35 2016, TD, Wednesday 24 February 2016
|
MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=5.770uA, ambient temperature +19.8 deg C
|