401
|
Mon Oct 24 02:46:57 2016 |
TD | Monday 24 October | 10.47 DAQ found stopped, no merger error messages
Most recent file R78_281, last accessed 06.46
R78_281 ADC data synchronised
DSSSD bias & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
System-wide checks - see attachments 4-11
Restart merger OK
DAQ STOP OK
DAQ RESET & SETUP OK
ReSYNC OK
System-wide checks OK
Disable H & data transfer #1
Restart merger & tapeserver
Enable H & data transfer #1
11.07 DAQ starts (R79)
18.39 DAQ OK
System wide checks OK *except* SYNC errors & SYNC pulses - see attachments 12-15
18.47 DAQ stalls
Most recent file R79_182, last accessed 18.46
System wide checks OK *except* SYNC errors & SYNC pulses - see attachments 16-19
18.56 DAQ starts (R80) |
|
877
|
Mon Jul 24 08:54:48 2023 |
TD | Monday 24 July |
09.54 DSSSD#1 bias +100V leakage current +9.065uA
ADC data item stats
per FEE64 1.8.W spectra 20us & 200us FSR - see attachments 1-5
10.02 aida04 asic#1 u/s - see attachments 6-7
10.39 AIDA PSU#1 FEE64s 1,2 3, 4, 5, 6, 7, 8 -> 1, 3, 2, 4, 5, 6, 7, 8
ADC data item stats ( slow comparator 0x64 & 0x14 )
per FEE64 1.8.W spectra 20us & 200us FSR - see attachments 8-12
11.08 Test + connected aida09, 01, 10, 12, 3 & 11
SSSD#1 bias +100V leakage current +12.150uA
ADC data item stats ( slow comparator 0x14 )
per FEE64 rate, 1.8.L & 1.8.W spectra 20us FSR - see attachments 13-18
aida01 pulser peak width 134 ch FWHM |
|
680
|
Mon Oct 23 02:24:14 2017 |
TD, OH | Monday 23 October | 10:18 Leakage and bias ok - Attachment 1
FEE temperatures ok - Attachment 2
Statistics - Attachment 3
DAQ continues ok
10:23 DAQ stopped on R7_10
Alpha run through night
10:28 MIDAS Sort spectra from the alpha run
Ep versus En - Attachment 4
x versus y 2D hit pattern - Attachment 5
Front versus back multiplicity - Attachment 6
2D Hit pattern indicates detection of alphas in all 4 regions of each of the 6 detectors
indicates cabling is ok
11:55 FEEs turned off after being unable to recover after crash
13:49 FEEs powered on again
Pi monitor indicates panic on nnaida22 the FEE that crashed
ReSync done on setup except for 23 and 24
Master ReSync done succesfully
ASIC settings loaded default to 2017Mar27
Restoring 2017Oct22
System wide checks ok
Leakage and bias ok - Attachment 7
FEE temperatures crashed - Attachment 8
Upon reboot all rates across the FEEs appear lower
In particular 13 and 14. Thresholds have been reduced back to 0xa for all ASICs on both FEEs
Layout 1 linear common y of 2000 - Attachment 9
Layout 1 log common y of 10 - Attachment 10
All front side FEE64 1.8.W waveforms - 20us range - Attachment 11
All back side FEE64 1.8.W waveforms - 20us range - Attachment 12
All front side FEE64 1.8.W waveforms - 2ms range - Attachment 13
All back side FEE64 1.8.W waveforms - 2m range - Attachment 14
FEE64 nnaida13 1.*.W waveforms - Ibias LF 0xf - 2ms range - Attachment 15
Statistics after switch and thresholds lowered - Attachment 16
Pulser widths measured. On boards changed 10 out of 12 FEEs show greater than 10% improvements
11 out of 24 total show greater than 10%
FEE stats post pulser width - Attachment 17
New ASIC settings DB key: 2017Oct23-14.38.03
Slow comparator 0xa for all FEE64s and all ASICs
IBias LF feedback at 0xf for all front side FEEs except nnaida 2 which is at 0xc
IBias LF feedback at 0xb for all back side FEEs
ASIC settings DB file checked OK
Ambient temperature +23.3 deg C, D.P +7.1 deg C, R.H. 35.2%
Julabo FL11006 set point +20.0 deg C, water le
15:38 Merger and TapeServer screenshots - Attachments 17-18
|
|
45
|
Mon Feb 23 07:03:10 2015 |
TD, AE & CG | Monday 23 February 2015 | 14.30 Saved nnaida1-16 ASIC + Waveform configurations to directory (Data Base Key)
2015Feb23-12.11.40
(directory aidas1:/MIDAS/DB/EXPERIMENTS/AIDA/2015Feb23-12.11.40)
nnaida 1-2, 5-6, 9-10, 13-14 positive input ( front p+n junction strips )
nnaida 3-4, 7-8, 11-12, 15-16 negative input ( back n+n ohmic strips )
Standard ASIC configurations except
- shaping time 0xb (6,5us)
- peaking time 0x8 (16us)
- slow comparator 0x10 (~160keV)
- fast comparator (LEC/MEC) 0xff
- fast comparator (HEC) 0xff
nnaida 1-2, 5-6, 9-10, 13-14 negative input ( AIDA ASIC preamp output inverts input,
i.e. preamp with positive input produces negative input for sampling ADC )
nnaida 3-4, 7-8, 11-12, 15-16 positive input
Standard Waveform configurations except
- threshold 1000
- pre-trigger 0x100
- all channels disabled
2015Feb23-12.11.40 is now the default configuration file
Note: The directory (Data Base Key) which will contain the saved settings is created uid: npg, gid: npgstaff, perms 755.
However, its contents are written uid: nfsnobody, gid nfsnobody, perms 777. The default directory (Data Base Key) permissions
do not work - it is necessary to change the directory (Data Base Key) to perms 777 after its creation and before saving the
individual FEE64 configuration directories/files to it.
|
|
582
|
Mon May 22 02:34:51 2017 |
CB, TD | Monday 22 May | 10.35 DAQ continues OK
Current file May2017/R19_253
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0xff
Detector biases; leakage currents OK - see attachment 1
FEE64 Temperatures OK - see attachment 2
good events; statistics OK - see attachment 3
No warnings/errors reported by merger since DAQ GO - see attachment 4
*all* system wide checks OK *except* ADC calibration of nnaida7
10.38 E17/F11 ambient temperature +24.7 deg C, RH 40.7%, d.p. +10.7 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)
10.39 DAQ stops OK
Current file May2017/R19_253
Deleted R19*
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0xff
BNC PB-4 settings
Amplitude 90,000
x5 attenuator IN
t_r 50ns, tau_d 50us
polarity -
rate 250Hz range
delay MIN, tail pulse
Check ASIC Control x2
Zero all spectra
10.47 DAQ starts
File May2017/R20
Ras Pi system console logs for USB0 attached - see attachment 5
Analysed R20_0. All appears to be working fine. Highest dead time ~0.3% - see attachment 6
Off-line analysis of R20_0 confirms ADC data and discriminator are synchronised - see attachments 7,8,9
16.55 DAQ continues OK
Current file May2017/R20_71
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0xff
Detector biases; leakage currents OK - see attachment 10
FEE64 Temperatures OK - see attachment 12
good events; statistics OK - see attachment 13
No warnings/errors reported by merger since DAQ GO - see attachment 14
*all* system wide checks OK *except* ADC calibration of nnaida7
17.02 E17/F11 ambient temperature +24.0 deg C, RH 40.3%, d.p. +9.7 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)
17.03 DAQ stops OK
Current file May2017/R20_73
Deleted R20*
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0xff
BNC PB-4 settings
Amplitude 90,000
x5 attenuator IN
t_r 50ns, tau_d 50us
polarity -
rate 250Hz range
delay MIN, tail pulse
Check ASIC Control x2
Zero all spectra
17.08 DAQ starts
File May2017/R21
Raspberry Pi system console logs for USB0 attached - see attachment 15
17.50 Changed permissions of options files /MIDAS/DB/EXPERIMENTS/Options/nnaida<n>/CONTENTS
from mode 644 to mode 777
18.09 DAQ stops OK
Current file May2017/R21_11
Deleted R21*
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0xff
BNC PB-4 settings
Amplitude 90,000
x5 attenuator IN
t_r 50ns, tau_d 50us
polarity -
rate 250Hz range
delay MIN, tail pulse
DAQ RESET
DAQ SETUP
*all* system wide checks OK *except* nnaida7, 9, 13 and 22 ADC calibration failed
Check ASIC Control x2
17.08 DAQ starts
File May2017/R22
Raspberry Pi system console logs for USB0 attached - see attachment 16 |
|
584
|
Mon May 22 08:43:10 2017 |
CB, TD, Patrick | Monday 22 May | > 10.35 DAQ continues OK
> Current file May2017/R19_253
>
> ASIC settings 2017Mar27-16.12.39 *except*
> slow comparator 0x0a
> LEC/MEC fast comparator 0xff
>
> Detector biases; leakage currents OK - see attachment 1
> FEE64 Temperatures OK - see attachment 2
> good events; statistics OK - see attachment 3
> No warnings/errors reported by merger since DAQ GO - see attachment 4
>
> *all* system wide checks OK *except* ADC calibration of nnaida7
>
> 10.38 E17/F11 ambient temperature +24.7 deg C, RH 40.7%, d.p. +10.7 deg C
> Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)
>
> 10.39 DAQ stops OK
> Current file May2017/R19_253
> Deleted R19*
>
> ASIC settings 2017Mar27-16.12.39 *except*
> slow comparator 0x0a
> LEC/MEC fast comparator 0xff
>
> BNC PB-4 settings
> Amplitude 90,000
> x5 attenuator IN
> t_r 50ns, tau_d 50us
> polarity -
> rate 250Hz range
> delay MIN, tail pulse
>
> Check ASIC Control x2
> Zero all spectra
>
> 10.47 DAQ starts
> File May2017/R20
> Ras Pi system console logs for USB0 attached - see attachment 5
> Analysed R20_0. All appears to be working fine. Highest dead time ~0.3% - see attachment 6
> Off-line analysis of R20_0 confirms ADC data and discriminator are synchronised - see attachments 7,8,9
Concerning nnaida7 and the failure of calibration..... have you tried to calibrate it using the FADC Align & Control ?
If it still fails then there is a question about the quality of the clock signal being received by this module.
I will log through and look at the console log for any anomolies but please try and calibrate.
Let me know if you wish me to take remote control. |
|
585
|
Mon May 22 08:53:42 2017 |
CB, TD, Patrick | Monday 22 May | > Concerning nnaida7 and the failure of calibration..... have you tried to calibrate it using the FADC Align & Control ?
> If it still fails then there is a question about the quality of the clock signal being received by this module.
> I will log through and look at the console log for any anomolies but please try and calibrate.
> Let me know if you wish me to take remote control.
Yes - there have been multiple attempts to re-calibrate nnaida7 ADCs
Note *all* ADCs were calibrated before the most recent FEE64 power cycle.
It is not unusual for ~1-3 FEE64s to fail ADC calibration tests following
a power cycle - there is no obvious pattern to which ADCs fail or which (if
any) will be successfully re-calibrated. |
|
461
|
Mon Nov 21 02:12:45 2016 |
CG & TD | Monday 21 November |
11.10 DSSSD bias & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
Rate, HitRate, per FEE64 1.8.L and 1.8.W spectra - see attachments 4-9
16.10 DSSD bias and leakage currents ok (attachments 10+11).
FEE temps ok. |
|
706
|
Mon Nov 20 00:53:27 2017 |
TD, CB | Monday 20 November | 09.50 Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
DAQ statistics OK - see attachment 4
7x FEE64s <10k
4x FEE64s 10-20k
5x FEE64s 20-30k
3x FEE64s 30-40k
1x FEE64s 40-50k
4x FEE64 >50k (nnaida13 255k)
PLL & DCM Clock Lock Monitor OK - see attachment 5 & https://elog.ph.ed.ac.uk/AIDA/704 attachment 14
*all* system wide checks OK *except* nnaida7 fails ADC calibration
16.39 Detector biases & leakage currents OK - see attachments 5 & 6
FEE64 Temperatures OK - see attachment 7
DAQ statistics OK - see attachment 8
7x FEE64s <10k
4x FEE64s 10-20k
5x FEE64s 20-30k
3x FEE64s 30-40k
1x FEE64s 40-50k
4x FEE64 >50k (nnaida13 255k)
*all* system wide checks OK *except* nnaida7 fails ADC calibration |
|
807
|
Mon May 20 02:16:58 2019 |
CA, TD | Monday 20 May 2019 | 10:18 DAQ Continues ok
Alpha run continues - Currently on file R8_2
System wide checks all ok
Detector bias/leakage currents ok - attachment 1
Temp +25.6 deg C, Humidity 34.5% AND D.P +8.8 deg C
FEE Temperatures ok - attachment 2
Good event statistics ok - attachment 3
Merger & TapeServer ok - attachments 4 & 5
17.15 DAQ continues OK to file May19/R8_2
alpha background
ASIC settings 2019May19-17.54.03
slow comparator 0x64
BNC PB-4 Pulser OFF
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 6 & 7
FEE64 temperatures OK - see attachment 8
good event statistics OK - see attachments 9
Merger OK - see attachment 10
TapeServer OK - see attachment 11
1.8.L spectra - see attachments 12 & 13
p+n FEE64s 1.8.W spectra 20us FSR - attachment 14
n+n FEE64s 1.8.W spectra 20us FSR - attachment 15
17.25 analysis of May19/R8_2 - see attachment 16
deadtime OK
ADC data for FEE64s nnaida5 & nnaida6 hot channels (Stat spectrum channel # 104)
Also observed at a much lower rate for nnaida1 & nnaida2
17.27 Check ASIC Control *all* ASICs & FEE64s
13.27 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 60% M-H markers - water topped up (1x jugful) to H level marker
Temperature +25.4 deg C dew point +9.6 deg C RH 36.3% |
|
342
|
Mon Aug 22 10:35:31 2016 |
TD | Monday 20 August | 11.00 Restore power to AIDA following Saturday's shutdoen in preparation for the planned maintenance
outage on Sunday.
Restart Julabo FL11006 chiller - set point +20 deg C
Restart AIDA
Temperature +24.3 deg C, DP +15.2 deg C, RH 57.5%
DSSSD #1-6 bias - see attachments 1-2
FEE64 temperatures - attachment 3
System-wide checks - see attachments 4-8
Note SYNC pulses received by *all* slave FEE64s = SYNC pulses sent by master FEE64 (nnaida6)
11.35 R6 starts
slow comparator 0x10
LEC/MEC fast comparator 0x10
All FEE64 discriminator masks -> 0x0
observe very high rates (~400-500kHz/FEE64) from majority ofp+n junction FEE64s
LEC/MEC fast comparator -> 0xff
18.45 DAQ running OK
No merger error messages since change of LEC/MEC fast comparator threshold -> 0xff
Offline analysis of R6_79 confirms that all FEE64s are synchronised
Added correlation trigger input
BNC PB-4 Pulser trigger output -> Ortec SCA + NIM output width 0.5us -> LeCroy 429A -> root MACB trigger
#4 input
21.05 R7 starts
Attachments 9-16 show FEE64 temperatures and results of each system-wide check.
Note the result of the Master Timestamp Control -> Check Clock Lock States
indicating clock state *not* locked
slow comparator 0x10
LEC/MEC fast comparator 0xff
All FEE64 discriminator masks -> 0x0 |
|
344
|
Mon Aug 22 13:37:30 2016 |
TD | Monday 20 August | > 11.00 Restore power to AIDA following Saturday's shutdoen in preparation for the planned maintenance
> outage on Sunday.
>
> Restart Julabo FL11006 chiller - set point +20 deg C
> Restart AIDA
>
> Temperature +24.3 deg C, DP +15.2 deg C, RH 57.5%
>
> DSSSD #1-6 bias - see attachments 1-2
>
> FEE64 temperatures - attachment 3
>
> System-wide checks - see attachments 4-8
> Note SYNC pulses received by *all* slave FEE64s = SYNC pulses sent by master FEE64 (nnaida6)
>
> 11.35 R6 starts
> slow comparator 0x10
> LEC/MEC fast comparator 0x10
>
> All FEE64 discriminator masks -> 0x0
>
> observe very high rates (~400-500kHz/FEE64) from majority ofp+n junction FEE64s
>
> LEC/MEC fast comparator -> 0xff
>
> 18.45 DAQ running OK
> No merger error messages since change of LEC/MEC fast comparator threshold -> 0xff
> Offline analysis of R6_79 confirms that all FEE64s are synchronised
>
> Added correlation trigger input
> BNC PB-4 Pulser trigger output -> Ortec SCA + NIM output width 0.5us -> LeCroy 429A -> root MACB trigger
> #4 input
>
> 21.05 R7 starts
>
> Attachments 9-16 show FEE64 temperatures and results of each system-wide check.
>
> Note the result of the Master Timestamp Control -> Check Clock Lock States
> indicating clock state *not* locked
>
> slow comparator 0x10
> LEC/MEC fast comparator 0xff
> All FEE64 discriminator masks -> 0x0
The status is OK. I changed the system wide check to reflect the new value but forgot about the individual check.
I will update the software.
Note attachment 5 shows all is well.
Updated version of /MIDAS/TclHttpd/Html/AIDA/MasterTS/MasterTS.tcl installed with correction. |
|
896
|
Mon Oct 2 14:03:42 2023 |
TD | Monday 2 October |
15.00 DSSSD bias +60V +6.44uA - breakdown above c. 60V, bias +100V leakage current +20-30uA
Slow comparator 0xa
aida03 LK3 fitted
CAEN N1419ET grounded outputs
HV#1 core aida04. braid aida10, aida1 & aida 9
BNC PB-5
Amplitide 10.0V
Attenuation x10
Decay time 1ms
Polarity +
Tail
Frequency 25Hz
test +/ test - daisy chains connected
test - disconnected at EG&G Ortec 534 output |
|
197
|
Mon May 2 12:13:25 2016 |
CG,PV,TD | Monday 2 May 2016 | |
|
735
|
Mon Jul 2 02:36:41 2018 |
OH | Monday 2 July | 10:35 Bias and leakage currents ok - attachment 1
FEE Temps ok - attachment 2
System wide checks all ok
|
|
390
|
Mon Oct 17 04:10:30 2016 |
TD | Monday 17 October | 12.05 DAQ found stopped
Most recent file R49_226, last accessed 06.26
Restart merger OK
DAQ STOP OK
System-wide checks - see attachments 1-2
SYNC errors and SYNC pulses fail
ReSYNC OK
System-wide checks OK
Restart merger & tapeserver
DSSSD bias & leakage currents OK - see attachments 3 & 4
FEE64 temperatures OK - see attachment 5
DAQ start
#1 nnaida15 data transfer #1 tag drops at DAQ GO
disable H & data transfer #1
restart merger & tapeserver
enable H & data transfer #1
12.22 #2 DAQ start (R51) OK
20.33 DAQ found stopped
Most recent file R51_145, last accessed 18.30
Restart merger OK
DAQ STOP OK
System-wide checks - see attachments 6 & 7
SYNC errors and SYNC pulses fail
DAQ RESET & SETUP OK
Disable H & data transfer #1
Restart merger & tapeserver
Enable H & data transfer #1
20.41 DAQ GO (R52) |
|
847
|
Sun Jun 16 16:00:58 2019 |
ML, CG, OH | Monday 17 June 2019 | It's a new day, it's a new log entry !!
01:25 DTAS run 46 started
BigRIPS run 2019
AIDA run R2_268
DAQ running smoothly
No change in Stat and Rate (stat zeroed at the beginning of this DTAS run)- See attachment 1
Merger rate: 630k item/sec - See Attachment 2
TapeServer rate still ~5 Mbytes/sec - See attachment 3
Temperature and HV ok - See attachment 5 & 6
System-wide checks OK except nnaida7 fails ADC calibration as usual.
!!!
Victor said that the run AIDA R2_249 is missing in the correlation merge data. This is the run when the
correlation scaler was reset (as DTAS was running out of range in some of their timing histograms) (See
elog end of Sunday)
!!!
02:23 Rate in DSSD#3 card 9 and 12 slowly reaching 100kHz while 14 and 15 ~ 50kHz. See attachment 7
TapeServer rate ~ 5.5 kHz
03:25 Change of DTAS run:
DTAS run 49 started
BigRIPS run 2020
AIDA run R2_286
Experiment still running smoothly. No significant change since last comments 2h ago.
Leakage current in detector 3 is increasing slowly see attachment 8 & 9
05:25 Change of DTAS run:
Note: DTAS run stopped during AIDA R2_305 and re-started during R2_306.
DTAS run 52 started
BigRIPS run 2021
AIDA run R2_306
Pretty quiet night shift.
Stat & Rate in attachment 10 - DSSD#3 FEE card 9 back down to 60kHz
Merger rate: 650k item/sec - See attachment 11
TapeServer rate still ~5.1 Mbytes/sec - See attachment 12
System-wide checks OK except nnaida7 fails ADC calibration as usual.
X vs Y HEC spectrum in attachment 13
LEC E vs HEC-LEC spectrum in attachment 14
Both spectra zeroed
Bias in attachment 15
Temperature unchanged
07:30 DTAS and BigRIPS stop runs
AIDA on file R2_323
07:33 New Run numbers
DTAS Run: 55
BigRIPS Run: 2022
AIDA on File: R2_324
07:56 System wide checks ok
Fee temps ok - attachment 16
100Sn setting, no degrader
ASIC settings 2019Jun13-08.56.35
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range, variable min. c. 2.5Hz
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
08:28 Bias and leakage ok - attachment 17
08:48 ASIC check successful
09:32 DTAS and BigRIPS stop
AIDA on file R2_342
DTAS starts file 58
BigRIPS run 2023
AIDA still on file R2_342 |
|
848
|
Mon Jun 17 02:31:53 2019 |
CA, TD, OH | Monday 17 June 08.00-16.00 | 10.41 DAQ continues OK file RIBF147/R2_352
100Sn setting, no degrader
ASIC settings 2019Jun13-08.56.35
DSSSD#1 slow comparator 0xa -> 0x1d
DSSSD#3 slow comparator 0xa -> 0x8
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range, variable min. c. 2.5Hz
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics - OK - see attachment 4
11x <20k 21x < 50k 24x <100k
Merger OK - see attachment 5
TapeServer OK - see attachment 6
p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
1.8.L spectra - see attachments 9 & 10
nnaida1 73 ch FWHM, nnaida3 78 ch FWHM
FEE64 Rate spectra - attachment 11
Note high rate issues observed for *some* strips in nnaida14 ASIC #3 & #4 fixed (temporarily) by Check
ASIC Control
FEE64 Rate spectra - attachment 12
10.58 ADC data synchronised - attachment 13
Online MIDASsort spectra - attachments 14-18
11.04 analysis of file RIBF147/R2_353 - attachment 19
deadtime <<1% *all* FEE64s
no timewarps
11.12 HitRate spectra - attachment 20
disable nnaida14 *hot* channels 29, 30 & 35
11.15 zero all histograms
zero DAQ stats
12.31 DTAS and BigRIPS stop
AIDA on file R2_368
DTAS starts file 63
BigRIPS run 2025
AIDA file R2_368
13:08 R2_374
The putty window for USB2 had frozen with a 1 in the voltage set as shown in attachment 1
The window was no longer updating.
When regaining the window the voltage limit was applied and the detector
began to ramp down
The set voltage was reset to 100V and bias was regained.
Total time with less than 100V was around 30sec
14:28 DTAS and BigRIPS stop
AIDA on run R2_386
DTAS Starts file 66
BigRIPS run 2026
AIDA on file R2_387
14:32 Bias and leakage currents ok - attachment 21
15:07 Rate spectra showing no spikes in DSSD 3
System wide checks
15:08 DTAS and BigRIPS stop
AIDA onfile R2_393
BigRIPS chaning slit positions
14:14 DTAS starts file:
AIDA on file R2_394
BigRIPS file 2027 (In control by bigRIPS)
15.30 TAS and BigRIPS stop
AIDA onfile R2_396 |
|
33
|
Mon Feb 16 06:17:31 2015 |
TD, AEV & CG | Monday 16th Feb progress | 16/02/2015
1500: AIDA moved into operational position.
Water cooling circuit visually checked, fittings, etc for loose fittings/snags/kinks. No problems
Julabo FL11006 chiller moved into position with mains transformer, filled and turned on without problem (small leak from rear overflow, due to overfilling).
Took roughly 2 ~30L water containers to fill reservoir.
Brief leak at connection between back of chiller and yellow hose, bolt tightened and it stopped.
Having been turned on and loop filled, reservoir at ~50% full. Pressure ~10 psi.
No leaks observed in main coolant loop.
When standing downstream of AIDA RH flow gauge blurred due to high rotational speed, LH flow gauge rotating at ~1 rev/s.
Two FEE modules closest to centre in each block checked for power cabling, TS distribution and network cables. All ok.
MACBs front panel cabling checked, no obvious problems. TS distribution tree in order.
Power on for NIM bins and fan tray, network switches and Raspberry Pi. Not yet for the FEE module PSUs.
1 corner of each new USB power supply damaged in transit.
Fuses changed from 10A -> 16A in each PSU. Old 10A fuses taped to back of each module.
1730: Turned on nnaida1-16. Temperatures stable. All on firmware version 0xa4ed006.
Input pulser into all FEE cards. All FEEs bar nnaida5 well-behaving. nnaida12 has several missing channels.
Typical pulser peak width for -ve input modules was ~14ch FWHM with shaping time of 1us.
Typical pulser peak width for +ve input modules was ~27ch FWHM with shaping time of 1us.
Changing shaping time of nnaida10 from 1->2.5us reduced FWHM from 27ch to 23ch.
Attachment 1: temperatures & FEE firmware revisions nnaida1-16
|
|
567
|
Mon May 15 02:01:02 2017 |
TD, OH | Monday 15 May | 10.01 DAQ continues OK
Current file May2017/R8_334
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0xff
Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
'good events' statistics OK - see attachment 4
No warnings/errors reported by merger since DAQ GO - see attachment 5
*all* system wide checks OK
10.06 Analysis of data file R8_333 - see attachment 6
dead-time/idle-time < 2%
AIDA SYNC100 variations ~1800/95000
10.10 E17/F11 ambient temperature +24.8 deg C, RH 36.4%, d.p. +8.9 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%)
10.15 DAQ stops OK
Current file May2017/R7_298
Files May2017/R8* *deleted*
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator *0x08*
LEC/MEC fast comparator 0xff
BNC PB-4 settings
Amplitude 90,000
x5 attenuator IN
t_r 50ns, tau_d 50us
polarity -
rate 250Hz range
delay MIN, tail pulse
Check ASIC Control x2
Zero all spectra
10.20 DAQ starts
File May2017/R8
Transient timeouts at DAQ GO for FEE64s nnaida1 and nnaida2 which resolved themselves without user
intervention
DAQ run control transitioned from 'RESET and STOP/GO' to 'RESET' to 'RESET and STOP/GO'
10.22 DAQ stops
File May2017/R9
Tapeserver data rates fluctuate 10-30Mb/s possibly due to parallel disk I/O activity (deleting R8 files)
Transient timeouts at DAQ STOP for FEE64s nnaida1 and nnaida2 which resolved themselves without user
intervention - see attachment 7
DAQ run control transitioned from 'RESET and STOP/GO' to 'RESET' to 'RESET and STOP/GO'
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator *0x0a*
LEC/MEC fast comparator 0xff
10.31 DAQ starts
File May2017/R10 |
|