Entry  Sat Nov 12 01:45:00 2016, TD, Saturday 12 November 6x
10.45 DSSSD bias & leakage currents OK - see attachments 1 & 2
      FEE64 temperatures OK - see attachment 3

14.45 DSSSD bias & leakage currents OK - see attachments 4 & 5

15.10 Download .xlsx file of AIDA leakage currents etc from
https://docs.google.com/spreadsheets/d/12hgbrywB10hGsKt5uc2HnLfZymh8_uvM6cEASbbqnBE/edit#gid=1357176413
      - see attachment 6
Entry  Fri Nov 11 00:50:00 2016, TD, Friday 11 November 11x
10.00 DAQ continues (R33) in no storage mode

      DSSSD bias & leakage currents OK - see attachments 1 & 2
      FEE64 temperatures OK - see attachment 3
      System wide checks OK *except* clock status - see attachments 4 & 5
      Merge statistics - see attachments 6-11

10.21 Ambient temperature +24.9 deg C, d.p. +1.4 deg C, RH 21.5% 

      
Entry  Thu Nov 10 12:02:24 2016, TD, OH, CG, Thursday 10 November 30.png31.png32.png34.png
21.17 DAQ starts (R30)

      Pulser walkthrough
      Amplitude 90,000 - 10,000 @ 10,000 step
      x5 attenuator IN
      t_r 50ns tau_d 50us
      polarity -, tail pulse, delay MIN
      frequency 250Hz range

21.30 DAQ stops OK 

21.32 DAQ starts (R31)

      Pulser walkthrough
      Amplitude 10,000 - 2,000 @ 2,000 step

21.40 DAQ stops OK

21.44 DAQ starts (R32)
     
      Background
      BNC PB-4 pulser OFF

21.59 DAQ stops OK

23.45 Started R35 so Jorge can do some tests.
      No storage mode.

00.20 Carried out manual fill on LN2 in advance of beam arriving later (~03.00).
      Took around 20mins.
      Continued writing to R35 after soft reset of DAQ.
Entry  Thu Nov 10 11:57:35 2016, TD, Data directory NP1412 renamed as RIBF123R1 
Data directory NP1412 renamed as RIBF123R1

... there are several NP1412 'experiments' in the current schedule ... 1412 may be a date
Entry  Thu Nov 10 09:56:20 2016, TD, OH, ML, NP1412 R26 R27 8x
start 18.55 10.11.16
stop 
AIDA filename R26_0
BRIKEN filename 161109_1852_767778Ni_045
BigRIPS filename 443
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents OK attachments 1 and 2 

Degrader all degraders out
F11 rate 430Hz
BigRIPS setting 767778Ni  F5  out, F7 CH2 in, F11 6mm C in

From offline analysis of R26_2

 *** DSSSD # 1 count:       459 old count:       225 dt:      1.02 s  HEC rate:    229.83 Hz
 *** DSSSD # 2 count:       385 old count:       191 dt:      1.02 s  HEC rate:    190.54 Hz
 *** DSSSD # 3 count:       461 old count:       228 dt:      1.02 s  HEC rate:    228.85 Hz
 *** DSSSD # 4 count:       403 old count:       197 dt:      1.02 s  HEC rate:    202.33 Hz
 *** DSSSD # 5 count:       218 old count:       121 dt:      1.02 s  HEC rate:     95.27 Hz
 *** DSSSD # 6 count:        87 old count:        42 dt:      1.02 s  HEC rate:     44.20 Hz

19.02 FEE64 temperature OK - see attachment 3

19.16 BRIKEN DAQ problems following restart - multiple correlation scaler reset requests issued - ignore data
before R26_5
      New BRIKEN data file 161109_1852_767778Ni_046

19.20 AIDA DAQ stalled (presumably *end* of LN2 fill cycle)

      Soft reset OK
      Clock lock state status 0xd - see attachment 6

19.29 Start DAQ (R27) 

19.35 BRIKEN DAQ problems starting SyncCheck program - - multiple correlation scaler reset requests issued -
ignore data before R27_2

19.37 All degraders (5.4mm Al) in for new Takechi et al. setting
      BigRIPS setting 767778Ni  F5 in, F7 out, F11 6mm C in

BRIKEN filename 161109_1852_767778Ni_045
AIDA fileame R27_3
BigRIPS filename 444

19.40 All degraders out

DSSSD bias & leakage currents OK attachments 4 & 5 

From offline analysis of R27_4

 *** DSSSD # 1 count:       281 old count:       247 dt:      2.31 s  HEC rate:    121.43 Hz
 *** DSSSD # 2 count:       247 old count:       219 dt:      2.31 s  HEC rate:    106.74 Hz
 *** DSSSD # 3 count:       286 old count:       252 dt:      2.31 s  HEC rate:    123.59 Hz
 *** DSSSD # 4 count:       287 old count:       253 dt:      2.31 s  HEC rate:    124.02 Hz
 *** DSSSD # 5 count:       201 old count:       178 dt:      2.31 s  HEC rate:     86.86 Hz
 *** DSSSD # 6 count:        33 old count:        29 dt:      2.31 s  HEC rate:     14.26 Hz

20.36 Noted x4 increase in F11 rate to c. 2kHz
      All degraders in (5.4mm Al)

      BRIKEN DAQ shows x4-5 increase for c. 6 minutes
      DSSSD bias & leakage currents OK attachments 7 & 8
Entry  Wed Nov 9 22:16:11 2016, CG, DK, TD, ML, NP1412 R24 R25 34x
07.15 R24 started. BRIKEN run 33.
      76~78Ni beam setting.
      F5 C target out. F7 CH2 in. F11 C out.
      F11 plastic rate ~500cps.
      1mm Al vairable degrader removed after ~3mins. Now have 4.4mm in.
      
07.26 R24 stopped for unknown reason - fill cycle had previously happened, similar scenario to yesterday.
      Soft reset performed successfully.

07.36 R25 started.
      No changes to beam settings above. F11 plastic rate ~500cps.
      Offline implant rates:
      *** DSSSD # 1 count:       894 old count:       682 dt:      1.02 s  HEC rate:    207.66 Hz
      *** DSSSD # 2 count:       781 old count:       590 dt:      1.02 s  HEC rate:    187.09 Hz
      *** DSSSD # 3 count:       900 old count:       682 dt:      1.02 s  HEC rate:    213.54 Hz
      *** DSSSD # 4 count:       790 old count:       594 dt:      1.02 s  HEC rate:    191.99 Hz
      *** DSSSD # 5 count:       421 old count:       317 dt:      1.02 s  HEC rate:    101.87 Hz
      *** DSSSD # 6 count:       162 old count:       128 dt:      1.02 s  HEC rate:     33.30 Hz

08.13 Beam stopped.
      corresponds to file R25_9.

08.15 Leakage currents attachments 1+2.

08.20 Takechi group changing settings: F5 C now in, F7 now empty, F11 C (6mm) now in.
      Beam returned to AIDA.
      Corresponding to 1.0Gb through file R25_10.
      4.4mm of Al variable degrader currently in place. Will gradually remove.

08.23 All variable degrader removed.
      Corresponds to start of file R25_11.
      Rates from offline implant finder:
      *** DSSSD # 1 count:       474 old count:       310 dt:      1.02 s  HEC rate:    160.08 Hz
      *** DSSSD # 2 count:       407 old count:       269 dt:      1.02 s  HEC rate:    134.70 Hz
      *** DSSSD # 3 count:       473 old count:       306 dt:      1.02 s  HEC rate:    163.01 Hz
      *** DSSSD # 4 count:       463 old count:       303 dt:      1.02 s  HEC rate:    156.18 Hz
      *** DSSSD # 5 count:       331 old count:       220 dt:      1.02 s  HEC rate:    108.35 Hz
      *** DSSSD # 6 count:        64 old count:        43 dt:      1.02 s  HEC rate:     20.50 Hz

09.18 Beam stopped 
      Corresponds to file R25_23

09.20 Takechi et al.change settings - F5, F7, F11 out
      Variable degrader 5.4mm Al in
      First AIDA data file R25_24, BRIKEN 161110_0923_767778Ni_036


from file R25_26 (offline):
 *** DSSSD # 1 count:     13306 old count:     13159 dt:      1.01 s  HEC rate:    145.63 Hz
 *** DSSSD # 2 count:      3578 old count:      3536 dt:      1.01 s  HEC rate:     41.61 Hz
 *** DSSSD # 3 count:      7699 old count:      7609 dt:      1.01 s  HEC rate:     89.16 Hz
 *** DSSSD # 4 count:      6191 old count:      6121 dt:      1.01 s  HEC rate:     69.35 Hz
 *** DSSSD # 5 count:      1891 old count:      1872 dt:      1.01 s  HEC rate:     18.82 Hz
 *** DSSSD # 6 count:      1377 old count:      1364 dt:      1.01 s  HEC rate:     12.88 Hz

      
10.08 All system wide checks OK *except* nnaida8 & 16 fail clock status, nnaida23 fails ADC calibration

10.14 Beam stopped 
      Corresponds to file R25_36

10.19 Takechi et al.change settings - F5 in, F7 out, F11 in
      Variable degrader all out
      First AIDA data file R25_37, BRIKEN 161110_1016_767778Ni_37

10.26 Rate, Hitrate, 1.8.L and 1.8.W spectra - see attachments 3-12
      All spectra cleared

10.30 Merge statistics - see attachments 13-17

10.32 Noted that nnaida19 histogram link has dropped - non-critical - continue DAQ

11.09 Change of target upstream: F5 empty, F7 CH2, F11 C 6mm
      Corresponds to AIDA data file R25_48
11:14 Change completed
      Corresponds to AIDA data file R25_49
11:21 All degraders out 
      Corresponds to AIDA data file R25_51

From AIDA R25_55:
 *** DSSSD # 1 count:       241 old count:         0 dt:      1.02 s  HEC rate:    235.75 Hz
 *** DSSSD # 2 count:       210 old count:         0 dt:      1.02 s  HEC rate:    205.42 Hz
 *** DSSSD # 3 count:       242 old count:         0 dt:      1.02 s  HEC rate:    236.72 Hz
 *** DSSSD # 4 count:       209 old count:         0 dt:      1.02 s  HEC rate:    204.44 Hz
 *** DSSSD # 5 count:       118 old count:         0 dt:      1.02 s  HEC rate:    115.43 Hz
 *** DSSSD # 6 count:        46 old count:         0 dt:      1.02 s  HEC rate:     45.00 Hz

11:40 lost beam 
      Current AIDA data R25_56
11:45 beam back
      Current AIDA data R25_57

12.19 Takechi et al. change settings - F5 in, F7 out, F11 C 6mm
      Corresponds to AIDA data file R25_65
      All degraders 5.4mm Al in

12.24 All degraders out 
      Corresponds to AIDA data file R25_67

      From offline analysis of R25_67

 *** DSSSD # 1 count:       584 old count:       472 dt:      3.91 s  HEC rate:    149.31 Hz
 *** DSSSD # 2 count:       516 old count:       410 dt:      3.91 s  HEC rate:    131.92 Hz
 *** DSSSD # 3 count:       607 old count:       488 dt:      3.91 s  HEC rate:    155.19 Hz
 *** DSSSD # 4 count:       595 old count:       475 dt:      3.91 s  HEC rate:    152.12 Hz
 *** DSSSD # 5 count:       439 old count:       355 dt:      3.91 s  HEC rate:    112.24 Hz
 *** DSSSD # 6 count:        87 old count:        69 dt:      3.91 s  HEC rate:     22.24 Hz

      ADC & disc data synchronised - see attachments 17-18

13.38 R25_84 Bias and leakage currents ok - see attachments 19 and 20

13.57 FEE settings checked by saving current settings to file and comparing against the last saved settings. No change was found in the ASIC settings.

14.33 BRIKEN now on run 40 file name 161110_1419. DSSD Rates

 *** DSSSD # 1 count:       138 old count:         0 dt:      1.02 s  HEC rate:    135.54 Hz
 *** DSSSD # 2 count:       119 old count:         0 dt:      1.02 s  HEC rate:    116.88 Hz
 *** DSSSD # 3 count:       139 old count:         0 dt:      1.02 s  HEC rate:    136.52 Hz
 *** DSSSD # 4 count:       134 old count:         0 dt:      1.02 s  HEC rate:    131.61 Hz
 *** DSSSD # 5 count:       101 old count:         0 dt:      1.02 s  HEC rate:     99.20 Hz
 *** DSSSD # 6 count:        20 old count:         0 dt:      1.02 s  HEC rate:     19.64 Hz

15:14 BigRIPS now on Run 439 (CH2 Target in at F7)
      AIDA file R25_104

15.17 BRIKEN now on run 41 file name 161110_1516

 *** DSSSD # 1 count:       380 old count:       169 dt:      1.05 s  HEC rate:    200.67 Hz
 *** DSSSD # 2 count:       326 old count:       144 dt:      1.05 s  HEC rate:    173.09 Hz
 *** DSSSD # 3 count:       381 old count:       172 dt:      1.05 s  HEC rate:    198.77 Hz
 *** DSSSD # 4 count:       336 old count:       150 dt:      1.05 s  HEC rate:    176.90 Hz
 *** DSSSD # 5 count:       191 old count:        86 dt:      1.05 s  HEC rate:     99.86 Hz
 *** DSSSD # 6 count:        73 old count:        35 dt:      1.05 s  HEC rate:     36.14 Hz

15.26 bias and leakage ok see attachment 21 and 22

16.04 BigRIPS Setting change now on 440 F5 In F7 Out F11 In

16.08 BRIKEN Now on run 42 File 161110_1608

16.09 Bias and leakage currents ok attachments 23 and 24

From offline r25_119

 *** DSSSD # 1 count:       255 old count:       117 dt:      1.04 s  HEC rate:    132.79 Hz
 *** DSSSD # 2 count:       234 old count:       111 dt:      1.04 s  HEC rate:    118.35 Hz
 *** DSSSD # 3 count:       259 old count:       120 dt:      1.04 s  HEC rate:    133.75 Hz
 *** DSSSD # 4 count:       250 old count:       115 dt:      1.04 s  HEC rate:    129.90 Hz
 *** DSSSD # 5 count:       181 old count:        84 dt:      1.04 s  HEC rate:     93.34 Hz
 *** DSSSD # 6 count:        33 old count:        18 dt:      1.04 s  HEC rate:     14.43 Hz

16.54 BRIKEN AND BigRIPS change runs. Corresponding to AIDA file R25_129
BigRIPS Now F5 In, F7 Out F11 Out
All variable Al degraders in 5.5mm

16.48 New BRIKEN run
      Run number 043
      File name: 161110_1658_767778Ni

17.02 Bias and leakage currents ok see attachment 25 and 26
      DSSD Rates from offline file R25_131

 *** DSSSD # 1 count:       269 old count:       127 dt:      1.04 s  HEC rate:    136.32 Hz
 *** DSSSD # 2 count:       224 old count:       107 dt:      1.04 s  HEC rate:    112.32 Hz
 *** DSSSD # 3 count:       257 old count:       118 dt:      1.04 s  HEC rate:    133.44 Hz
 *** DSSSD # 4 count:       160 old count:        71 dt:      1.04 s  HEC rate:     85.44 Hz
 *** DSSSD # 5 count:        29 old count:        13 dt:      1.04 s  HEC rate:     15.36 Hz
 *** DSSSD # 6 count:         3 old count:         2 dt:      1.04 s  HEC rate:      0.96 Hz

17.48 BigRIPS change setting putting back the thick Carbon at F11 6mm. Corresponds to AIDA file R25_141
      F5 In, F7 Out, F11 In C 6mm
      All variable Al degraders removed.
      BigRIPS run 442

17.52 BRIKEN starts run 44
      File name 161110_1752767778Ni
      AIDA file R25_142
      bias and leakage currents ok attachments 27 and 28

      DSSD rates offline from R25_142

 *** DSSSD # 1 count:       275 old count:       136 dt:      1.03 s  HEC rate:    134.63 Hz
 *** DSSSD # 2 count:       240 old count:       114 dt:      1.03 s  HEC rate:    122.04 Hz
 *** DSSSD # 3 count:       275 old count:       134 dt:      1.03 s  HEC rate:    136.57 Hz
 *** DSSSD # 4 count:       266 old count:       130 dt:      1.03 s  HEC rate:    131.73 Hz
 *** DSSSD # 5 count:       180 old count:        86 dt:      1.03 s  HEC rate:     91.05 Hz
 *** DSSSD # 6 count:        46 old count:        24 dt:      1.03 s  HEC rate:     21.31 Hz

18.16 System-wide checks OK *except* clock status (some/all of nnaida5,7,8,15&16 - see attachment 29) and ADC calibration (nnaida23) 

18.24 DAQ stops OK
      AIDA data filename R25_149

18.28 System wide checks show SYNC pulse and SYNC errors - see attachments 30-32

18.31 Stop & close merger & TapeServer
      DAQ reset
      DAQ setup - not complete - use Master Timestamp tab to ReSYNC 
      but note unusual(?) ReSYNC LSW value before second ReSYNC attempt- see attachments 33 & 34 before and after ReSYNC 







      



      
Entry  Wed Nov 9 11:07:14 2016, TD, CG, DK, NP1412 R22 R23 6x
DAQ start (R22) failed - merger remained in 'want first sync' state, no output to TapeServer -> soft reset OK

start 20.07
stop
AIDA filename R23
BRIKEN filename 161109_12012_767778Ni_23
BigRIPS filename 
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents OK see attachments 1 & 2
Degrader 5.4mm Al from start of run
F11 rate 450Hz
BigRIPS setting 76-78Ni

N.B. there are a number of correlation resets at the start of R23

Offline analysis of R23_2

 *** DSSSD # 1 count:       297 old count:       146 dt:      1.03 s  HEC rate:    146.07 Hz
 *** DSSSD # 2 count:       259 old count:       126 dt:      1.03 s  HEC rate:    128.66 Hz
 *** DSSSD # 3 count:       281 old count:       141 dt:      1.03 s  HEC rate:    135.43 Hz
 *** DSSSD # 4 count:       178 old count:        86 dt:      1.03 s  HEC rate:     89.00 Hz
 *** DSSSD # 5 count:        30 old count:        13 dt:      1.03 s  HEC rate:     16.45 Hz
 *** DSSSD # 6 count:         3 old count:         2 dt:      1.03 s  HEC rate:      0.97 Hz

00.04 R23 continuing.
      F5 C wedge in. F7 empty. F11 now empty. All variable degraders in place.
      F11 plastic rate ~400-500Hz.
      Change made around start of R23_70.
      Offline implant rates shown below for R23_72:
      *** DSSSD # 1 count:       562 old count:       414 dt:      1.06 s  HEC rate:    139.98 Hz
      *** DSSSD # 2 count:       507 old count:       372 dt:      1.06 s  HEC rate:    127.68 Hz
      *** DSSSD # 3 count:       552 old count:       404 dt:      1.06 s  HEC rate:    139.98 Hz
      *** DSSSD # 4 count:       350 old count:       262 dt:      1.06 s  HEC rate:     83.23 Hz
      *** DSSSD # 5 count:        46 old count:        37 dt:      1.06 s  HEC rate:      8.51 Hz
      *** DSSSD # 6 count:         4 old count:         3 dt:      1.06 s  HEC rate:      0.95 Hz

00:34 (corresponds to the beginning of R23_77) Remove 1 mm of Al variable degrader at F11 (now total Al is
2.2+2.2 = 4.4 mm)
      *** DSSSD # 1 count:       715 old count:       579 dt:      5.00 s  HEC rate:    142.96 Hz
      *** DSSSD # 2 count:       619 old count:       511 dt:      5.00 s  HEC rate:    123.77 Hz
      *** DSSSD # 3 count:       717 old count:       579 dt:      5.00 s  HEC rate:    143.36 Hz
      *** DSSSD # 4 count:       714 old count:       574 dt:      5.00 s  HEC rate:    142.76 Hz
      *** DSSSD # 5 count:       493 old count:       398 dt:      5.00 s  HEC rate:     98.57 Hz
      *** DSSSD # 6 count:        93 old count:        81 dt:      5.00 s  HEC rate:     18.59 Hz

00:43 R23_80 biases are shown as attachments 3 and 4

01.08 F11 C target installed.
      Corresponds to start of R23_90.
01.09 Final 4.4mm of variable Al degrader removed by start of R23_91.
      Implant rates sensible.

02.12 F5 C target removed. F7 CH2 target put in. F11 C target remains in place.
      Rate reduction during R23_115 as beam components changed.     
      implant rate >0 by start of R23_116 and BRIKEN run 30.
      F11 plastic rate ~500Hz.


3:15 Takechi group changes the condition.  This is beginning R23_132 
      F5 carbon, F7 empty.
      F11 pllastic rate is 450 cps
      F11 carbon remains in.

06.20 R23 stopped in anticipation of LN2 fill cycle.
      Takechi group changed setting > F5 C out,  F7 CH2 out, FF C in.
      All variable degraders in.
      
06.30 Leakage currents attachments 5+6.
Entry  Wed Nov 9 09:50:50 2016, TD, NP1412 R20, R21 1.png2.png3.png4.png
DAQ start (R20) failed - nnaida5 dropped data transfer #1 link

start 18.51
stop 19.30
AIDA filename R21
BRIKEN filename 161109_1858_767778Ni_21
BigRIPS filename 410
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents OK see attachments 1 & 2
Degrader 1mm Al from start of run
F11 rate 450Hz
BigRIPS setting 76-78Ni

19.35 Offline analysis of R21_12

 *** DSSSD # 1 count:       289 old count:       141 dt:      1.01 s  HEC rate:    146.23 Hz
 *** DSSSD # 2 count:       253 old count:       127 dt:      1.01 s  HEC rate:    124.49 Hz
 *** DSSSD # 3 count:       292 old count:       143 dt:      1.01 s  HEC rate:    147.22 Hz
 *** DSSSD # 4 count:       284 old count:       138 dt:      1.01 s  HEC rate:    144.25 Hz
 *** DSSSD # 5 count:       210 old count:        98 dt:      1.01 s  HEC rate:    110.66 Hz
 *** DSSSD # 6 count:        46 old count:        25 dt:      1.01 s  HEC rate:     20.75 Hz

Scalers - see attachment 4

1 F11
2 Upstream plastic top
3 Upstream plastic bottom
4 Si top
5 Si bottom
6 AIDA plastic veto
7 F11 and (upstream plastic top or upstream plastic bottom)
8 F11 and AIDA plastic veto 

19.42 Online analysis confirms ADC data synchronised
Entry  Wed Nov 9 08:47:03 2016, OH, TD, NP1412 R16, R17 and R18 r16bias1.pngr16bias2.png
start 17:44
stop 17:49
AIDA filename R16
BRIKEN filename 161105_1507
BigRIPS filename run405
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents ok attachments 1 and 2
Degrader All variable in at start (5.5mm Al)
F11 rate 430Hz
BigRIPS setting 76-78Ni

start 17:52
stop 17.55
AIDA filename R17
BRIKEN filename 
BigRIPS filename 
ASIC settings 2016Nov07_23.03.06
Degrader All variable in at start (5.5mm Al)
F11 rate 430Hz
BigRIPS setting 76-78Ni

start 17:56
stop 18:02
AIDA filename R18
BRIKEN filename 161105_1507
BigRIPS filename run405
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents ok
Degrader All variable in at start (5.5mm Al)
F11 rate 430Hz
BigRIPS setting 76-78Ni

start 18:04
stop 18.28
AIDA filename R19
BRIKEN filename 
BigRIPS filename 
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents ok
Degrader All variable in at start (5.5mm Al)
18.07 2x2.2mmAl degrader removed now running with 1.1mm.
F11 rate 430Hz
BigRIPS setting 76-78Ni

 *** DSSSD # 1 count:       474 old count:       224 dt:      1.03 s  HEC rate:    242.26 Hz
 *** DSSSD # 2 count:       414 old count:       194 dt:      1.03 s  HEC rate:    213.18 Hz
 *** DSSSD # 3 count:       480 old count:       225 dt:      1.03 s  HEC rate:    247.10 Hz
 *** DSSSD # 4 count:       414 old count:       196 dt:      1.03 s  HEC rate:    211.25 Hz
 *** DSSSD # 5 count:       232 old count:       111 dt:      1.03 s  HEC rate:    117.25 Hz
 *** DSSSD # 6 count:        95 old count:        42 dt:      1.03 s  HEC rate:     51.36 Hz

18.28 DAQ stopped - presumably ORNL HPGe clover LN2 refill cycle commences
Entry  Wed Nov 9 07:16:59 2016, OH, TD, NP1412 R15 6x
Run 15

start 16:14
stop 
AIDA filename 
BRIKEN filename 161109_1614
BigRIPS filename run406
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents ok attachments 1 and 2 
Degrader All variable in at start (5.5mm Al) at 
F11 rate 430Hz
BigRIPS setting 76-78Ni

16:16 removed all variable degrader

      Attachments 3, 4, 5 and 6 are plots for FEE 24 ASICs 1, 2, 3, 4

17.08 BRIKEN issues correlation scaler reset

17.32 Run stop
Entry  Wed Nov 9 02:14:08 2016, OH, TD, NP1412 R14 bias1.pngbias2.pngR14bias1.pngR14bias2.png
11.13 Still waiting to get beam back got access so took temp measurement.
      T=25.6 DP=0.0 H=18.1
      Biases and leakages good. Attachments 1 and 2

11.58 *All* system wide checks OK

14.29 *All* system wide checks OK



start 15:09
stop 16.04
AIDA filename R14
BRIKEN filename 161105_1507
BigRIPS filename run405
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents ok attachments 3 and 4
Degrader All variable in at start (5.5mm Al)
F11 rate 430Hz
BigRIPS setting 76-78Ni

 *** DSSSD # 1 count:       316 old count:       146 dt:      1.02 s  HEC rate:    166.21 Hz
 *** DSSSD # 2 count:       294 old count:       143 dt:      1.02 s  HEC rate:    147.64 Hz
 *** DSSSD # 3 count:       396 old count:       233 dt:      1.02 s  HEC rate:    159.37 Hz
 *** DSSSD # 4 count:       257 old count:       147 dt:      1.02 s  HEC rate:    107.55 Hz
 *** DSSSD # 5 count:        43 old count:        28 dt:      1.02 s  HEC rate:     14.67 Hz
 *** DSSSD # 6 count:         3 old count:         1 dt:      1.02 s  HEC rate:      1.96 Hz

15.24 Briefly reduce Al degrader from 5.4mm Al to 3.2mm Al
      Observe increased rate in DSSSDD #6, similar rates in  DSSSD #1 and increased
      rates in AIDA plastic veto 
15.26 Return to 5.4mm Al degrader

16.06 *All* system wide checks OK
Entry  Wed Nov 9 00:49:03 2016, TD, OH, NP1412 R13 r13bias1.pngr13bias2.png
09.50 Run 13 Start
      Thick C target (6mm) removed from F11. All variable Al degraders put in place
      AIDA settings file: 2016Nov07-23.03.06
      Keep alive script updated to current directory.

      Output from Tom's Implants.f script
 *** DSSSD # 1 count:      6316 old count:      6290 dt:      1.19 s  HEC rate:     21.89 Hz
 *** DSSSD # 2 count:      5435 old count:      5410 dt:      1.19 s  HEC rate:     21.05 Hz
 *** DSSSD # 3 count:      6314 old count:      6286 dt:      1.19 s  HEC rate:     23.57 Hz
 *** DSSSD # 4 count:      5211 old count:      5186 dt:      1.19 s  HEC rate:     21.05 Hz
 *** DSSSD # 5 count:      1383 old count:      1375 dt:      1.19 s  HEC rate:      6.73 Hz
 *** DSSSD # 6 count:        67 old count:        66 dt:      1.19 s  HEC rate:      0.84 Hz

Measured rates from file:

DSSD1:  309
DSSD2:  270
DSSD3:  308
DSSD4:  249
DSSD5:  67
DSSD6:  3.3

10.03 biases and leakage currents attachments 1 and 2

10.26 Run stopped as Takechi group optimise upstream lowering our rate.
Entry  Tue Nov 8 23:36:56 2016, DK, TD, NP1412 R12 10x

Takechi team changes the condition: F5 carbon; F7 empty; F11 carbon

  F11 Plastic is 1.1kHz

BRIKEN: 161109_0828_747576Ni_013.dlt

AIDA settings file: 2016Nov07-23.03.06

Initially we have all the Al variable degraders in, but then after the run began and we saw nothing, we reduce it to 1 mm of Aluminium.

Biases shown as attachments 1 and 2

Offline data rates (from R12_2)

 *** DSSSD # 1 count:      1022 old count:       709 dt:      2.91 s  HEC rate:    351.69 Hz
 *** DSSSD # 2 count:       863 old count:       597 dt:      2.91 s  HEC rate:    296.98 Hz
 *** DSSSD # 3 count:       995 old count:       693 dt:      2.91 s  HEC rate:    342.40 Hz
 *** DSSSD # 4 count:       827 old count:       575 dt:      2.91 s  HEC rate:    284.59 Hz
 *** DSSSD # 5 count:       246 old count:       164 dt:      2.91 s  HEC rate:     84.65 Hz
 *** DSSSD # 6 count:        11 old count:        10 dt:      2.91 s  HEC rate:      3.79 Hz
FEE64 temperatures OK - see attachment 3

08.55 Online analysis shows ADC data synchronised - see attachment 4

09.05 Rate, HitRate, 1.8.L and 1.8.H energy spectra - see attachmemnts 5-10

Run stop 9:46

 

Entry  Tue Nov 8 23:16:15 2016, DK, TD, NP1412 R11 R11_USB1.pngR11_USB2.png

Recovered AIDA after power cycle (unknown cause of crash from the end of R10)

Corresponding run in BRIKEN: 161109_0814_747576Ni_012

 

Run condition for Takechi has changed since our crash

F5 empty; F7 CH2; F11 carbon

At the beginning of our run, we had all Aluminium degraders in (1 mm was added back after our crash for safety)

We finally removed all but 1 mm Al on the variable degrader

SSD biases shown as attachments 1 and 2

8:20 They inform us that they want to take the F11 carbon out (in the middle of elog entry)

Entry  Tue Nov 8 22:04:21 2016, DK, NP1412 R10 R10_USB1.pngR10_USB2.png

Put the variable Al degraders all back in for safety (was 4.4 mm, now 5.4 mm)

Takechi team switches the condition: F5 empty; F7 CH2; F11 empty

 

7:00  Reset Run Control, run Setup, everything looks good.  Start the run

          Data rate around 10 MB/s

7:04  Checking the online data rate with MIDAS, the event rate is nearly zero...but it comes back (presume Takechi team was playing with the beam condition)

Events come back...

F11 Plastic rate is 1.2kHz

Here is a sample of the offline analysis of event rates with all Al degraders in:

 *** DSSSD # 1 count:       640 old count:       321 dt:      1.03 s  HEC rate:    309.51 Hz
 *** DSSSD # 2 count:       551 old count:       277 dt:      1.03 s  HEC rate:    265.85 Hz
 *** DSSSD # 3 count:       773 old count:       452 dt:      1.03 s  HEC rate:    311.45 Hz
 *** DSSSD # 4 count:       633 old count:       371 dt:      1.03 s  HEC rate:    254.20 Hz
 *** DSSSD # 5 count:       211 old count:       140 dt:      1.03 s  HEC rate:     68.89 Hz
 *** DSSSD # 6 count:         6 old count:         4 dt:      1.03 s  HEC rate:      1.94 Hz

7:11 Remove 1 mm of Al variable degrader (back to 4.4 mm total before AIDA)

Offline rates from R10_3:

 *** DSSSD # 1 count:      2596 old count:      2447 dt:      6.52 s  HEC rate:    398.43 Hz
 *** DSSSD # 2 count:      2193 old count:      2067 dt:      6.52 s  HEC rate:    336.58 Hz
 *** DSSSD # 3 count:      2641 old count:      2485 dt:      6.52 s  HEC rate:    405.34 Hz
 *** DSSSD # 4 count:      2626 old count:      2470 dt:      6.52 s  HEC rate:    403.04 Hz
 *** DSSSD # 5 count:      1940 old count:      1825 dt:      6.52 s  HEC rate:    297.75 Hz
 *** DSSSD # 6 count:       852 old count:       801 dt:      6.52 s  HEC rate:    130.76 Hz

7:18 Leak currents shown as attachments 1 and 2

7:20 Notice that the data rate has gone to zero...

 

Follow along with https://elog.ph.ed.ac.uk/AIDA/303 to fix it.  This looks good and the merger seems to come back successfully.

 

Trying to STOP on Run Control gave:

STATE for nnaida16 returned with an error
error: SOAP http transport timed out after 20000 ms
NONE
error: SOAP http transport timed out after 20000 ms
    while executing
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
    ("eval" body line 1)
    invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
    (procedure "DataAcquisitionControlClient__GetState" line 1)
    invoked from within
"DataAcquisitionControlClient__GetState"

 

Try a Reset on Run Control (this is the only option at present)

  nnaida16 showed "going" but after Reset shows "reset" like all the rest.

Now we get another error

JavaScript error occurred! 

Error description:     SyntaxError: missing ) after argument list
Page address:          http://localhost:8015/DataAcquisitionControl/DataAcquisitionControl.tml
Line number:           203

 

nnaida16 shows error

 

STATE for nnaida16 returned with an error
connect failed 
NONE
connect failed 
    while executing
"::http::geturl http://nnaida16:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 20000 -query {<?xml version="1.0" encoding="UTF-8..."
    ("eval" body line 1)
    invoked from within
"eval [list ::http::geturl $url] $args"
    (procedure "::http::geturl_followRedirects" line 4)
    invoked from within
"::http::geturl_followRedirects http://nnaida16:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 20000 -query {<?xml version="1.0"..."
    ("eval" body line 1)
    invoked from within
"eval ::http::geturl_followRedirects [list $url]  -headers [list $local_headers]  -type text/xml  -timeout $timeout  -query [list $request]  $local_pro..."
    (procedure "::SOAP::Transport::http::xfer" line 61)
    invoked from within
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
    ("eval" body line 1)
    invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
    (procedure "DataAcquisitionControlClient__GetState" line 1)
    invoked from within
"DataAcquisitionControlClient__GetState"
GET OPTION for nnaida16 returned with an error
connect failed 
NONE
connect failed 
    while executing
"::http::geturl http://nnaida16:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 60000 -query {<?xml version="1.0" encoding="UTF-8..."
    ("eval" body line 1)
    invoked from within
"eval [list ::http::geturl $url] $args"
    (procedure "::http::geturl_followRedirects" line 4)
    invoked from within
"::http::geturl_followRedirects http://nnaida16:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 60000 -query {<?xml version="1.0"..."
    ("eval" body line 1)
    invoked from within
"eval ::http::geturl_followRedirects [list $url]  -headers [list $local_headers]  -type text/xml  -timeout $timeout  -query [list $request]  $local_pro..."
    (procedure "::SOAP::Transport::http::xfer" line 61)
    invoked from within
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetOption HistEnable"
    ("eval" body line 1)
    invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetOption $args"
    (procedure "DataAcquisitionControlClient__GetOption" line 1)
    invoked from within
"DataAcquisitionControlClient__GetOption "$AcqGetOption""
GET OPTION for nnaida16 returned with an error
connect failed 
NONE
connect failed 
    while executing
"::http::geturl http://nnaida16:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 60000 -query {<?xml version="1.0" encoding="UTF-8..."
    ("eval" body line 1)
    invoked from within
"eval [list ::http::geturl $url] $args"
    (procedure "::http::geturl_followRedirects" line 4)
    invoked from within
"::http::geturl_followRedirects http://nnaida16:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 60000 -query {<?xml version="1.0"..."
    ("eval" body line 1)
    invoked from within
"eval ::http::geturl_followRedirects [list $url]  -headers [list $local_headers]  -type text/xml  -timeout $timeout  -query [list $request]  $local_pro..."
    (procedure "::SOAP::Transport::http::xfer" line 61)
    invoked from within
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetOption Xfer2Enable"
    ("eval" body line 1)
    invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetOption $args"
    (procedure "DataAcquisitionControlClient__GetOption" line 1)
    invoked from within
"DataAcquisitionControlClient__GetOption "$AcqGetOption""
 

TapeServer was still running, so maybe that was a mistake...?

 

Power cycle...

Entry  Tue Nov 8 21:01:40 2016, CG, DK, NP1412 R9 
06.02 R9 started with Takechi group C target out and all variable degraders in.
      F5 C in. F7 empty. F11 empty.
      F11 plastic rate 1.1kHz.
      Offline sort rates below:
      *** DSSSD # 1 count:       636 old count:       317 dt:      1.01 s  HEC rate:    314.93 Hz
      *** DSSSD # 2 count:       544 old count:       267 dt:      1.01 s  HEC rate:    273.47 Hz
      *** DSSSD # 3 count:       628 old count:       308 dt:      1.01 s  HEC rate:    315.92 Hz
      *** DSSSD # 4 count:       532 old count:       267 dt:      1.01 s  HEC rate:    261.62 Hz
      *** DSSSD # 5 count:       146 old count:        76 dt:      1.01 s  HEC rate:     69.11 Hz
      *** DSSSD # 6 count:         6 old count:         3 dt:      1.01 s  HEC rate:      2.96 Hz

06.09 Removed 1mm of variable degrader.
      Offline sort rates/below:
      *** DSSSD # 1 count:      1501 old count:      1316 dt:      4.61 s  HEC rate:    325.34 Hz
      *** DSSSD # 2 count:      1231 old count:      1079 dt:      4.61 s  HEC rate:    266.82 Hz
      *** DSSSD # 3 count:      1498 old count:      1314 dt:      4.61 s  HEC rate:    324.69 Hz
      *** DSSSD # 4 count:      1502 old count:      1323 dt:      4.61 s  HEC rate:    325.55 Hz
      *** DSSSD # 5 count:      1346 old count:      1182 dt:      4.61 s  HEC rate:    291.74 Hz
      *** DSSSD # 6 count:       487 old count:       430 dt:      4.61 s  HEC rate:    105.56 Hz

06.20 Stopped the run preemptively to Ge LN2 autofill which is known to crash the DAQ
Entry  Tue Nov 8 18:52:32 2016, CG, DK, NP1412 R7 R8 bias1.pngbias2.pngbias1.pngbias2.png
03.50 74-76Ni beam delivered to F11.
      F11 plastic rate ~1kHz.
      F5 C target in. F7 empty. F11 C target in.
      R7 started 03.51.
      Initial rate 0Hz in all DSSD.
      Removed all variable degraders and implant rate only up to ~1Hz in DSSD1.
      Will replace all variable degraders, enter B2F and remove fixed 5mm plate and work back from there.
      R7 stopped 04.00.

04.07 Beam returned to F11.
      F11 plastic rate 1.1kHz.
      R8 started normally.
      Will now gradually remove variable degraders.

04.18 Now removed all variable degraders.
      Implant rates from offline/full analysis below:
      *** DSSSD # 1 count:       697 old count:       354 dt:      1.03 s  HEC rate:    333.95 Hz
      *** DSSSD # 2 count:       592 old count:       308 dt:      1.03 s  HEC rate:    276.51 Hz
      *** DSSSD # 3 count:       696 old count:       349 dt:      1.03 s  HEC rate:    337.85 Hz
      *** DSSSD # 4 count:       688 old count:       348 dt:      1.03 s  HEC rate:    331.03 Hz
      *** DSSSD # 5 count:       609 old count:       312 dt:      1.03 s  HEC rate:    289.16 Hz
      *** DSSSD # 6 count:       229 old count:       116 dt:      1.03 s  HEC rate:    110.02 Hz

      Leakage currents @ 04.21 attachments 1+2.

05.55 Takechi group want to remove C target.
      Moving variable degraders back into place.
      R8 stopped 05.59.
      Leakage currents @ 05.55 attachments 3+4.
Entry  Tue Nov 8 15:10:16 2016, CG, TD, DK, OH, NP1412 R5 R6 bias1.pngbias2.png
00.15 Setting changed to 71,72,73Ni. All variable degraders in.
      Will remove while watching rates to settle on degrader setting.
      F5 CH2 in. F7 CH2 out. F11 C target in.
      F11 plastic rate 1.2kHz.
      R5 started.
      
00.30 Cut F2 momentum slit. F11 plastic rate down to ~400cps.

00.34 Increased primary beam intensity by x2.
      F11 plastic rate increased to ~800cps.
      This setting should run until the beam change ~03.00.
      
00.47 R6 started.
      All degraders gradually removed but implant rate in DSSD1 only ~10s Hz.

01.03 71,72,73Ni beam stopped for tuning to next setting.
      R6 stopped.
      Leakage currents attachments 1+2
Entry  Tue Nov 8 14:29:34 2016, TD, OH, DK, NP1412 Run 4 
23:28 Run 4 Start
      Thick 6mm C target in beam
      
      Running with 3.2mm (variable) + 8mm (fixed) of Al degrader. Fixed degrader in same position as R3.
      Variable degrader reduced to 1mm of Al.
Entry  Tue Nov 8 14:22:16 2016, TD, OH, DK, NP1412 R3 
11:21 Run 3 Started
      Degrader moved upstream. AIDA moved to centre of BRIKEN. Fixed degrader moved upstream from centre of BRIKEN.
11:25 Run 3 Stopped
ELOG V3.1.4-unknown