  409   Mon Nov 5 12:04:08 2018 FM, PA, ATLow threshold, 252Cf, z=15cm, Perpendicular 0deg
252Cf at z=15cm,ie center of moderator, source axis perpendicular at 0deg (typical orientation)
Low threshold.

18/11/05 START: 21:05
18/11/05 STOP:  22.30


** Configuration file: 
181016_Conf_BrikenFull_2153_test_thr.xlsx


ROOT file: 

181105_2105_252Cf_z15cm_2230.root


DLT files:

181105_2105_252Cf_z15cm_137.dlt
181105_2105_252Cf_z15cm_138.dlt
181105_2105_252Cf_z15cm_139.dlt




Comment: 
- Clovers are placed inside the moderator, front face ~5mm from AIDA hole border.
- AIDA not inside the moderator.
- In beam PE shielding placed.
  410   Mon Nov 5 13:43:57 2018 FM, PALow threshold, 252Cf, z=0cm, Parallel downstream
252Cf at z=0cm,ie center of moderator, source axis parallel downstream 
Low threshold.

18/11/05 START: 22:46
18/11/05 STOP:  00:01


** Configuration file: 
181016_Conf_BrikenFull_2153_test_thr.xlsx


ROOT file: 

181105_2245_252Cf_z0cm_0001.root


DLT files:

181105_2245_252Cf_z0cm_140.dlt
181105_2245_252Cf_z0cm_141.dlt
181105_2245_252Cf_z0cm_142.dlt



Comment: 
- Clovers are placed inside the moderator, front face ~5mm from AIDA hole border.
- AIDA not inside the moderator.
- In beam PE shielding placed.
  411   Mon Nov 5 15:10:43 2018 FM, PA,Low threshold, 252Cf, z=15cm, Parallel downstream
252Cf at z=15cm, i.e 15 cm from the center of the detector,  source axis parallel downstream 
Low threshold.

18/11/06 START: 00:10
18/11/06 STOP:  01:32


** Configuration file: 
181016_Conf_BrikenFull_2153_test_thr.xlsx


ROOT file: 

181106_0010_252Cf_z15cm_0132.root



DLT files:

181106_0010_252Cf_z15cm_143.dlt
181106_0010_252Cf_z15cm_144.dlt
181106_0010_252Cf_z15cm_145.dlt





Comment: 
- Clovers are placed inside the moderator, front face ~5mm from AIDA hole border.
- AIDA not inside the moderator.
- In beam PE shielding placed.
  412   Mon Nov 5 16:43:43 2018 FM, PA,Low threshold, 252Cf, z=25cm, Parallel downstream
252Cf at z=25cm, i.e 25 cm from the center of the detector,  source axis parallel downstream 
Low threshold.

18/11/06 START: 01:43
18/11/06 STOP:  03:03


** Configuration file: 
181016_Conf_BrikenFull_2153_test_thr.xlsx


ROOT file: 

181106_0143_252Cf_z25cm_0303.root



DLT files:

181106_0143_252Cf_z25cm_146.dlt
181106_0143_252Cf_z25cm_147.dlt





Comment: 
- Clovers are placed inside the moderator, front face ~5mm from AIDA hole border.
- AIDA not inside the moderator.
- In beam PE shielding placed.
  413   Mon Nov 5 18:05:48 2018 AT, JR, FM, PALow threshold, 252Cf, z=25cm, perpendicular 0 deg
252Cf at z=25cm, i.e 25 cm from the center of the detector,  source axis perpendicular  
Low threshold.

18/11/06 START: 03:18
18/11/06 STOP:  04:39

** Configuration file: 
181016_Conf_BrikenFull_2153_test_thr.xlsx


ROOT file: 

181106_0318_252Cf_z25cm_0439.root

DLT files:

181106_0318_252Cf_z25cm_148.dlt
181106_0318_252Cf_z25cm_149.dlt


Comment: 
- Clovers are placed inside the moderator, front face ~5mm from AIDA hole border.
- AIDA not inside the moderator.
- In beam PE shielding placed.
  414   Mon Nov 5 18:52:54 2018 MW, KRsuspicious clover G7 temp sensor

 

Quote:

We observed 6 pm LN2 filling of both clovers on Monday Nov 5th.

D4 went full within reasonable time of below 10 minutes.

Filling of G7 was stopped by "emergency sensor" after about 15 minutes of filling.

Likely the temp sensor at the LN2 overflow pipe of G7 should be changed.

The clock at LN2 fill system is off by about 25 minutes (filling comes later than set at the control unit.)

 

"The clock at LN2 fill system is off by about 25 minutes (filling comes later than set at the control unit.)"

Yes, This is how it always was.

 

"Filling of G7 was stopped by "emergency sensor" after about 15 minutes of filling. Likely the temp sensor at the LN2 overflow pipe of G7 should be changed."

You can put this on a different line of the fill system....

  415   Mon Nov 5 19:55:20 2018 AT, JR, FM, PALow threshold, 252Cf, z=0cm, perpendicular 90 deg
252Cf at z=0cm, i.e 0 cm from the center of the detector,  source axis perpendicular at 90 deg 
Low threshold.

18/11/06 START:04:56 
18/11/06 STOP: 06:16 

** Configuration file: 
181016_Conf_BrikenFull_2153_test_thr.xlsx


ROOT file: 

181106_0456_252Cf_z0cm_0616.root

DLT files:

181106_0456_252Cf_z0cm_150.dlt
181106_0456_252Cf_z0cm_151.dlt
181106_0456_252Cf_z0cm_152.dlt

Comment: 
- Clovers are placed inside the moderator, front face ~5mm from AIDA hole border.
- AIDA not inside the moderator.
- In beam PE shielding placed.
  416   Mon Nov 5 21:24:20 2018 AT, JR, FM, PALow threshold, 252Cf, z=25cm, perpendicular 90 deg
252Cf at z=25cm, i.e 25 cm from the center of the detector,  source axis perpendicular at 90 deg 
Low threshold.

18/11/06 START: 06:25
18/11/06 STOP:  07:45

** Configuration file: 
181016_Conf_BrikenFull_2153_test_thr.xlsx


ROOT file: 

181106_0625_252Cf_z25cm_0745.root

DLT files:

181106_0625_252Cf_z25cm_153.dlt
181106_0625_252Cf_z25cm_154.dlt

Comment: 
- Clovers are placed inside the moderator, front face ~5mm from AIDA hole border.
- AIDA not inside the moderator.
- In beam PE shielding placed.
  417   Mon Nov 5 22:47:40 2018 AT, JR, FM, PALow threshold, 252Cf, z=0cm, perpendicular 0 deg
252Cf at z=0cm, i.e 0 cm from the center of the detector,  source axis perpendicular at 0 deg 
Low threshold.

18/11/06 START:08:00 
18/11/06 STOP: 09:11 

** Configuration file: 

181030_Conf_BrikenFull_LowThrAdj_masterCLK.xlsx

ROOT file: 

181106_0800_252Cf_z0cm_nopulser_0911.root

DLT files:

181106_0800_252Cf_z0cm_nopulser_155.dlt
181106_0800_252Cf_z0cm_nopulser_156.dlt
181106_0800_252Cf_z0cm_nopulser_157.dlt

Comment: 
- Pulser OFF, using configuration file 181030_Conf_BrikenFull_LowThrAdj_masterCLK.xlsx
- Clovers are placed inside the moderator, front face ~5mm from AIDA hole border.
- AIDA not inside the moderator.
- In beam PE shielding placed.
- In the online in RateBRIKEN a weird fluctuation is observed (screenshot, attachement 1)
Attachment 1: Screenshot_from_2018-11-06_08-32-50.png
Screenshot_from_2018-11-06_08-32-50.png
  418   Tue Nov 6 21:51:54 2018 NT BrewerNew D4Off.conf for HV also on github
V A V/s V/s 1/0
------ --------------- --------- ----- ---- ------- ------- ------ ------ -------------
0 192.168.13.239 0 0 PA1 1450 0.0001 50 10 1
0 192.168.13.239 1 0 PA2 1450 0.0001 50 10 1
0 192.168.13.239 2 0 PA3 1550 0.0001 50 10 1
0 192.168.13.239 3 0 PA4 1550 0.0001 50 10 1
0 192.168.13.239 4 0 PA5 1550 0.0001 50 10 1
0 192.168.13.239 5 0 OFF 0 0.0001 50 10 0
0 192.168.13.239 6 0 OFF 0 0.0001 50 10 0
0 192.168.13.239 7 0 OFF 0 0.0001 50 10 0
0 192.168.13.239 0 1 PA6 1200 0.0001 50 10 1
0 192.168.13.239 1 1 PA7 1750 0.0001 50 10 1
0 192.168.13.239 2 1 PA8 1750 0.0001 50 10 1
0 192.168.13.239 3 1 PA9 1750 0.0001 50 10 1
0 192.168.13.239 4 1 PA10 1750 0.0001 50 10 1
0 192.168.13.239 5 1 OFF 0 0.0001 50 10 0
0 192.168.13.239 6 1 OFF 0 0.0001 50 10 0
0 192.168.13.239 7 1 OFF 0 0.0001 50 10 0
0 192.168.13.239 0 3 OFF 3000 0.0001 2 5 0
0 192.168.13.239 1 3 OFF 3000 0.0001 2 5 0
0 192.168.13.239 2 3 OFF 3000 0.0001 2 5 0
0 192.168.13.239 3 3 OFF 3000 0.0001 2 5 0
0 192.168.13.239 4 3 G70 2500 0.0001 2 5 1
0 192.168.13.239 4 3 G71 2500 0.0001 2 5 1
0 192.168.13.239 6 3 G72 2500 0.0001 2 5 1
0 192.168.13.239 7 3 G73 2500 0.0001 2 5 1
-1
 
=================================================
Everything below type = -1 is treated as a comment
(Clover Names are DetectorName+(0,1,2,3=Black,Red/Orange,Green,Blue)
 
type = 0 => MPOD HV modules => accessed thru snmp
type = 1 => CAEN x527 HV modules => accessed thru CAENHVwrapper
 
With MPOD one can use IP address and channel numbers based on slot occupation
With CAEN one needs IP address and SLOT and CHAN
 
Type IP address Slot Chan Name vSet iSet uRamp dRamp Switch On/Off
(V) (uA) (V/s) (V/s) 1/0
------ -------------- ---- ---- -------------- ------ ------ ------ ------ -------------
0 192.168.13.239 0 3 D40 3000 0.5 2 5 0
0 192.168.13.239 1 3 D41 3000 0.5 2 5 0
0 192.168.13.239 2 3 D42 3000 0.5 2 5 0
0 192.168.13.239 3 3 D43 3000 0.5 2 5 0
0 192.168.13.239 4 3 G70 2500 0.5 2 5 0
0 192.168.13.239 5 3 G71 3000 0.5 2 5 0
0 192.168.13.239 6 3 G72 2500 0.5 2 5 0
0 192.168.13.239 7 3 G73 2500 0.5 2 5 0

-1


 
Type IP address channel slot name voltage Current RampUp RampDn Switch On/Off
  419   Thu Nov 22 07:01:36 2018 ATPID high mass seting (165Pm) - RIBF148
PID plot for the high mass setting (165Pm), run 2018, rare earth peak experiment.
Partial statistics (~30h) from identification at F7.
Please use for internal reports about the BRIKEN activities.
Attachment 1: F7pid_RIBF148_165PmSetting_labels.png
F7pid_RIBF148_165PmSetting_labels.png
ELOG V3.1.4-unknown