AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC, Page 29 of 37  ELOG logo
Entry  Tue Mar 9 10:23:45 2021, TD, Analysis of R14 (w/ NewMerger with min info code 4 & 5 data items) R14_0R14_19

Per https://elog.ph.ed.ac.uk/DESPEC/183 R14 data files have been analysed
Entry  Thu Jan 24 13:29:40 2019, TD, Analysis of GSIJan19/R1_0 WR_sample_R1_0_24jan19WR_sample_R1_0_24jan19.loganalyser.fanalyser.f

Analysis of file GSIJan19/R1_0 (white rabbit data)
Entry  Sat Oct 13 17:26:53 2018, TD, Analysis of 3 October 2018 T4 tests 7x
 
Entry  Mon Mar 8 09:45:20 2021, CA, TD, Analysis R7_20 (new version of NewMerger with min info code 4 & 5 data items) R7_20R7_20_verbose

An analysis of file R6_70 can be found at https://elog.ph.ed.ac.uk/DESPEC/177 attachment 19
We observe c. 80M ADC data items per 2Gb file and c. 160M info code 4 & 5 data items.
Entry  Thu Jan 24 14:53:26 2019, TD, Analyses GSIJan19/R2_0 - R4_0 WR_sample_R2_0_24Jan19WR_sample_R3_0_24Jan19WR_sample_R4_0_24Jan19WR_sample_R4_0_24Jan19.log

Attachment 1 - summary analysis R2_0
Entry  Tue May 4 21:26:43 2021, OH, Alpha run 10x
22:26 Tape server set up to write to directory May21
      DAQ stopped
      Slow comparator set to 0x64 for all FEE64 and an ASIC check performed
Entry  Mon Nov 4 08:48:18 2019, NH, Alpha Status 04.11.2019 alphas-bias.pngalphas-temp.pngalpha-aida10.pngstats-0511.pngspec-0511.png
Alpha run has been running over the weekend. Stats looking positive with notable exception that aida10 is not sending much data. Will attempt a powercycle.

Temperatures and Leakage Currents very good to excellent

Noted that the slow rate of aida10 seems to be slowing down the rate data is sent to tape - perhaps being buffered until an ADC event arrives?
Entry  Thu Oct 31 17:13:54 2019, NH, Alpha Run 
Beginning Alpha run

Thresholds set to 100 (LEC/Slow) 255 (LEC/Fast) 2 (HEC)
Entry  Sat Feb 29 10:56:38 2020, NH, Alpha Hit Pattern alphas_1d.pdf
Circa 700 GB of data

A few channels in DSSD 1 missing? Otherwise seems OK
Entry  Thu Apr 28 18:45:09 2022, NH, Alpha Background Run  AnyDeskMSI_2022-04-28_19-47-33.pngAnyDeskMSI_2022-04-28_19-47-56.pngAnyDeskMSI_2022-04-28_19-48-13.png
Runs are in /TapeData/BGAPR22 located on 8TB disk "SecondDrive" (N.B. 2A experiment is using ThirdDrive)

Old runs and test files were removed from disk. Left is an unknown S481 folder (fibre implant test?) and Patrick
Entry  Fri Dec 20 10:19:29 2019, NH, Alpha Analysis (Pre Run) Alphas.pngAlphas_E.png
======================================                                                                                                                 
                                                                                                                      
AIDA Unpacker Analysis                                                                                                                                
Entry  Mon Nov 11 10:34:19 2019, NH, Alpha Analysis (11.11.19) 9x
Analysis of approx 20GB of data since Thursday night (not continuous - estimate of hours incominG)

Figures 1-3: Energy Front vs Energy Back (No F-B energy gate)

Figures 4-6: 2D hit pattern
Entry  Sun Apr 18 19:05:33 2021, ML-OH, All spectra reset 
 All spectra reset at 8pm CET.

 
Entry  Tue Nov 12 14:10:18 2019, NH, OH, CA, Aida10 Unusual Pulser Peak aida10-pulser-121119.pngpulsers121119.pngpulsersodd121119.png
aida10 (no alpha rate recorded) has a strange pulser peak - looks like double peak?

Pulser terminator checked and OK...

All others "OK" but noise quite bad (see previous entry)
    Reply  Wed Nov 13 01:54:00 2019, TD, Aida10 Unusual Pulser Peak 
Double peaking normally indicates periodic baseline variations with period comparable to shaping time




Quote:



aida10 (no alpha rate recorded) has a strange pulser peak - looks like double peak?
Entry  Tue Apr 23 10:52:39 2024, PP, Aida02 down and back 
We lost aida02 at about 11:25 again. Nick fixed it about 11:50.

It is now back in the DAQ.
Entry  Fri Aug 16 15:09:05 2019, NH, Aida Waves (All FEEs) 12x
Included waves for all 12 FEEs

FEES 1, 5, 9 all show this HF noise on it, and are all located at the top of the AIDA mount.
Connections: HV from DSSD (-ve) is in FEE9
Entry  Fri Aug 9 12:33:14 2019, NH, Aida Test 9/8/19 - FEES 1-6 
The water issue has been resolved - merely incorrect temperature guages (replaced). Water is 20 C as required.

S4 conditions at 13:12 CEST

26.5 C / 47.2 % / Td = 14.3 C
Entry  Fri Jun 28 15:05:49 2019, NH, Aida 28/6/2019 - WATER OFF 
The humidity sensor hasn't reported humdities under 90% yet. In order to try and help dry the pipes and turn AIDA the water has been turned off the
weekend.
Sunday temperatures might reach 40 C which means it may struggle anyway.
    Reply  Mon Jul 8 16:02:52 2019, NH, PJCS, Aida 28/6/2019 - WATER OFF 
It occurs to to wonder if there is now water inside the modules. If the metal of the cooling plates is damp on the outside then it is safe to assume
there is water on the inside too . Can I suggest that one  of the modules is dismounted, disassembled and checked?

Patrick
ELOG V3.1.3-7933898