Mon May 16 07:03:16 2022, MS, OH, Monday 16th May 08:00-16:00 17x
|
08:00 Took over the night shift from Tom (Attachments 1-3)
10:00: Attachement 4-6
around 10:30: Beam went off (Attachment 7) |
Wed May 19 09:09:27 2021, MS, JS, OH, May 16 08:00-24:00 21x
|
16th May 08:00 - 12:00 shift
Author: MS
|
Sat Apr 17 15:29:27 2021, MS, Saturday 17 April 16:00-20:00 2021 6x
|
16:21
|
Mon Apr 19 06:46:15 2021, MS, Monday 19 April 07:47-12:00 2021 9x
|
07:47
|
Wed Apr 21 06:56:42 2021, MS, Wednesday 21 April 08:00-12:00 2021 10x
|
8:00
|
Wed Jun 22 07:12:46 2022, MS, Wednesday 22 June 8:00-16:00 16x
|
Took over the night shift from Tom.
8:10 : attachments 1-3
10:00 : attachments 4-6 |
Sat Jun 25 06:55:07 2022, MS, Saturday 25 June 2022 8:00-16:00 15x
|
Took over the night shift from Tom.
7:00 The beam is back.
8:00 : attachments 1-3 |
Thu Oct 30 10:06:21 2025, MP, NH, Setup and data taking with 207Bi    
|
- Initialised the Merger and Tape server
- loading of settings: EXPERIMENTS/AIDA/2025Oct29-16.11.06
- aida16 very noisy (attachments 3,4), slow comparator thresholds was risen to 500 keV, until reducing its deadtime to 0 |
Thu Oct 30 16:43:36 2025, MP, NH, Setup and data taking with 207Bi
|
Data taking and DAQ stopped now (17.40)
Quote:
- Initialised the Merger and Tape server |
Tue Nov 4 16:34:47 2025, MP, NH, Bi207 source data, 700keV
|
Some FEES were not acquiring data for some reason we could not understand. We lowered the thresholds to 200 keV, then rose them again to 700 keV. It
seems they are now acquiring. We start data taking again.
Quote:
|
Tue Jul 22 17:00:26 2025, MP, CC, Status 
|
Test with shielded HV cables for both detectors.
The FWHM for ch1 is now 184 ch.
Yesterday cables for cards 5, 6, 10, 11, 12, 14, 16 were partially connected (only 1 out of 2 flat cables was in), now both are connected. |
Mon Jul 21 13:01:19 2025, MP, Status 
|
| Status before grounding checks |
Wed Oct 29 08:12:35 2025, MP, Further setup for 207Bi test    
|
Inspection of side FEEs (2,6,4,8) and possible grounding improvements:
- all screws and nuts for copper connections were tightenedFee
- FEE 6 was isolated from the copper connection grounding FEE2 with insulating tape (attachment 1) |
Tue Nov 4 14:57:19 2025, MP, Bi207 source data, 700keV   
|
- Bias and basic checks: the pulser FWHM appears to be ~170, which is the same we have obtained at the end of last week.
- Initialised the Merger and Tape server
- loading of settings: EXPERIMENTS/AIDA/2025Nov03-16.21.44 |
Tue Nov 4 17:29:28 2025, MP, Bi207 source data, 700keV
|
Anydesk has crashed. I am not able to set it up again. I switch off bias and fees
Quote:
Some FEES were not acquiring data for some reason we could not understand. We lowered the thresholds to 200 keV, |
Fri Nov 7 10:47:29 2025, MP, 207Bi source test - data taking
|
DAQ stopped at ~17.30 due to anydesk failure
|
Sun Apr 18 17:27:44 2021, ML-OH, system wide checks 17x
|
system wide checks at 6pm CET
##################################################
Difference noted from previous shift: |
Sun Apr 18 19:05:33 2021, ML-OH, All spectra reset
|
All spectra reset at 8pm CET.
|
Sun Apr 18 21:43:01 2021, ML-OH, system wide checks 20x
|
New System wide check 10pm CET
Clock status test result: Passed 12, Failed 0
Understand status as follows |
Fri Apr 16 05:13:09 2021, ML, system wide checks 16x
|
system wide checks:
Done around 4am.
################################# |