ID |
Date |
Author |
Subject |
125
|
Sun Jun 16 00:52:27 2019 |
Ela, Sonja, Victor | root file 6:30-8:30 |
|
Attachment 1: 190616_0833_100Sn_020.root
|
128
|
Sun Jun 16 03:42:17 2019 |
Ela, Sonja, Victor | root file 8:30-10:30 |
|
Attachment 1: 190616_1032_100Sn_023.root
|
129
|
Sun Jun 16 04:44:05 2019 |
Ela, Sonja, Victor | root file 10:30-12:30 |
The beam stopped a couple of times due to problems before RRC |
Attachment 1: 190616_1233_100Sn_026.root
|
Attachment 2: Runs_24-26_breaks.png
|
|
130
|
Sun Jun 16 06:55:31 2019 |
Ela, Sonja, Victor | root file 12:30-14:30 |
|
Attachment 1: 190616_1435_100Sn_029.root
|
99
|
Fri Jun 7 01:55:24 2019 |
CA, JA , AT | merger test |
Setup: On 17th-May (11:18 - 11:38), AIDA was inside DTAS. Measurement of 60Co source.
Files:
AIDA: R6_{3..7}.gz
1,2G may 17 11:21 R6_3.gz
982M may 17 11:25 R6_4.gz
926M may 17 11:28 R6_5.gz
1,1G may 17 11:32 R6_6.gz
1,2G may 17 11:38 R6_7.gz
20M may 17 11:39 R6_8.gz
DTAS: AidaMergerTest_2_0{23..29}.dlt
955M may 17 11:21 AidaMergerTest_2_023.dlt
955M may 17 11:24 AidaMergerTest_2_024.dlt
956M may 17 11:27 AidaMergerTest_2_025.dlt
956M may 17 11:31 AidaMergerTest_2_026.dlt
955M may 17 11:33 AidaMergerTest_2_027.dlt
956M may 17 11:36 AidaMergerTest_2_028.dlt
732M may 17 11:38 AidaMergerTest_2_029.dlt
Working dir: ~/Corrigan
Test 1: merge R6_5 with AidaMergerTest_2_025.dlt |
98
|
Thu Jun 6 09:36:52 2019 |
Alvaro, Jorge, Ale | new config file |
New config file: 20190606_Conf.xlsx
Changed with respect to the earlier file (20190605_Conf.xlsx):
-The threshold of the Reference detector was changed from 2800 to 3500.
-A new gated histogram was added: ECryA gated by F11PLR in the energy range of the implanted ions, named as
EcryAGF11
Issue in AIDA PL (see screenshot). A lot of small signals; top, samples, bottom output of EFIR (picture taken
when integration = 50 samples; value in configuration = 20) |
Attachment 1: 20190606_Conf.xlsx
|
Attachment 2: Screenshot_from_2019-06-06_16-54-15.png
|
|
144
|
Wed Jun 19 05:10:15 2019 |
Alvaro | Aida plastic - gain change |
TAC AIDAPL - CryA changed
After increasing the gain, x2 peaks in TAC |
Attachment 1: changeGainAPL.png
|
|
10
|
Wed May 15 11:53:17 2019 |
Algora, Nacher, Kiss | Background |
Start: 20:05
Stop: 09:44
Background rate 2.9 kHz (Dynodes), 2.3 (Anodes) (Threshold of detector 13 was increased)
Files: Background_20190515_006.dlt ... _014.dlt
Root file: 201905160944_bckg_overnight.root (attached)
Pulser self-triggered at 100Hz |
Attachment 1: 201905160944_bckg_overnight.root
|
9
|
Wed May 15 11:50:04 2019 |
Algora, Nacher | 137Cs listmode, and rootfile |
137Cs source in the center of the DTAS, with the IFIC holder
list mode files: Cs137_Source_20190515_000.dlt to Cs137_Source_20190515_005.dlt
Root file: 201905151942_137Cs.root
|
Attachment 1: 201905151942_137Cs.root
|
4
|
Wed May 15 06:40:16 2019 |
Algora, Kiss, Nacher | DTAS aligned and calibrated |
Signal cables (long cables, from preamp to digitizer) of channels 13 and 16 were changed. For the moment they have not been relabelled.
We assumed that it was not a HV supply or cable problem, since the anods were aligned and only the dynodes moved.
After changing the cables we realigned the dynodes and recalibrated all signals. Screen shots are attached.
File (root file) with the two sources (Co and Cs) saved as 201905151432_137Cs60Co.root
Configuration file: 20190515_Conf.xlsx
Rate with two sources: 52 kHz (Dynodes), 47 kHz (Anodes)
|
Attachment 1: Screenshot_from_2019-05-15_14-29-57.png
|
|
Attachment 2: Screenshot_from_2019-05-15_14-30-57.png
|
|
Attachment 3: Screenshot_from_2019-05-15_14-31-25.png
|
|
Attachment 4: Screenshot_from_2019-05-15_14-31-55.png
|
|
11
|
Thu May 16 02:02:24 2019 |
Algora Kiss Nacher | detector 3 moved over night |
After the long background measurement we have measured 60Co to check alignment.
Everything looks fine except for detector #3 that has moved to lower gain (seen in dynodes and anodes).
See figures attached. |
Attachment 1: Screenshot_from_2019-05-16_10-00-53.png
|
|
Attachment 2: Screenshot_from_2019-05-16_10-01-29.png
|
|
Attachment 3: Screenshot_from_2019-05-16_10-01-57.png
|
|
Attachment 4: Screenshot_from_2019-05-16_10-09-16.png
|
|
14
|
Thu May 16 06:59:51 2019 |
Algora Kiss Nacher | Det 3 |
Det 3 moved, both Anode and Dynode. So we have changed the HV cable to test if there are no more jumps. We used cable 9 and it was not
relabeled. The HV of the detector was corrected. It was changed to 527.5 V |
108
|
Mon Jun 10 04:14:39 2019 |
Ale, Jose | Background measurement overnight |
Start:~14:00 (June 9)
Stop:10:30
File: 201906101030_background_AIDA_in.root |
109
|
Mon Jun 10 04:18:24 2019 |
Ale, Jose | New Configuration File |
File: 20190610_1145_Conf_DTAS.xlsx
Several changes and corrections
Naming: V2A8C1 -> Aux01, ...
New or corrected
Aux13 -> LP_out (puse out of light pulser)
Aux15 -> DTpulser (Out of GSI clock triggering LP)
Aux16 -> SyncPulser (Out of GSI clock send out for synchronization with AIDA & BigRIPS)
Ranges changed
Aux09, ERefAn
We observed that LP_out was counting less than the other two
so we tried to check DACQ parameters. We basically change nothing
Only activated "CFD time"
Note that pulse out is a very strange signal:
Negative(?),-400mV offset, 200mV high
In the DACQ is treated as positive signal with a large negative offset. |
Attachment 1: 20190610_1145_Conf_DTAS.xlsx
|
110
|
Mon Jun 10 08:07:17 2019 |
Ale, Jose | Test of pulser count rates |
File: 20190610_1602_background_AIDA_in.root
SyncPulse: 61777
DTPulser: 61277
LP_out: 61289 (!!!?)
EDy01: 61247
EDy13: 61226
LP_out counts more now |
Attachment 1: 20190610_1602_background_AIDA_in.root
|
113
|
Mon Jun 10 10:44:36 2019 |
Ale, Jose | New Conf file with new calibration |
File: 20190610_1835_Conf_DTAS.xlsx |
114
|
Mon Jun 10 12:31:52 2019 |
Ale, Jose | New calibration |
New calibration with more statistics
File: 20190610_2030_60Co_137Cs_AIDA_in.root
An01_EFIR 306750 536250 606750
An02_EFIR 281250 489750 554250
An03_EFIR 282750 492750 558750
An04_EFIR 263250 456750 518250
An05_EFIR 270750 471750 534750
An06_EFIR 308250 537750 609750
An07_EFIR 290250 506250 572250
An08_EFIR 284250 494250 560250
An11_EFIR 365250 636750 720750
An12_EFIR 263250 459750 519750
An13_EFIR 321750 560250 635250
An14_EFIR 327750 572250 648750
An15_EFIR 299250 521250 590250
An16_EFIR 287250 500250 566250
An17_EFIR 296250 516750 585750
An18_EFIR 273750 479250 542250
Dy01_EFIR 34296.9 59609.4 67578.1
Dy02_EFIR 33828.1 59140.6 66953.1
Dy03_EFIR 34140.6 59453.1 67421.9
Dy04_EFIR 34140.6 59453.1 67265.6
Dy05_EFIR 34453.1 59921.9 67890.6
Dy06_EFIR 33984.4 59453.1 67421.9
Dy07_EFIR 34453.1 60078.1 68046.9
Dy08_EFIR 33828.1 58984.4 66796.9
Dy11_EFIR 34453.1 60078.1 68046.9
Dy12_EFIR 34296.9 59921.9 67890.6
Dy13_EFIR 34296.9 59609.4 67578.1
Dy14_EFIR 34296.9 59921.9 68046.9
Dy15_EFIR 33984.4 59453.1 67421.9
Dy16_EFIR 34296.9 59921.9 67578.1
Dy17_EFIR 34296.9 59921.9 67734.4
Dy18_EFIR 33984.4 59453.1 67265.6
An01_EFIR 306750 536250 606750 0.00223435 -23.9647
An02_EFIR 281250 489750 554250 0.00245634 -29.3018
An03_EFIR 282750 492750 558750 0.00243187 -25.7883
An04_EFIR 263250 456750 518250 0.00263384 -31.3306
An05_EFIR 270750 471750 534750 0.00254201 -26.4752
An06_EFIR 308250 537750 609750 0.00222596 -24.3613
An07_EFIR 290250 506250 572250 0.00237623 -28.3674
An08_EFIR 284250 494250 560250 0.00243187 -29.4361
An11_EFIR 365250 636750 720750 0.00188632 -27.4314
An12_EFIR 263250 459750 519750 0.00261235 -26.3813
An13_EFIR 321750 560250 635250 0.00214105 -27.0487
An14_EFIR 327750 572250 648750 0.00209042 -23.3901
An15_EFIR 299250 521250 590250 0.00230503 -28.1551
An16_EFIR 287250 500250 566250 0.00240375 -28.9034
An17_EFIR 296250 516750 585750 0.00231789 -24.9283
An18_EFIR 273750 479250 542250 0.00249619 -21.9428
Dy01_EFIR 34296.9 59609.4 67578.1 0.0201694 -29.8925
Dy02_EFIR 33828.1 59140.6 66953.1 0.0202412 -23.2189
Dy03_EFIR 34140.6 59453.1 67421.9 0.0201694 -26.741
Dy04_EFIR 34140.6 59453.1 67265.6 0.0202412 -29.5443
Dy05_EFIR 34453.1 59921.9 67890.6 0.020068 -29.6613
Dy06_EFIR 33984.4 59453.1 67421.9 0.020068 -20.2545
Dy07_EFIR 34453.1 60078.1 68046.9 0.0199676 -26.3096
Dy08_EFIR 33828.1 58984.4 66796.9 0.0203445 -26.6048
Dy11_EFIR 34453.1 60078.1 68046.9 0.0199676 -26.3096
Dy12_EFIR 34296.9 59921.9 67890.6 0.0199676 -23.1897
Dy13_EFIR 34296.9 59609.4 67578.1 0.0201694 -29.8925
Dy14_EFIR 34296.9 59921.9 68046.9 0.0198974 -20.435
Dy15_EFIR 33984.4 59453.1 67421.9 0.020068 -20.2545
Dy16_EFIR 34296.9 59921.9 67578.1 0.0201077 -28.6651
Dy17_EFIR 34296.9 59921.9 67734.4 0.0200376 -25.9333
Dy18_EFIR 33984.4 59453.1 67265.6 0.020139 -23.0111 |
115
|
Mon Jun 10 12:48:01 2019 |
Ale, Jose | New Config file |
Conf file with new calibration:
20190610_2035_Conf_DTAS.xlsx
Check root file:
20190610_2050_60Co_137Cs_AIDA_in.root
See screen shots: |
Attachment 1: 20190610_2035_Conf_DTAS.xlsx
|
Attachment 2: 20190610_2050_60Co_137Cs_AIDA_in.root
|
Attachment 3: Screenshot_from_2019-06-10_20-54-43.png
|
|
Attachment 4: Screenshot_from_2019-06-10_20-55-42.png
|
|
Attachment 5: Screenshot_from_2019-06-10_20-56-39.png
|
|
Attachment 6: Screenshot_from_2019-06-10_20-57-29.png
|
|
119
|
Fri Jun 14 16:22:25 2019 |
AT, OH, JA, CG, TD | AIDA Sync |
Problem with the sync checker appeared AIDA was not its notmal 14us out from DTAS.
Check with source shows we are in sync still
AIDA files R34_1 and R34_2 |
Attachment 1: AIDACorrelation.png
|
|
118
|
Wed Jun 12 02:04:07 2019 |
AA, EN, PM, AT | background during in-beam exp. |
In-beam experiment going on. Only 12h of beam in the last 21h.
Start: ~1pm (11th-June)
Stop: ~10am (12th-June) |
Attachment 1: 20190612_0959_background_AIDA_in.root
|