ID Date Author Subjectdown
  401   Mon Oct 24 02:46:57 2016 TDMonday 24 October
10.47 DAQ found stopped, no merger error messages
      Most recent file R78_281, last accessed 06.46

      R78_281 ADC data synchronised

      DSSSD bias & leakage currents OK - see attachments 1 & 2
      FEE64 temperatures OK - see attachment 3
      System-wide checks - see attachments 4-11 

      Restart merger OK
      DAQ STOP OK

      DAQ RESET & SETUP OK
      ReSYNC OK
      System-wide checks OK
      Disable H & data transfer #1
      Restart merger & tapeserver
      Enable H & data transfer #1

11.07 DAQ starts (R79)

18.39 DAQ OK

      System wide checks OK *except* SYNC errors & SYNC pulses - see attachments 12-15

18.47 DAQ stalls
      Most recent file R79_182, last accessed 18.46
      System wide checks OK *except* SYNC errors & SYNC pulses - see attachments 16-19

18.56 DAQ starts (R80)
  877   Mon Jul 24 08:54:48 2023 TDMonday 24 July
09.54 DSSSD#1 bias +100V leakage current +9.065uA

      ADC data item stats
      per FEE64 1.8.W spectra 20us & 200us FSR - see attachments 1-5


10.02 aida04 asic#1 u/s - see attachments 6-7

10.39 AIDA PSU#1 FEE64s 1,2 3, 4, 5, 6, 7, 8 -> 1, 3, 2, 4, 5, 6, 7, 8

      
      ADC data item stats ( slow comparator 0x64 & 0x14 )
      per FEE64 1.8.W spectra 20us & 200us FSR - see attachments 8-12

11.08 Test + connected aida09, 01, 10, 12, 3 & 11
      SSSD#1 bias +100V leakage current +12.150uA

      ADC data item stats ( slow comparator 0x14 )
      per FEE64 rate, 1.8.L & 1.8.W spectra 20us FSR - see attachments 13-18
      aida01 pulser peak width 134 ch FWHM
  680   Mon Oct 23 02:24:14 2017 TD, OHMonday 23 October
10:18 Leakage and bias ok - Attachment 1
      FEE temperatures ok - Attachment 2
      Statistics - Attachment 3
      DAQ continues ok

10:23 DAQ stopped on R7_10
      Alpha run through night

10:28 MIDAS Sort spectra from the alpha run
      Ep versus En - Attachment 4
      x versus y 2D hit pattern - Attachment 5
      Front versus back multiplicity - Attachment 6

      2D Hit pattern indicates detection of alphas in all 4 regions of each of the 6 detectors
      indicates cabling is ok

11:55 FEEs turned off after being unable to recover after crash

13:49 FEEs powered on again
      Pi monitor indicates panic on nnaida22 the FEE that crashed

      ReSync done on setup except for 23 and 24
      Master ReSync done succesfully
      ASIC settings loaded default to 2017Mar27
      Restoring 2017Oct22
      
      System wide checks ok
      Leakage and bias ok - Attachment 7
      FEE temperatures crashed - Attachment 8

      Upon reboot all rates across the FEEs appear lower
      In particular 13 and 14. Thresholds have been reduced back to 0xa for all ASICs on both FEEs

      Layout 1 linear common y of 2000 - Attachment 9
      Layout 1 log common y of 10 - Attachment 10

      All front side FEE64 1.8.W waveforms - 20us range - Attachment 11
      All back side FEE64 1.8.W waveforms - 20us range - Attachment 12

      All front side FEE64 1.8.W waveforms - 2ms range - Attachment 13
      All back side FEE64 1.8.W waveforms - 2m range - Attachment 14

      FEE64 nnaida13 1.*.W waveforms - Ibias LF 0xf - 2ms range - Attachment 15

      Statistics after switch and thresholds lowered - Attachment 16
      Pulser widths measured. On boards changed 10 out of 12 FEEs show greater than 10% improvements
      11 out of 24 total show greater than 10%
      FEE stats post pulser width - Attachment 17

      New ASIC settings DB key: 2017Oct23-14.38.03
      Slow comparator 0xa for all FEE64s and all ASICs
      IBias LF feedback at 0xf for all front side FEEs except nnaida 2 which is at 0xc
      IBias LF feedback at 0xb for all back side FEEs
      ASIC settings DB file checked OK

      Ambient temperature +23.3 deg C, D.P +7.1 deg C,  R.H. 35.2%
      Julabo FL11006 set point +20.0 deg C, water le

15:38 Merger and TapeServer screenshots - Attachments 17-18
      
  45   Mon Feb 23 07:03:10 2015 TD, AE & CGMonday 23 February 2015

14.30 Saved nnaida1-16 ASIC + Waveform configurations to directory (Data Base Key)

           2015Feb23-12.11.40

          (directory aidas1:/MIDAS/DB/EXPERIMENTS/AIDA/2015Feb23-12.11.40)

          nnaida 1-2, 5-6, 9-10, 13-14 positive input ( front p+n junction strips )

          nnaida 3-4, 7-8, 11-12, 15-16 negative input ( back n+n ohmic strips )

 

           Standard ASIC configurations except

          - shaping time 0xb (6,5us)

          - peaking time 0x8 (16us)

         - slow comparator 0x10 (~160keV)

         - fast comparator (LEC/MEC) 0xff

         - fast comparator (HEC) 0xff

 

          nnaida 1-2, 5-6, 9-10, 13-14 negative input ( AIDA ASIC preamp output inverts input,

                                                                                 i.e. preamp with positive input produces negative input for sampling ADC )

          nnaida 3-4, 7-8, 11-12, 15-16 positive input

         Standard Waveform configurations except

         - threshold 1000

         - pre-trigger 0x100

         - all channels disabled

 

         2015Feb23-12.11.40 is now the default configuration file

 

Note:  The directory (Data Base Key) which will contain the saved settings is created uid: npg, gid: npgstaff, perms 755.

However, its contents are written uid: nfsnobody, gid nfsnobody, perms 777. The default directory (Data Base  Key) permissions

do not work - it is necessary to change the directory (Data Base Key)  to perms 777 after its creation and before saving the

individual FEE64 configuration directories/files to it.

 

 

 

 

        

        

 

  582   Mon May 22 02:34:51 2017 CB, TDMonday 22 May
10.35 DAQ continues OK 
      Current file May2017/R19_253

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      Detector biases; leakage currents OK - see attachment 1
      FEE64 Temperatures OK - see attachment 2
      good events; statistics OK - see attachment 3
      No warnings/errors reported by merger since DAQ GO - see attachment 4

      *all* system wide checks OK *except* ADC calibration of nnaida7

10.38 E17/F11 ambient temperature +24.7 deg C, RH 40.7%, d.p. +10.7 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)

10.39 DAQ stops OK
      Current file May2017/R19_253
      Deleted R19*

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate 250Hz range
      delay MIN, tail pulse

      Check ASIC Control x2
      Zero all spectra

10.47 DAQ starts
      File May2017/R20
      Ras Pi system console logs for USB0 attached - see attachment 5
      Analysed R20_0. All appears to be working fine. Highest dead time ~0.3%  - see attachment 6
      Off-line analysis of R20_0 confirms ADC data and discriminator are synchronised - see attachments 7,8,9

16.55 DAQ continues OK 
      Current file May2017/R20_71

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      Detector biases; leakage currents OK - see attachment 10
      FEE64 Temperatures OK - see attachment 12
      good events; statistics OK - see attachment 13
      No warnings/errors reported by merger since DAQ GO - see attachment 14

      *all* system wide checks OK *except* ADC calibration of nnaida7

17.02 E17/F11 ambient temperature +24.0 deg C, RH 40.3%, d.p. +9.7 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)

17.03 DAQ stops OK
      Current file May2017/R20_73
      Deleted R20*

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate 250Hz range
      delay MIN, tail pulse

      Check ASIC Control x2
      Zero all spectra

17.08 DAQ starts
      File May2017/R21
      Raspberry Pi system console logs for USB0 attached - see attachment 15

17.50 Changed permissions of options files /MIDAS/DB/EXPERIMENTS/Options/nnaida<n>/CONTENTS 
      from mode 644 to mode 777

18.09 DAQ stops OK
      Current file May2017/R21_11
      Deleted R21*

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate 250Hz range
      delay MIN, tail pulse

      DAQ RESET
      DAQ SETUP

      *all* system wide checks OK *except* nnaida7, 9, 13 and 22 ADC calibration failed

      Check ASIC Control x2

17.08 DAQ starts
      File May2017/R22
      Raspberry Pi system console logs for USB0 attached - see attachment 16
  584   Mon May 22 08:43:10 2017 CB, TD, PatrickMonday 22 May
> 10.35 DAQ continues OK 
>       Current file May2017/R19_253
> 
>       ASIC settings 2017Mar27-16.12.39 *except*
>       slow comparator 0x0a
>       LEC/MEC fast comparator 0xff
> 
>       Detector biases; leakage currents OK - see attachment 1
>       FEE64 Temperatures OK - see attachment 2
>       good events; statistics OK - see attachment 3
>       No warnings/errors reported by merger since DAQ GO - see attachment 4
> 
>       *all* system wide checks OK *except* ADC calibration of nnaida7
> 
> 10.38 E17/F11 ambient temperature +24.7 deg C, RH 40.7%, d.p. +10.7 deg C
>       Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)
> 
> 10.39 DAQ stops OK
>       Current file May2017/R19_253
>       Deleted R19*
> 
>       ASIC settings 2017Mar27-16.12.39 *except*
>       slow comparator 0x0a
>       LEC/MEC fast comparator 0xff
> 
>       BNC PB-4 settings 
>       Amplitude 90,000
>       x5 attenuator IN
>       t_r 50ns, tau_d 50us
>       polarity -
>       rate 250Hz range
>       delay MIN, tail pulse
> 
>       Check ASIC Control x2
>       Zero all spectra
> 
> 10.47 DAQ starts
>       File May2017/R20
>       Ras Pi system console logs for USB0 attached - see attachment 5
>       Analysed R20_0. All appears to be working fine. Highest dead time ~0.3%  - see attachment 6
>       Off-line analysis of R20_0 confirms ADC data and discriminator are synchronised - see attachments 7,8,9

Concerning nnaida7 and the failure of calibration..... have you tried to calibrate it using the FADC Align  & Control ?
If it still fails then there is a question about the quality of the clock signal being received by this module.
I will log through and look at the console log for any anomolies but please try and calibrate.
Let me know if you wish me to take remote control.
  585   Mon May 22 08:53:42 2017 CB, TD, PatrickMonday 22 May
> Concerning nnaida7 and the failure of calibration..... have you tried to calibrate it using the FADC Align  & Control ?
> If it still fails then there is a question about the quality of the clock signal being received by this module.
> I will log through and look at the console log for any anomolies but please try and calibrate.
> Let me know if you wish me to take remote control.

Yes - there have been multiple attempts to re-calibrate nnaida7 ADCs

Note *all* ADCs were calibrated before the most recent FEE64 power cycle.

It is not unusual for ~1-3 FEE64s to fail ADC calibration tests following
a power cycle - there is no obvious pattern to which ADCs fail or which (if
any) will be successfully re-calibrated.
  461   Mon Nov 21 02:12:45 2016 CG & TDMonday 21 November
11.10 DSSSD bias & leakage currents OK - see attachments 1 & 2
      FEE64 temperatures OK - see attachment 3

      Rate, HitRate, per FEE64 1.8.L and 1.8.W spectra - see attachments 4-9

16.10 DSSD bias and leakage currents ok (attachments 10+11).
      FEE temps ok.
  706   Mon Nov 20 00:53:27 2017 TD, CBMonday 20 November
09.50 Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      DAQ statistics OK - see attachment 4
      7x FEE64s <10k
      4x FEE64s 10-20k
      5x FEE64s 20-30k
      3x FEE64s 30-40k
      1x FEE64s 40-50k
      4x FEE64 >50k (nnaida13 255k)
      PLL & DCM Clock Lock Monitor OK - see attachment 5 & https://elog.ph.ed.ac.uk/AIDA/704 attachment 14
      *all* system wide checks OK *except* nnaida7 fails ADC calibration

16.39 Detector biases & leakage currents OK - see attachments 5 & 6
      FEE64 Temperatures OK - see attachment 7
      DAQ statistics OK - see attachment 8
      7x FEE64s <10k
      4x FEE64s 10-20k
      5x FEE64s 20-30k
      3x FEE64s 30-40k
      1x FEE64s 40-50k
      4x FEE64 >50k (nnaida13 255k)
      *all* system wide checks OK *except* nnaida7 fails ADC calibration
  807   Mon May 20 02:16:58 2019 CA, TDMonday 20 May 2019
10:18 DAQ Continues ok
      Alpha run continues - Currently on file R8_2

      System wide checks all ok

      Detector bias/leakage currents ok - attachment 1
     
      Temp +25.6 deg C, Humidity 34.5% AND D.P +8.8 deg C
      
      FEE Temperatures ok - attachment 2

      Good event statistics ok - attachment 3

      Merger & TapeServer ok - attachments 4 & 5


17.15 DAQ continues OK to file May19/R8_2

      alpha background

      ASIC settings 2019May19-17.54.03
      slow comparator 0x64

      BNC PB-4 Pulser OFF

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 6 & 7

      FEE64 temperatures OK - see attachment 8
      good event statistics OK - see attachments 9

      Merger OK - see attachment 10
      TapeServer OK - see attachment 11

      1.8.L spectra - see attachments 12 & 13
      p+n FEE64s 1.8.W spectra 20us FSR - attachment 14 
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 15

17.25 analysis of May19/R8_2 - see attachment 16
      deadtime OK
      ADC data for FEE64s nnaida5 & nnaida6 hot channels (Stat spectrum channel # 104)
      Also observed at a much lower rate for nnaida1 & nnaida2

17.27 Check ASIC Control *all* ASICs & FEE64s

13.27 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 60% M-H markers - water topped up (1x jugful) to H level marker
      Temperature +25.4 deg C dew point +9.6 deg C RH 36.3%
  342   Mon Aug 22 10:35:31 2016 TDMonday 20 August
11.00 Restore power to AIDA following Saturday's shutdoen in preparation for the planned maintenance
      outage on Sunday.

      Restart Julabo FL11006 chiller - set point +20 deg C
      Restart AIDA

      Temperature +24.3 deg C, DP +15.2 deg C, RH 57.5%

      DSSSD #1-6 bias - see attachments 1-2
     
      FEE64 temperatures - attachment 3

      System-wide checks - see attachments 4-8
      Note SYNC pulses received by *all* slave FEE64s = SYNC pulses sent by master FEE64 (nnaida6)

11.35 R6 starts
      slow comparator 0x10
      LEC/MEC fast comparator 0x10

      All FEE64 discriminator masks -> 0x0

      observe very high rates (~400-500kHz/FEE64) from majority ofp+n junction FEE64s

      LEC/MEC fast comparator -> 0xff

18.45 DAQ running OK
      No merger error messages since change of LEC/MEC fast comparator threshold -> 0xff
      Offline analysis of R6_79 confirms that all FEE64s are synchronised

      Added correlation trigger input
      BNC PB-4 Pulser trigger output -> Ortec SCA + NIM output width 0.5us -> LeCroy 429A -> root MACB trigger
#4 input

21.05 R7 starts

      Attachments 9-16 show FEE64 temperatures and results of each system-wide check. 

      Note the result of the Master Timestamp Control -> Check Clock Lock States 
      indicating clock state *not* locked

      slow comparator 0x10
      LEC/MEC fast comparator 0xff
      All FEE64 discriminator masks -> 0x0
  344   Mon Aug 22 13:37:30 2016 TDMonday 20 August
> 11.00 Restore power to AIDA following Saturday's shutdoen in preparation for the planned maintenance
>       outage on Sunday.
> 
>       Restart Julabo FL11006 chiller - set point +20 deg C
>       Restart AIDA
> 
>       Temperature +24.3 deg C, DP +15.2 deg C, RH 57.5%
> 
>       DSSSD #1-6 bias - see attachments 1-2
>      
>       FEE64 temperatures - attachment 3
> 
>       System-wide checks - see attachments 4-8
>       Note SYNC pulses received by *all* slave FEE64s = SYNC pulses sent by master FEE64 (nnaida6)
> 
> 11.35 R6 starts
>       slow comparator 0x10
>       LEC/MEC fast comparator 0x10
> 
>       All FEE64 discriminator masks -> 0x0
> 
>       observe very high rates (~400-500kHz/FEE64) from majority ofp+n junction FEE64s
> 
>       LEC/MEC fast comparator -> 0xff
> 
> 18.45 DAQ running OK
>       No merger error messages since change of LEC/MEC fast comparator threshold -> 0xff
>       Offline analysis of R6_79 confirms that all FEE64s are synchronised
> 
>       Added correlation trigger input
>       BNC PB-4 Pulser trigger output -> Ortec SCA + NIM output width 0.5us -> LeCroy 429A -> root MACB trigger
> #4 input
> 
> 21.05 R7 starts
> 
>       Attachments 9-16 show FEE64 temperatures and results of each system-wide check. 
> 
>       Note the result of the Master Timestamp Control -> Check Clock Lock States 
>       indicating clock state *not* locked
> 
>       slow comparator 0x10
>       LEC/MEC fast comparator 0xff
>       All FEE64 discriminator masks -> 0x0

The status is OK. I changed the system wide check to reflect the new value but forgot about the individual check.
I will update the software.
Note attachment 5 shows all is well.

Updated version of /MIDAS/TclHttpd/Html/AIDA/MasterTS/MasterTS.tcl installed with correction.
  896   Mon Oct 2 14:03:42 2023 TDMonday 2 October
15.00 DSSSD bias +60V +6.44uA - breakdown above c. 60V, bias +100V leakage current +20-30uA

      Slow comparator 0xa 

      aida03 LK3 fitted

      CAEN N1419ET grounded outputs
      HV#1 core aida04. braid aida10, aida1 & aida 9 

      BNC PB-5 
      Amplitide 10.0V
      Attenuation x10
      Decay time 1ms
      Polarity +
      Tail
      Frequency 25Hz

      test +/ test - daisy chains connected
      test - disconnected at EG&G Ortec 534 output
  197   Mon May 2 12:13:25 2016 CG,PV,TDMonday 2 May 2016
  735   Mon Jul 2 02:36:41 2018 OHMonday 2 July
10:35 Bias and leakage currents ok - attachment 1
      FEE Temps ok - attachment 2
      System wide checks all ok
      
  390   Mon Oct 17 04:10:30 2016 TDMonday 17 October
12.05 DAQ found stopped
      Most recent file R49_226, last accessed 06.26
      Restart merger OK
      DAQ STOP OK

      System-wide checks - see attachments 1-2 
      SYNC errors and SYNC pulses fail

      ReSYNC OK
      System-wide checks OK
      Restart merger & tapeserver

      DSSSD bias & leakage currents OK - see attachments 3 & 4
      FEE64 temperatures OK - see attachment 5

      DAQ start
      #1 nnaida15 data transfer #1 tag drops at DAQ GO
         disable H & data transfer #1
         restart merger & tapeserver
         enable H & data transfer #1

12.22 #2 DAQ start (R51) OK

20.33 DAQ found stopped
      Most recent file R51_145, last accessed 18.30
      Restart merger OK
      DAQ STOP OK

      System-wide checks - see attachments 6 & 7
      SYNC errors and SYNC pulses fail

      DAQ RESET & SETUP OK
      Disable H & data transfer #1
      Restart merger & tapeserver
      Enable H & data transfer #1

20.41 DAQ GO (R52)
  847   Sun Jun 16 16:00:58 2019 ML, CG, OHMonday 17 June 2019
It's a new day, it's a new log entry !!

01:25 DTAS run 46 started
      BigRIPS run 2019
      AIDA run R2_268

      DAQ running smoothly
      No change in Stat and Rate (stat zeroed at the beginning of this DTAS run)- See attachment 1
      Merger rate: 630k item/sec - See Attachment 2
      TapeServer rate still ~5 Mbytes/sec - See attachment 3
      
      Temperature and HV ok - See attachment 5 & 6

      System-wide checks OK except nnaida7 fails ADC calibration as usual.

      !!!
      Victor said that the run AIDA R2_249 is missing in the correlation merge data. This is the run when the
      correlation scaler was reset (as DTAS was running out of range in some of their timing histograms) (See
      elog end of Sunday)
      !!!

02:23 Rate in DSSD#3 card 9 and 12 slowly reaching 100kHz while 14 and 15 ~ 50kHz. See attachment 7
      TapeServer rate ~ 5.5 kHz
      

03:25 Change of DTAS run:
      DTAS run 49 started
      BigRIPS run 2020
      AIDA run R2_286

      Experiment still running smoothly. No significant change since last comments 2h ago.
      Leakage current in detector 3 is increasing slowly see attachment 8 & 9  

05:25 Change of DTAS run:
      Note: DTAS run stopped during AIDA R2_305 and re-started during R2_306.  

      DTAS run 52 started
      BigRIPS run 2021
      AIDA run R2_306

      Pretty quiet night shift.
      Stat & Rate in attachment 10 - DSSD#3 FEE card 9 back down to 60kHz
      Merger rate: 650k item/sec - See attachment 11
      TapeServer rate still ~5.1 Mbytes/sec - See attachment 12

      System-wide checks OK except nnaida7 fails ADC calibration as usual.
      
      X vs Y HEC spectrum in attachment 13
      LEC E vs HEC-LEC spectrum in attachment 14
      Both spectra zeroed

      Bias in attachment 15
      Temperature unchanged

07:30 DTAS and BigRIPS stop runs
      AIDA on file R2_323
      
07:33 New Run numbers
      DTAS Run: 55
      BigRIPS Run: 2022
      AIDA on File: R2_324

07:56 System wide checks ok
      Fee temps ok - attachment 16
      
      100Sn setting, no degrader

      ASIC settings 2019Jun13-08.56.35

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

08:28 Bias and leakage ok - attachment 17
      
08:48 ASIC check successful

09:32 DTAS and BigRIPS stop
      AIDA on file R2_342

      DTAS starts file 58
      BigRIPS run 2023
      AIDA still on file R2_342
  848   Mon Jun 17 02:31:53 2019 CA, TD, OHMonday 17 June 08.00-16.00
10.41 DAQ continues OK file RIBF147/R2_352

      100Sn setting, no degrader

      ASIC settings 2019Jun13-08.56.35
      DSSSD#1 slow comparator 0xa -> 0x1d
      DSSSD#3 slow comparator 0xa -> 0x8

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3

      good event statistics - OK - see attachment 4
      11x <20k 21x < 50k 24x <100k
      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
      1.8.L spectra - see attachments 9 & 10
      nnaida1 73 ch FWHM, nnaida3 78 ch FWHM

      FEE64 Rate spectra - attachment 11
      Note high rate issues observed for *some* strips in nnaida14 ASIC #3 & #4 fixed (temporarily) by Check
ASIC Control
      FEE64 Rate spectra - attachment 12

10.58 ADC data synchronised - attachment 13

      Online MIDASsort spectra - attachments 14-18

11.04 analysis of file RIBF147/R2_353 - attachment 19
      deadtime <<1% *all* FEE64s
      no timewarps

11.12 HitRate spectra - attachment 20
      disable nnaida14 *hot* channels 29, 30 & 35 

11.15 zero all histograms
      zero DAQ stats

12.31 DTAS and BigRIPS stop
      AIDA on file R2_368

      DTAS starts file 63
      BigRIPS run 2025
      AIDA file R2_368

13:08 R2_374
      The putty window for USB2 had frozen with a 1 in the voltage set as shown in attachment 1
      The window was no longer updating.
      When regaining the window the voltage limit was applied and the detector
      began to ramp down
      The set voltage was reset to 100V and bias was regained.
      Total time with less than 100V was around 30sec

14:28 DTAS and BigRIPS stop
      AIDA on run R2_386

      DTAS Starts file 66
      BigRIPS run 2026
      AIDA on file R2_387

14:32 Bias and leakage currents ok - attachment 21

15:07 Rate spectra showing no spikes in DSSD 3
      System wide checks

15:08 DTAS and BigRIPS stop
      AIDA onfile R2_393
      BigRIPS chaning slit positions

14:14 DTAS starts file: 
      AIDA on file R2_394
      BigRIPS file 2027 (In control by bigRIPS)

15.30 TAS and BigRIPS stop
      AIDA onfile R2_396
  33   Mon Feb 16 06:17:31 2015 TD, AEV & CGMonday 16th Feb progress

16/02/2015

1500: AIDA moved into operational position.

           Water cooling circuit visually checked, fittings, etc for loose fittings/snags/kinks. No problems

           Julabo FL11006 chiller moved into position with mains transformer, filled and turned on without problem (small leak from rear overflow, due to overfilling).

           Took roughly 2 ~30L water containers to fill reservoir.

           Brief leak at connection between back of chiller and yellow hose, bolt tightened and it stopped.

           Having been turned on and loop filled, reservoir at ~50% full. Pressure ~10 psi.

           No leaks observed in main coolant loop.

           When standing downstream of AIDA RH flow gauge blurred due to high rotational speed, LH flow gauge rotating at ~1 rev/s.

           Two FEE modules closest to centre in each block checked for power cabling, TS distribution and network cables. All ok.

            MACBs front panel cabling checked, no obvious problems. TS distribution tree in order.

           Power on for NIM bins and fan tray, network switches and Raspberry Pi. Not yet for the FEE module PSUs.

 

           1 corner of each new USB power supply damaged in transit.

           Fuses changed from 10A -> 16A in each PSU. Old 10A fuses taped to back of each module.

 

1730: Turned on nnaida1-16. Temperatures stable. All on firmware version 0xa4ed006.

            Input pulser into all FEE cards. All FEEs bar nnaida5 well-behaving. nnaida12 has several missing channels.

            Typical pulser peak width for -ve input modules was ~14ch FWHM with shaping time of 1us.

            Typical pulser peak width for +ve input modules was ~27ch FWHM with shaping time of 1us.

            Changing shaping time of nnaida10 from 1->2.5us reduced FWHM from 27ch to 23ch.

          

Attachment 1: temperatures & FEE firmware revisions nnaida1-16

 

 

  567   Mon May 15 02:01:02 2017 TD, OHMonday 15 May
10.01 DAQ continues OK 
      Current file May2017/R8_334

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      'good events' statistics OK - see attachment 4
      No warnings/errors reported by merger since DAQ GO - see attachment 5

      *all* system wide checks OK

10.06 Analysis of data file R8_333 - see attachment 6
      dead-time/idle-time < 2%
      AIDA SYNC100 variations ~1800/95000

10.10 E17/F11 ambient temperature +24.8 deg C, RH 36.4%, d.p. +8.9 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%)

10.15 DAQ stops OK
      Current file May2017/R7_298

      Files May2017/R8* *deleted* 

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator *0x08*
      LEC/MEC fast comparator 0xff

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate 250Hz range
      delay MIN, tail pulse

      Check ASIC Control x2
      Zero all spectra

10.20 DAQ starts
      File May2017/R8

      Transient timeouts at DAQ GO for FEE64s nnaida1 and nnaida2 which resolved themselves without user
intervention
      DAQ run control transitioned from 'RESET and STOP/GO' to 'RESET' to 'RESET and STOP/GO'

10.22 DAQ stops
      File May2017/R9

      Tapeserver data rates fluctuate 10-30Mb/s possibly due to parallel disk I/O activity (deleting R8 files)

      Transient timeouts at DAQ STOP for FEE64s nnaida1 and nnaida2 which resolved themselves without user
intervention - see attachment 7
      DAQ run control transitioned from 'RESET and STOP/GO' to 'RESET' to 'RESET and STOP/GO'

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator *0x0a*
      LEC/MEC fast comparator 0xff

10.31 DAQ starts 
      File May2017/R10