  DESPEC, Page 22 of 37  ELOG logo
Entry  Wed Jan 18 13:40:33 2023, PJCS TD, MACB settings with either Emulator or VITAR macb_apr20.jedmacb_apr20.vhdzybo.jpgMACB.jpg
When using the VETAR connected to the HDMI port of the root MACB the settings for all the MACB in the system 0x3.

This setting takes the Clock and Data line from the input HDMI and outputs it via teh HDMI output ports.

When using the Emulator connected via the SMA connectors on the back of the root MACB then the setting for the root MACB
Entry  Wed Jan 23 09:25:55 2019, CA, TD, NH, VP, MACB and setup pics 23.01.19 11x
Pictures of leaking cooling pipe

MACB switch settings
Entry  Tue May 4 14:05:48 2021, NH, MACB Cables macb_front.jpgmacb_side.jpg
During the installation of the 4 new HDMI cables for aidas13-16 the HDMI cables between the MACB Root and the MACB leaves were looped through the cable
guide at the top to support them more
This may improve their connection and reduce the chance of WR errors seen in S460/S452
Entry  Thu May 12 22:49:39 2022, BA, MA, Low Rates RatesLow2022-05-12_23-48-01.png
The rate is very low in all aida, not sure .
    Reply  Fri May 13 06:56:14 2022, OH, BA, MA, Low Rates 
 




Quote:



The rate is very low in all aida, not sure .
Entry  Mon May 16 14:07:06 2022, TD, Lost activity monitor Screenshot_from_2022-05-16_15-04-56.png
 
Entry  Fri Mar 15 16:29:57 2024, NH, Leakage currents 
The behaviour of the DSSSD leakage current at low voltages and during biases is unusual and varies depending on how the adapter boards are connected
To summarise the behaviour I have observed
Entry  Thu Jan 9 16:25:51 2025, TD, LEC fast comparator threshold too low Screenshot_from_2025-01-09_17-28-45.png

LEC fast comparator has been set to 0x2 (200keV) which is *very* low and means the fast comparator will be triggering on noise.
    Reply  Thu Jan 9 17:36:50 2025, TD, LEC fast comparator threshold too low 

> LEC fast comparator has been set to 0x2 (200keV) which is *very* low and means the fast comparator will be triggering on noise.
    Reply  Thu Jan 9 17:56:14 2025, TD, LEC fast comparator threshold too low 
> > 
> > LEC fast comparator has been set to 0x2 (200keV) which is *very* low and means the fast comparator will be triggering on noise.
> > 
    Reply  Fri Jan 10 10:03:43 2025, TD, LEC fast comparator threshold too low 
> > > 
> > > LEC fast comparator has been set to 0x2 (200keV) which is *very* low and means the fast comparator will be triggering on noise.
> > > 
Entry  Thu Jan 24 10:21:26 2019, CA, TD, NH, VP, January 24th 2019 7x
10.30 AIDA setup and DAQ start complete
      writing to file TapeData/NULL/R1
Entry  Wed Jan 23 17:03:08 2019, CA, TD, NH, VP, January 23rd 2019 6x
Post White Rabbit update

Attachment 1 - 1.8.W pulser waveform spectra for all FEE64
Entry  Wed Jan 23 08:12:01 2019, CA, TD, NH, VP, January 22nd 2019 220119_18W.png220119_pulserpeaks.png220119_stats.png220119_Temp.png
Attachment 1 - 1.8.W waveform spectra with pulser signal

Attachment 2 - 1.8.L pulser peak spectra
Entry  Tue Apr 2 12:36:25 2024, JB, CC, NH, Installing FEE64s of DSSSD2 cont. 
Some additional checks

- check 'top hat' electrical isolators are correctly installed at each of the 4x mounting points of the AIDA snout
assembly
Entry  Wed Mar 27 14:22:35 2024, JB, NH, Installing FEE64s of DSSSD2 14x
Mounted on frame:

DSSD 1 (Upstream) :     3208-2/3208-5/3208-8
DSSD 2 (Downstream): 3208-3/3208-21/3208-22
    Reply  Thu Mar 28 09:18:53 2024, TD, Installing FEE64s of DSSSD2 
Some additional checks

- check 'top hat' electrical isolators are correctly installed at each of the 4x mounting points of the AIDA snout assembly 

- check snout is electrically isolated with respect to AIDA support assembly/stand, bPlas PCBs/cabling/ground/drain wires and BB7
Entry  Wed Dec 11 10:58:52 2019, NH, Initital Dec. Analysis 11x
Fig 1: 2D hit pattern of each AIDA DSSD for MBS run 24 (40Ar primary beam, centred)

Low statistics in DSSD2 partly due to low energy deposition 
Entry  Fri Apr 16 02:15:12 2021, NH, Implants/Punchthrough 53_AM.png
Rate with beam < 1000 Hz still

Not Uranium but some fragments possibly produced in the thick 9g degrader
Entry  Thu Sep 20 09:16:31 2018, TD, Images of AIDA setup for DESPEC 9x
Attachment 1 - detail of MSL type BB18(DS)-1000 p+n junction side strips, bias ring, multiple guard rings, filed 
plates and bond wires