  646   Thu Jun 8 16:06:21 2017 OHFriday 9 June 00:00-08:00
00:06 System wide checks all ok
      Bias and leakage currents ok - attachment 1
      Good event statistics - attachment 2
      FEE temperatures - attachment 3

00:29 BRIKEN and BigRIPS stop their DAQs
      AIDA on file R7_494

00:31 BigRIPS starts run 3080
      BRIKEN starts run 77
      AIDA on file R7_494

01:31 BigRIPS and BRIKEN stop DAQs
      AIDA on file R7_508

01:33 BRIKEN starts run 78
      BigRIPS starts run 3081
      AIDA on file R7_508

02:10 System wide checks all ok
      Bias and leakage ok - attachment 4
      Good event statistics - attachment 5
      FEE temperatures - attachment 6

02:30 BRIKEN and BigRIPS stop their DAQs
      AIDA on file R7_211

02:32 BigRIPs started Run 3082
      BigRIPS stopped

02:33 Beam stopped to put and extra 1mm of degrader in to shift beam
      Current degrader settings:
      F7 = 2mm Al
      F11 = 4.5mm Al

02:39 biGRIPS starts run 3083
      BRIKEN starts run 79
      AIDA on file R7_523

      DSSD rates from R7_235
 *** DSSSD # 1 count:    119355 old count:     78874 dt:      1.05 s  LEC rate:  38696.78 Hz
 *** DSSSD # 2 count:    188044 old count:    124828 dt:      1.05 s  LEC rate:  60429.72 Hz
 *** DSSSD # 3 count:     53151 old count:     34736 dt:      1.05 s  LEC rate:  17603.35 Hz
 *** DSSSD # 4 count:     61647 old count:     40919 dt:      1.05 s  LEC rate:  19814.40 Hz
 *** DSSSD # 5 count:     14623 old count:      9739 dt:      1.05 s  LEC rate:   4668.74 Hz
 *** DSSSD # 6 count:     41905 old count:     27687 dt:      1.05 s  LEC rate:  13591.33 Hz
 *** DSSSD # 1 count:         1 old count:         1 dt:      1.05 s  HEC rate:      0.00 Hz
 *** DSSSD # 2 count:        11 old count:         7 dt:      1.05 s  HEC rate:      3.82 Hz
 *** DSSSD # 3 count:        19 old count:        10 dt:      1.05 s  HEC rate:      8.60 Hz
 *** DSSSD # 4 count:        15 old count:        10 dt:      1.05 s  HEC rate:      4.78 Hz
 *** DSSSD # 5 count:         0 old count:         0 dt:      1.05 s  HEC rate:      0.00 Hz
 *** DSSSD # 6 count:         3 old count:         2 dt:      1.05 s  HEC rate:      0.96 Hz

03:37 bigRIPS and BRIKEN stop their daqs
      AIDA on file R7_537

03:39 BigRIPS starts run 3084
      BRIKEN starts run 80
      AIDA on file R7_537

04:16 System wide checks ok
      Leakage and bias ok - attachment 7
      Good events - attachment 8
      FEE temperatures - attachment 9

04:33 BRIKEN AND BigRIPS stop daqs
      AIDA on file R7_549

04:34 BRIKEN starts run 81
      BigRIPS starts run 3085
      AIDA on file R7_550

05:30 BigRIPS and BRIKEN stop their DAQs
      AIDA on file R7_563

05:40 bigRIPS starts run 3086
      BRIKEN starts run 82
      AIDA on R7_565

06:14 System wide checks all ok
      Bias and leakages ok - attachment 10
      Good events - attachment 11
      FEE temperatures - attachment 12

06:33 BigRIPS and BRIKEN stop their DAQs
      AIDA on file R7_577

06:34 BigRIPS starts run 3087
      BRIKEN starts run 83
      AIDA on file R7_577

07:30 BRIKEN and BigRIPS stop daqs
      AIDA on file R7_590

07:31 BigRIPS starts run 3088
      BRIKEN starts run 84
      AIDA on file R7_591
  654   Sun Jun 11 00:07:24 2017 OHSunday 11 June 08:00-16:00
08:10 BRIKEN and BigRIPS stops DAQs
      AIDA on file R7_1287

08:11 Beam going for half an hour. Problem with vacuum around RRC

08:24 Beam is back

08:25 BigRIPS starts run 4016
      BRIKEN starts run 126
      AIDA on file R7_1291

08:32 RF down should be 5 minutes

08:44 Beam back

09:25 RF Down

09:26 BigRIPS and BRIKEN stop DAQs
      AIDA on file R7_1306

09:35 Beam back
      BigRIPS starts run 4017
      BRIKEN starts run 127
      AIDA on file R7_1308

09:53 System wide checks all ok
      Bias and leakage currents ok - attachment 1
      Good events - attachment 2
      FEE temperatures - attachment 3

10:29 BRIKEN and BigRIPS stop DAQs
      AIDA on file R7_1322

10:31 BigRIPS starts run 4018
      BRIKEN starts run 128
      AIDA on file R7_1323

11:31 BigRIPS and BRIKEN stop their DAQs
      AIDA on file R7_1338

11:32 BigRIPS starts run 4019
      BRIKEN starts run 129
      AIDA on file R7_1338

11.49 Detector biases & leakage currents OK - see attachments 4 & 5
      FEE64 Temperatures OK - see attachment 6
      good events statistics stable & OK - see attachment 7
      No warnings/errors reported by merger since DAQ GO - see attachment 8

      *all* system wide checks OK (zero fails)

11.53 online analysis
      per DSSSD x versus y (HEC) - see attachment 9
      per DSSSD E_p versus E_n (HEC) - see attachment 10
      ADC & disc data synchronised - see attachment 11

11.55 analysis of file RIBF128/R7_1344 - dead time < 0.5%, no timewarps, 1 MBS timewarps - see attachment 12

12.03 offline analysis RIBF128/R7_1346 rates

 *** DSSSD # 1 count:    942048 old count:    885423 dt:      1.04 s  LEC rate:  54296.31 Hz
 *** DSSSD # 2 count:   1402896 old count:   1319759 dt:      1.04 s  LEC rate:  79718.00 Hz
 *** DSSSD # 3 count:    422965 old count:    397383 dt:      1.04 s  LEC rate:  24529.94 Hz
 *** DSSSD # 4 count:    356558 old count:    336037 dt:      1.04 s  LEC rate:  19677.08 Hz
 *** DSSSD # 5 count:    124978 old count:    117233 dt:      1.04 s  LEC rate:   7426.49 Hz
 *** DSSSD # 6 count:    277397 old count:    260863 dt:      1.04 s  LEC rate:  15854.04 Hz
 *** DSSSD # 1 count:         6 old count:         6 dt:      1.04 s  HEC rate:      0.00 Hz
 *** DSSSD # 2 count:       150 old count:       139 dt:      1.04 s  HEC rate:     10.55 Hz
 *** DSSSD # 3 count:        98 old count:        96 dt:      1.04 s  HEC rate:      1.92 Hz
 *** DSSSD # 4 count:        12 old count:        10 dt:      1.04 s  HEC rate:      1.92 Hz
 *** DSSSD # 5 count:        36 old count:        33 dt:      1.04 s  HEC rate:      2.88 Hz
 *** DSSSD # 6 count:       155 old count:       145 dt:      1.04 s  HEC rate:      9.59 Hz

      all SortSas spectra & histograms zero'd, waveforms set x-range = 1000ch = 2ms

12.04 BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency 25Hz range 2.6Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Mar27-16.12.39

      BigRIPS 149Xe setting, F1 slits 51mm,  degrader 2mm Al (at F7), 1.3mm Al fixed + 2mm Al variable (at F11)

12.15 layout spectra - see attachments 13-20

12.20 FEE64# 1-16 power fail
      AIDA on file R7_1351

12.30 requested beam off, replaced 250V/16A/fast fuse

      nnaida13 good events rate ~250k, cycled DSSSD bias reduced to ~140k

13:06 BRIKEN starts run 130
      BigRIPS starts run 4020
      AIDA on file R9_0

13:07 Ambient Temperature +23.9 - Humidity 29.6% - Dew point +5.1C
      Leakage currents and bias ok - attachment 21
      Water level in chiller ok

14:06 BigRIPS and BRIKEN stop their daqs
      AIDA on file R9_16

14:08 BigRIPS starts run 4021
      BRIKEN starts run 131
      AIDA on file R9_16

14:31 System wide checks ok
      Good events - attachment 22
      FEE temperatures - attachment 23
      Analysis output for R9_16 - attachment 24

15:10 BRIKEN and BigRIPS stop daqs
      AIDA on file R9_32

15:11 BRIKEN starts run 132
      BigRIPS starts run 4022
      AIDA on R9_32

15:13 Leakage and bias ok - attachment 25

15:30 DSSD Rates from R9_36
 *** DSSSD # 1 count:     95157 old count:     45985 dt:      1.03 s  LEC rate:  47603.61 Hz
 *** DSSSD # 2 count:    153477 old count:     76756 dt:      1.03 s  LEC rate:  74273.90 Hz
 *** DSSSD # 3 count:     40379 old count:     20036 dt:      1.03 s  LEC rate:  19694.14 Hz
 *** DSSSD # 4 count:     46339 old count:     23014 dt:      1.03 s  LEC rate:  22581.02 Hz
 *** DSSSD # 5 count:     14241 old count:      7012 dt:      1.03 s  LEC rate:   6998.42 Hz
 *** DSSSD # 6 count:     36061 old count:     18008 dt:      1.03 s  LEC rate:  17477.18 Hz
 *** DSSSD # 1 count:         0 old count:         0 dt:      1.03 s  HEC rate:      0.00 Hz
 *** DSSSD # 2 count:        16 old count:         8 dt:      1.03 s  HEC rate:      7.74 Hz
 *** DSSSD # 3 count:        14 old count:         8 dt:      1.03 s  HEC rate:      5.81 Hz
 *** DSSSD # 4 count:         1 old count:         0 dt:      1.03 s  HEC rate:      0.97 Hz
 *** DSSSD # 5 count:         4 old count:         1 dt:      1.03 s  HEC rate:      2.90 Hz
 *** DSSSD # 6 count:        26 old count:        13 dt:      1.03 s  HEC rate:     12.59 Hz

16:10 BigRIPS and BRIKEN stop DAQs
      AIDA on file R9_47

16:12 BRIKEN starts run 133
      BigRIPS starts run 4023
      AIDA on file R9_47

16:18 BigRIPS slit is closing
     DSSD rates for R9_49

 *** DSSSD # 1 count:    121641 old count:     64167 dt:      1.00 s  LEC rate:  57287.24 Hz
 *** DSSSD # 2 count:    179018 old count:     93277 dt:      1.00 s  LEC rate:  85462.38 Hz
 *** DSSSD # 3 count:     52138 old count:     28005 dt:      1.00 s  LEC rate:  24054.58 Hz
 *** DSSSD # 4 count:     53927 old count:     27419 dt:      1.00 s  LEC rate:  26421.86 Hz
 *** DSSSD # 5 count:     16617 old count:      9291 dt:      1.00 s  LEC rate:   7302.19 Hz
 *** DSSSD # 6 count:     38751 old count:     19986 dt:      1.00 s  LEC rate:  18704.02 Hz
 *** DSSSD # 1 count:         1 old count:         0 dt:      1.00 s  HEC rate:      1.00 Hz
 *** DSSSD # 2 count:        52 old count:        30 dt:      1.00 s  HEC rate:     21.93 Hz
 *** DSSSD # 3 count:        12 old count:         7 dt:      1.00 s  HEC rate:      4.98 Hz
 *** DSSSD # 4 count:         3 old count:         3 dt:      1.00 s  HEC rate:      0.00 Hz
 *** DSSSD # 5 count:         9 old count:         5 dt:      1.00 s  HEC rate:      3.99 Hz
 *** DSSSD # 6 count:        37 old count:        19 dt:      1.00 s  HEC rate:     17.94 Hz
  659   Mon Jun 12 15:51:15 2017 OHTuesday 13 June 00:00-08:00
00:02 System wide checks ok
      Good events - attachment 1
      FEE temperatures - attachment 2
      Bias and leakages ok - attachment 3

00:31 BRIKEN and BigRIPS stop their DAQs
      AIDA on file R3_236

00:32 BRIKEN starts run 40
      BigRIPS starts run 2512
      AIDA on file R3_236

01:30 BRIKEN and BigRIPS stop their DAQs
      AIDA on file R3_254

01:31 BRIKEN starts run 41
      BigRIPS starts run 2513
      AIDA on file R3_255

01:41 DSSD rates from R3_257
 *** DSSSD # 1 count:    119356 old count:     62524 dt:      2.02 s  LEC rate:  59170.96 Hz
 *** DSSSD # 2 count:    168540 old count:     87963 dt:      2.02 s  LEC rate:  83554.01 Hz
 *** DSSSD # 3 count:     61578 old count:     32845 dt:      2.02 s  LEC rate:  30527.41 Hz
 *** DSSSD # 4 count:     49914 old count:     25818 dt:      2.02 s  LEC rate:  24744.96 Hz
 *** DSSSD # 5 count:     25977 old count:     13821 dt:      2.02 s  LEC rate:  12878.15 Hz
 *** DSSSD # 6 count:     43384 old count:     22072 dt:      2.02 s  LEC rate:  21507.70 Hz
 *** DSSSD # 1 count:         0 old count:         0 dt:      2.02 s  HEC rate:      0.00 Hz
 *** DSSSD # 2 count:         6 old count:         3 dt:      2.02 s  HEC rate:      2.97 Hz
 *** DSSSD # 3 count:         1 old count:         1 dt:      2.02 s  HEC rate:      0.50 Hz
 *** DSSSD # 4 count:         1 old count:         1 dt:      2.02 s  HEC rate:      0.50 Hz
 *** DSSSD # 5 count:        14 old count:        10 dt:      2.02 s  HEC rate:      6.94 Hz
 *** DSSSD # 6 count:       211 old count:       118 dt:      2.02 s  HEC rate:    104.60 Hz

01:43 Analysis of R3_257 - attachment 4
      No timewarps

02:10 System wide checks ok
      Leakage and bias ok - attachment 5
      Good events - attachment 6
      FEE temperatures - attachment 7

02:30 BRIKEN and BigRIPS stop their DAQs
      AIDA on file R3_274

02:32 BRIKEN starts run 42
      BigRIPS starts run 2514
      AIDA on file R3_274

03:30 BRIKEN and BigRIPS stop their DAQs
      AIDA on file R3_292

03:32 BRIKEN starts run 43
      BigRIPS starts run 2514
      AIDA on file R3_293

04:28 BRIKEN and BigRIPS stop their DAQs
      AIDA on file R3_311

04:29 BRIKEN starts run 44
      BigRIPS starts run 2515
      AIDA on file R3_311

04:54 System wide checks ok
      Bias and leakage ok - attachment 8
      Good Events - attachment 9
      FEE temperatures - attachment 10

05:00 DSSD rates from R3_319
 *** DSSSD # 1 count:    255278 old count:    193923 dt:      1.01 s  LEC rate:  60935.40 Hz
 *** DSSSD # 2 count:    360584 old count:    272321 dt:      1.01 s  LEC rate:  87659.38 Hz
 *** DSSSD # 3 count:    133027 old count:    100845 dt:      1.01 s  LEC rate:  31961.91 Hz
 *** DSSSD # 4 count:    108047 old count:     82416 dt:      1.01 s  LEC rate:  25455.71 Hz
 *** DSSSD # 5 count:     51776 old count:     38962 dt:      1.01 s  LEC rate:  12726.37 Hz
 *** DSSSD # 6 count:     88202 old count:     66648 dt:      1.01 s  LEC rate:  21406.59 Hz
 *** DSSSD # 1 count:         1 old count:         1 dt:      1.01 s  HEC rate:      0.00 Hz
 *** DSSSD # 2 count:        11 old count:         8 dt:      1.01 s  HEC rate:      2.98 Hz
 *** DSSSD # 3 count:         2 old count:         2 dt:      1.01 s  HEC rate:      0.00 Hz
 *** DSSSD # 4 count:         3 old count:         2 dt:      1.01 s  HEC rate:      0.99 Hz
 *** DSSSD # 5 count:        32 old count:        25 dt:      1.01 s  HEC rate:      6.95 Hz
 *** DSSSD # 6 count:       439 old count:       335 dt:      1.01 s  HEC rate:    103.29 Hz

05:30 BRIKEN and BigRIPS stop DAQs
      AIDA on file R3_330

05:31 BigRIPS starts run 2516
      BRIKEN starts run 45
      AIDA on file R3_331

06:31 BigRIPS and BRIKEN stop DAQs
      AIDA on file R3_350

06:32 BRIKEN starts run 46
      BigRIPS starts run 2517
      AIDA on file R3_350

06:50 System wide checks ok
      Bias and leakage currents ok - attachment 11
      Good events - attachment 12
      FEE temperatures - attachment 13

07:09 Merger still running ok - attachment 14

07:31 BigRIPS and BRIKEN stop their DAQs
      AIDA on file R3_369

07:32 BRIKEN starts run 47
      BigRIPS starts run 2518
      AIDA on file R3_369
  663   Thu Jun 15 07:44:41 2017 OHRIKEN Cable Inventory
List of the type and quantity of cables we have at RIKEN that are not currently in use.
  677   Fri Oct 20 01:43:40 2017 OHFriday 20th October
09:43 DAQ continues ok
      System wide checks all ok
      Detector bias and leakage ok - Attachment 1
      FEE temperatures ok - Attachment 2
      On file R3_279 writing at ~11Mb/s

10:07 Waveforms for nnaida 13, 11, 20 and 22 - Attachments 3-7
      nnaida 13 & 22 p-side
      nnaida 11 & 20 n-side
      Preamp feedback settings for ASIC are still default

10:49 DAQ Stopped
      Detector bias taken off
  679   Sun Oct 22 02:10:03 2017 OHSunday 22 October
10:10 DAQ found crashed
      Stopped writing to file at 05:31
      System wide checks ok
      Statistics - Attachment 1
      FEE temperatures - Attachment 2

      Error message when stopping the DAQ - Attachment 3

      Bias and leakage currents ok - Attachment 4

10:18 Error when trying to reset and setup DAQ - Attachment 5

      Followed by:
STATE for nnaida1 returned with an error
connect failed 
NONE
connect failed 
    while executing
"::http::geturl http://nnaida1:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 20000
-query {<?xml version="1.0" encoding="UTF-8"..."
    ("eval" body line 1)
    invoked from within
"eval [list ::http::geturl $url] $args"
    (procedure "::http::geturl_followRedirects" line 4)
    invoked from within
"::http::geturl_followRedirects http://nnaida1:8015/DataAcquisitionControlServer -headers {} -type text/xml
-timeout 20000 -query {<?xml version="1.0" ..."
    ("eval" body line 1)
    invoked from within
"eval ::http::geturl_followRedirects [list $url]  -headers [list $local_headers]  -type text/xml  -timeout
$timeout  -query [list $request]  $local_pro..."
    (procedure "::SOAP::Transport::http::xfer" line 61)
    invoked from within
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
    ("eval" body line 1)
    invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
    (procedure "DataAcquisitionControlClient__GetState" line 1)
    invoked from within
"DataAcquisitionControlClient__GetState"
STATE for nnaida2 returned with an error
connect failed 
NONE
connect failed 
    while executing
"::http::geturl http://nnaida2:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 20000
-query {<?xml version="1.0" encoding="UTF-8"..."
    ("eval" body line 1)
    invoked from within
"eval [list ::http::geturl $url] $args"
    (procedure "::http::geturl_followRedirects" line 4)
    invoked from within
"::http::geturl_followRedirects http://nnaida2:8015/DataAcquisitionControlServer -headers {} -type text/xml
-timeout 20000 -query {<?xml version="1.0" ..."
    ("eval" body line 1)
    invoked from within
"eval ::http::geturl_followRedirects [list $url]  -headers [list $local_headers]  -type text/xml  -timeout
$timeout  -query [list $request]  $local_pro..."
    (procedure "::SOAP::Transport::http::xfer" line 61)
    invoked from within
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
    ("eval" body line 1)
    invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
    (procedure "DataAcquisitionControlClient__GetState" line 1)
    invoked from within
"DataAcquisitionControlClient__GetState"
STATE for nnaida3 returned with an error
connect failed 
NONE
connect failed 
    while executing
"::http::geturl http://nnaida3:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 20000
-query {<?xml version="1.0" encoding="UTF-8"..."
    ("eval" body line 1)
    invoked from within
"eval [list ::http::geturl $url] $args"
    (procedure "::http::geturl_followRedirects" line 4)
    invoked from within
"::http::geturl_followRedirects http://nnaida3:8015/DataAcquisitionControlServer -headers {} -type text/xml
-timeout 20000 -query {<?xml version="1.0" ..."
    ("eval" body line 1)
    invoked from within
"eval ::http::geturl_followRedirects [list $url]  -headers [list $local_headers]  -type text/xml  -timeout
$timeout  -query [list $request]  $local_pro..."
    (procedure "::SOAP::Transport::http::xfer" line 61)
    invoked from within
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
    ("eval" body line 1)
    invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
    (procedure "DataAcquisitionControlClient__GetState" line 1)
    invoked from within
"DataAcquisitionControlClient__GetState"
STATE for nnaida4 returned with an error
connect failed 
NONE
connect failed 
    while executing
"::http::geturl http://nnaida4:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 20000
-query {<?xml version="1.0" encoding="UTF-8"..."
    ("eval" body line 1)
    invoked from within
"eval [list ::http::geturl $url] $args"
    (procedure "::http::geturl_followRedirects" line 4)
    invoked from within
"::http::geturl_followRedirects http://nnaida4:8015/DataAcquisitionControlServer -headers {} -type text/xml
-timeout 20000 -query {<?xml version="1.0" ..."
    ("eval" body line 1)
    invoked from within
"eval ::http::geturl_followRedirects [list $url]  -headers [list $local_headers]  -type text/xml  -timeout
$timeout  -query [list $request]  $local_pro..."
    (procedure "::SOAP::Transport::http::xfer" line 61)
    invoked from within
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
    ("eval" body line 1)
    invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
    (procedure "DataAcquisitionControlClient__GetState" line 1)
    invoked from within
"DataAcquisitionControlClient__GetState"
STATE for nnaida5 returned with an error
connect failed 
NONE
connect failed 
    while executing
"::http::geturl http://nnaida5:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 20000
-query {<?xml version="1.0" encoding="UTF-8"..."
    ("eval" body line 1)
    invoked from within
"eval [list ::http::geturl $url] $args"
    (procedure "::http::geturl_followRedirects" line 4)
    invoked from within
"::http::geturl_followRedirects http://nnaida5:8015/DataAcquisitionControlServer -headers {} -type text/xml
-timeout 20000 -query {<?xml version="1.0" ..."
    ("eval" body line 1)
    invoked from within
"eval ::http::geturl_followRedirects [list $url]  -headers [list $local_headers]  -type text/xml  -timeout
$timeout  -query [list $request]  $local_pro..."
    (procedure "::SOAP::Transport::http::xfer" line 61)
    invoked from within
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
    ("eval" body line 1)
    invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
    (procedure "DataAcquisitionControlClient__GetState" line 1)
    invoked from within
"DataAcquisitionControlClient__GetState"
STATE for nnaida6 returned with an error
connect failed 
NONE
connect failed 
    while executing
"::http::geturl http://nnaida6:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 20000
-query {<?xml version="1.0" encoding="UTF-8"..."
    ("eval" body line 1)
    invoked from within
"eval [list ::http::geturl $url] $args"
    (procedure "::http::geturl_followRedirects" line 4)
    invoked from within
"::http::geturl_followRedirects http://nnaida6:8015/DataAcquisitionControlServer -headers {} -type text/xml
-timeout 20000 -query {<?xml version="1.0" ..."
    ("eval" body line 1)
    invoked from within
"eval ::http::geturl_followRedirects [list $url]  -headers [list $local_headers]  -type text/xml  -timeout
$timeout  -query [list $request]  $local_pro..."
    (procedure "::SOAP::Transport::http::xfer" line 61)
    invoked from within
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
    ("eval" body line 1)
    invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
    (procedure "DataAcquisitionControlClient__GetState" line 1)
    invoked from within
"DataAcquisitionControlClient__GetState"
STATE for nnaida7 returned with an error
connect failed 
NONE
connect failed 
    while executing
"::http::geturl http://nnaida7:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 20000
-query {<?xml version="1.0" encoding="UTF-8"..."
    ("eval" body line 1)
    invoked from within
"eval [list ::http::geturl $url] $args"
    (procedure "::http::geturl_followRedirects" line 4)
    invoked from within
"::http::geturl_followRedirects http://nnaida7:8015/DataAcquisitionControlServer -headers {} -type text/xml
-timeout 20000 -query {<?xml version="1.0" ..."
    ("eval" body line 1)
    invoked from within
"eval ::http::geturl_followRedirects [list $url]  -headers [list $local_headers]  -type text/xml  -timeout
$timeout  -query [list $request]  $local_pro..."
    (procedure "::SOAP::Transport::http::xfer" line 61)
    invoked from within
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
    ("eval" body line 1)
    invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
    (procedure "DataAcquisitionControlClient__GetState" line 1)
    invoked from within
"DataAcquisitionControlClient__GetState"
STATE for nnaida8 returned with an error
connect failed 
NONE
connect failed 
    while executing
"::http::geturl http://nnaida8:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 20000
-query {<?xml version="1.0" encoding="UTF-8"..."
    ("eval" body line 1)
    invoked from within
"eval [list ::http::geturl $url] $args"
    (procedure "::http::geturl_followRedirects" line 4)
    invoked from within
"::http::geturl_followRedirects http://nnaida8:8015/DataAcquisitionControlServer -headers {} -type text/xml
-timeout 20000 -query {<?xml version="1.0" ..."
    ("eval" body line 1)
    invoked from within
"eval ::http::geturl_followRedirects [list $url]  -headers [list $local_headers]  -type text/xml  -timeout
$timeout  -query [list $request]  $local_pro..."
    (procedure "::SOAP::Transport::http::xfer" line 61)
    invoked from within
"$transport $procVarName $url $req"
    (procedure "::SOAP::invoke" line 18)
    invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControl

10:26 Unable to recover from crash by restarting the merger
      FEEs turned off. Will leave off whilst changing the adapter boards

11.28 Power up all FEE64s - 20s between each relay 

11.46 Pi Monitor Browser reports all FEE64s 'Finished OK', all FEE64s '0 panics'

15:12 Waveforms of FEEs 5,6,9,10,17,18,13,14 showing the change in Ibias LF feedback value
      First attachment will be waveform with LF 0x8 and second will be 0xF
      FEE13 has only one as was on the rails at anything above 0x8 - Attachments 6-20

      Waveforms for FEEs 19 and 20 for comparisons - Attachments 21 & 22

      Attachment      FEE       Change
           6           5        default
           7           5        LF 0xf
           8           6        default
           9           6        LF 0xf
           10          9        default
           11          9        LF 0xf
           12          10       default
           13          10       LF 0xf
           14          17       default
           15          17       LF 0xf
           16          18       default
           17          18       LF 0xf
           18          13       default
           19          14       default
           20          14       LF 0xf
           21          19       default
           22          20       default

16:02 New ASIC settings folder 2017Oct22-15.59.25
      Settings changed related to A and C FEEs in which Ibias LF Feedback has been changed to LF0x8 for all
      with the exception of FEE 2 which is set to 0xC and 14 which is set to 0x8
      ASIC 1 of nnaida14 has a SC threshold of 0x20
      All of nnaida 13 has a SC threshold of 0xf

      Stats running in these setting - Attachment 23
      System wide checks ok
      FEE temperatures ok - Attachment 24
      Leakage and bias ok - Attachment 25

16.16 Start Oct17/R7

      Background alpha activity

      BNC PB-4 Pulser OFF

      ASIC settings per 2017Oct22-15.59.25
  694   Wed Nov 1 02:24:20 2017 OHWednesday 1 November
11:24 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1
      FEE64 Temperatures OK - see attachment 2
      DAQ statistics OK - see attachment 3
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
  696   Thu Nov 2 22:57:53 2017 OHFriday 3 November
10.56 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1
      FEE64 Temperatures OK - see attachment 2
      DAQ statistics OK - see attachment 3
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
  698   Sat Nov 4 23:49:14 2017 OHSunday 5 November
08.09 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1
      FEE64 Temperatures OK - see attachment 2
      DAQ statistics OK - see attachment 3
      *all* system wide checks OK *except* nnaida21 fails ADC calibration

13:35 DAQ running ok
      Fee temperatures - attachment 4
      Stats - attachment 5
      Leakage currents and bias ok - Attachment 6

13:39 Bias taken off detectors
      Server windows closed and servers halted
      FEE power relays turned off
      Chiller remains on
  726   Tue Jun 26 05:44:44 2018 OHTuesday 26 June
04:12 TD found the nnaida14 appeared to have become noisy at some point. Biases/;eakages/temps ok
      ASIC check load appears to have fixed problem
      Noise appears to affect R9_0 to R9_51

13:45 Biases and leakages all ok - attachment 1
      FEE temperatures ok
      System wide checks ok
      T = 25.1C H = 35.7% DP 8.4C
      Currently up to file R9_54

21:53 Biases and leakages all ok - attachment 2
      FEE Temps ok
      System wide checks ok
      Currently up to file R9_56
  727   Wed Jun 27 02:23:03 2018 OHWednesday 26 June
10:20 Bias and leakage currents ok - attachment 1
      FEE temperatures ok
      System wide checks all ok
      Currently on file R9_58

15:25 Bias and leakage currents ok - attachment 2
      FEE temperatures ok - attachment 3
      System wide checks ok
      Currently on file R9_59
  729   Thu Jun 28 02:19:30 2018 OHThursday 28th June
10:18 Bias and leakage currents ok - attachment 1
      FEE Temperatures ok
      System wide checks all ok
      Currently on file R9_63
      Humidity 43.3%, Temperature 25.1C, Dew Point +11.8C
      Filler level ok

12:51 Bias and leakage currents ok - attachment 2
      FEE Temperatures ok
      System wide checks all ok
      Currently on file R9_64

19:20 Bias and leakage currents - attachment 3
      FEE temperatures ok
      System wide checks all ok
      Currently on file R9_65
      ASIC 2 is still missing from FEE29. ASIC checks earlier in the afternoon have failed to recover it. -
attachment 4
      This ASIC has been observed to be working since it was installed last week. Once the alpha run is over
further attempts
      will be made to test it is back.
  730   Thu Jun 28 05:22:31 2018 OHAnalysis of R7 -> R9
R7_0 -> R7_1 High Energy pulser run
R8_0 -> R8_1 Low energy puler run
R9_52 -> R9_62 Alpha run

On analysing all three runs it appears an ASIC is missing in FEE 29 - attachments 1 - 3

This ASIC does not appear to be missing in the alpha run carried out on June 24th - attachment 4

I do not believe that this is a broken ASIC.

I have performed an ASIC check load on the FEE module to try and regain it. I will check the files created since
later to see if it has been regained.

Using the pulser runs two different sets of offsets were calculated. One using the low energy run and one using
the high energy.
To compare the two different sets of offsets calculated Ex vs Ey plots were created for the first detector.

Attachment 5 shows these plots on the left is the low energy offsets and on the right the high energy pulser offsets
It can be seen that the grouping is much better with the events being tighter to the y = x line on the run sorted
using the low energy pulser offsets.
  731   Fri Jun 29 02:26:34 2018 OHFriday 29 June
10:27 Bias and leakage currents ok - attachment 1
      Fee temperatures ok - attachment 2
      Temperature 24.7C, Humdidity 44.5%, D.P +11.8C
      System wide checks all ok
      Currently on file R9_69

      Vandle guys are about to start building. They have been made aware to avoid the power cables for the
chiller and AIDA
      Also of the network cables and HDMI cables.

14:46 Bias and leakage currents ok - attachment 3
      System wide checks all ok
      FEE temperatures all ok
      Currently on R9_70

19:55 Bias and leakage currents ok - attachment 4
      Fee temperatures ok
      System wide checks ok
      
  732   Sat Jun 30 03:05:36 2018 OHSaturday 30 June
11:05 Bias and leakage currents ok - attachment 1
      FEE temperatures ok
      System wide checks ok
      Currently on file R9_74
      Temperature 25.3C, Humidity 42.9%, D.P +12.1C

13:27 Bias and leakage currents ok - attachment 2
      FEE Temperatures ok
      System wide checks ok

      Analysis of the pulser run R9_52 -> R9_74 has shown no major losses in terms of apparent broken bond wires.
      The large gap in FEE64 29 can be explained by an ASIC dropping out: See https://elog.ph.ed.ac.uk/AIDA/730
      Another odd trend appears in the channels relating to DSSD 8 on the x side. - attachment 5
      The pattern shows one strip on one strip off for the first 32 strips.
      nnaida18 is also the FEE which is running at a much higher rate during this run. - attachment 3
      I'm not sure that I believe that these channels are actually relate to broken bond wires.

      Here are the hit patterns for the 8 DSSD - attachment 4
      Attached is also the x and y histograms for DSSD 8 which shows a weird pattern - attachment 5
  733   Sun Jul 1 02:26:50 2018 OHSunday 1st July
10:25 Bias and leakage currents ok - attachment 1
      FEE Temperatures ok
      System wide checks ok
      Alpha run R9 stopped on file R9_74
  734   Sun Jul 1 08:25:24 2018 OHInventory
Attached is an updated inventory of the cables and adapter boards we have in RIKEN.
  735   Mon Jul 2 02:36:41 2018 OHMonday 2 July
10:35 Bias and leakage currents ok - attachment 1
      FEE Temps ok - attachment 2
      System wide checks all ok
      
  736   Tue Jul 3 02:24:45 2018 OHTuesday 3 July
10:23 Bias and leakage currents ok - attachment 1
      Fee temperatures ok - attachment 2
      System wide checks ok

14:07 Bias and leakage currents ok - attachment 3
      FEE temperatures ok
      System wide checks ok

      Have been unable to recover ASIC 2 in nnaida29. Have checked removing the detector cabling and can still
not recover the ASIC
      Also noise in DSSD1 and 2 is still high - attachment 4
      However in DSSD3-8 it is very good

      Humidity 44.7%, Temperature 25.5C, D.P +13C

15:57 FEE modules turned off via rly16
      DSSD bias removed

      AIDA has been put in a safe state.
  749   Wed Oct 3 10:47:06 2018 OHAnalysis of R2_0 -> R2_37 Alpha Run
Data was analysed using offsets calculated based on pulser walkthroughs carried out in June 2018
Offsets appear to still be good. Ex V Ey plots - attachment 1

Hit pattern across the FEEs looks good - attachment 2
x axis is (FEE-1)*64 + Channel

No obvious dead areas in x vs y plots either - attachment 3