AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA, Page 37 of 46  ELOG logo
Entry  Fri May 27 06:38:27 2016, TD, AE & CG, Friday 27 May 2016 13x
14.30 DSSSD stack moved upstream to position DSSSD#3 at centre of EURICA.

      Initial tests of detector bias with DSSSD #1-6 were Ok. Unfortunately we were unable to bias
      DSSSD #7 obtaining either open or short circuit depending on which of the two p+n junction 
      side and n+n ohmic side detector bias bus lines we connected to.

      Demonstrated that short circuit was upstream, of AIDA adaptor PCB by disconnecting Kapton
      LH couplers.

      Remove DSSSD stack and replace DSSSD #7. In progress.

15.50 Startup AIDA with DSSSDs disconnected

      Temperatures & firmware version - see attachment 1
      AIDA SYNCs - attachment 2
      System wide checks - attachments 3-6 - note persistent nnaida19 ASIC clock and SYNC errors
      Statistics nnaida19, 20 - attachments 7-8
      
      nnaida19, 20, 27 & 28 hit pattern and waveform spectra - see attachments 9-13

      Some nnaida19 system wide errors above and noisy ASIC#4 hit channel for nnaida28 - otherwise OK 
Entry  Fri Oct 26 03:28:10 2018, OH, CA, Friday 26 October 181026_1150_bias.png181026_1151_FEETemp.png181026_1334_Bias.png181026_1335_FEETemp.png181026_1400_Stats.png
October 25th:
In anticipation of losing access in the evening on arrival at RIKEN the following was done:
- Chiller was refilled
- All power cables reconnected after being removed in October for the RIKEN power down
- All NIM bins, fan trays and RPis were turned on
- 50MHz AIDA clock was set up and connected
- Remote connections to AC Mains relay and RPi for CAEN bias was tested
- Chiller was powered up

October 26th:

11:31 Still have access to the F11 area due to a bigRIPS delay.
      Double checked wiring from yesterday and all still seems good
      Chiller is still running with set point at +20 deg C 

11:36 FEEs powered up. All 24 mounted

11:57 Bias and leakage currents all ok - attachment 1
      FEE temperatures all ok - attachment 2
      System wide checks all ok

13:26 Bias and leakage currents ok - attachment 3
      FEE temperatures all ok - attachment 4
      System wide checks all ok

14:00 Stats are ok - attachment 5
      Appear similar to last time: https://elog.ph.ed.ac.uk/AIDA/747
      However FEE24 and 24 appear much noisier than last time with rates around 100k

19:24 Pulser walkthrough run from 90000 to 10000 in steps of 10000 for R2
      Opened file Oct18/R3, alpha run, threshold at 100kev, pulser on.
      
Entry  Fri May 26 02:20:42 2017, CB, OH, NN, Friday 26 May 15x
10.20 DAQ continues OK 
      Current file May2017/R27_22

      Detector biases; leakage currents OK - see attachment 1
      FEE64 Temperatures OK - see attachment 2
      good events; statistics OK - see attachment 3
      No warnings/errors reported by merger since DAQ GO - see attachment 4

      *all* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (=6)

10.30 E17/F11 ambient temperature +24.2 deg C, RH 39.7%, d.p. +9.7 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)

      Left DAQ running on alpha background calibration run
      
      ttyUSB7 log (nnaida1) attached
      GUI got stuck while attempting to drag-and-drop ttyUSB7 to ELOG to upload. Terminal still working and
confirmed Merger working OK.
      Killed nautilus process to get GUI unstuck. Uploaded ttyUS7 via Browse -> Upload button on ELOG instead.

15.51 DAQ continues OK 
      Current file May2017/R27_28

      Detector biases; leakage currents OK - see attachment 6
      FEE64 Temperatures OK - see attachment 7
      good events; statistics OK - see attachment 8
      No warnings/errors reported by merger since DAQ GO - see attachment 9

      *all* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (=6)

15.53 E17/F11 ambient temperature +24.1 deg C, RH 40.1%, d.p. +9.8 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)

      Left DAQ running on alpha background calibration run
      
      ttyUSB7 log (nnaida1) attached

19.20 DAQ continues OK 
      Current file May2017/R27_31

      Detector biases; leakage currents OK - see attachment 11 & 12
      FEE64 Temperatures OK - see attachment 13
      good events; statistics OK - see attachment 14
      No warnings/errors reported by merger since DAQ GO - see attachment 15

      *all* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)

19.30 E17/F11 ambient temperature +24.2 deg C, RH 38.5%, d.p. +9.2 deg C
Entry  Fri Aug 26 02:10:31 2016, TD, Friday 26 August 28x
10.11 R9 continues

      ASIC settings 2016Aug20-15.34.53
      Discriminator masks -> 0x0
      From Correlation Status & Control tab change Trigger delay from 0x180 to 0x5

      BNC PB-4 Pulser
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns tau_d 50us 
      - polarity
      INT ref, delay MIN, tail pulse
      External (random) trigger from EG&G Ortec 462 Start output (Range 81.92us) - frequency c. 120Hz

      B2F/F11 temp +24.8 deg C, d.p. +12.9 deg C, RH 46.8%

      DSSSD bias/leakage current OK - see attachments 1-2
      FEE64 temperatures OK - see attachment 3

      Some error messages observed for merger - see attachment 4-12
      All related to Q5 (nnaida6)

      System-wide checks OK - see attachments 13-18

      AIDA SYNC, PAUSE, RESUME, Correlation L, SYNC time warp, Good Events, nnaida6, nnaida5 & nnaida20
statistics - see attachments
19-28

      Analysis of R9_561 shows data synchronised - see attachmen

      FEE # versus timestamp (0.1ms / channel)
      Note nnaida19 not currently configured

10.49 R9 stops
Entry  Fri Nov 25 07:33:44 2016, TD, Friday 25 November 9x
16.35 DSSSD bias & leakage currents OK - see attachments 1 & 2
      note - the significant increase observed yesterday appears to be plateau'ing
      FEE64 temperatures OK - see attachment 3

17.10 Rate, HitRate, 1.8.L & 1.8.W spectra - see attachments 4-9
Entry  Fri Nov 24 10:36:06 2023, NH, Friday 24 November Screenshot_from_2023-11-24_11-47-29.pngScreenshot_from_2023-11-24_11-47-39.pngScreenshot_from_2023-11-24_11-48-14.pngScreenshot_from_2023-11-24_12-00-33.png
Pulser LEMOs removed from p+n fees
Pulser connected to n+n fees: aida02 and aida04 
BNC PB-5 Polarity --> NEGATIVE

HV and FEE power up

Rates look similar to usual (Attachments 1 and 2)

Turn on pulser, see them in n+n spectra as expected  (attachment 3) 

Pulser run R6

See peaks in ADC spectra (attachment 4)

Remove pulser from aida02, left on aida04 with 50Ohm terminator 

AIDA powered off, HV off
Entry  Fri May 24 03:04:26 2019, TD, Friday 24 May 15x
11.04 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 75% M-H markers
      Temperature +26.1 deg C dew point +4.2 deg C RH 24.4%

      DAQ continues OK to file May19/R9_5

      alpha background

      ASIC settings 2019May19-17.54.03
      slow comparator 0x64

      BNC PB-4 Pulser OFF

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3
      good event statistics OK - see attachments 4

      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      1.8.L spectra - see attachments 7 & 8
      p+n FEE64s 1.8.W spectra 20us FSR - attachment 9
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 10

14.17 analysis of May19/R9_5 - see attachment 11
      deadtime OK *except* nnaida22 
      ADC data for FEE64s nnaida5 & nnaida6 hot channels (Stat spectrum channel # 104)
      Also observed at a much lower rate for nnaida1 & nnaida2
      Statistics imply nnaida22 running at high rate for some part of R9_5

      analysis of May19/R9_4 - see attachment 12
      No indication of higher rates for nnaida22

11.19 Check ASIC Control *all* ASICs & FEE64s

11.29 zero DAQ statistics

14.42 DAQ stop file May19/R9_6

      Run stopped to move AIDA out of DTAS prior to changes to Pb shielding upstream of (CH2)n shield

15.01 AIDA removed from inside DTAS

      Slow comparator 0x64 -> 0xa

      Check ASIC Control *all* ASICs & FEE64s

      good event statistics OK - see attachments 13
      13x < 20k, 22 < 50k, 0 > 100k

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 14
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 15

      all system wide checks OK *except* nnaida7 fails ADC calibration

      zero all histograms

      Slow comparator 0xa -> 0x64

      Check ASIC Control *all* ASICs & FEE64s

15.08 DAQ start file May19/R10

      nnaida1 'good event' rate changed from c. 380Hz to 0Hz at Start ASIC Readout - DAQ stopped

15.10 DAQ start file May19/R11  

      alpha background

      ASIC settings 2019May19-17.54.03
      slow comparator 0x64

      BNC PB-4 Pulser OFF
Entry  Fri Jun 22 02:50:37 2018, OH, TD, Friday 22 June 2018 60.png70.png71.pngstartup.tcldhcpd.conf
10:50 Yesterday 4 FEE modules installed for FEE64s 25-32
      dhcpd.conf updated to reflect MAC addresses of new FEE64s

      AIDA config file updated
      located at /MIDAS/config/TclHttpd/aidas1/

      Firmware updated to /MIDAS/Aida/FEE_GF_Nov16_3.bin
      Using run control GUI failed for nnaida28 - see attachment 1
      Re-run manually OK

      FEE64 temperatures OK, all firmware versions OK - see attachment 2

      Power cycle all FEE64 mounted. 5 minute delay in nnaida30 being mounted

      Updated ASIC, DISC (but not WAVEFORM) settings for FEE64s 25-32
        DB key 2018Jun22-11.28.18

      Update Options for FEE64s 25-32 (now same as FEE64s 1-24)

      Power cycle OK

      Histogramming tag (H) drops for FEE64s 25-32 at DAQ GO - see attachment 3 

      *all* FEE64 statistics (except histogram-related counters) OK

      AIDA config & dhcpd.conf files attached - see attachments 4-5
Entry  Fri Oct 21 01:41:54 2016, TD, Friday 21 October 14x
09.38 DAQ (R66) continues OK

      Detector bias and leakage currents OK - see attachments 1&2
      FEE64 temperatures OK - see attachment 3

      System wide checks SYNC count & SYNC errors fail - see attachments 4-7 - other checks OK

      R66_367 ADC data synchronised

      AIDA SYNC stats - attachment 8

      Some merger errors - see attachment 9

13.53 DAQ found stalled
      Most recent files R66_396 last accessed 10.55

      System wide checks SYNC count & SYNC errors fail - see attachments 10-13 - other checks OK

      AIDA SYNC stats - attachment 14

      No further merger error messages

      Restart merger OK
      DAQ STOP OK

      DAQ RESET & SETUP OK
      ReSYNC OK
      System-wide checks OK
      Disable H & data transfer #1
      Restart merger & tapeserver
      Enable H & data transfer #1      

14.02 DAQ starts (R67)

18.27 DAQ found stalled
      Most recent file R67_121 last accessed 18.26

      Restart merger
      DAQ STOP
      System wide checks 
      DAQ RESET & SETUP
      System-wide checks

18.39 DAQ start (R68)
      Disable H & data transfer #1
      Restart merger & tapeserver
      Enable H & data transfer #1

18.44 DAQ stalled
      Most recent file R68_1 last accessed 18.44

      Restart merger
      DAQ STOP
      System wide checks 
      DAQ RESET & SETUP
      System-wide checks

18.53 DAQ starts (R70)
Entry  Fri Jun 21 01:08:35 2019, CA, CB, Friday 21 June 21junSync.png20june_2319_temps.png21jun_bias.png
09.08 all system wide checks okay *except* naida7 fails adc calibration
      mismatch between sync pulses received from master and FEE64 - attachment 1

      FEE64 temperatures ok - attachment 2

      detector bias/leakage currents ok - attachment 3
      leakage currents slightly ramping since last night

      RH 50.7%, Temperature 22.2 deg, dew point 11.6 deg

09.30 AIDA powered off and now in safe state
Entry  Fri Oct 20 01:43:40 2017, OH, Friday 20th October 6x
09:43 DAQ continues ok
      System wide checks all ok
      Detector bias and leakage ok - Attachment 1
      FEE temperatures ok - Attachment 2
      On file R3_279 writing at ~11Mb/s

10:07 Waveforms for nnaida 13, 11, 20 and 22 - Attachments 3-7
      nnaida 13 & 22 p-side
      nnaida 11 & 20 n-side
      Preamp feedback settings for ASIC are still default

10:49 DAQ Stopped
      Detector bias taken off
Entry  Fri Sep 2 06:22:26 2016, TD, Friday 2 September 17x
14.20 R11 continues

      DSSSD bias/leakage current OK - see attachments 1-2
      FEE64 temperatures OK - see attachment 3

      System-wide checks OK *except* nnaida19 fails ASIC timestamp check (no change)
      Free memory - see attachment 4

      Merger errors messages for links 4 & 5 - see attachment 5

     AIDA SYNC, PAUSE, RESUME, Correlation L, SYNC time warp, Good Events, AIDA disc, nnaida6, nnaida5 &
nnaida20 statistics - see 
attachments 6-16

Analysis of R11_645 - FEE64 data synchronised - see attachment 17

aidas1> ./analyser2 /TapeData/Aug2016/R11_645
 *** GREAT format 3.2.0 analyser - TD - May 2014
 *** ERROR: READ I/O error:       5002
                   blocks:      32000
          ADC data format:  156678476 ( 1098367.0 Hz)
        Other data format:  104409524 (  731944.6 Hz)
 Sample trace data format:          0 (       0.0 Hz)
         Undefined format:          0 (       0.0 Hz)
   Other data format type:      PAUSE:         92 (       0.6 Hz)
                               RESUME:         92 (       0.6 Hz)
                              SYNC100:    1303563 (    9138.4 Hz)
                           FEE64 disc:  103104691 (  722797.4 Hz)
                             MBS info:       1086 (       7.6 Hz)
                           Other info:          0 (       0.0 Hz)

   ADC data range bit set:        650 (       4.6 Hz)

                Timewarps:        ADC:          0 (       0.0 Hz)
                                PAUSE:          0 (       0.0 Hz)
                               RESUME:          0 (       0.0 Hz)
                              SYNC100:          0 (       0.0 Hz)
                           FEE64 disc:          0 (       0.0 Hz)
                             MBS info:          0 (       0.0 Hz)
                            Undefined:          0 (       0.0 Hz)
                         Sample trace:          0 (       0.0 Hz)

 Timestamp elapsed time:      142.647 s
 FEE module #:   1 elapsed dead time       0.186 s
 FEE module #:   2 elapsed dead time       0.427 s
 FEE module #:   3 elapsed dead time       0.000 s
 FEE module #:   4 elapsed dead time       0.000 s
 FEE module #:   5 elapsed dead time       1.001 s
 FEE module #:   6 elapsed dead time       3.508 s
 FEE module #:   7 elapsed dead time       0.000 s
 FEE module #:   8 elapsed dead time       0.000 s
 FEE module #:   9 elapsed dead time       0.053 s
 FEE module #:  10 elapsed dead time       0.000 s
 FEE module #:  11 elapsed dead time       0.000 s
 FEE module #:  12 elapsed dead time       0.000 s
 FEE module #:  13 elapsed dead time       0.100 s
 FEE module #:  14 elapsed dead time       0.164 s
 FEE module #:  15 elapsed dead time       0.000 s
 FEE module #:  16 elapsed dead time       0.000 s
 FEE module #:  17 elapsed dead time       0.264 s
 FEE module #:  18 elapsed dead time       0.000 s
 FEE module #:  19 elapsed dead time       0.000 s
 FEE module #:  20 elapsed dead time       0.000 s
 FEE module #:  21 elapsed dead time       0.028 s
 FEE module #:  22 elapsed dead time       0.171 s
 FEE module #:  23 elapsed dead time       0.000 s
 FEE module #:  24 elapsed dead time       0.000 s
 FEE module #:  25 elapsed dead time       0.000 s
 FEE module #:  26 elapsed dead time       0.000 s
 FEE module #:  27 elapsed dead time       0.000 s
 FEE module #:  28 elapsed dead time       0.000 s
 FEE module #:  29 elapsed dead time       0.000 s
 FEE module #:  30 elapsed dead time       0.000 s
 FEE module #:  31 elapsed dead time       0.000 s
 FEE module #:  32 elapsed dead time       0.000 s

 *** Program elapsed time:   15.730s ( 2034.269 blocks/s, 127.142 Mb/s)

14.54 R11 stopped (two DAQ STOP commands required)

15.00 New firmware version 8.24 installed (/MIDAS/Aida/FEE_GF_Feb16_24.bin) 
      - guarantee SYNC is the last data item with timestamp ending 0x000000A0 to be output (merger assumes this)
      - ASIC clock runs when output disabled (ASIC requires periodic reset)

18.40 R13 starts

      ASIC settings 2016Sep02-18.17.01
      Discriminator masks -> 0x0
      Correlation scaler trigger delay 0x180 -> 0x5
Entry  Fri Nov 2 00:57:24 2018, CA, CB, TD, OH, Friday 2 November 14x
09.58 DAQ continues OK - current file Oct18/R15_6
      alpha background run

09.20 all system wide checks OK *except*
        nnaida7 fails ADC calibration
        SYNC errors since baseline - see attachment 1 - baseline not zero'd following reboot of nnaida2 and
setup of all FEE64s yesterday
        SYNC errors OK following new baseline check

      detector biases & leakage currents OK - see attachments 2 & 3

      FEE64 temperatures OK - see attachment 4
      good event statistics OK - see attachment 5
      nnaida2 rate due to one hot channel

      Merger OK - see attachment 6
      TapeServer OK - see attachment 7
      no errors or warnings shown by server terminal sessions 

      1.8.L spectra - see attachments 8 & 9

       N.B. evidence of minimally ionising particles (presumably protons) from beam to F11 E detector
            ~10-15cm Pb bricks were placed within BRIKEN (CH2)n stack to stop heavy particles

      1.8.W spectra - see attachments 10 & 11 - trace updates for 9/12 p+n FEE64s have stopped, n+n FEE64s OK

15.54 DAQ continues - current file Oct18/R15_15
      nnaida17 good event rate c. 400kHz


15.57 FEE64 reset, setup, system wide checks
      all ASIC clocks sync'd using ReSYNC
      all spectra zero'd
      ASIC check load
      
      DAQ go - data transfer disabled

      good events statistics - attachment 12

      1.8.W spectra - see attachments 13 & 14
      changed LED threshold 7700 -> 8500 for 9/12 p+n FEE64s
Entry  Fri Jun 2 00:01:48 2017, OH, PW, CB, Friday 2 June 08:00-16:00 9x
08:05 Briken and BigRIPS stopping daqs
      AIDA on file R1_1099

08:07 BigRIPS starts run 1088
      BRIKEN starts run  77

08:13 System wide checks all ok
      FEE temperatures all ok - attachment 1
      Good events statistics - attachment 2

08:50 Bias and leakage currents ok - attachment 3

09:04 BigRIPS and BRIKEN stop daqs
      AIDA on file R1_1113

09:05 BigRIPS starts run 1089
      BRIKEN starts run 78
      AIDA on file R1_1113

09:22 Beam lost AIDA on file R1_1117

09:26 Beam back AIDA on file R1_1118

10:06 BigRIPS and BRIKEN stop daqs
      AIDA on file R1_1128

10:07 BigRIPS starts run 1090
      BRIKEN starts run 79
      AIDA on file R1_1128

10:12 System wide checks ok

10:24 DSSD rates from R1_1131
 *** DSSSD # 1 count:     88384 old count:     45978 dt:      1.01 s  LEC rate:  42084.25 Hz
 *** DSSSD # 2 count:    119845 old count:     62135 dt:      1.01 s  LEC rate:  57272.13 Hz
 *** DSSSD # 3 count:     40809 old count:     21150 dt:      1.01 s  LEC rate:  19509.84 Hz
 *** DSSSD # 4 count:     37059 old count:     18986 dt:      1.01 s  LEC rate:  17935.87 Hz
 *** DSSSD # 5 count:     14947 old count:      7772 dt:      1.01 s  LEC rate:   7120.56 Hz
 *** DSSSD # 6 count:     29809 old count:     15349 dt:      1.01 s  LEC rate:  14350.29 Hz
 *** DSSSD # 1 count:        35 old count:        16 dt:      1.01 s  HEC rate:     18.86 Hz
 *** DSSSD # 2 count:       108 old count:        56 dt:      1.01 s  HEC rate:     51.61 Hz
 *** DSSSD # 3 count:        77 old count:        38 dt:      1.01 s  HEC rate:     38.70 Hz
 *** DSSSD # 4 count:        76 old count:        35 dt:      1.01 s  HEC rate:     40.69 Hz
 *** DSSSD # 5 count:        27 old count:        15 dt:      1.01 s  HEC rate:     11.91 Hz
 *** DSSSD # 6 count:         4 old count:         2 dt:      1.01 s  HEC rate:      1.98 Hz

11:07 BigRIPS and BRIKEN stop their daqs
      AIDA on file R1_1142

11:08 BigRIPS starts run 1091
      BRIKEN starts run 80
      AIDA on file R1_1142

11:10 Leakages and biases ok - attachment 4

11.36 DAQ continues OK
      Attached HEC x vs y - attach 5
      Zero'd HEC x vs y

11.59 Analysed R1_1153 - see attachment 6
      No timewarps
      Dead time negligible
      All OK

12:06 BigRIPS and BRIKEN stop their daqs
      AIDA on file R1_1156

12:07 BigRIPS starts run1092
      BRIKEN starts run 81
      AIDA on file R1_1156

13:07 BigRIPS and BRIKEN stop daqs
      AIDA on file R1_1170

13:08 BRIKEN starts run 82
      BigRIPS starts run 1093
      AIDA on file R1_1170

13:11 System wide checks ok

13:55 BRIKEN and BigRIPS stop their daqs
      AIDA on file R1_1181
      Beam gone

14:11 Bias and leakage currents ok - attachment 7
      Ambient Temperature 23.9C - Humidity 37.3% - D.P +8.3

15:26 Beam is back
      AIDA on file R1_1201

15:07 BigRIPS starts run 1094
      BRIKEN starts run 83
      AIDA on file R1_1201

15:34 System wide checks ok
      FEE temperatures ok - attachment 8
      Good Events statistics - attachment 9

15:39 No beam - beam back

15:46 Beam gone for couple of minutes

15:50 Beam back
Entry  Thu Jun 1 16:06:03 2017, NN, Friday 2 June 00:00-08:00  14x
00:35   Beam is on
        current AIDA file R1_996

00:36   BRIKEN start run 72
        BigRIPS start run 1083
        AIDA file RIBF127R1/R1_996

00:55   Beam is off but DAQ of BRIKEN and BigRIPS are still running)
        expecting it will be back within 10 minutes
        AIDA file R1_1001

00:57   Analysis of file RIBF127R1/R1_1000 - dead time < 0.4% - see attachment 1

01:03   Beam is on
        DAQ is running continuously

01:15   statistics normal
        system wide checks ok 
        leakage current and bias ok see attachment 2
        merger and tapeserver ok

01:25   BigRIPS and BRIKEN stop their runs
        current AIDA run R1_1007

01:26   BRIKEN start run 73
        BigRIPS start run 1084
        AIDA on file R1_1008

01:36   FEE temperature is normal see attachment 3

02:22   BigRIPS and BRIKEN stop their runs
        current AIDA run R1_1021

02:23   high energy correlation between x and y attachment 4
        AIDA run R1_1021

02:26   BRIKEN start run 74
        BigRIPS start run 1085
        AIDA on file R1_1023 

03:00   System wide checks all ok      
        FEE temperatures ok - attachment 5
        Good event statistics - attachment 6
        Merger and tape server still running - attachment 7

03:06   Leakages and bias ok - attachment 8

03:09   Beam is OFF

03:11   BigRIPS and BRIKEN stop their runs
        current AIDA run R1_1033

04:10   Beam is still OFF 
        Investigation is still going on, on the ion source. 
        (Ion source were getting down more and more frequently before)

04:49   FEE's temperature normal
        statistics normal
        system wide checks ok 
        leakage current and bias ok 
        merger and tapeserver ok
        (waiting beam) 

06:01   Beam is ON
        AIDA run R1_1069

06:02   BRIKEN start run 75
        BigRIPS start run 1086
        AIDA on file R1_1070

06:08   FEE's temperature normal
        statistics normal
        system wide checks ok 
        leakage current and bias ok see attachment 9 
        merger and tapeserver ok

06:20   Analysis of file RIBF127R1/R1_1072 - dead time < 0.6% (most of the FEE cards have 0.0%) - see attachment 10

07:01   high energy correlation between x and y attachment 11
        AIDA run R1_1084

07:02   BigRIPS and BRIKEN stop their runs
        current AIDA run R1_1084

07:04   BRIKEN start run 76
        BigRIPS start run 1087
        AIDA on file R1_1085

07:16 per DSSSD rates file RIBF127R1/R1_1086
 *** DSSSD # 1 count:     92842 old count:     41280 dt:      1.05 s  LEC rate:  49071.92 Hz
 *** DSSSD # 2 count:    124879 old count:     56250 dt:      1.05 s  LEC rate:  65314.71 Hz
 *** DSSSD # 3 count:     44779 old count:     19466 dt:      1.05 s  LEC rate:  24090.56 Hz
 *** DSSSD # 4 count:     37971 old count:     18208 dt:      1.05 s  LEC rate:  18808.59 Hz
 *** DSSSD # 5 count:     15612 old count:      7434 dt:      1.05 s  LEC rate:   7783.06 Hz
 *** DSSSD # 6 count:     31207 old count:     14708 dt:      1.05 s  LEC rate:  15702.22 Hz
 *** DSSSD # 1 count:        27 old count:        11 dt:      1.05 s  HEC rate:     15.23 Hz
 *** DSSSD # 2 count:        89 old count:        43 dt:      1.05 s  HEC rate:     43.78 Hz
 *** DSSSD # 3 count:        82 old count:        44 dt:      1.05 s  HEC rate:     36.16 Hz
 *** DSSSD # 4 count:        65 old count:        32 dt:      1.05 s  HEC rate:     31.41 Hz
 *** DSSSD # 5 count:        29 old count:        17 dt:      1.05 s  HEC rate:     11.42 Hz
 *** DSSSD # 6 count:         6 old count:         3 dt:      1.05 s  HEC rate:      2.86 Hz

07:22   FEE's temperature normal - see attachment 12
        statistics normal - see attachment 13
        system wide checks ok 
        leakage current and bias ok
        merger and tapeserver ok - see attachment 14
Entry  Thu Dec 1 16:01:10 2016, CG, PV, Friday 2 December 14x
00.07 BigRIPS data file 1135

01.07 BigRIPS data file 1136.

02.09 BigRIPS data file 1137.

02.13 Beam off. ~1/2 way through R44_137

02.16 Beam back. ~600MB into R44_139
      BigRIPS data file 1138.

03.17 BigRIPS data file 1139.

03.37 Beam stopped. End of R44_167.

03.43 Beam returned to F11. ~700MB through R44_170.
      BigRIPS data file 1140.

04.00 Beam off for stripper foil change. End of R44_175.

04.06 Beam returned to F11. 1.2GB into R44_178.
 
04.08 BigRIPS data file 1141.

05.09 BigRIPS data file 1142.

06.02 BigRIPS data fill 1143.

06.26 R44 stopped @ R44_227 ahead of LN2 fill.

07.02 BigRIPS data file 1144.

07.43 Run R45 started.

08.02 BigRIPS data file 1145.

08.05 Local Elog and BigRIPS log uploaded (attachments 3-6).

09.00 Machine time ended, the last data file is R45_18

13.40 Biases are shown as attachments 7 and 8

15.04 DAQ starts (RIBF03R1/R46)
      Pulser walkthrough
      Amplitude 90,000 - 10,000 @ 10,000 step
      x5 attenuator IN
      t_r 50ns tau_d 50us
      frequency 250Hz range
      polarity -, tail pulse, delay MIN

15.19 Walk through finishes.  See attachments 9 and 10 for histograms

15.26 DAQ starts (RIBF03R1/R47)
      Pulser walkthrough
      Amplitude 12,000 - 2,000 @ 2,000 step

15.35 R47_2 finishes.
      See attachments 11 and 12 for histograms

15.41 DAQ starts (RIBF03R1/R48)
      Background - BNC PB-4 pulser OFF

16.35 DAQ stops OK (RIBF03R1/R48_13)

16.55 DSSSD bias & leakage currents Ok - see attachment 13 & 14
      Ambient temperature +25.2 deg C, -4.2 deg C, RH 14.2%
Entry  Fri May 19 02:44:14 2017, TD, CB, OH, Friday 19th May 190517_bias1045.png190517_feetemp1045.png190517_goodevents1056.png
10:44 System wide checks all ok
      Leakages and biases all ok (see attachment 1)
      Ambient Temperature 24.4C - Humidity 38.1% - Dew Point +9.1
      FEE temperatures ok (see attachment 2)
      DAQ, Merger and Tape server all still running ok
      Good events statistics (see attachment 3)

      Orientation of RS-232 cards was wrong. They have now been corrected will enable logging then do a halt and
go on the daq.
      Power supply also moved onto AIDA circuit.

11:00 Current file R15_333.
      
11:04 halt and go performed
      24 Log files produced on Rasp Pi console monitor.
Entry  Fri Aug 19 07:42:22 2016, TD, PV, Friday 19 August 6x
14.00 Both AIDA NIM bins U/S 
      Upper NIM bin - BNC PB-4, Ortec 433, 2x CAEN N1419, 1x CAEN N1419B - ac mains fuse 100V/8A/slow? blowing
      Lower NIM bin - 11x MACB - not powering up

      Replaced both NIM bins.
      Upper NIM bin - 11x MACB
      Lower NIM bin - BNC PB-4, Ortec 433, 2x CAEN N1419, 1x CAEN N1419B

15.50 Julabo FL11006 chiller ON - set point +20 deg C
      B2F/F11 ambient temp +24.0 deg C, RH 54%, DP +13.8 deg C

      Bias 6x MSL type BB18 DSSSDs - see attachments 1-2

16.10 Startup DAQ nnaida1-24 *except* nnaida19

      System-wide checks

      Check clock status OK
      Check ADC calibration - nnaida7 failed
      Check ASIC clock timestamp - multiple failures - see attachment 3
      Check SYNC errors not counted Since Baseline - see attachment 4
      Check SYNC pulses received - see attachment 5

      BNC PB-4 

      Amplitude 90,000, x5 attenuator IN
      t_r 50ns, tau_d 50us
      Polarity -
      INT ref, delay MIN, tail pulse

      FEE64 temperatures - see attachment 6

16.34 R1 starts (directory Aug2016)

      slow comparator 0x10, LEC/MEC fast comparator 0x10

      merge 555k data items/s, tapeserver 4430 kb/s

18.28 R4 starts

      Settings per R1

18.31 DAQ/merge stalls with (many) messages as follows

MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
Merge state now Paused
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in pending data, found 1146,
last 575
copyItems: queue 14, maybe copied in wrong buffer: sequence 1146, last 575
Merge_addItems: queue 14, maybe added wrong buffer: sequence 1146, last 575
MERGE Program (24316): check_for_data: index 14, unexpected header sequence number in new data, found 689, last 1258
copyItems: queue 14, maybe copied in wrong buffer: sequence 689, last 1258
Merge_addItems: queue 14, maybe added wrong buffer: sequence 689, last 1258
MERGE Data Link (24254): client closed connection
MERGE Data Link (24254): listening on port 11001
MERGE Data Link (24255): client closed connection
MERGE Data Link (24255): listening on port 11002
MERGE Data Link (24257): client closed connection
MERGE Data Link (24257): listening on port 11004
MERGE Data Link (24258): client closed connection
MERGE Data Link (24258): listening on port 11005
MERGE Data Link (24261): client closed connection
MERGE Data Link (24261): listening on port 11008
MERGE Data Link (24260): client closed connection
MERGE Data Link (24260): listening on port 11007
MERGE Data Link (24259): client closed connection
MERGE Data Link (24259): listening on port 11006
MERGE Data Link (24256): client closed connection
MERGE Data Link (24256): listening on port 11003
MERGE Data Link (24262): client closed connection
MERGE Data Link (24262): listening on port 11009
MERGE Data Link (24263): client closed connection
MERGE Data Link (24263): listening on port 11010
MERGE Data Link (24265): client closed connection
MERGE Data Link (24265): listening on port 11012
MERGE Data Link (24266): client closed connection
MERGE Data Link (24266): listening on port 11013
MERGE Data Link (24267): client closed connection
MERGE Data Link (24267): listening on port 11014
MERGE Data Link (24269): client closed connection
MERGE Data Link (24269): listening on port 11016
MERGE Data Link (24268): client closed connection
MERGE Data Link (24268): listening on port 11015
MERGE Data Link (24270): client closed connection
MERGE Data Link (24270): listening on port 11017
MERGE Data Link (24264): client closed connection
MERGE Data Link (24264): listening on port 11011
MERGE Data Link (24273): client closed connection
MERGE Data Link (24273): listening on port 11020
MERGE Data Link (24271): client closed connection
MERGE Data Link (24271): listening on port 11018
MERGE Data Link (24277): client closed connection
MERGE Data Link (24277): listening on port 11024
MERGE Data Link (24275): client closed connection
MERGE Data Link (24275): listening on port 11022
MERGE Data Link (24274): client closed connection
MERGE Data Link (24274): listening on port 11021
MERGE Data Link (24276): client closed connection
MERGE Data Link (24276): listening on port 11023
    Reply  Fri Aug 19 09:43:27 2016, TD, VP, reply PJCS, Friday 19 August 
> 
> 14.00 Both AIDA NIM bins U/S 
>       Upper NIM bin - BNC PB-4, Ortec 433, 2x CAEN N1419, 1x CAEN N1419B - ac mains fuse 100V/8A/slow? blowing
>       Lower NIM bin - 11x MACB - not powering up
> 
>       Replaced both NIM bins.
>       Upper NIM bin - 11x MACB
>       Lower NIM bin - BNC PB-4, Ortec 433, 2x CAEN N1419, 1x CAEN N1419B
> 
> 15.50 Julabo FL11006 chiller ON - set point +20 deg C
>       B2F/F11 ambient temp +24.0 deg C, RH 54%, DP +13.8 deg C
> 
>       Bias 6x MSL type BB18 DSSSDs - see attachments 1-2
> 
> 16.10 Startup DAQ nnaida1-24 *except* nnaida19
> 
>       System-wide checks
> 
>       Check clock status OK
>       Check ADC calibration - nnaida7 failed
>       Check ASIC clock timestamp - multiple failures - see attachment 3
>       Check SYNC errors not counted Since Baseline - see attachment 4
>       Check SYNC pulses received - see attachment 5
> 
>       BNC PB-4 
> 
>       Amplitude 90,000, x5 attenuator IN
>       t_r 50ns, tau_d 50us
>       Polarity -
>       INT ref, delay MIN, tail pulse
> 
>       FEE64 temperatures - see attachment 6
> 
> 16.34 R1 starts (directory Aug2016)
> 
>       slow comparator 0x10, LEC/MEC fast comparator 0x10
> 
>       merge 555k data items/s, tapeserver 4430 kb/s
Looks like a problem with the inter-MACB wiring.
The screen shots show groups of 4 with the problems.
I suggest re-seating the short HDMI cables and reset the system.
    Reply  Fri Aug 19 11:07:44 2016, TD, VP, reply PJCS, Friday 19 August 7.png
> > 
> > 14.00 Both AIDA NIM bins U/S 
> >       Upper NIM bin - BNC PB-4, Ortec 433, 2x CAEN N1419, 1x CAEN N1419B - ac mains fuse 100V/8A/slow? blowing
> >       Lower NIM bin - 11x MACB - not powering up
> > 
> >       Replaced both NIM bins.
> >       Upper NIM bin - 11x MACB
> >       Lower NIM bin - BNC PB-4, Ortec 433, 2x CAEN N1419, 1x CAEN N1419B
> > 
> > 15.50 Julabo FL11006 chiller ON - set point +20 deg C
> >       B2F/F11 ambient temp +24.0 deg C, RH 54%, DP +13.8 deg C
> > 
> >       Bias 6x MSL type BB18 DSSSDs - see attachments 1-2
> > 
> > 16.10 Startup DAQ nnaida1-24 *except* nnaida19
> > 
> >       System-wide checks
> > 
> >       Check clock status OK
> >       Check ADC calibration - nnaida7 failed
> >       Check ASIC clock timestamp - multiple failures - see attachment 3
> >       Check SYNC errors not counted Since Baseline - see attachment 4
> >       Check SYNC pulses received - see attachment 5
> > 
> >       BNC PB-4 
> > 
> >       Amplitude 90,000, x5 attenuator IN
> >       t_r 50ns, tau_d 50us
> >       Polarity -
> >       INT ref, delay MIN, tail pulse
> > 
> >       FEE64 temperatures - see attachment 6
> > 
> > 16.34 R1 starts (directory Aug2016)
> > 
> >       slow comparator 0x10, LEC/MEC fast comparator 0x10
> > 
> >       merge 555k data items/s, tapeserver 4430 kb/s
> Looks like a problem with the inter-MACB wiring.
> The screen shots show groups of 4 with the problems.
> I suggest re-seating the short HDMI cables and reset the system.

Short HDMI cables re-seated. See attachment 1
ELOG V3.1.3-7933898