Entry  Thu Oct 6 14:43:47 2016, JL Tain, A Tarifeno, A Tolosa, J Agramunt, N Brewer, S Go, ..., Electronics 
We mount back all preamplifiers
Initially we used all the PA coming from US, the repaired and the new ones.
There was an issue with the new ones (strong noise and cross talk)
Entry  Thu Oct 6 02:03:32 2016, JL Tain, START 6x
We started the assembly of BRIKEN and AIDA detectors in their final position
on Tuesday October 4:
Entry  Thu Aug 18 23:32:59 2016, NT Brewer, New preamps arrived at Oak Ridge and tested working. IMG_20160818_145801.jpgIMG_20160803_160430.jpg
Collaborators,

Attached is a phone photo of example signals from the new preamps with 3 3He tubes connected to a scope. Scope channels 1,2,3 (color y,c,m respectively)
are preamp channels 1,2,3 while 4 (green) is open. Trigger is on channel 3, at 500mV/div and other channels are at 500,200,and 20 mV/Div for channels 2,1,4
Entry  Thu Aug 18 22:39:35 2016, NT Brewer, New preamps arrived at Oak Ridge and tested working. IMG_20160818_145801.jpg
Collaborators,

Attached is a phone photo of example signals from the new preamps with 3 3He tubes connected to a scope. Scope channels 1,2,3 (color y,c,m respectively)
are preamp channels 1,2,3 while 4 (green) is open. Trigger is on channel 3, at 500mV/div and other channels are at 500,200,and 20 mV/Div for channels 2,1,4
Entry  Fri Jul 29 09:52:28 2016, Keishi Matsui, accuracy of Cf-252's intensity in RIKEN IMG_9350.JPG
 According to the reply from Japan Radioisotope Association, the accuracy of the Cf-252's intensity is +0%/-15%.
( "+0%" is important, because 3.7 MBq is a standard limit for radio-active sources  in Japan.)
Entry  Fri Jul 22 12:50:29 2016, J.L. Tain, END 
20/JUL/2016 20:00 
End of the test measurements
Entry  Wed Jul 20 17:21:59 2016, J. Agramunt, A. Tolosa, DACQ 160718Conf_BrikenFull_cal5.xlsx
DACQ configuration parameters

The last configuration file used was: 160718Conf_BrikenFull_cal5.xlsx
Entry  Wed Jul 20 17:18:59 2016, J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, K. Gabor, S. Go, K. Matsui..., Electronics Bkg-afterPAdrying.pngV2A4C3-SpuriousSignal.png
Measurement with preamplifiers after drying.

Amplifiers are brought back after ~12h and connected.
Entry  Wed Jul 20 17:05:59 2016, J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, ..., Background Bkg-with95tubes.png
Background measurement with remaining tubes (95) overnight

Listmode: 160719_2343_Bkg_*.dlt, Run: 031-044
Entry  Wed Jul 20 16:55:48 2016, K.Gabor, K. Matsui, A. Tarifeņo, Electronic 
Drying of preamplifiers

PA8, PA9 and PA10 are removed and brought to a dryer box
Entry  Wed Jul 20 16:50:12 2016, J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, S. Go..., 252Cf 
Measurement of the 252Cf source at the central position

Efficiency and neutron-gamma correlations
Entry  Wed Jul 20 16:40:57 2016, S. Go, J.Agramunt, A.Tolosa, J.L. Tain,..., CLOVER 152Eu.png
Calibration and resolution check

CLOVER detectors were calibrated in energy with 60Co and 137Cs
Entry  Wed Jul 20 16:12:58 2016, S. Go, ..., CLOVER DSC_2522.JPGDSC_2523.JPGDSC_2524.JPG
Mounting of the CLOVER detectors in position

CLOVER G7: right of beam
Entry  Wed Jul 20 16:07:58 2016, A.Tarifeņo, J.L. Tain, J. Agramunt, A. Tolosa, ..., 252Cf 
19/Jul/2016
Scanning of the efficiency as a function of longitudinal position
Entry  Wed Jul 20 12:00:15 2016, J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, ..., 252Cf 
Measurement with the source at the center

Listmode: 160719_BRIKEN_252Cf_009.dlt
Entry  Wed Jul 20 11:49:00 2016, J.L. Tain, J. Agramunt, R. Baeturia, A. Tarifeņo, A. Tolosa, ..., Background 160718_FullBrikenFirstData_bkg.zip
18/07/2016 
First background measurement overnight.
Entry  Wed Jul 20 11:40:00 2016, J.L. Tain, J. Agramunt, R. Baeturia, A. Tarifeņo, A. Tolosa, ..., 252Cf RIKEN-252Cf-V1A1.pngUPC-252Cf-V1A3.pngORNL1-252Cf-V1A5.pngORNL2-252Cf-V1A6.png
First measurement with full 140 tube version 
of the BRIKEN neutron detector
Entry  Wed Jul 20 11:16:18 2016, J.L. Tain, Electronics 
Pulser settings

We use two pulsers, Pulser1 for RIKEN, ORNL1 and ORNL2 tubes (7 PA)
Entry  Wed Jul 20 08:58:13 2016, J.L. Tain, I.Dillman, R. Caballero, J. Agramunt, A. Tarifeņo, A. Tolosa, ..., Electronics TubeDistribution-new.xlsxTubeDistribution-new.pdf
Distribution of connections of tubes in preamplifiers and digitizers

Entry  Tue Jul 19 13:47:55 2016, A. Tarifeņo-Saldivia, Neutron efficiency - Scanning on Z preliminary result Preliminar_Zscaning_BRIKEN_PremLabel_vsmall.png
 We have carried out a scanning of the neutron efficiency with respect to the position along the Z-axis (AIDA and BEAM hole) of a 252Cf neutron
source. A preliminary analysis is presented based on data from the online. A plot with the comparison of the simulations and experimental data is attached. The
experimental results are in good agreement with the GEANT4 simulations of the studies for design of the neutron detector. The simulations for the