AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA, Page 33 of 46  ELOG logo
Entry  Mon Dec 4 10:34:48 2023, NH, AIDA Todo for 11.12.23 
This is a short list of things todo or check before the test beam time on 11.12.23 
Updated after 15:00 CET meeting

1. Replace water interlock
- Also allows testing of AIDA for 24h+ 

2. Decide on snout configuration: 1 or 2 BB18s (plus BB7 plus bPlastic)
  It is U so if we put 2 we could damage 2 instead of just 1 (of course we try to be careful)
  To decide which bPlas is installed (only 1 will be bothered): Upstream is preferred by HA
  Install Peter's light blocking frame between bPlastic and AIDA

3. Replace aida04 with a working FEE64
  Important as aida04 is in deadtime and not very useful

4. Investigate/replace aida06 
  Less urgent because 2 strips and runs with 0 dead time

5. If 2 BB18s, add another 8 working FEE64s to the system
  Note: Cannot do 2BB18s and 1BB7 (cannot power/cool 17 FEEs)

6. Test AIDA firmware fix for individual ASIC controls (especially if BB7<->AIDA will be tested)
  BB7<->AIDA will not be tested probably, as PCB will not arrive.
  Still good to check if possible

7. Decide on FEE to take pulser (assuming no distribution/inverter available)
  Test implant response with pulser and SC41 trigger?
Entry  Wed Dec 6 16:32:03 2023, NH, Replacement of faulty Honeywell HSS-DPS sensor Honeywell_HSS-DPS_63-2711.pdf1000008603.jpg1000008604.jpg1000008605.jpg
Honeywell HSS-DPS sensor is faulty - attachments 1 & 2

Replace with Sontay Condensation Prevention Switch - attachments 3 & 4

Wiring

      Honeywell             Sontay
      V+(*) red             V+(*) red
      GND   blue            0V    blue
      COM   green           COM   black
      DRY   yellow          N/O   yellow

Honeywell V+ 24V ac/dc +/- 20%
Sontay V+ 24V dc +/-5%, 24V ac +/-10%

V+ measured as +16.5V dc with DMM

Interlock still not working

- V value too low ?
- Honeywell DRY not equivalent to Normally Open ?
Entry  Tue Jan 30 14:18:16 2024, NH, [HowTo] Update MACB Firmware  MACB_GSI_Simple.jed
https://elog.ph.ed.ac.uk/CARME/482
Entry  Wed Mar 13 14:50:04 2024, NH, [HowTo] Backup and Restore Raspberry Pi SD Cards 
Micro SD cards used by the raspberry pis are cheap but prone to failing

If the system won't boot and the SD card is read only it may have failed. Best way is to buy a new micro SD card (16 or 32GB) and restore a backup

I have uploaded an image of nnrpi2 (AIDA@GSI) to aida-3 (/home/npg/nnrpi2.img)

To backup an SD card

dd if=/dev/mmcblk0 of=nnrpi2.img status=progress

To restore a backup to a new SD card

dd if=nnrpi2.img of=/dev/mmcblk0 status=progress

(/dev/mmcblk0 name may differ on different machines, is for a laptop with built in SD card reader)
Entry  Sun Jun 16 16:00:58 2019, ML, CG, OH, Monday 17 June 2019 17x
It's a new day, it's a new log entry !!

01:25 DTAS run 46 started
      BigRIPS run 2019
      AIDA run R2_268

      DAQ running smoothly
      No change in Stat and Rate (stat zeroed at the beginning of this DTAS run)- See attachment 1
      Merger rate: 630k item/sec - See Attachment 2
      TapeServer rate still ~5 Mbytes/sec - See attachment 3
      
      Temperature and HV ok - See attachment 5 & 6

      System-wide checks OK except nnaida7 fails ADC calibration as usual.

      !!!
      Victor said that the run AIDA R2_249 is missing in the correlation merge data. This is the run when the
      correlation scaler was reset (as DTAS was running out of range in some of their timing histograms) (See
      elog end of Sunday)
      !!!

02:23 Rate in DSSD#3 card 9 and 12 slowly reaching 100kHz while 14 and 15 ~ 50kHz. See attachment 7
      TapeServer rate ~ 5.5 kHz
      

03:25 Change of DTAS run:
      DTAS run 49 started
      BigRIPS run 2020
      AIDA run R2_286

      Experiment still running smoothly. No significant change since last comments 2h ago.
      Leakage current in detector 3 is increasing slowly see attachment 8 & 9  

05:25 Change of DTAS run:
      Note: DTAS run stopped during AIDA R2_305 and re-started during R2_306.  

      DTAS run 52 started
      BigRIPS run 2021
      AIDA run R2_306

      Pretty quiet night shift.
      Stat & Rate in attachment 10 - DSSD#3 FEE card 9 back down to 60kHz
      Merger rate: 650k item/sec - See attachment 11
      TapeServer rate still ~5.1 Mbytes/sec - See attachment 12

      System-wide checks OK except nnaida7 fails ADC calibration as usual.
      
      X vs Y HEC spectrum in attachment 13
      LEC E vs HEC-LEC spectrum in attachment 14
      Both spectra zeroed

      Bias in attachment 15
      Temperature unchanged

07:30 DTAS and BigRIPS stop runs
      AIDA on file R2_323
      
07:33 New Run numbers
      DTAS Run: 55
      BigRIPS Run: 2022
      AIDA on File: R2_324

07:56 System wide checks ok
      Fee temps ok - attachment 16
      
      100Sn setting, no degrader

      ASIC settings 2019Jun13-08.56.35

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

08:28 Bias and leakage ok - attachment 17
      
08:48 ASIC check successful

09:32 DTAS and BigRIPS stop
      AIDA on file R2_342

      DTAS starts file 58
      BigRIPS run 2023
      AIDA still on file R2_342
Entry  Sun Nov 27 19:33:00 2016, Liu, PV, AE, F11 degrader setting (night Nov27-Nov28) run1038_imp_profile.pngrun1038_imp_pid.pngrun104x_imp_profile.pngrun104x_imp_pid.png
Analysis of files for two more settings of F11 degrader to optimize implantation of 40Mg.

+ setting 1: from Nov 27th, 22:16
    Fixed degrader: 3 mm Pb + 1 mm Al, Variable degrader: 1.3 mm W + 2 mm Al
    AIDA run: R26_0 to R26_38
    BigRIPS run: 40Mg1037 (partial) to 40Mg1040
    see attachments: run1038_imp*png

+ setting 2: from Nov 28th, 1:38.
    Entered F11 during stripper foil change and removed 1mm Al fixed degrader (same as during Nov 26th - ~ run
40Mg1012).
    Fixed degrader: 3 mm Pb, Variable degrader: 1.3 mm W + 2 mm Al
    AIDA run: R26_39 to ...
    BigRIPS run: from 40Mg1041 to ...
    see attachments: run104x_imp*png (2 hours of data: run 1041 and 1042)

Setting 2 moved implantation depth one or two strips upstream (1.2 according to calculations). 40Mg seems to be
implanted about the third strip; a conclusion with LOW STATISTICS, but consistent with the trend of how the
implantation position depends on neutron number for the higher-count isotopes. 

Setting 2 seems to be our best option (at least without going into F11 and reconfiguring degraders). We will
continue accumulating data with it.

Attachments: the 1D plot shows the ion implant position (DSSD) gated for all ions and for Al and Mg gates in the
PID. The 2D plots show the PID with gates on implants in different DSSDs.

Range calculation for degrader's configuration:
https://docs.google.com/spreadsheets/d/1QJp8t98sT3pExAQb3xtHtAvk4TUDVO6alhbLIaNtx8A/edit?usp=sharing
Entry  Sun Nov 27 22:35:37 2016, Liu, PV, AE, Ion hit pattern in AIDA after degrader changing aida_ion_hitpattern_run26_51to63.png

After changing degrader to optimize 40Mg, hit pattern in AIDA is shown in figure below.

The implantation position changed from right side position of  first DSSD to center position of last DSSD.

Entry  Sat Nov 26 20:26:06 2016, Liu, AE, Analysis of implantation profile in AIDA imp.jpgpid_corr.png

Here we show the PID plot and implantation depth in AIDA from BigRIPS file 40Mg1012 (1 hr of data from 20:30 in Nov 26th), merged with AIDA files R10_8 to R10_29

The more neutron-rich isotopes have a shorter range in AIDA, and seem to be implanted towards the first layers of the stack. For 40Mg - most exciting case(?) - there is only one count at DSSD1. But comparing to the position of 38Mg, we could deduce that the implantation depth is already good. need more data analyzed to decide if we need to remove a bit of degrader.

Degrader settings: 3mm Pb (at BRIKEN PE shielding), variable of 1mm W (+1mmAl) + 0.3mmW (+1 mm Al)

Attachment 1: DSSD of detected implant for cuts in the PID.

Attachment 2: PID from F8 to F11.

BigRIPS trees produced with macro at gamma16@a04:exp/users/alfredo/macro/MakePIDTree.C (or MakeAllTree.C, depending on version of merger)

Merging code in: ur16@ribfana03:analysis/users/liujj/BRIKEN_161125Parasitic/
 

Entry  Mon Nov 12 08:59:43 2018, LS, CA, TD, Monday 12 November 6pm onwards 28x

17:59 DAQ continues OK
      Stats OK - attachment 1
      FEE temperatures OK - attachment 2
      Bias and leakage currents OK - attachment 3
      System-wide checks OK *except* nnaida7 fails ADC calibration


19:07 DAQ continues OK
      Stats OK - attachment 4
      FEE temperatures OK - attachment 5
      Bias and leakage currents OK - attachment 6
      System-wide checks OK *except* nnaida7 fails ADC calibration

20.19 DAQ starts
      AIDA file RIBF148/R5_4
      BRIKEN file 16
      BigRIPS 1001

      165Pm setting + 3mm Al degrader

20.21 all online & MIDASsort histograms zero'd

20.23 FEE64 good event statistics - see attachment 7
      Merger - see attachment 8
      TapeServer - see attachment 9

      HEC x versus y spectra - see attachment 10
       SN thinks DSSSD #3 are the ions of interest, #4 are the ions of interest at different charge state
       and #5 & #6 are lighter ions which do not stop in DSSSD #1-6

21.21 DAQ stops 

      DAQ starts
      AIDA file RIBF148/R5_14
      BRIKEN file 18
      BigRIPS 1003

21.26 HEC x vs y multiplicity spectra - attachment 11

21.29 HEC Ex vs Ey spectra - attachment 12
      FEE64 rate spectra - attachment 13

21.31 1.8.L front/rear energy spectra (attachments 14/15) LEC

21.33 1.8.H front/rear energy spectra (attachments 16/17) HEC

21.46 system wide checks ok, except nnaida7 ADC calibration failure
      FEE64 good event statistics ok (attachment 18)
      FEE64 temperatures ok (attachment 19)
      Leakage currents/detector bias ok (attachment 20)

21.49 Merger and TapeServer ok, no error messages in terminal
      DAQ still running (RIBF148/R5_18)

22.10 attachment 21 - pulser spectra in all channels fired at same time
                    - FEE64 in sync

22.12 SYNC100 at 9153.8 Hz
      FEE64 discriminator at ~15000 Hz - higher than expected
      ~1 second deadtime in FEE19 - due to nnaida12 running fast
      no timewarps observed in FEE64
      all of above in attachment 22
   
22:32 DAQ continues OK
      Stats OK - attachment 24
      FEE temperatures OK - attachment 23
      Bias and leakage currents OK - attachment 25
      System-wide checks OK *except* nnaida7 fails ADC calibration   

23:30 DAQ continues OK
      Stats OK - attachment 26
      FEE temperatures OK - attachment 27
      Bias and leakage currents OK - attachment 28
      System-wide checks OK *except* nnaida7 fails ADC calibration 
      
Entry  Tue Nov 13 14:27:38 2018, LS, Wednesday 14 November 00:00 to 08:00 9x
00:00 BRIKEN stop AIDA file RIBF148/R5_269 

---------------------------------------------------------------- 

00.02 BRIKEN start 
      BRIKEN file 45 
      BigRIPS file 1030 
      AIDA file RIBF148/R5_270 

00:07 system wide checks ok, except nnaida7 ADC calibration failure 
      FEE64 temperatures ok (attachment 1) 
      good event statistics ok (attachment 2) 
      detector bias/leakage currents ok (attachment 3) 

01:01 BRIKEN stop 
      AIDA file RIBF148/R5_279 

--------------------------------------------------------------- 

01.02 BRIKEN start 
      BRIKEN file 46 
      BigRIPS file 1031 
      AIDA file RIBF148/R5_279 

02:00 BRIKEN stop 
      AIDA file RIBF148/R5_289 

---------------------------------------------------------------- 

02.01 BRIKEN start 
      BRIKEN file 47 
      BigRIPS file 1032 
      AIDA file RIBF148/R5_289 

03:00 BRIKEN stop 
      AIDA file RIBF148/R5_298 

---------------------------------------------------------------- 

03.01 BRIKEN start 
      BRIKEN file 48 
      BigRIPS file 1033 
      AIDA file RIBF148/R5_298 

03:04 system wide checks ok, except nnaida7 ADC calibration failure 
      FEE64 temperatures ok (attachment 4) 
      good event statistics ok (attachment 5) 
      detector bias/leakage currents ok (attachment 6) 

04:00 BRIKEN stop 
      AIDA file RIBF148/R5_308 

----------------------------------------------------------------

04.03 BRIKEN start 
      BRIKEN file 49 
      BigRIPS file 1034 
      AIDA file RIBF148/R5_308 

05:00 BRIKEN stop 
      AIDA file RIBF148/R5_317 

---------------------------------------------------------------- 

05.01 BRIKEN start 
      BRIKEN file 50 
      BigRIPS file 1035 
      AIDA file RIBF148/R5_318 

06:00 BRIKEN stop 
      AIDA file RIBF148/R5_327 

---------------------------------------------------------------- 

06.02 BRIKEN start 
      BRIKEN file 51 
      BigRIPS file 1036
      AIDA file RIBF148/R5_327 

06:07 system wide checks ok, except nnaida7 ADC calibration failure 
      FEE64 temperatures ok (attachment 7) 
      good event statistics ok (attachment 8) 
      detector bias/leakage currents ok (attachment 9)

07:00 BRIKEN stop 
      AIDA file RIBF148/R5_337 

---------------------------------------------------------------- 

07.01 BRIKEN start 
      BRIKEN file 52 
      BigRIPS file 1037
      AIDA file RIBF148/R5_327 
Entry  Wed Nov 14 22:54:11 2018, LS, Thursday 15th November 08:00 - 16:00 10x
08.10 BRIKEN stop
      AIDA file RIBF148/R5_587

08.11 BRIKEN start
      AIDA file RIBF148/R5_588
      BRIKEN run 77
      BigRIPS run 1062

08.16 system wide checks ok, except nnaida7 ADC calibration failure
      FEE64 temperatures ok (attachment 1)
      good event statistics ok (attachment 2)
      detector bias/leakage currents ok (attachment 3)
      Merger and TapeServer still ok

09.11 BRIKEN stop
      AIDA file RIBF148/R5_598

09.13 BRIKEN start
      AIDA file RIBF148/R5_598
      BRIKEN run 78
      BigRIPS run 1063

10.10 BRIKEN stop
      AIDA file RIBF148/R5_608

10.12 BRIKEN start
      AIDA file RIBF148/R5_608
      BRIKEN run 79
      BigRIPS run 1064

11.10 BRIKEN stop
      AIDA file RIBF148/R5_618

11.12 BRIKEN start
      AIDA file RIBF148/R5_618
      BRIKEN run 80
      BigRIPS run 1065

11.17 system wide checks ok, except nnaida7 ADC calibration failure
      FEE64 temperatures ok (attachment 4)
      good event statistics ok (attachment 5)
      detector bias/leakage currents ok (attachment 6)
      Merger and TapeServer still ok

12:10 BigRIPS and BRIKEN stop
      AIDA on file R5_628

12:11 BRIKEN start
      AIDA on file R5_628
      BigRIPS run 1066
      BRIKEN run 81 181115_1211_165Pm_081.dlt

13.10 BRIKEN stop
      AIDA file RIBF148/R5_638

13.12 BRIKEN start
      AIDA file RIBF148/R5_638
      BRIKEN run 82
      BigRIPS run 1067

13.32 AIDA file RIBF148/R5_642
      BRIKEN stopped
      Turning beam off to access area

13.50 AIDA file RIBF148/R5_644
      beam back on
      BRIKEN run 83
      BigRIPS run 1068

14.26 MIDASsort analysis file RIBF148/R5_648 - all ADC data synchronised - attachment 7
      analysis of file RIBF148/R5_648 - attachment 8
       nnaida19 dead time ~0.5% - all other deadtimes lower or 0% - no timewarps

14.35 1.8.W spectra layout ID 8 (attachment 9)

14.50 BRIKEN stop
      AIDA file RIBF148/R5_654

14.51 BRIKEN start
      AIDA file RIBF148/R5_654
      BRIKEN run 83
      BigRIPS run 1068

15.10 1.8.H high energy channel spectra (attachment 10)

15.35 beam off
      AIDA file RIBF148/R5_662
      BRIKEN and BigRIPS stopped run           
Entry  Thu Nov 15 15:05:33 2018, LS, Friday 16 November 00.00 to 08.00 9x
00.25 System wide checks ok * except nnaida 7 fails adc*
      FEE Temps ok - attachment 1
      Stats ok - attachment 2 
      Leakage currents ok - attachment 3
      tape server and merger ok

00:57 BigRIPS and BRIKEN stop
      AIDA on file R5_753

00.59 BigRIPS starts run 1079
      BRIKEN starts run 94
      AIDA on file R5_753

01:56 BigRIPS and BRIKEN stop
      AIDA on file R5_762

01.57 BigRIPS starts run 1080
      BRIKEN starts run 95
      AIDA on file R5_762

02:47 BigRIPS and BRIKEN stop
      AIDA on file R5_770

02.48 BigRIPS starts run 1081
      BRIKEN starts run 96
      AIDA on file R5_770

03:45 BigRIPS and BRIKEN stop
      AIDA on file R5_780

03.47 BigRIPS starts run 1082
      BRIKEN starts run 97
      AIDA on file R5_780

04:45 BigRIPS and BRIKEN stop
      AIDA on file R5_790

04.47 BigRIPS starts run 1083
      BRIKEN starts run 98
      AIDA on file R5_790

04.50 System wide checks ok * except nnaida 7 fails adc*
      FEE Temps ok - attachment 4
      Stats ok - attachment 5
      Leakage currents ok - attachment 6
      tape server and merger ok

05.46 BigRIPS and BRIKEN stop
      AIDA on file R5_800

05.47 BigRIPS starts run 1084
      BRIKEN starts run 99
      AIDA on file R5_800

06.45 BigRIPS and BRIKEN stop
      AIDA on file R5_810

06.46 BigRIPS starts run 1085
      BRIKEN starts run 100
      AIDA on file R5_810

07.25 System wide checks ok * except nnaida 7 fails adc*
      FEE Temps ok - attachment 7
      Stats ok - attachment 8
      Leakage currents ok - attachment 9
      tape server and merger ok

07.44 BigRIPS and BRIKEN stop
      AIDA on file R5_820

07.45 BigRIPS starts run 1086
      BRIKEN starts run 101
      AIDA on file R5_820
Entry  Sat Jun 3 00:00:11 2017, LJB, PW, TD, OH, Saturday 3 June 08.00-16.00 19x
08.29 BRIKEN 100; BigRIPS 1111 stopping run.
      AIDA on run R1_1429

08.30 BigRIPS starts run 1112
      BRIKEN starts run 101
      AIDA on run R1_1430

09.08 System wide checks OK
      FEE64 temperatures OK - attachment 1
      Good events statistics OK - attachment 2
      Leakage currents OK - attachment 3    
      No new merger errors - attachment 4
      HEC x vs y plot OK - attachment 6
      Zero SortSas

09.30 BRIKEN 101; BigRIPS 1112 stopping run. AIDA on run R1_1442

09.30 BigRIPS starts run 1113
      BRIKEN starts run 102
      AIDA on run R1_1443

09.55 Beam off

09.57 Beam on

10.32 BRIKEN 102; BigRIPS 1113 stopping run. AIDA on run R1_1457

10.33 BigRIPS starts run 1114
      BRIKEN starts run 103
      AIDA on run R1_1457

10.37 Online analysis of R1_1456, dead time below 0.3%, see attachment 7

11.08 System wide checks OK
      FEE64 temperatures OK - attachment 8
      Good events statistics OK - attachment 9
      Leakage currents OK - attachment 10
      No new merger errors - attachment 11 
      HEC x vs y plot OK - attachment 12
      Zero SortSas

11.30 BRIKEN 103; BigRIPS 1114 stopping run. AIDA on run R1_1470

11.35 BigRIPS starts run 1115
      BRIKEN starts run 104
      AIDA on run R1_1472

12.25 BRIKEN 104; BigRIPS 1115 stopping run. AIDA on run R1_1483, BRIKEN timestamp reset

12.29 DAQ starts
      AIDA file RIBF127R1/R1_1485
      BigRIPS file 1116
      BRIKEN file 105

      DAQs synchronised

13.11 Analysis of R1_1484 - deadtime <0.5%, no timewarps, no MBS timewarps - see attachment 13

13.24 Beam lost AIDA file RIBF127R1/R1_1497
      BRIKEN, BigRIPS DAQs stop

13.28 Beam returns AIDA file RIBF127R1/R1_1497

13.30 DAQ starts
      AIDA file RIBF127R1/R1_1497
      BigRIPS file 1117
      BRIKEN file 106

13.54 Beam lost AIDA file RIBF127R1/R1_1504

13.57 Beam returns AIDA file RIBF127R1/R1_1505

14:30 BigRips and BRIKEN stop their daqs
      AIDA on file R1_1513

14:31 BRIKEN starts file 107
      BigRIPS starts file 1118
      AIDA on file R1_1513

14:36 System wide checks ok

15.29 System wide checks OK
      FEE64 temperatures OK - attachment 14
      Good events statistics OK - attachment 15
      Leakage currents OK - attachment 16
      No new merger errors - attachment 17
      HEC x vs y plot OK - attachment 18
      Zero SortSas

15.31 BigRips 1118 and BRIKEN 107 stop their daqs
      AIDA on file R1_1527

15:31 BRIKEN starts file 108
      BigRIPS starts file 1119
      AIDA on file R1_1527

14.57 Analysis of R1_1532 - deadtime <0.5%, no timewarps, no MBS timewarps - see attachment 19
Entry  Sun Jun 4 08:13:10 2017, LHB, PV, NN, 4th June 16.00-00.00 28x
16.25  *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
       FEE64 Temperatures OK - see attachment 1
       Good events statistics OK - see attachment 2 
       Detector biases & leakage currents OK - see attachment 3
       No warnings/errors reported by merger since DAQ GO - see attachment 4

16.30  Analysis file RIBF148/R2_122 - nnaida2 deadtime 0.9%, all other deadtimes <0.3%, zero timewarps, one MBS
timewarp - see attachment 5

16.44  BRIKEN and BigRIPS stop daqs
       AIDA on file R2_127

16.47  BRIKEN starts run 9
       BigRIPS starts run 2008
       AIDA on file R2_128

17.08  Online analysis - per DSSSD HEC x versus y attachment 6
                         per DSSSD HEC E_p versus E_n - see attachment 7

17.44  BRIKEN and BigRIPS stop daqs
       AIDA on file R2_143

17.46  BRIKEN starts run 10
       BigRIPS starts run 2009
       AIDA on file R2_144

18.00  BRIKEN and BigRIPS stop daqs
       AIDA on file R2_147

18.02  BRIKEN starts run 11
       BigRIPS starts run 2010
       AIDA on file R2_148

18.15  *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
       FEE64 Temperatures OK - see attachment 8
       Good events statistics OK - see attachment 9
       Detector biases & leakage currents OK - see attachment 10
       No warnings/errors reported by merger since DAQ GO - see attachment 11

18.47  Online analysis - per DSSSD HEC x versus y attachment 12
                         per DSSSD HEC E_p versus E_n - see attachment 13
       Zero online sort spectra

18.50  Analysis of file RIBF148/R2_159 - all deadtimes <0.5%, zero timewarps, four MBS timewarps - see attachment 14

19.04  BRIKEN and BigRIPS stop daqs
       AIDA on file R2_165

19.07  BRIKEN starts run 12
       BigRIPS starts run 2011
       AIDA on file R2_166

20:00  BRIKEN and BigRIPS stop daqs
       AIDA on file R2_180

20.02  BRIKEN starts run 13
       BigRIPS starts run 2012
       AIDA on file R2_180
       Zero online sort spectra so we can monitor the influence on HEC x versus y due to the degrader just inserted.

20.15  *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
       FEE64 Temperatures OK - see attachment 15
       Good events statistics OK - see attachment 16
       Detector biases & leakage currents OK - see attachment 17
       No warnings/errors reported by merger since DAQ GO - see attachment 18

20.28  Analysis of file RIBF148/R2_185 - all deadtimes <0.4%, zero timewarps, three MBS timewarps - see attachment 19

20.58  Online analysis - per DSSSD HEC x versus y attachment 20
                         per DSSSD HEC E_p versus E_n - see attachment 21
       Zero online sort spectra
     
21.02  BRIKEN and BigRIPS stop daqs
       AIDA on file R2_197

21.03  BRIKEN starts run 14
       BigRIPS starts run 2013
       AIDA on file R2_197

22.02  BRIKEN and BigRIPS stop daqs
       AIDA on file R2_213

22.03  BRIKEN starts run 15
       BigRIPS starts run 2014
       AIDA on file R2_214

22.17  *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
       FEE64 Temperatures OK - see attachment 22
       Good events statistics OK - see attachment 23
       Detector biases & leakage currents OK - see attachment 24
       No warnings/errors reported by merger since DAQ GO - see attachment 25

22.22  Analysis of file RIBF148/R2_218 - all deadtimes <0.5%, zero timewarps, four MBS timewarps - see attachment 26

22.45  Online analysis - per DSSSD HEC x versus y attachment 27
                         per DSSSD HEC E_p versus E_n - see attachment 28
       Zero online sort spectra

23.04  BRIKEN and BigRIPS stop daqs
       AIDA on file R2_230

23.05  BRIKEN starts run 16
       BigRIPS starts run 2015
       AIDA on file R2_231
Entry  Thu Jun 1 08:49:39 2017, LHB, PV, Thursday 1 June Evening Shift 27x
16.44 DAQ stop BigRIPS, BRIKEN
      AIDA file RIBF127R1/R1_886

16.47 BRIKEN start run 64
      BigRIPS start run 1075
      AIDA file RIBF127R1/R1_887

      *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)

17.10 FEE64 Temperatures OK - see attachment 1
      Detector biases & leakage currents OK - see attachments 1
      Good events statistics OK - see attachment 3
      No new merger error messages - see attachment 4

      *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
      all SortSas spectra zero'd

17.40 Analysis of file RIBF127R1/R1_898 - dead time < 0.5% - see attachment 5

17.44 DAQ stop BigRIPS, BRIKEN
      AIDA file RIBF127R1/R1_900

17.46 BRIKEN start run 65
      BigRIPS start run 1076
      AIDA file RIBF127R1/R1_901

18.27 Online analysis: DSSSD HEC x versus y spectra - see attachment 6

18.46 DAQ stop BigRIPS, BRIKEN
      AIDA file RIBF127R1/R1_915

18.47 BRIKEN start run 66
      BigRIPS start run 1077
      AIDA file RIBF127R1/R1_915

18.54 FEE64 Temperatures OK - see attachment 7
      Detector biases & leakage currents OK - see attachments 8
      Good events statistics OK - see attachment 9
      No new merger error messages - see attachment 10
      *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
      All SortSas spectra zero'd

19.50 DAQ stop BigRIPS, BRIKEN
      AIDA file RIBF127R1/R1_930

19.52 BRIKEN start run 67
      BigRIPS start run 1078
      AIDA file RIBF127R1/R1_930

20.01 Analysis of file RIBF127R1/R1_931 - dead time < 0.5% - see attachment 11

20.41 FEE64 Temperatures OK - see attachment 12
      Detector biases & leakage currents OK - see attachment 13
      Good events statistics OK - see attachment 14
      No new merger error messages - see attachment 15
      *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)

20.47 Online analysis: DSSSD HEC x versus y spectra - see attachment 16
      All SortSas spectra zero'd

20.52 DAQ stop BigRIPS, BRIKEN
      AIDA file RIBF127R1/R1_944

20.54 BRIKEN start run 68
      BigRIPS start run 1079 
      AIDA file RIBF127R1/R1_945

21.51 DAQ stop BigRIPS, BRIKEN
      AIDA file RIBF127R1/R1_958

21.52 BRIKEN start run 69
      BigRIPS start run 1080 
      AIDA file RIBF127R1/R1_959

21.59 Analysis of file RIBF127R1/R1_959 - dead time < 0.5% - see attachment 17

22.40 FEE64 Temperatures OK - see attachment 18
      Detector biases & leakage currents OK - see attachment 19
      Good events statistics OK - see attachment 20
      No new merger error messages - see attachment 21
      *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)

22.42 Online analysis: DSSSD HEC x versus y spectra - see attachment 22
      All SortSas spectra zero'd

22.55 DAQ stop BigRIPS, BRIKEN
      AIDA file RIBF127R1/R1_974

22.56 BRIKEN start run 70
      BigRIPS start run 1081 
      AIDA file RIBF127R1/R1_974

23.18 Beam is off (SRC is down)
      AIDA file RIBF127R1/R1_979
      DAQ stop BigRIPS, BRIKEN

23.24 Beam is back
      Current AIDA file RIBF127R1/R1_980

23.25 BRIKEN start run 71      
      BigRIPS start run 1082
      AIDA file RIBF127R1/R1_981

23.47 *All* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (status=6)
      FEE64 Temperatures OK - see attachment 23
      Detector biases & leakage currents OK - see attachment 24
      Good events statistics OK - see attachment 25
      No new merger error messages - see attachment 26

23.52 Analysis of file RIBF127R1/R1_986 - dead time < 0.5% - see attachment 27

00:01 Beam is off
      DAQ stop BigRIPS, BRIKEN
      AIDA file RIBF127R1/R1_989
Entry  Fri Jun 2 08:25:34 2017, LHB, 2nd June Evening Shift 20x
16.34  Beam is off
       Aida R_1216, BRIKEN (83)/BigRIPS (1094) continue to run for background data

16.40  BRIKEN run 83 and BigRIPS run 1094 stopping daqs
       AIDA on file R1_1218

16.44  Still no beam, running background
       BigRIPS starts run 1095
       BRIKEN starts run  84
       Aida on file R1_1218

17.55  Still no beam, running background
       System wide checks all ok
       FEE temperatures all ok - attachment 1
       Good events statistics - attachment 2
       Bias and leakage currents ok - attachment 3
       Merger ok - attachment 4

17.58  Beam on, AIDA file R1_1234

17.59  Briken run 84 and BigRIPS run 1095 stopping daqs 
       AIDA on file R1_1234

18.00  BigRIPS starts run 1096
       BRIKEN starts run  85
       Aida on file R1_1234
       Zero'd online SAS spectra

19.00  BRIKEN run 85 and BigRIPS run 1096 stopping daqs 
       AIDA on file R1_1248

19.02  BigRIPS starts run 1097
       BRIKEN starts run  86
       Aida on file R1_1248
 
19.04  Online analysis: DSSSD HEC x versus y spectra - see attachment 5

20.00  BRIKEN run 86 and BigRIPS run 1097 stopping daqs 
       AIDA on file R1_1261

20.01  BigRIPS starts run 1098
       BRIKEN starts run  87
       Aida on file R1_1261

20.04  Beam is off
       Aida R_1262, BRIKEN (87)/BigRIPS (1098) continue to run for background data

20.15  System wide checks all ok
       FEE temperatures all ok - attachment 6
       Good events statistics - attachment 7
       Bias and leakage currents ok - attachment 8
       Merger ok - attachment 9

20.21  BRIKEN run 87 and BigRIPS run 1098 stopping daqs because beam about to be switched back on
       AIDA on file R1_1265

20.22  Beam on
       BigRIPS starts run 1099
       BRIKEN starts run  88
       Aida on file R1_1266

21.18  Online analysis: DSSSD HEC x versus y spectra - see attachment 10
       Zero'd SAS spectra

21.19  BRIKEN run 88 and BigRIPS run 1099 stopping daqs 
       AIDA on file R1_1278

21.20  BigRIPS starts run 1100
       BRIKEN starts run  89
       Aida on file R1_1279

22.05  System wide checks all ok
       FEE temperatures all ok - attachment 11
       Good events statistics - attachment 12
       Bias and leakage currents ok - attachment 13
       Merger ok - attachment 14

22.17  Analysis of file RIBF127R1/R1_1290 - dead time < 0.4% - see attachment 15

22.18  BRIKEN run 89 and BigRIPS run 1100 stopping daqs 
       AIDA on file R1_1291

22.20  BigRIPS starts run 1101
       BRIKEN starts run  90
       Aida on file R1_1292

22.24  Online analysis: DSSSD HEC x versus y spectra - see attachment 16
       Zero'd SAS spectra

23.20  BRIKEN run 90 and BigRIPS run 1101 stopping daqs 
       AIDA on file R1_1305

23.22  BigRIPS starts run 1102
       BRIKEN starts run  91
       Aida on file R1_1305

23.48  System wide checks all ok
       FEE temperatures all ok - attachment 17
       Good events statistics - attachment 18
       Bias and leakage currents ok - attachment 19
       Merger ok - attachment 20
Entry  Sat May 2 02:04:50 2015, JA, TD, AE, CG, VP & JLT, Saturday 2 May 2015 50x

0055: Degrader changed. 1mm removed.

            1.4mm total

0537: R38 ends nnaida11 slow comparator 0x20 -> 0x40 LEC fast comparator -> 0xff

0530: R39 10.00 AIDA status screenshots

1159: R39 ends HEC fast comparator 0x2 -> 0xf (all FEE64s)

1202: R40 nnaida8 no SYNCS see attachments 19-22

           VFEP toggled merge pause state check merge activity spinner accelerates nnaida8 GO Electronics (from select required function menu of Run Control) nnaida8 started OK toggled merge pause state

1500: Attachments 23-28 show waveforms, low E and high E spectra for nnaida6+11.

           Increasing fast comp threshold to 0xf shifts position of energy cut off away from zero point, as expected, but does not remove low E peak around ch 32500.

1530: Increase bias to +150V on det 1 (ch 1). Attachments 29-34 show waveforms, low E and high E spectra with increased bias voltage. Diode link threshold at 0x23

           V high E events in .L spectra show reduction over those before bias increase.

           In .H spectra from both nnaida6+11, increased bias reduced threshold. Finite threshold still seen on n-side however (greater than on p-side).

           All references/comparisons in this ELog entry re. nnaida11 are w.r.t. to spectra/waveforms seen at this 'standard' setting.

1544: Changed diode link thresh on nnaida11 to 0x11. Attachments 35-37.

           Saved run R42.

           High energy spectra saw threshold increase of ~700ch, but no other changed were observed.

1555: Change nnaida11 diode link thresh to 0x29. Saved run R43. Attachments 38-40.

            Saw reduction in high E threshold by ~700 ch but no other obvious changes.

1610: Changed diode link thresh on nnaida11 to 0x14 and preAmp reference to 0x20. Saved run R44. Attachments 41-44.

            Waveforms look much worse, more flatlining that previously seen.

            High E spectra show marginally higher threshold. No significant change in low E spectra.

1630: Changed nnaida11 preAmp ref to 0x40 and diode link thresh to 0x38. Saved run R45. Attachments 45-47

           Waveforms look similar to original setting. High E spectra show no significant change.

          Low E spectra have small peak near 32k and v. wide stretching to top of FSR.

1643: Changed nnaida11 preAmp ref to 0x60 and diode link thresh to 0x58. Saved run R46. Attachments 48-50.

            Waveforms settled on ~7500 much more frequently, with most signals beginning and returning to ~7500.

            High E spectra thresh at ~33500 as it was with 'standard' setting.

            Low E spectra show vast reduction in high E rubbish, looking more like an exponential decay compared to previous wide 'mound'.

            Big improvement over previous settings. Will try all n-side FEE with these settings later.

1700: Changed nnaida11 diode link thresh to 0x5a. Saved run R47 (~2hr run during tea break and other work).

           0x60 HEC fast comparator threshold (all FEE64s except nnaida30) 0x02.

19.04 R47 ends Change BigRIPS setting from 110Zr to 88Ge

19.15 R48 Pulser walkthough BNC PB-4 amplitude 90,000-10,000 @ 10,000 step, freq 25Hz.

            nnaida7 showed no SYNCS.

            Used procedure described above (1202) - nnaida7 recovered OK

19.28 Started run R48 - pulser walkthrough.

            R48 ends OK..

1944: Changed fast comparator threshold on LEC for...

  • nnaida 1: 0xb -> 0x30
  • nnaida2: 0x18 -> 0x30
  • nnaida6: 0x10 -> 0x30
  • nnaida10: 0x10 -> 0x30
  • nnaida13: 0x12 -> 0x30
  • nnaida14: 0x14 -> 0x30

          Change, for all n-side FEE64s, preAmp reference from previous value to 0x60 and diode link thresh to 0x5a.

          Saved settings to /MIDAS/DB/EXPERIMENTS/AIDA/2015Apr30-diode-mod3

 (around) 20:30 pm: started data taking for 88Ge setting. First runs are with empty H target at MINOS/DALI for transmission measurement.

                                     Degrader also changed during run R50. Initially: 0.6mm + 0.4mm + (0.3 mm?) + 1mm. The 1mm degrader was removed after ~ 1hr of run.

 

2300: NNAIDA#11: changed Fast Comparator of Low Energy channels to 0x40. It was set at 0xFF, and no discriminator data being produced. Rate of ~12 kHz for discriminator events in NNAIDA#11 for this setting (comparable to other FEE64s).

 

         RUN: R51_* started after adjusting the NNAIDA11 setting

Entry  Sat May 2 14:41:20 2015, JA, TD, AE, CG, VP & JLT, MUSIC detector signals MUSIC_lowE.pngMUSIC_highE_log.png

Adjusted setting of MUSIC detector signals during break

 

MUSIC ---> (16dB att)  --> Lemo split --->  J13

                                                                  ---> J14

 

MUSIC ---> (16dB att)  --> Lemo split --->  J15

                                                                   ---> J16

 

MUSIC ---> (BNC att.)  --> Lemo split --->  J17

                                                                  ---> J18

 

MUSIC ---> (BNC att.)  --> Lemo split --->  J19

                                                                  ---> J20

 

A couple seem to bee too much attenuated.  Spectra for 88Ge setting attached.

Entry  Sun May 3 01:32:18 2015, JA, TD, AE, CG, VP & JLT, Sunday 3 May 2015 
09.19 Beam off on arrival, now back
      R51 running

      BigRIPS 88Ge setting

      Degraders 4mm Al + 1.3mm Al variable, MINOS target filled 

      FEE64 temperatures OK
      Julabo FL11006 recirculating chiller OK, operating temp +20 deg C
      Detector HV OK
      AIDA SYNC/PAUSE/RESUME rates OK

10.05 R51 ends

10.51 Change BigRIPS setting to 94Se

      OR MACB triggers from nnaida1, 2, 6, 10, 13 & 14 input EURICA trigger module input #7
      All nnaida1, 2, 6, 10, 13 & 14 fast discriminators currently disabled.

11.33 R53

      BigRIPS 94Se setting

      c. 15.54 EURICA DAQ LUPO timestamp reset

14.00 Degrader changed t0 4mm fixed + 0.7mm variable (Al)
Entry  Mon May 4 18:09:42 2015, JA, TD, AE, CG & JLT, Tuesday 5 May 2015 
02.09 R61 starts
      nnaida1, 2, 6, 10, 13 & 14 fast discs disabled

10.10 R61 ends
      background + pulser run
ELOG V3.1.3-7933898