| ID |
Date |
Author |
Subject |
|
12
|
Sun May 3 09:04:07 2015 |
J.L. Tain, J. Agramunt, G. Kiss | Summary of neutron measurements |
U: (Up: 2 RIKEN tubes at 40cm from AIDA)
D: (Down: 12 RIKEN tubes at 80cm from AIDA)
US: (2 ORNL 2" tubes close to the floor at 1.8m from AIDA)
Natural background:
File: 150426_Bckg_1425_1805.root
Rate per tube (n/s): U=0.0071, D=0.0045, US=0.0222
Notes:
-This rate seems quite small but we should remember
that we are underground.
-The fact that U is larger than D could be explained as
related that the amount of PE around tubes is larger.
-For comparison:
BELEN-20@JYFL=0.6n/s (no beam), 0.9n/s (beam)
BELEN-30@GSI=1.2n/s (beam)
BELEN-48@JYFL=1.5n/s
110Zr setting:
File:150501_16111810_110Zr.root
Rate per tube (n/s): U=0.483, D=0.139, US=0.292
Ratio to background: U=68.3, D=31.1, US=13.2
Individual tube distribution:
U1 0.280 D1 0.087 D7 0.084 US1 0.132
U2 0.249 D2 0.092 D8 0.084 US2 0.188
D3 0.083 D9 0.077
D4 0.075 D10 0.070
D5 0.073 D11 0.069
D6 0.065 D12 0.054
Notes:
-There is huge increase of with beam
-The drop of increase with distance is consistent
with either beam induced or decay of implants
110Zr setting:
File: 150501_1733_1900_110Zr.root (actually 02/05)
Rate per tube (n/s): U=0.869, D=0.242, US=0.516
Ratio to 1st 110Zr: U=1.80, D=1.74, US=1.77
Notes:
-We don't quite understand the reason for the near factor two
difference since apparently the settings are the same and beam was
rather stable
In order to investigate the point of origin of the neutrons
we add extra PE covering D5,D6,D11,D12 from the top and
D1,D2,D7,D8 from the front
88Ge setting:
File: 150502_21502333_88Ge.root
Rate per tube (n/s): U=1.271, D=0.349, US=0.706
Ratio to 1st 110Zr: U=2.63, D=2.51, US=2.42
Individual rates:
U1 1.340 D1 0.424 D7 0.402 US1 0.576
U2 1.202 D2 0.446 D8 0.381 US2 0.835
D3 0.408 D9 0.342
D4 0.393 D10 0.315
D5 0.359 D11 0.250
D6 0.275 D12 0.193
Notes:
-One notices a relative drop of rate for D6, D11 and D12
pointing to a source located in the vicinity of AIDA
94Se setting:
File: 150503_1219_1319_94Se.root
Rate per tube (n/s): U=1.267, D=0.349, US=0.717
Ratio to 1st 110Zr: U=2.62, D=2.51, US=2.46
Notes:
-Very similar to 88Ge measurement |
|
13
|
Sun May 3 10:07:25 2015 |
J.L. Tain, J. Agramunt, G. Kiss | Photos of new PE shielding |
Photos of the general arrangement and details of the new PE blocks |
| Attachment 1: 20150502_195020.jpg
|
|
| Attachment 2: 20150502_194340.jpg
|
|
| Attachment 3: 20150502_192253.jpg
|
|
|
14
|
Sun May 3 10:10:13 2015 |
J.L. Tain, J. Agramunt, G. Kiss | 94Se setting measurement |
94Se setting:
File: 150503_1500_1554_94Se.root
Rate per tube (n/s): U=0.966, D=0.263, US=0.538
Ratio to 1st 110Zr: U=2.00, D=1.89, US=1.84
Notes:
-Compared to the 1st 94Se setting measurement (12:19-13:19)
there is a drop of 25% in the rate without apparent reason |
|
21
|
Tue May 5 05:04:00 2015 |
J.L. Tain, J. Agramunt, G. Kiss | Change of degrader thickness |
For EURICA measurements degrader thickness were increased
trying to decrease MACI and AIDA trigger rate
Before we had 5.7 mm of Al
Now we have:
Degrader: 4.0 + 1.0 + 0.6 + 0.4 +0.3 = 6.3 mm of Al
also 2x 0.8mm Al plates were added before
the MACI start detector (thus covering partially AIDA)
Trigger from AIDA will be taken from 3rd DSSSD |
|
22
|
Tue May 5 05:17:36 2015 |
J.L. Tain, J. Agramunt, G. Kiss | Ambient background measurement |
Failure of a quadrupole triplet at BigRIPS (some time after earthquake)
We take the chance of measuring background:
File: 150504_1500_2047_bckg.root
Rate per tube (n/s): U=0.098, D=0.0062, US=0.0200
Ratio to 1st 110Zr: U=1.38, D=1.99, US=0.90
Notes:
-Compared to the 1st background measurement there are changes
larger than expected statistical fluctuations:
perhaps due to slight changes in noise
|
|
23
|
Tue May 5 05:29:26 2015 |
J.L. Tain, J. Agramunt, G. Kiss | Ambient background measurement |
Beam still down. New background measurement
with new RIKEN tube below the MUSIC chamber
(call it B)
File: 150505_0040_1130_aprox_bckg.root
Average rate per tube (n/s): U=0.0073, D=0.0065, B=0.0070 |
|
24
|
Wed May 6 04:57:24 2015 |
J.L. Tain, J. Agramunt, G. Kiss | New degrader thickness |
The Al degrader thickness was reduced to 4.8mm |
|
25
|
Wed May 6 05:10:14 2015 |
J.L. Tain, J. Agramunt, G. Kiss | 94Se setting measurement |
After the long shutdown the beam is back at 00:30 (06/05)
Intensity is slightly lower 26pnA and somewhat unstable (short time)
Measurement in the morning:
File: 150506_0846_0915_94Se.root
Average rate per tube (n/s): U=3.39 D=0.93B=0.40
Notes:
-The rate has increased by 3.5 in the detectors close to EURICA (!?)
-The rate at the detector below the MUSIC is substantial.
From the different magnitudes it is clear that neutrons are originated
both outside and inside EURICA. |
|
54
|
Wed Jul 20 16:50:12 2016 |
J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, S. Go... | 252Cf |
Measurement of the 252Cf source at the central position
Efficiency and neutron-gamma correlations
Listmode: 160719_2233_252Cf_XX.dlt, Run: 029-030
Online: 160719_2233_252Cf_029_030.root |
|
57
|
Wed Jul 20 17:18:59 2016 |
J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, K. Gabor, S. Go, K. Matsui... | Electronics |
Measurement with preamplifiers after drying.
Amplifiers are brought back after ~12h and connected.
Measurement of background for 1hour
Online: 160720_1350_1452_Bkg.root
Listmode: No
There might have been a slight improvement on the PA behaviour
but is not clear.
Attached a screen shot of the spectrum
The spurious signals show very little noise superimposed
as can be seen in the attached capture of the sample by the DACQ
but is not so obvious how this can help to veto them. |
| Attachment 1: Bkg-afterPAdrying.png
|
|
| Attachment 2: V2A4C3-SpuriousSignal.png
|
|
|
50
|
Wed Jul 20 12:00:15 2016 |
J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, ... | 252Cf |
Measurement with the source at the center
Listmode: 160719_BRIKEN_252Cf_009.dlt
Online: 160719_BRIKEN_252Cf.root
BRIKEN: 1.03697E7
Pulser: 5985
Rate: 17326 n/s |
|
56
|
Wed Jul 20 17:05:59 2016 |
J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, ... | Background |
Background measurement with remaining tubes (95) overnight
Listmode: 160719_2343_Bkg_*.dlt, Run: 031-044
Online: 160719_2343_1030_Bkg.root
Unfortunately the 252Cf source remained ind the hall (shielded and far away but still)
Screen shot of neutron spectrum |
| Attachment 1: Bkg-with95tubes.png
|
|
|
27
|
Thu May 7 08:36:56 2015 |
J.L. Tain, J. Agramunt, A. Estrade, G.Kiss, T. Davinson, C.Griffin | First BRIKEN_v.0 beta-delayed neutron correlations? |
During the 98,100Kr setting run we observe
in the time correlation plot (middle right panel)
of beta signals (marked by AIDA, with amplitude larger than ch=600)
with Riken 3He tube signals (noise and pulser cleaned)
a small peak around t=0 which might be indication
of beta-delayed neutrons.
The spectrum was measured for about 2h. |
| Attachment 1: Screenshot_from_2015-05-07_16_01_27.png
|
|
|
46
|
Wed Jul 20 08:58:13 2016 |
J.L. Tain, I.Dillman, R. Caballero, J. Agramunt, A. Tarifeņo, A. Tolosa, ... | Electronics |
Distribution of connections of tubes in preamplifiers and digitizers
Because of the need to use 10 PAs instead of 9 PAs we had to redistribute
the signal connections. Now:
PA1(RIKEN), PA2(RIKEN), PA3(UPC), PA4(UPC), PA5(UPC) -> V1A1-V1A4
PA6(ORNL1) -> V1A5
PA7(ORNL2) -> V1A6
PA8(ORNL2) -> V2A1,V2A2
PA9(ORNL2) -> V2A3,V2A4
PA10(ORNL2) -> V2A5,V2A6
So V1A1 to V1A4 contain now signals from 2 PAs.
We also redistributed the tubes within the PAs
We also changed the numbering or RIKEN Tubes. Tube 1 to 12 are now
those on the AIDA side (PA1 attached to the PE block)
A figure showing the distribution of tubes an numbering is included.
Tube 1 is behind tube 19 and so forth.
Attached also is an excel file with tube-channel distribution |
| Attachment 1: TubeDistribution-new.xlsx
|
| Attachment 2: TubeDistribution-new.pdf
|
|
|
10
|
Sat May 2 03:52:14 2015 |
J.L. Tain | 3He tube description |
3He tube drawings and technical details:
RIKEN, UPC, GSI, ORNL-1, ORNL-2 |
| Attachment 1: He3_GE_30cm.pdf
|
|
| Attachment 2: He3-252248.pdf
|
|
| Attachment 3: 1inch3He.pdf
|
|
| Attachment 4: 2inch3He.pdf
|
|
|
30
|
Sat Jul 9 20:44:31 2016 |
J.L. Tain | BRIKEN neutron detector |
START
Assembly and commissioning of BRIKEN neutron detector
The hybrid configuration of the BRIKEN neutron detector will be assembled at BF2 for the first time with 140 3He
tubes and 2 CLOVER detectors, connected to the Gasific digital data acquisition system and tested with 252Cf
sources. |
|
31
|
Sat Jul 9 21:00:54 2016 |
J.L. Tain | Hardware |
A list of the available hardware |
| Attachment 1: Hardware.xlsx
|
| Attachment 2: Hardware.pdf
|
|
|
32
|
Sat Jul 9 21:08:08 2016 |
J.L. Tain | Neutron moderator |
Drawings of the polyethylene neutron moderator and table assembly |
| Attachment 1: Moderator-Table.pdf
|
|
| Attachment 2: Table.pdf
|
|
| Attachment 3: Moderator-Shielding.pdf
|
|
|
33
|
Sat Jul 9 21:14:12 2016 |
J.L. Tain | 3He tubes |
Drawings and specs of 3He tubes |
| Attachment 1: He3_GE_30cm-RIKEN.pdf
|
|
| Attachment 2: 1inch3He-ORNL.pdf
|
|
| Attachment 3: 2inch3He_ORNL.pdf
|
|
| Attachment 4: He3-252248-UPC-GSI.pdf
|
|
|
34
|
Sat Jul 9 21:21:38 2016 |
J.L. Tain | 3He tubes |
Tube distribution in the moderator, pre-amplifiers and digitizers |
| Attachment 1: TubeDistribution.pdf
|
|
| Attachment 2: TubeDistribution.xlsx
|