AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
DESPEC |
 |
|
Message ID: 239
Entry time: Sun Apr 18 11:18:21 2021
|
Author: |
TD |
Subject: |
Sunday 18 April 12.00-16.00 |
|
|
12.18 DAQ continues file NULL/R30_490
ASIC settings 2019Dec19-16.19.51
slow comparator 0x64 -> 0xa
BNC PB-5
amplitude 2V
attenuation 1x
decay time 1ms
frequency 2Hz
12.22 all histograms zero'd
c. 12.30 219Rn setting
12.35 analysis of R30_490 - see attachment 1
13.15 system wide checks
Clock status test result: Passed 12, Failed 0
Understand status as follows
Status bit 3 : firmware PLL that creates clocks from external clock not locked
Status bit 2 : always logic '1'
Status bit 1 : LMK3200(2) PLL and clock distribution chip not locked to external clock
Status bit 0 : LMK3200(1) PLL and clock distribution chip not locked to external clock
If all these bits are not set then the operation of the firmware is unreliable
FEE64 module aida06 failed
FEE64 module aida07 failed
FEE64 module aida10 failed
Calibration test result: Passed 9, Failed 3
If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module
Base Current Difference
aida01 fault 0x7686 : 0x768a : 4
aida02 fault 0x941d : 0x9421 : 4
aida03 fault 0x7cd7 : 0x7cdb : 4
aida04 fault 0xb86d : 0xb871 : 4
aida05 fault 0x1a59 : 0x1a5f : 6
aida06 fault 0x4f45 : 0x4f4b : 6
aida07 fault 0x3bfc : 0x3c04 : 8
aida08 fault 0xc7ce : 0xc7d4 : 6
aida09 fault 0xb33b : 0xb33c : 1
White Rabbit error counter test result: Passed 3, Failed 9
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Base Current Difference
aida12 fault 0xa : 0xb : 1
FPGA Timestamp error counter test result: Passed 11, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last
Returned 0 0 0 0 0 0 0 0 0 0 0 0
Mem(KB) : 4 8 16 32 64 128 256 512 1k 2k 4k
aida01 : 23 10 3 1 1 2 1 3 3 3 6 : 36156
aida02 : 2 11 7 4 2 2 2 4 2 3 6 : 36048
aida03 : 21 8 6 4 5 4 3 3 2 3 6 : 36276
aida04 : 19 11 14 3 2 3 2 4 2 3 6 : 36324
aida05 : 25 4 5 4 4 2 1 3 2 4 6 : 37460
aida06 : 13 9 15 1 3 5 1 3 3 3 6 : 36812
aida07 : 19 10 6 1 3 3 2 2 4 3 6 : 37212
aida08 : 23 6 4 2 1 1 3 4 2 3 6 : 36044
aida09 : 11 5 5 1 2 4 3 3 2 3 6 : 35908
aida10 : 14 5 6 5 3 1 3 3 2 3 6 : 35744
aida11 : 16 5 3 0 3 3 1 4 2 3 6 : 35800
aida12 : 12 10 0 2 4 4 1 3 3 3 6 : 36544
Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same
FEE : aida01 => Options file size is 1025 Last changed Sat Apr 17 06:14:30 CEST 2021
FEE : aida02 => Options file size is 1014 Last changed Fri Apr 16 00:56:20 CEST 2021
FEE : aida03 => Options file size is 1014 Last changed Wed Apr 14 21:52:04 CEST 2021
FEE : aida04 => Options file size is 1025 Last changed Sat Apr 17 06:07:36 CEST 2021
FEE : aida05 => Options file size is 1025 Last changed Fri Apr 16 00:53:25 CEST 2021
FEE : aida06 => Options file size is 1014 Last changed Wed Apr 14 21:52:04 CEST 2021
FEE : aida07 => Options file size is 1014 Last changed Wed Apr 14 21:52:04 CEST 2021
FEE : aida08 => Options file size is 1014 Last changed Wed Apr 14 21:52:04 CEST 2021
FEE : aida09 => Options file size is 1014 Last changed Wed Apr 14 21:52:05 CEST 2021
FEE : aida10 => Options file size is 1014 Last changed Wed Apr 14 21:52:06 CEST 2021
FEE : aida11 => Options file size is 1014 Last changed Wed Apr 14 21:52:05 CEST 2021
FEE : aida12 => Options file size is 1025 Last changed Wed Apr 14 21:58:54 CEST 2021
*no faults detected by ~/oh/OptionsCheck.py
13.10 Detector biases & leakage currents OK - see attachment 2
Statistics - good events, adc data, disc data, info code 4 & 5, pause, resume, correlation info - attachments 3-12
per FEE64 rate spectra - attachments 13-14
1.8.L spectra - attachments 15-18
pulser peak widths
1 80
2 113
3 308
4 105
5 50
6 91
7 61
8 90
9 200
10 133
11 187
12 107
1.8.H spectra - attachments 19-20
1.8.W spectra - attachments 21-22
ucesb - attachment 23
implant rates DSSSD#1-3 to c. 1kHz
grafana - AIDA DSSSD leakage currents for previous 2 days - attachment 24
merger & merger/tape server/mbs data tranfser - attachment 25
no recent merger warning/error messages
13.30 225At setting |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|