Mon Jun 5 08:43:24 2017, CB, TD, PW, DK, Monday 5 June 16:00 to 24:00 38x
|
16:43 Leakage currents OK - see attachment 1
FEE64 temperatures OK - see attachment 2
Statistics show rate increase in nnaida13, 21. Reason unclear - see attachment 3
|
Sun Jun 4 23:50:24 2017, CB, PW, DK, TD, Monday 5 June 08:00 to 16:00 24x
|
07:50 Leakage currents OK - attachment 1
FEE64 temperatures OK - attachment 2
Good event statistics OK - attachment 3
|
Sun Jun 4 16:00:49 2017, NN, OH, Monday 5 June 00:00 to 08:00 19x
|
23:59 BigRIPS and Briken stop their DAQ
AIDA run file R2_245
|
Mon Dec 5 10:51:08 2016, TD, Monday 5 December    
|
19.50 DSSSD bias & leakage currents OK - see attachments 1 & 2
FEE64 temperatures HIGH - see attachment 3
FL11006 E01 low fluid level (actual halfway between minimum and middle visual markers)
|
Mon May 4 01:27:56 2015, JA, TD, AE, CG & JLT, Monday 4 May 2015 16x
|
09.18 SN changed degrader to 5.7mm Al (4mm + all variable degraders + additional 1mm)
|
Mon Oct 31 02:00:01 2016, TD, Monday 31 October 11x
|
11.00 DAQ continues OK
DAQ STOP OK
Most recent files R91_568 last accessed 10.59
|
Mon Jul 31 09:43:37 2023, TD, Monday 31 July 44x
|
10.43 DSSSD bias +100V leakage current +7.550uA
|
Mon May 30 02:50:19 2016, CG, Monday 30th May 
|
10.30 Pulser walkthrough carried out 90,000 - 10,000 @ 10,000 step
10.40 .msf spectra saved /MIDAS/SpecData/AIDA
|
Mon Oct 30 01:10:28 2017, TD, OH, Monday 30 October 8x
|
10.10 DAQ running OK - data transfer not enabled
BNC PB-4
|
Mon Jun 3 05:40:19 2019, TD, Monday 3 June 25x
|
13.40 DAQ continues OK
ASIC settings 2019May19-17.54.03
|
Sun Oct 28 23:34:28 2018, TD, OH, CA, Monday 29 October 15x
|
08.34 DAQ continues OK - current file Oct18/R9_6
08.37 all system wide checks OK *except* nnaida7 fails ADC calibration
|
Mon May 29 00:59:27 2017, TD, PV, CB, NN, OH, Monday 29 May 10x
|
09.00 DAQ continues OK
Current file May2017/R32_239
|
Mon Aug 29 06:57:32 2016, TD, Monday 29 August 7x
|
Per the instructions provided by https://elog.ph.ed.ac.uk/AIDA/331 added Servisol Heat Sink Compound
to each of the 24x FEE64s currently installed (plus spare FEE module stored in the B3F AIDA storage
|
Sun Nov 27 16:29:51 2016, CG, AE, DK, Monday 28 November 12x
|
01.28 Beam stopped temporarily.
~750MB into R26_38.
Bias and leakage currents ok, as are FEE temps (attachments 1-3).
|
Tue May 28 01:37:14 2019, TD, Monday 28 May 8x
|
09.30 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 75% M-H markers
Temperature +25.1 deg C dew point +9.7 deg C RH 37.9%
|
Mon May 27 02:50:20 2019, TD, Monday 27 May 30x
|
10.50 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 60% M-H markers - topped up 1x jugful to 100% M-H markers
Temperature +25.6 deg C dew point +9.6 deg C RH 36.2%
|
Mon Mar 27 03:29:05 2017, OH, TD, DK, CG, Monday 27 March 15x
|
11:28 Biases and leakages okm (see attachment 1)
14:00 Removed jumpers on cooling rack D. Noise was around 50 channels worse (FWHM of pulser peak) for all
|
Mon Apr 27 01:48:59 2015, JA, LHB, TD, VP, AE & CG, Monday 27 April 2015 8x
|
DSSSD #1 +100V, +3.8uA
DSSSD #2 +100V, +5.5uA
|
Mon Jun 25 01:56:08 2018, OH, TD, Monday 25th June 21x
|
09:55 Alpha run stopped
10:45 Investigating the large noise in FEE28
|
Mon Apr 25 10:11:40 2016, CG, KM, PV & TD, Monday 25 April
|
AIDA stand, Julabo FL11006 recirculating chiller (with 110VAC/50Hz - 240VAC/50Hz 3ph transformer)
moved from F11 downstream wall to ~2m downstream of EURICA
|