ID |
Date |
Author |
Subject |
Text |
|
41
|
Tue Jul 19 03:19:34 2016 |
A. Tarifeņo, I. Dillman, J. Agramunt, A. Tolosa, J.L. Tain... | Electronics | Preamplifier and signal cable testing
We check all signal cables,the flat cable |
|
42
|
Tue Jul 19 04:09:32 2016 |
JL Tain A Tolosa R. Caballero A. Tarifeņo J. Agramunt | Last tube fit in the matrix!!! | The photo of great moment |
|
43
|
Tue Jul 19 04:34:57 2016 |
JL Tain A Tolosa R. Caballero A. Tarifeņo J. Agramunt | Photos of the electronics | Racks, crates, rear part of the preamps and
matrix with all tubes. |
9x |
44
|
Tue Jul 19 11:54:03 2016 |
JL Tain A Tolosa R. Caballero A. Tarifeņo J. Agramunt | Photos of the AIDA insertion hole with the 3He tubes | RIKEN tubes on the AIDA side with the SHV
right angle connector will protrude
by ~95mm in the HYBRID configuration (with |
|
45
|
Tue Jul 19 13:47:55 2016 |
A. Tarifeņo-Saldivia | Neutron efficiency - Scanning on Z preliminary result | We have carried out a scanning of
the neutron efficiency with respect to the
position along the Z-axis (AIDA and BEAM |
|
46
|
Wed Jul 20 08:58:13 2016 |
J.L. Tain, I.Dillman, R. Caballero, J. Agramunt, A. Tarifeņo, A. Tolosa, ... | Electronics | Distribution of connections of tubes in preamplifiers
and digitizers
|
|
47
|
Wed Jul 20 11:16:18 2016 |
J.L. Tain | Electronics | Pulser settings
We use two pulsers, Pulser1 for RIKEN, ORNL1 |
|
48
|
Wed Jul 20 11:40:00 2016 |
J.L. Tain, J. Agramunt, R. Baeturia, A. Tarifeņo, A. Tolosa, ... | 252Cf | First measurement with full 140 tube version
of the BRIKEN neutron detector
|
|
49
|
Wed Jul 20 11:49:00 2016 |
J.L. Tain, J. Agramunt, R. Baeturia, A. Tarifeņo, A. Tolosa, ... | Background | 18/07/2016
First background measurement overnight.
|
|
50
|
Wed Jul 20 12:00:15 2016 |
J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, ... | 252Cf | Measurement with the source at the center
Listmode: 160719_BRIKEN_252Cf_009.dlt
|
|
51
|
Wed Jul 20 16:07:58 2016 |
A.Tarifeņo, J.L. Tain, J. Agramunt, A. Tolosa, ... | 252Cf | 19/Jul/2016
Scanning of the efficiency as a function
of longitudinal position
|
|
52
|
Wed Jul 20 16:12:58 2016 |
S. Go, ... | CLOVER | Mounting of the CLOVER detectors in position
CLOVER G7: right of beam
|
|
53
|
Wed Jul 20 16:40:57 2016 |
S. Go, J.Agramunt, A.Tolosa, J.L. Tain,... | CLOVER | Calibration and resolution check
CLOVER detectors were calibrated in energy |
|
54
|
Wed Jul 20 16:50:12 2016 |
J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, S. Go... | 252Cf | Measurement of the 252Cf source at the central
position
|
|
55
|
Wed Jul 20 16:55:48 2016 |
K.Gabor, K. Matsui, A. Tarifeņo | Electronic | Drying of preamplifiers
PA8, PA9 and PA10 are removed and brought |
|
56
|
Wed Jul 20 17:05:59 2016 |
J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, ... | Background | Background measurement with remaining tubes
(95) overnight
|
|
57
|
Wed Jul 20 17:18:59 2016 |
J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, K. Gabor, S. Go, K. Matsui... | Electronics | Measurement with preamplifiers after drying.
Amplifiers are brought back after ~12h and |
|
58
|
Wed Jul 20 17:21:59 2016 |
J. Agramunt, A. Tolosa | DACQ | DACQ configuration parameters
The last configuration file used was: 160718Conf_BrikenFull_cal5.xlsx
|
|
59
|
Fri Jul 22 12:50:29 2016 |
J.L. Tain | END | 20/JUL/2016 20:00
End of the test measurements |
|
60
|
Fri Jul 29 09:52:28 2016 |
Keishi Matsui | accuracy of Cf-252's intensity in RIKEN | According to the reply from Japan
Radioisotope Association, the accuracy of
the Cf-252's intensity is +0%/-15%. |
|