ID |
Date |
Author |
Subject |
|
107
|
Mon Dec 16 14:48:02 2019 |
NH TD | AIDA 16.12.19 |
DSSD Snout was reassembled to install the second bPlast (downstream)
(fig 1)
Ribbon cable for aida10 was replaced (DSSD3 upper)
All DSSDs bias again (see fig 2)
Alpha run in progress - channels OK with 2 exceptions (fig 3)
- aida10 still recording minimal to no counts
- channel 55 aida08 reading lots of data even with comparator 0x64 - corresponds to DSSD1 strip 0 (bottom strip)
possible ground/cable issue? will check when access possible (fig 4)
Unusual MIDAS error noticed (fig 5) - restart of HTTPd fixed it |
Attachment 1: 20191213_103710.jpg
|
|
Attachment 2: 161219_aida_DSSDs.png
|
|
Attachment 3: 161219_rates.png
|
|
Attachment 4: 161219_aida0855.png
|
|
Attachment 5: midasmessage131219.png
|
|
|
106
|
Wed Dec 11 10:58:52 2019 |
NH | Initital Dec. Analysis |
Fig 1: 2D hit pattern of each AIDA DSSD for MBS run 24 (40Ar primary beam, centred)
Low statistics in DSSD2 partly due to low energy deposition
Fig 2: Position according to TPC42 (ca. 2 metres before AIDA)
Fig 3: 2D hit pattern of DSSD 1 for MBS run 37 (34Si fragments, no PID gate)
-
Bottom part of AIDA noticably cut off - rotation in DSSD also noticed (not an issue)
Snout was investigated and platform was discovered to not be fully level by Bogdan using a laser level - he has adjusted it so that the AIDA snout is now level to within 1 mm.
(The beam right side was previous 4-6mm below the beam left side)
Y cut in fragment run is currently of unknown origin. Optics are being investigated
-
Fig 4: Hit pattern of 40Ar with strip offsets fixed (no missing strips)
Fig 5: Energy loss in DSSD1 of 40Ar primary beam
Fig 6: TPC 41, TPC 42, AIDA X comparison
Fig 7: " Y comparison
Fig 8-11: As above for 34Si run
Y cut unknown still.
Statistics of 40Ar file
======================================
AIDA Unpacker Analysis
--------------------------------------
First event: Fri Dec 6 05:21:42 2019
Last event : Fri Dec 6 05:26:17 2019
Duration : 274.604 seconds
--------------------------------------
MBS Events (M): 131808426 (481 kHz)
AIDA Words (W): 110608065 (403 kHz)
ADC Words (A): 36869349 (134 kHz)
ADC Decays (D): 36782621 (134 kHz)
ADC Implants (I): 86728 (316 Hz)
Info Words (F): 73738716 (269 kHz)
SYNC48 Words (S): 36869354 (134 kHz)
SYNC63 Words (s): 36869354 (134 kHz)
PAUSE Words (P): 4 (0 Hz)
RESUME Words (R): 4 (0 Hz)
Discrim Words(D): 0 (0 Hz)
Unknown Words(U): 0 (0 Hz)
Timewarps (T): 0 (0 Hz)
Dead Time (X): 0.113631 s (0.0413799%)
======================================
======================================
Individual FEE Overview
======================================
FEE | W A D I F S s P R D U T X
1 | 4383018 1461006 1439401 21605 2922012 1461006 1461006 0 0 0 0 0 0
2 | 11468351 3822783 3814577 8206 7645568 3822783 3822783 1 1 0 0 0 0.00733666
3 | 4353624 1451208 1428500 22708 2902416 1451208 1451208 0 0 0 0 0 0
4 | 8355002 2785000 2755870 29130 5570002 2785001 2785001 0 0 0 0 0 0
5 | 1133544 377848 376489 1359 755696 377848 377848 0 0 0 0 0 0
6 | 9617940 3205980 3205767 213 6411960 3205980 3205980 0 0 0 0 0 0
7 | 4385477 1461825 1460962 863 2923652 1461826 1461826 0 0 0 0 0 0
8 | 15105135 5035043 5032985 2058 10070092 5035045 5035045 1 1 0 0 0 0.0126429
9 | 37487211 12495735 12495422 313 24991476 12495736 12495736 2 2 0 0 0 0.0936512
10 | 110316 36772 36772 0 73544 36772 36772 0 0 0 0 0 0
11 | 12329904 4109968 4109763 205 8219936 4109968 4109968 0 0 0 0 0 0
12 | 1878543 626181 626113 68 1252362 626181 626181 0 0 0 0 0 0
Statistics of 34Si file
======================================
AIDA Unpacker Analysis
--------------------------------------
First event: Sat Dec 7 05:20:22 2019
Last event : Sat Dec 7 05:23:51 2019
Duration : 208.635 seconds
--------------------------------------
MBS Events (M): 148639738 (714 kHz)
AIDA Words (W): 132033086 (634 kHz)
ADC Words (A): 44011024 (211 kHz)
ADC Decays (D): 43874168 (210 kHz)
ADC Implants (I): 136856 (657 Hz)
Info Words (F): 88022062 (423 kHz)
SYNC48 Words (S): 44011028 (211 kHz)
SYNC63 Words (s): 44011028 (211 kHz)
PAUSE Words (P): 3 (0 Hz)
RESUME Words (R): 3 (0 Hz)
Discrim Words(D): 0 (0 Hz)
Unknown Words(U): 0 (0 Hz)
Timewarps (T): 0 (0 Hz)
Dead Time (X): 0.0806864 s (0.0386734%)
======================================
======================================
Individual FEE Overview
======================================
FEE | W A D I F S s P R D U T X
1 | 6814953 2271651 2255653 15998 4543302 2271651 2271651 0 0 0 0 0 0
2 | 14012582 4670860 4668280 2580 9341722 4670860 4670860 1 1 0 0 0 0.0088226
3 | 7618290 2539430 2523614 15816 5078860 2539430 2539430 0 0 0 0 0 0
4 | 8928934 2976310 2950380 25930 5952624 2976312 2976312 0 0 0 0 0 0
5 | 1631415 543805 531957 11848 1087610 543805 543805 0 0 0 0 0 0
6 | 10908021 3636007 3633830 2177 7272014 3636007 3636007 0 0 0 0 0 0
7 | 5644296 1881432 1864053 17379 3762864 1881432 1881432 0 0 0 0 0 0
8 | 14863755 4954585 4931620 22965 9909170 4954585 4954585 0 0 0 0 0 0
9 | 36745143 12248379 12244135 4244 24496764 12248380 12248380 2 2 0 0 0 0.0718638
10 | 198573 66191 66191 0 132382 66191 66191 0 0 0 0 0 0
11 | 15101900 5033966 5018830 15136 10067934 5033967 5033967 0 0 0 0 0 0
12 | 9565224 3188408 3185625 2783 6376816 3188408 3188408 0 0 0 0 0 0
All good (no deadtime issues) |
Attachment 1: AIDA_HitDec19.png
|
|
Attachment 2: TPCs.png
|
|
Attachment 3: AIDA-fragments.png
|
|
Attachment 4: AIDA_XY.png
|
|
Attachment 5: AIDA_E.png
|
|
Attachment 6: TPC_41_42_AIDA_X.png
|
|
Attachment 7: TPC_41_42_AIDA_Y.png
|
|
Attachment 8: AIDA_XY_34Si.png
|
|
Attachment 9: AIDA_E_34Si.png
|
|
Attachment 10: TPC_41_42_AIDA_X_34Si.png
|
|
Attachment 11: TPC_41_42_AIDA_Y_34Si.png
|
|
|
105
|
Sat Dec 7 02:39:02 2019 |
DK, NH et al. | 34Si Fragments PID |
See attached
Tentative PID (DK) as attachment #3
DESPEC LISE++ file as attachment #4 |
Attachment 1: 2019-12-07-PID.png
|
|
Attachment 2: 59_AM.png
|
|
Attachment 3: pid.png
|
|
Attachment 4: Ar-300_S4_updated_matter_DEPSEC_34Si.lpp
|
Version 11.0.88
{============================= Main Part ======================================}
[general]
File = C:\Users\Nic\Documents\Ar-300_S4_updated matter_DEPSEC 34Si.lpp
Date = 06-12-2019
Time = 23:05:23
Configuration = GSI\FRS-TA2-S4_2014.lcn
Optionsfile = GSI_FRS_2012.lopt
Title = GSI FRS - TA2 -2014
BlockStructure = MMMDSWMDMSAMMMMMMWMMDMSDMMMMMMMMMMMMMMMMMMSMMMWMMMMMMMMMMMMMMMMMM
[settings]
A,Z,Q = 40Ar18+ ; Mass ElementName Charge+ Beam
Energy = 300 MeV/u
Intensity = 1e+7 pps ; enA,pna,pps,kW
RF frequency = 20 MHz
Bunch length = 1 ns
Settings on A,Z = 34Si ; Mass ElementName Charge+ Beam
[OpticsBeam]
BX = 1 (±)mm ; one-half the horisontal beam extent (x)
BT = 5 (±)mrad ; one-half the horisontal beam divergence(x')
BY = 1.5 (±)mm ; one-half the vertical beam extent (y)
BF = 3.33 (±)mrad ; one-half the vertical beam divergence (y')
BL = 0 (±)mm
BD = 0.05 (±)% ; one-half of the momentum spread (dp/p)
ShiftX = 0 mm ; beam respect to the spectrometer axis
AngleX = 0 mrad ; beam respect to the spectrometer axis
ShiftY = 0 mm ; beam respect to the spectrometer axis
AngleY = 0 mrad ; beam respect to the spectrometer axis
Scheme Angle = 0 degrees
ShapeX = 1
ShapeT = 1
ShapeY = 1
ShapeF = 1
ShapeL = 1
ShapeD = 1
OptBeam_X = 1 (±)mm
OptBeam_T = 30 (±)mrad
OptBeam_Y = 1 (±)mm
OptBeam_F = 30 (±)mrad
OptBeam_L = 0 (±)mm
OptBeam_D = 1.5 (±)%
[options]
NP simple = 32 ; Number of points in distribution
NP charge states = 32 ; Number of points in distribution
NP wedge = 32 ; Number of points in distribution
Charge states = No ; No & Yes
CutEdgeEffect = 1 ; 1-Yes. Default, 0-no - for extended configurations
Prim.beam scatter = 0 ; 0-without, 1-with
Delta peak = 0 ; 0-without, 1-with
BrhoMeanMax = 1 ; 0-Max, 1-Mean
BrhoMeMaLeRi = 3 ; 0-Max, 1-Mean, 2-Left, 3-Right /for fission/
[target]
Target contents = 0,4,1,9.012 ; Nomer,Z,Atoms,Mass
Target thickness = 1,2500.7,1.848,0,0,0 ; State,Thickness,density,angle,SolidGas,..
Target fusion compound = 0
Targ use for Q-states = 1
Target Defect = 1,0 ; [0] choice - % or micron at 0 degree, [1]=value;
Degrader contents = 0,22,1,47.9 ; Nomer,Z,Atoms,Mass
Degrader thickness = 0,35,4.519,0,0,0 ; State,Thickness,density,angle,SolidGas,..
Degra use for Q-states = 1
Degrader Defect = 1,0 ; [0] choice - % or micron at 0 degree, [1]=value;
[mechanism]
Reaction = 0 ; 0 - fragm, 1 - fusion-resid, 2 - fusion-fission
CalcOther = 1 ; calculate other reactions
V calculation = 5 ; 0 - constant, 1 - Borrel, 2 - Rami, 3-convolution, 4-two body reaction
V_opt/Vbeam = 1 ; default 1
Velocity_exceed = 1 ; 0 - without, 1-with - two-body recations velocity corrections
Binding Energy for Vf/Vp = 8 MeV ; Binding energy for Borrel's expression
Shift for Vf/Vp calc = 0
Prefragment_Rami = 1 ; 1-Yes, 0-No
Sigma0 = 90 MeV/c ; default 90
SigmaD = 0 MeV/c ; default 200
SigmaM = 87 MeV/c ; default 87
Asymmetry = 0 % ; default 0
Method v-sigma = 0 ; 0 - Goldhaber, 1-Morrissey,2-Friedman,3-Convolution
G_Surface = 0.95 MeV/fm^2
Symmetry around half_Ab = 1 ; 1 - yes, 0-no
Pfaff pickup correction = 0 ; 1 - yes, 0-no
ChargeExchangePfaff = 0 ; 1 - exclude, 0-forget
Sigma corr 0 = 0 ; Coulomb energy
Sigma corr 1 = 0 ; Projectile mass
Friedman mode = 2 ; 0-Qgg, 1-Surface, 2-Qgg+Surface
Prefragment_Fri = 1 ; 1-Yes, 0-No
Coulomb_Friedman = 1 ; 1-Yes, 0-No
K_Morr = 8 MeV/A ; E/A=8MeV/A default; D.Morrissey coef.
K_MorHalf = 8 MeV/A ; E/A=8MeV/A default at Afrag=Aproj/2; D.Morrissey coef.
AA_fast = 0 ; 1-Yes, 0-No
BarrierShape = 1 ; 0-classical, 1-quantum mech.
H_omega = 5 MeV ; default 3
Probabilty_CN = 0 ; 0/1 use Prbabilty for CN formation
UseVanishing = 1 ; 1-Yes, 0-No
VanishMode = 0 ; 0-Sierk, 1-Cohen
NuclPotential = 1 ; 0-Bass, 1-WS
WS_V0 = 105 MeV
WS_R0 = 1.12 fm
WS_a = 0.75 fm
FusDiffuseness = 1
Width Coef = 1 ; default 1; for Leon's charge state distribution
gZt Correction = 1 ; default 1; Leon's C.S.D.
PowerCoefLeon = 0.477 ; default 0.477; Leon's C.S.D.
Cross section = File ; Fit & File
Charge method = 3 ; charge calculations method 0-5
EPAX Cross Section = 4 ; cross section calculations method 0-4
SR Cross Section = 2 ; EPAX for SR 0-2
Energy Loss = 2 ; energy loss calculation method 0-3
Anglular straggling = 1 ; 0-LISE, 1-ATIMA
StragglingCoef1 = 0.217
StragglingCoef2 = 1.12
Energy straggling = 1 ; 0-LISE, 1-ATIMA
EnergyStragMethod = 0 ; 0-integrate, 1-table
EnergyStragShape = 0 ; 0-Gauss, 1-Landau-Vavilov
EquilThickness = 1 ; 0-Charge, 1-Global
MassMethod = 0 ; 0-DB+calcul, 1 + just calcul
MassDataBase = 0 ; 0-A&W, 1-User ME
Mass formula = 2 ; 0-LDM, 1-Myerer, 2: 1+corrections
UseChargeForEnergy = 2 ; 0-No, 1-Yes, 2-Auto
EnergyValueAuto = 30 ; default value 30 MeV/u
EquilibriumMode = 1 ; 0-Equil, 1-NonEquil
UB_Global = 70 ; default 70 MeV/u
MinZ_Global = 2 ; default Z>=29
ChargeStateOptim = 1 ; 0-No, 1-Yes
ZmQ_AfterReactn = 0 ; default 0 (full stripped)
EPAX_p_Norm = 1
EPAX_p_Un = 1.65
EPAX_p_Up0 = 1.79
EPAX_p_Up1 = 0.00472
EPAX_p_Up2 = -1.3e-5
EPAX_p_H = 1
[fission]
FisAngDistShape = 0 ; 0-isotropic; 1-anisotropic
FisMomCutForAngDist = 2 ; 0-dont use; 1-use just MatrixKinematics; 2-use for all; (default 2)
OddEvenCorrections = 1 ; 0-dont use; 1-use
PostScissionEvaporation = 1 ; 0-dont use; 1-use
DeexcitFunctionPoints = 0 ; 0- average deexcitation energy; 1- 3 points; 2 - manually
FisEXmanually = 20 ; Excitation energy manually
FisCSmanually = 1000 ; Cross section manually
FisTXEmethod = 1 ; 0-from Edissipated, 1 from Q-value
Fis_f = 0.0045 ; default 0.0045
FisEXsigma = 5.5 MeV ; default 5.5
FisCS_Global = 1e-11
FisCS_TKE = 1e-8
FisBeta1 = 0.625 ; deformation of light fragment
FisBeta2 = 0.625 ; deformation of heavy fragment
FisTKE_d = 2 fm ; d-param in Wilkins formula
FisBetaFit = 1 ; 0-manual, 1-fit
N0 = 83 ; default 82
dU0 = -2.65 ; default -2.5
C0 = 0.7 ; default 1.4
cpol0 = 0.65 ; default 0.65
width0 = 0.63 ; default 0.63
N1 = 90 ; default 90
dU1 = -3.8 ; default -5.5
C1 = 0.15 ; default 0.16
cpol1 = 0.55 ; default 0.55
width1 = 0.97 ; default 0.97
[charge_suppression]
FragInd = 1e-3
FragTotal = 1e-5
BeamInd = 1e-20
BeamTotal = 1e-20
[convolution]
Convolution mode = 1 ; 0-Qgg, 1-Surface, 2-Qgg+Surface
SigmaConv = 91.5 MeV/c ; default 90 for Convolution
CoefConv_0 = 3.344
CoefConv_1 = 3
CoefConv_2 = 2.936
ShiftConv_0 = 0.158
ShiftConv_1 = 0.149
ShiftConv_2 = 0.153
[evaporation]
NP evaporation = 32 ; Number of points in distribution
EvapMethod = 2
StateDensityMode = 2 ; 0, 1+pairing, 2+shell
EvapUnstableNuclei = 1 ; 0 - only stable,1 +unstable
Tunnelling = 1 ; 1-Yes, 0-No
AvoidUnboundCS = 1 ; 1-Yes, 0-No
ProtectedChannels = 1 ; 1-Yes, 0-No
R_Evaporation = 5.7 fm ; correction for the effective Coulomb barrier
Mode_Apf_manual = 0 ; 1-manual, 0-auto
Energy_in_T = 2 ; default 2
EvaporationVelocity = 0 ; 0 - quality, 1 -fast
DeltaOddEvenEvap = 12
DeltaOddEvenFission = 14
BreakupTemperature250 = 4.7
BreakupTemperature150 = 5.9
BreakupTemperature050 = 8
BreakupDiffuseness = 0.05
DissipationKramers = 0 ; 0 - no, 1 - use
DissipationStepFunction = 1 ; 0 - no, 1 - use
DissipationBeta = 1 ; default 2.0
mode_1n = 1 ; 1-Yes, 0-No
mode_2n = 0 ; 1-Yes, 0-No
mode_1p = 1 ; 1-Yes, 0-No
mode_2p = 0 ; 1-Yes, 0-No
mode_a = 1 ; 1-Yes, 0-No
mode_d = 0 ; 1-Yes, 0-No
mode_t = 0 ; 1-Yes, 0-No
mode_3he = 0 ; 1-Yes, 0-No
mode_fis = 1 ; 1-Yes, 0-No
mode_brk_up = 1 ; 1-Yes, 0-No
mode_gamma = 0 ; 1-Yes, 0-No
[fission_barrier]
FissionBarrierFactor = 1
FissionBarrierMode = 1 ; #0-4
OddEvenCorrections = 1 ; 1-Yes, 0-No
ShellCorrections = 1 ; 1-Yes, 0-No
FB_InOutMax = 2 ; #0-2 - in/out/max
ModeForUser = 1 ; #0-2
NdeltaOddEven = 2.5
ZdeltaOddEven = 9
[excitation_energy]
GeomAA_Correction = 1 ; 0 - don't use,1 - use -default
Thermalization = 0 ; 1-Yes, 0-No
ThermaTimeCoef = 3e+0 ; 2.1e-22 MeV *s/e(t)
Friction = 0 ; 0 - off,1 - on
Ev_A_SigmaCoef = 9.6
G_FrictionCoef1 = 6.5
G_FrictionCoef2 = 0
G_FactorCoef1 = 1.5
G_FactorCoef2 = 2.5
DepthHole = 40
EnergyCoef_CB0 = 0
EnergyCoef_CB1 = 13.3
EnergyCoef_CB2 = 0
SigmaCoef_CB0 = 0
SigmaCoef_CB1 = 9.6
SigmaCoef_CB2 = 0
D_MeanTemp = 13
AA_factor = 1
ApplyLimitTemp = 0 ; 1-Yes, 0-No
[evapauto]
tun_a0 = -0.61392
tun_a1 = 0.44559
tun_a2 = 0.12008
A_Bound = 300 ; mass
A_Pairing = 70 ; mass
[plot]
Start target = Detector ; Detector & RF
Start of TOF = M7
Stop of TOF = M34
dE-detector-1st = M28
dE-detector-2nd = M7
TKE-detector = M41
PlotBrho = D1
PlotWedge = S4
X-detector = M19
Y-detector = M19
Tilting = M1
Stopper = W1
RO_Wedge = n
ConditionBlock = A10
Plot threshold = 1e-10 pps ; minimal value for plot scale
Shift of TOF for RF = 0 ns ; for dE-TOF plot with RF
Fraction of RF trigger = 2
UseCondition = 0
TKE_calibration = 1,1,0,MeV ; Input PV(0) or CH(1), A, B, dimension
[cs_file]
UserDiffCS = 0 ; Number of User Diff CS saved in this file
AppendOverwrite = 1
AttachedInside = 1
ShowCSinPlot = 1
Chi2 = 1
CSfilename =
[sec_reactions]
NP sec.reactions = 16 ; Number of points in distribution
Secondary reactions = 0 ; 0/1 - use secondary reactions in calculations
fiss_FilterUse0 = 1
fiss_FilterUse1 = 1
fiss_FilterUse2 = 1
fiss_ellipse = 5
fiss_NdeltaTop = 0
fiss_ZdeltaTop = 0
fiss_NdeltaBot = 25
fiss_ZdeltaBot = 20
frag_FilterUse0 = 3
frag_FilterUse1 = 3
frag_FilterUse2 = 3
frag_ellipse = 4
frag_NdeltaTop = 5
frag_ZdeltaTop = 5
frag_NdeltaBot = 6
frag_ZdeltaBot = 6
[abrasion_fission]
... 2087 more lines ...
|
|
104
|
Fri Dec 6 20:34:56 2019 |
CA, DK, NH | Dec 6 Night Shift |
21.35 AIDA DAQ currently running, TapeService on, not writing to storage
System wide checks ok
Attachments 1-3: temperatures, good event stats, detector bias/leakage currents ok
22.04 waiting for beam. Possibility of fragments, further news to come
22.43 expecting fragments with 300 MeV/u 40Ar primary beam
histograms zeroed
00.31 still waiting for beam
detector bias/leakage currents ok (see google spreadsheet)
FEE64 temps and good event statistics ok
02:02 still awaiting beam - issues with interlock
03:17 beam tuning - centering primary beam on target
03:43 beam reached AIDA, fragmentation beam (see elog entry by DK for PID)
primary beam 300 Mev/u 40Ar, fragments in region of 34Si
started writing to file Dec19/R2 at rate ~ 4000 kb/s
histograms zeroed
03.49 system wide checks all ok
attachments 4-6:
temperatures ok
bias/leakage currents ok
good event statistics ok, except aida09 running slightly fast
04:03 attachments 7-8 - 1.8.H energy spectra
- no high energy events seen in aida10 -> checked ASIC control
- still no high energy events in aida10
04:09 attachment 9 - rate spectra for all FEE modules
04.16 attachment 10 - 1D hits for all FEE modules
04.19 beam stop - file R2_3
04:52 beam back on - file R2_6 - rate of ~ 5000 ions per spill
06:13 temperatures ok
good event statistics ok (aida09 still running fast) - attachment 11
detector bias/ leakage currents ok - see google spreadsheet
all system wide checks ok
1.8.H spectra - attachments 12/13
06:20 attachment 14 - rate spectra for all FEE modules
attachment 15 - 1D hits for all FEE modules
still no high energy events in aida10
|
Attachment 1: 191206_2139_temp.png
|
|
Attachment 2: 191206_2140_bias.png
|
|
Attachment 3: 191206_2140_stats.png
|
|
Attachment 4: 191107_0351_temp.png
|
|
Attachment 5: 191207_0352_bias.png
|
|
Attachment 6: 191207_0352_stats.png
|
|
Attachment 7: 191207_18h_2.png
|
|
Attachment 8: 191207_18H_1.png
|
|
Attachment 9: 191207_0408_rates.png
|
|
Attachment 10: 191207_0415_statspec.png
|
|
Attachment 11: 191207_0614_stats.png
|
|
Attachment 12: 191207_0606_18H_1.png
|
|
Attachment 13: 191207_0609_18h_2.png
|
|
Attachment 14: 191207_0618_rate.png
|
|
Attachment 15: 191207_0619_hits.png
|
|
|
103
|
Thu Dec 5 22:00:11 2019 |
CA, DK, NH | Dec 5 Night Shift |
15:30
TD found the merger was running at ~43 Mb/s
The odd-numbered FEE64s all had significant (>200k) disk rates
ALL LEC/MEC fast discriminators and increased slow comparator from 0xa to 0x14
Merger rate dropped to a more nominal value like 2.3 Mb/s
Settings not saved
40Ar beam 300 MeV/u expected, but SIS problem at the moment.
23:00
Leak currents are stable as previously (20-21 Nov) and entered to Google spreadsheet
S4 vault is closed.
Presently AIDA is running without tape storage
System wide checks:
Master clock / aida01 sync reports error as expected;
All white rabbit decoders fail, but they are just off by 1 bit (no problem)
Everything else okay.
Stats look okay, but keep in mind above discriminator threshold -- see attachment 1
Temperatures fine -- see attachment 2
MBS looks okay -- we see a possible hint for the 1 bit difference with an "unexpected data word" -- see attachment 3
Data rate is 1.7 Mb/s, merger okay -- see attachment 4
23:30
Now we wait for the beam
00:05
Attachments 5-11: results of system wide checks after collecting sync errors and WR errors from baseline - all ok
00:25
Histograms zeroed at 00:25.
00:35
1.8.L pulser peak widths
FEE64 width(ch)
1 92
2 131
3 75
4 97
5 89
6 117
7 90
8 87
9 223
10 58
11 112
12 126
no peak observed on aida09
For comparison, pulser peaks measured on 13/11/19:
pulser peak widths:
FEE width (ch)
aida01 - 130
aida02 - 168
aida03 - 83
aida04 - 83
aida05 - 200
aida06 - 148
aida07 - 250
aida08 - 81
aida09 - 134
aida10 - 190
aida11 - 333
aida12 - 149
2:00
40Ar beam tuning is finished. Intensity at S4 is about 3 kHz per spill. 1 spill per / 10 seconds
MBS system cannot be used for online merging tonight, because NeuLAND group left their connection open
We can merge offline using the White Rabbit timestamps
Remove the beam blocking steel flange from in front of AIDA et al.
Writing to file Dec19/R1
03.20 Rate, Stat, 1.8.L, 1.8.H, 1.8.W spectra - attachments 13-20
Observe
light ions in 1.8.L spectra
low energy (<1.5GeV) heavy ions in 1.8.H spectra with most heavy ions stopping in DSSSDs 1 & 2
Stat hit patterns indicate broad (c. 5cm FWHM) centred heavy ion peak
03.24 Detector biases & leakage currents OK - attachment 21
FEE64 temperatures OK - attachment 22
Good event statistics (12 of 12 < 50k) - attachment 23
System wide checks
List All Sync Error Counters - attachment 24
List all Sync Received Counters - attachment 25
Baseline invoked earlier c. 00.05
Collect all SYNC Error Counters for Baseline
Collect All White Rabbit status Error Counters for Baseline
04:07 2.6 mm Al degrader. 300Hz beam. 300 MeV/u 40Ar beam.
04.14 pulser peak seen in aida09 after performing ASIC check-load earlier in shift - width = 219 ch
04.45 1.8.H energy spectra (attachment 26/27)
06:08 attachments 28-30: detector bias/leakage currents ok
temperatures ok
good event statistics ok
all system wide checks ok
08:49 DAQ stop/start. TapeServer writing to No Storage. End of beam.
|
Attachment 1: 05-Dec-Stats.png
|
|
Attachment 2: 05-Dec-Temps.png
|
|
Attachment 3: 05-Dec-MBS.png
|
|
Attachment 4: 05-Dec-MergerRates.png
|
|
Attachment 5: adccalib.png
|
|
Attachment 6: clockstatus.png
|
|
Attachment 7: meminfo.png
|
|
Attachment 8: syncerrors.png
|
|
Attachment 9: syncpulses.png
|
|
Attachment 10: wrdecoder.png
|
|
Attachment 11: 1.png
|
|
Attachment 12: 2.png
|
|
Attachment 13: 3.png
|
|
Attachment 14: 4.png
|
|
Attachment 15: 5.png
|
|
Attachment 16: 6.png
|
|
Attachment 17: 7.png
|
|
Attachment 18: 8.png
|
|
Attachment 19: 9.png
|
|
Attachment 20: 10.png
|
|
Attachment 21: 11.png
|
|
Attachment 22: 12.png
|
|
Attachment 23: 13.png
|
|
Attachment 24: 18hOdd.png
|
|
Attachment 25: even18h.png
|
|
Attachment 26: cja_bias.png
|
|
Attachment 27: cja_stats.png
|
|
Attachment 28: cja_temp.png
|
|
Attachment 29: pulser_peaks1.png
|
|
Attachment 30: pulser_peaks2.png
|
|
|
102
|
Wed Nov 27 16:33:50 2019 |
NH | New Bias & Waveforms |
HV core (-160 V) now attached to bottom 3 FEEs (with grounded kapton)
DSSDs bias OK, leakages unchanged
No appreciable change in waveforms
Some estimates
aida01/aida05/aida09 VHF noise: 5 channels frequency = 10 MHz ?
All FEEs:
Oscillation at ~25-35 channels = 1.4 to 2 MHz ?
Varies a bit with each FEE.
No obvious issue with upper FEEs unsure why 10 MHz is present - fee09 has been replaced but showed it before as well.
Lower frequency noise on every FEE may be ground noise from e.g. switching PSU?
No more access to S4 to check thoroughly again.
|
Attachment 1: waves_new_odd.png
|
|
Attachment 2: waves_new_even.png
|
|
|
101
|
Mon Nov 25 13:25:51 2019 |
NH | Waveforms |
> AIDA Waveforms immediately after beam at 21.11.19 8am or so.
Update 25.11.19 with all FEE64 waveforms added
Things to note:
aida01, aida05, aida09 are all on TOP - show VHF noise, esp. aida01?
TOP is where HV is connected (-ve core)
Everything noisy in general though |
Attachment 1: wave_odd.png
|
|
Attachment 2: wave_even1.png
|
|
|
100
|
Fri Nov 22 08:55:31 2019 |
NH | Waveforms |
AIDA Waveforms immediately after beam at 21.11.19 8am or so. |
Attachment 1: waves_21119_even.png
|
|
Attachment 2: 21119_waves_odd.png
|
|
Attachment 3: waves_21119_even_z.png
|
|
Attachment 4: waves_21119_odd_z.png
|
|
|
99
|
Wed Nov 20 20:36:14 2019 |
CA, CB, DK, NH | 20/21 November Night Shift |
21:35 (NH) - A few merger timestamp errors were appearing - Decided to powercycle all FEE64s, all checks OK.
No Screenshot due to forgetting before I closed merger
FEEs were often not stopping properly and needing merger reset as well.
Beam Plan:
40Ar 300 MeV/u Primary Beam
and/or
34Si Fragment from Ar
22:45
FEES were restart near these timestamps in the TapeData, evidently
-rw-rw-r--. 1 npg npg 2.0G Nov 20 21:17 R1_361
-rw-rw-r--. 1 npg npg 2.0G Nov 20 21:36 R1_362
Run number was kept the same (Tape Server was not stopped @ FEE restart?)
All system wide checks are okay (except of course the master clock fails)
Temps look okay (attachment 1, dated 21/Nov by mistake)
Leak currents all around 0.9 uA as before (attachment 2)
Stats are behaved. We don't have beam yet (attachment 3)
Merger and tape server happy (see attachment 4)
Data rate histos while beam is still off (see attachment 5)
==Nov 21==
0:00
TD noted that the data rates have gone up at some point (about double).
Now we increased the slow comparator thresholds from 0xa to 0xb (data rate drops from 24 Mb/s to 16 Mb/s).
We increased it again from 0xb to 0xc, but near that time the beam was also turned on, and the data rate was about 25 Mb/s.
Beam was turned on near R1_472 or a bit earlier.
TD requested waveforms, but at the moment they are not appearing. Will enable WFs after a condition change.
Beam seen in all three DSSSDs, see attachment 6. Beam was turned off before I could get more screencaps.
Degrader change ... near R1_481
00:32 waveforms now enabled under Run Control. They also appear to be enabled on ASIC1 undre LED and Waveform controls. However, we are unable to see the waveforms in the relevant histograms.
00:44 3 g/cm^2 inserted R1_488
00:45 3.2 g/cm^2 degrader in R1_489.
1:00 Several degraders and combinations are being tested since last note.
1:10 It seems 1 g/cm^2 of energy loss is not accounted for by LISE++. Trying to understand the reason.
1:28 Satisfied with a total degrader thickness of 2.6 g/cm^2 -- seems to calculate that fragments should implant mainly in DSSSD2? Issue seems to be resolved updating the LISE++ file.
Near R1_507
Can see the implant samples attached as #7 and #8
1:43 According to the stats, most of the beam energy is being deposited in DSSSD1. See #9.
Present beam burst repetition is around 0.1 Hz (each ~10 sec)
2:00 better example of stats histos (see #10)
2:45 tuning the FRS degraders (ones for RIB separation rather than AIDA implantation depth).
3:17 Just started fragments. Disregard previous filenames with "34Si". Finally no degrader in the FRS was needed to get Z separation.
R1_559
As the DAQs are not synchronized, a quick call to the BNC PB-5 pulser shifiting from 2 -> 50 -> 2 Hz was used to make a signal burst.
S4 degrader is 5.2 g/cm^2
3:22 Updated biases in Google Sheets.
Having trouble seeing the fragments in the AIDA histograms. Since we just started fragments, zero the histos.
3:35 Attachment 11, with 34Si fragments as Stats. Difficult to click Update on the Rates at the right time.
From the beam monitors, fragments should be at around 3000 cps.
3:45 (Up to now, some optics tuning with slits, etc)
3:46 Optics settings accepted, AIDA at R1_572, accept this condition for data runs on ~34Si fragments.
3:47 Sent a pulser burst 2 -> 50 -> 2 Hz
3:50 zero histograms again.
Temperatures attachment 12
Biases attachment 13
Statistics attachment 14
Merger / Tape Server attachment 15
3:56
StatHistos attachment 16
4:09 FRS starts new run #56, AIDA R1_583. Pulser 2-> 50 -> 2 Hz
4:56 Stat histograms in attachment 17. Not zeroed since noted at 3:50 |
Attachment 1: 21-11-temps.png
|
|
Attachment 2: 20-11-biases.png
|
|
Attachment 3: 20-11-stats.png
|
|
Attachment 4: 20-11-merger-ts.png
|
|
Attachment 5: 20-11-statshistos.png
|
|
Attachment 6: 21-11-layout5histos.png
|
|
Attachment 7: 34si-implants-130-1.png
|
|
Attachment 8: 34si-implants-130-2.png
|
|
Attachment 9: 34si-stats-histos.png
|
|
Attachment 10: stats-histos-2am.png
|
|
Attachment 11: 34si-fragments-stats.png
|
|
Attachment 12: temperatures-fragments.png
|
|
Attachment 13: biases-fragments.png
|
|
Attachment 14: stats-fragments.png
|
|
Attachment 15: merger-tapeserver-fragments.png
|
|
Attachment 16: stathistos-fragments.png
|
|
Attachment 17: stathistos-5am.png
|
|
|
98
|
Tue Nov 19 21:22:06 2019 |
CA, CB, DK, NH | 19/20 November 2019 - overnight shift |
19 November
Waiting for 40Ar beam to be delivered for tests.
Objective today is to assess degrader thickness suitable to stop 40Ar in 2nd DSSD layer.
Energy expected at 300 MeV/A, around 1E3-4 beam intensity - to be determined.
Tomorrow fragments could be expected.
DSSD / FEE64 pairing as follows
DSSD3 (most downstream): FEEs 9,10,11,12
DSSD2 (middle): FEEs 5,6,7,8
DSSD1 (most upstream): FEEs 1,2,3,4
+plastic scintillator furthest upstream
New AIDA Google spreadsheet created with leakage current, to monitor them during beamtime. First entry added. Bias attached.
https://docs.google.com/spreadsheets/d/12hgbrywB10hGsKt5uc2HnLfZymh8_uvM6cEASbbqnBE/edit#gid=813167023
DESPEC platform has already been moved. No effect on rates - see attached. FEE1 is missing one channel. FEE9, 10 noisy as expected. See attached.
System wide checks *all OK* except
Master clock failed (normal)
ADC calibration fails on 6 and 12. Recalibrated via FADC Align. All OK now.
MBS transfer OK. See attached.
Good event statistics OK. See attached.
FEE64 temperatures OK. See attached.
Merger and data rate OK. See attached.
Waiting for beam.
20 November
01:07 - Still no beam.
System wide checks OK.
Stats OK.
Temps OK.
MBS OK.
Merger and data rate OK.
02:35 - Beam is expected soon
Tried enabling writing data to disk. Merger crashed. FEE64 stopped responding to commands.
Reset Merger. FEE64 now respond to commands.
Tried enabling writing to disk again. Merger crashed once more. Reset merger.
Reset Tape Server interface. Now writing to disk.
No data currently being sent to MBS, but operators say they "are adding a new detector". May be normal.
DAQ & Merger going OK. Writing to Nov19/R1
No beam yet.
Temperatures OK
System wide checks OK
Stats OK. See attached.
MBS not working.
Merger and data rate OK. See attached.
~02:48 Beam on. 40Ar. 300 MeV/A, 500 pps. Degrader settings unclear.
03:54 Degrader settings have been changing between 4.5, 1 and 6 ug/cm2
Currently on 6 g/cm2.
AIDA seen through MBS shows no HE on detector 3, and Det 1 and 2 show no counts on positive y (FEE6 and 1). That is not the case from MIDAS.
04:30 The aim of the tests above was to have no 40Ar at all in DSSD3. It does not appear that 6 g/cm2 is sufficient, as we still see events. The reduction between 1 & 2 vs. 3 is about a factor of 2-10 depending on ADC channel. See attached.
~04:30 Moved to 6.2 g/cm2 degrader. See attached.
~04:45 Moved to 5.8 g/cm2. See attached.
~05:09 Moved to 6.4 g/cm2. See attached.
05:31 High energy channel spectra (1.8.H) for all FEE64 w/ 6.4 g/cm2 degrader - See attachments 13 & 14
05:38: Moved to 1.0 g/cm2. See attached. Difference is not huge which may indicate these spectra are misleading, as partly expected. A more detailed analysis will be carried out off line.
06:00: Beam stopped. ~R77. AIDA keeps running background over the day. Beam should restart around 11 pm later today. |
Attachment 1: Bias2235.png
|
|
Attachment 2: Rates2236.png
|
|
Attachment 3: 19112241MBS.png
|
|
Attachment 4: 19112242Stats.png
|
|
Attachment 5: 19112244Temperatures.png
|
|
Attachment 6: 19112246Merger.png
|
|
Attachment 7: 20110239.png
|
|
Attachment 8: 20110240DataTransfer.png
|
|
Attachment 9: 6gdegrader.png
|
|
Attachment 10: 6.2degrader.png
|
|
Attachment 11: 5.8degrader.png
|
|
Attachment 12: 6.4degrader.png
|
|
Attachment 13: 18h_0530.png
|
|
Attachment 14: 18H_0528.png
|
|
Attachment 15: 1gcm2degrader.png
|
|
|
97
|
Wed Nov 13 15:38:20 2019 |
NH | [HowTo] MBS for AIDA (Nov 19) |
For MBS integration AIDA forwards its data to an MBS Foreign Data Receiver (FDR), currently the PC assigned for this task is x86l-94.
To start the MBS system do the following:
Connect to the MBS FDR: ssh despec@x86l-94 (Ask local for password)
cd to the AIDA directory: cd mbsrun/nov19/aida_to_mbs
Start mbs: mbs
Startup the receiver: @startup
Now open the MBS relay for MIDAS (far right icon on the top)
Check both terminals (Relay & MBS) report a connection
In another terminal ssh into the FDR and run rate to monitor the data rate in MBS
For data to be transferred the Tape Server must be GOING but can be in No Storage mode if no MIDAS storage is required.
--
For experiments MBS file saving is handled by the DESPEC time sorter which merges all the subsystems together.
For testing you can write files from x86l-94 using the following commands
connect rfio XXX -disk
Where XXX is an Linux PC (ask local for a PC)
open file /path/to/file_ first=1 size=2000 -auto -rfio
This opens the file, writing in 2GB chunks
clo file
This will close the file when you are done
--
It is possible to restart the MBS relay at any point if the connection seems to have failed, also confirm the Tape Server is GOing |
|
96
|
Wed Nov 13 09:10:37 2019 |
CA, NH, OH | 13th November 2019 |
10:10 moved HV cables from 11 and 12 to 10 and 9 - all detectors biased in same configuration
lowered slow comparator threshold to 0xa
DAQ start
pulser on
zero'd histograms
pulser peak widths:
FEE width (ch)
aida01 - 130
aida02 - 168
aida03 - 83
aida04 - 83
aida05 - 200
aida06 - 148
aida07 - 250
aida08 - 81
aida09 - 134
aida10 - 190
aida11 - 333
aida12 - 149
For comparison, peak widths from yesterday's measurement:
pulser peak widths - FEE width(ch)
1 128
2 145
3 84
4 74
5 207
6 125
7 172
8 80
9 193
10 183
11 238
12 138
Attachments 1 & 2 : pulser peak spectra
Attachment 3 - good event statistics
Attachment 4 - detector bias/ leakage currents
|
Attachment 1: 191113_pulserpeaksOdd.png
|
|
Attachment 2: 191113_pulserpeaksEven.png
|
|
Attachment 3: 191113_goodevents.png
|
|
Attachment 4: 191113_bias.png
|
|
|
95
|
Wed Nov 13 01:54:00 2019 |
TD | Aida10 Unusual Pulser Peak |
Double peaking normally indicates periodic baseline variations with period comparable to shaping time
Quote: |
aida10 (no alpha rate recorded) has a strange pulser peak - looks like double peak?
Pulser terminator checked and OK...
All others "OK" but noise quite bad (see previous entry)
|
|
|
94
|
Tue Nov 12 14:10:18 2019 |
NH, OH, CA | Aida10 Unusual Pulser Peak |
aida10 (no alpha rate recorded) has a strange pulser peak - looks like double peak?
Pulser terminator checked and OK...
All others "OK" but noise quite bad (see previous entry) |
Attachment 1: aida10-pulser-121119.png
|
|
Attachment 2: pulsers121119.png
|
|
Attachment 3: pulsersodd121119.png
|
|
|
93
|
Tue Nov 12 13:59:07 2019 |
NH, OH, CA | Awfully noisy waveforms |
Rates in all FEEs much higher than before - waveform shows very noisy in all systems. - attachment 1
Good event statistics - attachment 2
DISC in 7 and 3 above 300k
ASIC check load performed.
Good event statistics - attachment 3
Thesholds at 0xa for all FEE
Disc rate now 0 across all FEE
Pulser settings: attachment 4
pulser peak widths - FEE width(ch)
1 128
2 145
3 84
4 74
5 207
6 125
7 172
8 80
9 193
10 183
11 238
12 138
Attachments 5 & 6: pulser peaks for all FEE64
16.24: DAQ stop
slow comparator threshold changed to 0x64
ASIC check performed
Alpha run/DAQ start (w/ data transfer enabled) - writing to file 31Oct19/R13
Merger and TapeService ok - attachments 7 & 8 |
Attachment 1: awfulwaves121119.png
|
|
Attachment 2: 191112_GoodEvents.png
|
|
Attachment 3: 191112_StatsPostLoad.png
|
|
Attachment 4: pulsersettings191112.png
|
|
Attachment 5: pulserpeaksOdd.png
|
|
Attachment 6: pulserpeaks_Even.png
|
|
Attachment 7: merger191112.png
|
|
Attachment 8: tapeservice191112.png
|
|
|
92
|
Tue Nov 12 13:22:19 2019 |
OH, NH, CA | AIDA09 Replacement |
ASIC 1 on FEE9 was not producing signals.
A new HDMI cable was installed but still no signals.
AIDA09 was replaced with a new FEE.
Current list of FEE to serial numbers installed attachment 1
MAC address of new FEE was obtained.
Backup of DHCPD.conf was made dhcpd.confBACKUP191112
dhcpd.conf was updated with new MAC address.
All FEEs powered on and 09 was seen to mount and is seen by AIDAServer.
AIDA09 flashed to newest version of firmware. (0x18430701) - attachment 2 |
Attachment 1: FEE-SerialNumber.txt
|
FEE# Messanie Serial No. MAC Address
4 026-1690101-0016 d8-80-39-41-f6-b7
8 026-1690101-0013 d8-80-39-41-ba-2b
12 026-1690101-0030 d8-80-39-41-cf-ac
2 026-1690101-0011 d8-80-39-41-ba-22
6 026-1690101-0035 d8-80-39-41-ee-72
10 026-1690101-0032 d8-80-39-41-ee-10
1 026-1690101-0027 d8-80-39-41-ba-8a
5 026-1690101-0008 d8-80-39-41-d7-cc
9 026-1690101-0015 d8-80-39-41-f6-ee
7 Not checked d8-80-39-41-b4-0c
11 026-1690101-0006 d8-80-39-41-f6-5a
3 026-1690101-0019 d8-80-39-41-d8-21
Dead FEE Messanie Serial No. MAC Address
ASIC 1 Dead 026-1690101-0005 d8-80-39-42-0d-0c
|
|
91
|
Mon Nov 11 10:34:19 2019 |
NH | Alpha Analysis (11.11.19) |
Analysis of approx 20GB of data since Thursday night (not continuous - estimate of hours incominG)
Figures 1-3: Energy Front vs Energy Back (No F-B energy gate)
Figures 4-6: 2D hit pattern
Figures 7-9: 1D hit pattern
aida10 does not seem to be reading out still, unsure of reason? Looks fine with pulser.
Slow comparator not setting or unusual gain?
Otherwise good |
Attachment 1: DSSD1-Energy.png
|
|
Attachment 2: DSSD2-Energy.png
|
|
Attachment 3: DSSD3-Energy.png
|
|
Attachment 4: DSSD1-2DHP.png
|
|
Attachment 5: DSSD2-2DHP.png
|
|
Attachment 6: DSSD3-2DHP.png
|
|
Attachment 7: DSSD1-1DHP.png
|
|
Attachment 8: DSSD2-1DHP.png
|
|
Attachment 9: DSSD3-1DHP.png
|
|
|
90
|
Thu Nov 7 10:22:26 2019 |
Nic, Patrick | Reply: WR Timestamps |
> > > All 12 FEEs have valid WR Timestamps
> > > Had to powercycle aida09 once as before raw readout was displaying upper 12 bits of WR timestamp as 0. Unsure of other method.
> > >
> > > HDMI cables in aida09 checked and good.
> >
> > The problem would be the cable , one end or the other.
> > I think ( if I recall ) a setup would restart the WR decoder.
> >
> > I notice you have set the WR info word rate to be quite high , 6123/sec typ, is this intentional ?
>
> Cable will be replaced once a spare is available.
>
> Setup did not restart the WR decoder when this problem occurred beforehand - Reset/Setup tried.
>
> WR rate was chosen by Vic I believe, I am unsure of reasoning myself.
An update/clarification:
Although the upper bits are zero I believe actually it is a total failure to synchronise to WR:
aida09
WR Time Item 0x80500000 0x0fbd8000; Time (48:63)=0x0; Time (28:47)=0x249; Time (0:27)=0x0fbd8000
WR Time Item 0x80400249 0x0fbd8000; Time (28:47)=0x249; Time (0:27)=0x0fbd8000
WR Time Item 0x80500000 0x0fbdc000; Time (48:63)=0x0; Time (28:47)=0x249; Time (0:27)=0x0fbdc000
WR Time Item 0x80400249 0x0fbdc000; Time (28:47)=0x249; Time (0:27)=0x0fbdc000
WR Time Item 0x80500000 0x0fbe0000; Time (48:63)=0x0; Time (28:47)=0x249; Time (0:27)=0x0fbe0000
WR Time Item 0x80400249 0x0fbe0000; Time (28:47)=0x249; Time (0:27)=0x0fbe0000
aida10
WR Time Item 0x8050022e 0x0bde8000; Time (48:63)=0x22e; Time (28:47)=0xe29d5; Time (0:27)=0x0bde8000
WR Time Item 0x804e29d5 0x0bde8000; Time (28:47)=0xe29d5; Time (0:27)=0x0bde8000
WR Time Item 0x8050022e 0x0bdec000; Time (48:63)=0x22e; Time (28:47)=0xe29d5; Time (0:27)=0x0bdec000
WR Time Item 0x804e29d5 0x0bdec000; Time (28:47)=0xe29d5; Time (0:27)=0x0bdec000
WR Time Item 0x8050022e 0x0bdf0000; Time (48:63)=0x22e; Time (28:47)=0xe29d5; Time (0:27)=0x0bdf0000
WR Time Item 0x804e29d5 0x0bdf0000; Time (28:47)=0xe29d5; Time (0:27)=0x0bdf0000 |
|
89
|
Tue Nov 5 11:15:41 2019 |
NH PJCS | Report: aida10 database corruption |
If ypou now check the actual values in the aida10 Options screen are the correct ?
It can happen that corruptions once loaded are propagated when Options are generally saved.
Looking at the dates the files were modified does that make sense to you. It doesn't look as if the corrupted database was changed except when they all were ? Can you check ? Have you checked the other Options files ? I seem to recall that RIKEN AIDA had a program to to this ?
Quote: |
Hi, see the attached files in the original comment. I think the BACKUPCorrupt is quite old now.
The Options data gets corrupted and for example the ASIC folder gets set to 'undefined' which I think is why the ASIC data loads incorrectly.
Quote: |
This looks like the Options not the ASIC data. That will be in the CONTENTS file of the FEE64 named directory within the dated EXPERIMENT directory. Looks like the latest is at /MIDAS/DB/EXPERIMENTS/AIDA/2019Oct31-13.24.23/aida10 ?
A quick look at this file and it looks normal , not all 0xad ?
Looked at the Options file in the /MIDAS/DB/EXPERIMENTS/AIDA/Options.BACKUPCorrupt/aida10 and the worst bit is the name of the Data Acquistion program. That could welll cause you problems with readout as it won't understand some of the new data.
Quote: |
Often aida10 database seems to get corrupted (ASIC values are all 0xad)
Unsure why it's only aida10, might be related to weird alpha rate behaviour at the moment.
Attached is corrupted database and correct database
|
|
|
|
|
88
|
Tue Nov 5 10:32:21 2019 |
NH PJCS | Report: aida10 database corruption |
Hi, see the attached files in the original comment. I think the BACKUPCorrupt is quite old now.
The Options data gets corrupted and for example the ASIC folder gets set to 'undefined' which I think is why the ASIC data loads incorrectly.
Quote: |
This looks like the Options not the ASIC data. That will be in the CONTENTS file of the FEE64 named directory within the dated EXPERIMENT directory. Looks like the latest is at /MIDAS/DB/EXPERIMENTS/AIDA/2019Oct31-13.24.23/aida10 ?
A quick look at this file and it looks normal , not all 0xad ?
Looked at the Options file in the /MIDAS/DB/EXPERIMENTS/AIDA/Options.BACKUPCorrupt/aida10 and the worst bit is the name of the Data Acquistion program. That could welll cause you problems with readout as it won't understand some of the new data.
Quote: |
Often aida10 database seems to get corrupted (ASIC values are all 0xad)
Unsure why it's only aida10, might be related to weird alpha rate behaviour at the moment.
Attached is corrupted database and correct database
|
|
|