ID Date Authordown Subject
  422   Fri Nov 4 08:10:46 2016 CGPb bricks upstream

Stack of Pb bricks upstream of AIDA+BRIKEN to protect from incoming beam.

  456   Wed Nov 16 03:13:25 2016 CGWednesday 16 November
11.30 Leakage currents and FEE temps ok (attachments 1-3)
  460   Sun Nov 20 02:24:53 2016 CGSunday 20 November
11.20 FEE temps and bias leakages ok (attachments 1-3).
      However, stats screen shows no SYNCs being issued and no "Good Events" (attachment 4).
      Stop and Go had no effect -> SYNCs output but SYNC related system checks failed (attachment 5+6).
      Carried out Reset and Setup and returned to normal operation. 

17.20 FEE temps and bias leakage currents ok (attachments 7+8).
      DAQ still running ok.
      BigRIPS tuning through ZDS for next expt.
      15cm Pb upstream of BRIKEN. 
  531   Wed Feb 1 16:54:40 2017 CGNP1306 run summary
  855   Wed Jun 19 08:30:36 2019 CGWed Jun 19 16.00 - end of run
16.10 DAQ continues as before.
      Merger rate ~700k items/s and ~5.3Kb/s
      Individual FEE rates remain roughly constant
      Temps and bias ok.

17.33 DTAS run 149 stopped

      DTAS run 150 started
      BigRIPS run 2057 
      AIDA run ~1.3GB in R2_865

19.30 DTAS run 152 stopped

      DTAS run 153 started
      BigRIPS run 2058
      AIDA run ~1.8GB in R2_884

      Temps and bias ok.
      Rates in merger, tape server and individual FEEs as before.
      
19.50 In looking at previous data, AIDA is seeing ~30% fewer implants that what is observed at F11Plastic
      Beam spot is not central on AIDA and some ions will be lost over sides of DSSD, but does not look
      like it would account for a ~30% reduction
      AIDA implant profile shown in Attachment 1 - from file
merged_a2_687_708_br190618_2315_100Sn118_121bgrun2049_sw_Tsi.root

21.01 Beam stopped.
      DTAS run 155 stopped.
      AIDA file R2_899.

      DTAS and AIDA will continue with background measurements.

21.04 DTAS run 156 started
      AIDA run R2_899
      
      Chiller is full.
      Lead bricks upstream of moderator.

      That's all folks.
  860   Fri Jun 21 09:57:15 2019 CGB2F network cables
AIDA network cables in B2F disconnected, looped and cable tied to hand rail, out of the way.

Labelled AIDA1/AIDA2 according to which network switch they belong to.
  914   Wed Oct 25 12:42:18 2023 CCAIDA test
Started AIDA at 1:41 pm, writing to disk: file BG18OCT2/R6, stopped at  17:26
pict1: HV supply
pict2: temperature
pict3: spectrum in stat, layout2
pict4: statistics
  767   Tue Nov 6 00:13:00 2018 CB. PJW, TD, OH, CATuesday 6 November
09:13 DAQ still running R18_1
      Merger and TapeServer OK
      System wide checks ok, except for nnaida7 ADC calibration failure
      Temperature OK (attachment 1)
      Bias and leakage current OK (attachment 2)
      Good event statistics OK (attachment 3)
      
11.20 1.8.W spectra - attachment 4 & 5

      analysis file Oct18/R18_1 - attachment 6


15:35 Inserted AIDA snout in BRIKEN. Lowered threshold to 100 keV. Rate looks OK.
      R19 pulser walkthrough 90 0000 ... 10 0000 (10 000 steps) - attach 7&8
      R21 pulser walkthrough 12 000 ... 2 000 (2 000 steps)
      Pulser back to 90 000. Frequency set to zero.
      R22 background run starts. 
      Rates OK (attach 9)

16.07 Raised slow comparator threshold to 100 from 10
      DAQ start (file R23)
      Merger and TapeServer ok, no error messages in terminal
      Good event statistics ok (attachment 10)
      System wide checks ok, except nnaida7 failing ADC calibration 
      FEE64 temperatures ok (attachment 11)
      bias/leakage currents ok (attachment 12)

16.58 Repeat pulser walkthroughs with additional ASIC check loads

      R25 pulser walkthrough 90 0000 ... 10 0000 (10 000 steps) - attach 7&8
      R26 pulser walkthrough 12 000 ... 2 000 (2 000 steps)


18.26 DAQ start - file Oct18/R27
      alpha background
      LEC/MEC slow comparator 0xa -> 0x64
      BNC PB-4 pulser OFF @ 18.29!

18.34 1.8.W spectra 20us FSR - see attachments 13 & 14
  584   Mon May 22 08:43:10 2017 CB, TD, PatrickMonday 22 May
> 10.35 DAQ continues OK 
>       Current file May2017/R19_253
> 
>       ASIC settings 2017Mar27-16.12.39 *except*
>       slow comparator 0x0a
>       LEC/MEC fast comparator 0xff
> 
>       Detector biases; leakage currents OK - see attachment 1
>       FEE64 Temperatures OK - see attachment 2
>       good events; statistics OK - see attachment 3
>       No warnings/errors reported by merger since DAQ GO - see attachment 4
> 
>       *all* system wide checks OK *except* ADC calibration of nnaida7
> 
> 10.38 E17/F11 ambient temperature +24.7 deg C, RH 40.7%, d.p. +10.7 deg C
>       Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)
> 
> 10.39 DAQ stops OK
>       Current file May2017/R19_253
>       Deleted R19*
> 
>       ASIC settings 2017Mar27-16.12.39 *except*
>       slow comparator 0x0a
>       LEC/MEC fast comparator 0xff
> 
>       BNC PB-4 settings 
>       Amplitude 90,000
>       x5 attenuator IN
>       t_r 50ns, tau_d 50us
>       polarity -
>       rate 250Hz range
>       delay MIN, tail pulse
> 
>       Check ASIC Control x2
>       Zero all spectra
> 
> 10.47 DAQ starts
>       File May2017/R20
>       Ras Pi system console logs for USB0 attached - see attachment 5
>       Analysed R20_0. All appears to be working fine. Highest dead time ~0.3%  - see attachment 6
>       Off-line analysis of R20_0 confirms ADC data and discriminator are synchronised - see attachments 7,8,9

Concerning nnaida7 and the failure of calibration..... have you tried to calibrate it using the FADC Align  & Control ?
If it still fails then there is a question about the quality of the clock signal being received by this module.
I will log through and look at the console log for any anomolies but please try and calibrate.
Let me know if you wish me to take remote control.
  585   Mon May 22 08:53:42 2017 CB, TD, PatrickMonday 22 May
> Concerning nnaida7 and the failure of calibration..... have you tried to calibrate it using the FADC Align  & Control ?
> If it still fails then there is a question about the quality of the clock signal being received by this module.
> I will log through and look at the console log for any anomolies but please try and calibrate.
> Let me know if you wish me to take remote control.

Yes - there have been multiple attempts to re-calibrate nnaida7 ADCs

Note *all* ADCs were calibrated before the most recent FEE64 power cycle.

It is not unusual for ~1-3 FEE64s to fail ADC calibration tests following
a power cycle - there is no obvious pattern to which ADCs fail or which (if
any) will be successfully re-calibrated.
  633   Mon Jun 5 08:43:24 2017 CB, TD, PW, DKMonday 5 June 16:00 to 24:00
16:43  Leakage currents OK - see attachment 1
       FEE64 temperatures OK - see attachment 2
       Statistics show rate increase in nnaida13, 21. Reason unclear - see attachment 3
       Waveforms for nnaida13 look OK - see attachment 4
       No merger errors - see attachment 5

       All system-wide checks OK, except ADC calibration for nnaida1, 8, 19

       Current writing speed 10 Mb/s, 30% higher than before - see attachment 6

       Current AIDA run /TapeData/RIBF128/R5_6

17:50  Leakage currents OK - see attachment 7
       FEE64 temperatures OK - see attachment 8
       Statistics as before - see attachment 9
       No merger errors - see attachment 10

       Analysed /TapeData/RIBF128/R5_25 - attachment 11
       Rate high as expected, no timewarps, negligible dead time, idle time in agreement with dead time


19.34 DAQ starts
      AIDA file RIBF128/R5_58
      BigRIPS file 2088
      BRIKEN file 0

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency 25Hz range
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Mar27-16.12.39

      130Ag BigRIPS setting, degrader setting - fixed 6mm (in F11) + variable all out = 6 mm Al

19.38 DAQ stops
      AIDA file RIBF128/R5_59

20.04 Detector biases & leakage currents OK - see attachments 12 & 13
      FEE64 Temperatures OK - see attachment 14
      good events statistics OK - see attachment 15
      No warnings/errors reported by merger since DAQ GO - see attachment 16

      *all* system wide checks OK (zero fails)

      analysis of file RIBF128/R5_67 - dead time < 0.5%, no timewarps, 2 MBS timewarps - see attachment 17

20.14 online analysis - ADC & disc data synchronised - see attachment 18

20.28 layout spectra - attachments 19-28
      no pulser peaks observed in p+n junction strip FEE64s - to be checked when access is obtained

21.01 E17/F11 ambient temperature +24.3 deg C, RH 41.0%, d.p. +10.0 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)

21.07 DAQ starts
      AIDA file RIBF128/R5_85
      BigRIPS file 3003
      BRIKEN file 3

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency 25Hz range
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Mar27-16.12.39

      130Ag BigRIPS setting, degrader setting - fixed 2mm Al at F7, fixed 3mm + variable 1mm Al at F11 = 6 mm Al

21.15 DAQ stop
      AIDA file RIBF128/R5_88

21.31 DAQ starts
      AIDA file RIBF128/R5_93
      BigRIPS file 3004
      BRIKEN file 4

      DAQ sync check OK

      All SortSas spectra & histograms zero'd

21.45 DAQ stop
      AIDA file RIBF128/R5_98

21.31 DAQ starts
      AIDA file RIBF128/R5_98
      BigRIPS file 3005
      BRIKEN file 5

22.00 Detector biases & leakage currents OK - see attachments 29 & 30
      FEE64 Temperatures OK - see attachment 31
      good events statistics OK - see attachment 32
      No warnings/errors reported by merger since DAQ GO - see attachment 33

      *all* system wide checks OK (zero fails)

      analysis of file RIBF128/R5_67 - dead time < 0.7%, no timewarps, 1 MBS timewarp - see attachment 34

20.14 online analysis - ADC & disc data synchronised - see attachment 35

22.08 offline analysis file RIBF128/R5_106 - rates

 *** DSSSD # 1 count:    403476 old count:    321247 dt:      1.03 s  LEC rate:  80056.89 Hz
 *** DSSSD # 2 count:    559009 old count:    446189 dt:      1.03 s  LEC rate: 109839.82 Hz
 *** DSSSD # 3 count:    262328 old count:    206447 dt:      1.03 s  LEC rate:  54404.88 Hz
 *** DSSSD # 4 count:    159863 old count:    127957 dt:      1.03 s  LEC rate:  31063.19 Hz
 *** DSSSD # 5 count:     66930 old count:     52878 dt:      1.03 s  LEC rate:  13680.81 Hz
 *** DSSSD # 6 count:    109482 old count:     87597 dt:      1.03 s  LEC rate:  21306.90 Hz
 *** DSSSD # 1 count:         1 old count:         1 dt:      1.03 s  HEC rate:      0.00 Hz
 *** DSSSD # 2 count:        72 old count:        56 dt:      1.03 s  HEC rate:     15.58 Hz
 *** DSSSD # 3 count:       278 old count:       214 dt:      1.03 s  HEC rate:     62.31 Hz
 *** DSSSD # 4 count:       218 old count:       178 dt:      1.03 s  HEC rate:     38.94 Hz
 *** DSSSD # 5 count:        22 old count:        18 dt:      1.03 s  HEC rate:      3.89 Hz
 *** DSSSD # 6 count:        27 old count:        15 dt:      1.03 s  HEC rate:     11.68 Hz

     online analysis - per DSSSD x versus y (HEC) - attachment 36
                       per DSSSD E_p versus E_n (HEC) - attachment 37

     offline analysis file RIBF128/R5_106 - ADC & disc data synchronised - attachment 38
     N.B. pulser to p+n strips appears to be low amplitude - absence of secondary ADC peak, low multiplicity for
disc data


23.01 DAQ stop
      AIDA file RIBF128/R5_124

23.02 DAQ starts
      AIDA file RIBF128/R5_124
      BigRIPS file 3006
      BRIKEN file 6

00:01 DAQ stop
      AIDA file R5_145
  571   Thu May 18 02:16:05 2017 CB, TD, OHThursday 18 May
10.10 DAQ continues OK 
      Current file May2017/R10_346

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      Detector biases &amp; leakage currents OK - see attachments 1 &amp; 2
      FEE64 Temperatures OK - see attachment 3
      &#39;good events&#39; statistics OK - see attachment 4
      No warnings/errors reported by merger since DAQ GO - see attachment 5

      *all* system wide checks OK

10.20 E17/F11 ambient temperature +24.3 deg C, RH 39.0%, d.p. +9.3 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%)

10.25 DAQ stops OK
      Current file May2017/R13_294

      Files May2017/R13* *deleted* 

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate 250Hz range
      delay MIN, tail pulse

      Check ASIC Control x2
      Zero all spectra

10.30 DAQ starts
      File May2017/R14

17.34 DAQ stops OK
      Current file May2017/R14_113

      Files May2017/R14* *deleted* 

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate 250Hz range
      delay MIN, tail pulse

      Check ASIC Control
      Zero all spectra

      *all* system wide checks OK

17.37 DAQ starts
      File May2017/R15

      Detector biases & leakage currents OK - see attachments 6 & 7
      FEE64 Temperatures OK - see attachment 8
      good events statistics - higher rates for several FEE64s - due to addition of system consoles? - see
attachment 9
      No warnings/errors reported by merger since DAQ GO - ~ +30% data rate - see attachment 10
  577   Fri May 19 09:55:54 2017 CB, TD, OHLogs from Raspberry Pi System Console
17:50 Rebooted Server and FEE64. Connected remotely to Ras Pi System Console. Repeated USB connection procedure. All 24 modules can now be seen. Sample log attached
  578   Fri May 19 10:16:27 2017 CB, TD, OHDAQ sever rebooted
18:16 Finished rebooting DAQ server and acquisition. All OK, except ADC calibration on #7.
      Started run May2017/R16
      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      Detector biases; leakage currents OK
      FEE64 Temperatures OK
      good events; statistics OK
      No warnings/errors reported by merger since DAQ GO
  580   Sun May 21 02:54:22 2017 CB, TD, OHSunday 21 May
10.54 DAQ continues OK 
      Current file May2017/R18_208

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      Detector biases; leakage currents OK - see attachment 1
      FEE64 Temperatures OK - see attachment 2
      good events; statistics OK - see attachment 3
      No warnings/errors reported by merger since DAQ GO - see attachment 4

      *all* system wide checks OK *except* ADC calibration of nnaida7

10.58 E17/F11 ambient temperature +24.6 deg C, RH 40.5%, d.p. +10.3 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 50%)

11.00 DAQ stops OK
      Current file May2017/R18_208
      Deleted R17*

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate 250Hz range
      delay MIN, tail pulse

      Check ASIC Control x2
      Zero all spectra

11.16 DAQ starts
      File May2017/R19
      Ras Pi system console logs for USB0 attached - see attachment 5

18.26 DAQ continues OK 
      Current file May2017/R19_77

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      Detector biases & leakage currents OK - see attachment 6 & 7
      FEE64 Temperatures OK - see attachment 8
      good events statistics OK - see attachment 9
      No warnings/errors reported by merger since DAQ GO - see attachment 10

      *all* system wide checks OK *except* ADC calibration of nnaida7

18.30 E17/F11 ambient temperature +24.7 deg C, RH 38.9%, d.p. +9.8 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 60%)
  603   Sat May 27 05:38:39 2017 CB, TD, OHSaturday 27 May
13.38 DAQ continues OK 
      Current file May2017/R27_49

      Detector biases; leakage currents OK - see attachment 1
      FEE64 Temperatures OK - see attachment 2
      good events; statistics OK - see attachment 3
      No warnings/errors reported by merger since DAQ GO - see attachment 4

      *all* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (=6)

10.30 E17/F11 ambient temperature +23.8 deg C, RH 40.9%, d.p. +9.6 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)

13:40 Stopped DAQ running on alpha background calibration run
      Current file May2017/R49
      tyUSB7 log (nnaida1) attached

13:43 Detector bias OFF

14.16 DAQ starts
      File May2017/R28

      Pulser walkthrough

      BNC PB-4
      Amplitude 90,000-10,000 @ 10,000 step
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency 250Hz range
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0x20

      See attachment 6

14.25 DAQ stop OK

14.27 DAQ starts
      File May2017/R29

      Pulser walkthrough

      BNC PB-4
      Amplitude 12,000-2,000 @ 2,000 step

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0x20

      See attachment 7

14.34 DAQ stop OK

14.38 DAQ starts
      File May2017/R30

      Background, pulser OFF

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0x20

15.16 DAQ stop OK
      May2017/R30_8

16:17 DAQ starts okay 
      current run May2017/R31_0
      leakage current and temperature fine
      see attachment 8 and 9


  
  590   Tue May 23 03:01:52 2017 CB, TD, NN, OHTuesday 23 May
11.02 DAQ continues OK 
      Current file May2017/R22_192

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      Detector biases; leakage currents OK - see attachment 1
      FEE64 Temperatures OK - see attachment 2
      good events; statistics OK - see attachment 3
      No warnings/errors reported by merger since DAQ GO - see attachment 4

      *all* system wide checks OK *except* ADC calibration of nnaida7, 9, 13, 22

11.05 E17/F11 ambient temperature +23.9 deg C, RH 41.3%, d.p. +10.5 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)
      Voltage on BRIKEN 3He tubes now ON

11.07 DAQ stops OK
      Current file May2017/R22_193
      Deleted R22*

      Loaded new firmware to test ADC calibration clock issues - attachment 6
      Firmware loaded - attachment 7
      Restarted Firefox - temperatures still look fine - see attachment 7
      System wide checks - nnaida1 clock failed, other clocks OK - attachment 8
                           ADC calibration failed for nnaida1, 3, 21 - attachment 9
                           Other system wide checks OK
      Statistics - few SYNC pulses received by nnaida 9 and 12. See attachement 10


      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate 250Hz range
      delay MIN, tail pulse

13.10 DAQ started
      File May2017/R23
      Run Control GO - Tape server GO - Merger Unpause - Merger data rate 0, but Tapeserver sees events and
Merger links are green - see attachment 11
      Start ASIC data readout - Merger sees data rate of ~800 000 items/s (OK!)
      FPGA clock monitor page (http://localhost:8015/AIDA/MONLOCK/MONLOCK.tml) screenshot attached (12)

      Ras Pi system console logs for USB0 attached - see attachment 13

18.00 Merger found stopped. We were unable to stop the DAQ. Powered off FEEs. Power on, only eight were supplied
power. 
      Blown fuse in USB bias controller - replaced.
      Power ON the FEEs. All 24 FEEs mounted.
       
      Current file May2017/R23_27 - stopped at 15:27

18.31 nnaida20 rebooted on its own
      
      
18.37 E17/F11 ambient temperature +23.7 deg C, RH 39.8%, d.p. +9.5 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 70%)

18.41 Restarted DAQ 
      
      Detector biases; leakage currents OK - see attachment 14
      System wide checks - OK *except* several modules failed ADC calibration - see attachment 15
      FPGA clock monitor screenshot attached (16)

18.51 Stopped DAQ. Flashed previous firmware on FEEs (FEE_GF_Feb16_24.bin)
      Power cycled. All 24 modules mounted OK.
      nnaida22 rebooted on its own
      Time out Setting up the DAQ on nnaida2 and others. Power cycled again.
      nnaida17 rebooted on its own.
      Time out Setting up the DAQ on nnaida1. Power cycled again.
      
19.56 DAQ finally started.
      Current run R24

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x64
      LEC/MEC fast comparator 0x20

      *all* system wide checks OK *except* some ADC calibrations and clock on nnaida1
      Temperatures OK
      Statistics OK
      
      Logs for nnaida 17,20,22 attached

      
  834   Sun Jun 9 03:43:00 2019 CB, TD, CG, OHSunday 9 June
11.43 DAQ continues OK

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 1 & 2

      FEE64 temperatures OK - see attachment 3

      good event statistics - DSSSD#1 noisy - see attachments 4
      10x <20k, 19x <50k, 20x <100k cf. 13x <20k 22x < 50k 24x <100k yesterday afternoon
      Merger OK - see attachment 5
      TapeServer OK - see attachment 6

      p+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 8
      1.8.L spectra - see attachments 9 & 10
      nnaida1 76 ch FWHM, nnaida3 83 ch FWHM - cf. 104 & 111 ch FWHM respectively yesterday aftenoon

      FEE64 Rate spectra - attachment 11
      Check ASIC Control *all* FEE64s & ASICs x2
      FEE64 Rate spectra - attachment 12

11.46 ADC data synchronised - attachment 13

      p+n FEE64s 1.8.W spectra 20us, 200us, 2ms & 20ms FSR - DSSSD#1 noisy - attachments 14-17

      Check ASIC Control *all* FEE64s & ASICs x2   - no change DSSSD#1 noise

      DSSSD#1 bias cycle - no change DSSSD#1 noise

      press Kapton Samtec into adaptor PCBs - reduces DSSSD#1 noise to previous levels

12.17 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
      Water level c. 60% M-H markers
      Temperature +25.3 deg C dew point +7.6 deg C RH 32.2%

12.23 temporary fix for DSSSD#1 noise? check by
      zero all histograms
      zero DAQ stats

      ...and wait

      good event statistics - DSSSD#1 noisy - see attachments 18
      11x <20k, 21x <50k, 23x <100k

13.30 DAQ stop

      OH replaces LH kapton PCB for DSSSD#1 n+n FEE64s 20 & 23 which we suspect may be damaged due to twist
      placed on kpaton cables by weight of all ribbon cables + cu foil shielding - particularly pronounced
      for nnaida20

      good event statistics - OK - see attachments 19
      7x <20k, 20x <50k, 24x <100k
      p+n FEE64s 1.8.W spectra 20us FSR - OK - attachment 20

      good event statistics - OK - see attachment 21
      7x <20k, 22x <50k, 24x <100k
      p+n FEE64s 1.8.W spectra 20us FSR - OK - attachment 22

17.50 detector biases & leakage currents OK - see attachments 23 & 24

      FEE64 temperatures OK - see attachment 25

      good event statistics - OK - see attachments 26
      9x <20k, 21x < 50k, 24x <100k cf. 10x <20k, 19x <50k, 20x <100k earlier today

      p+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 7
      n+n FEE64s 1.8.W spectra 20us FSR - DSSSD#1 noisy - attachment 8

17.57 all histograms zero'd
      DAQ stats zero'd
  582   Mon May 22 02:34:51 2017 CB, TDMonday 22 May
10.35 DAQ continues OK 
      Current file May2017/R19_253

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      Detector biases; leakage currents OK - see attachment 1
      FEE64 Temperatures OK - see attachment 2
      good events; statistics OK - see attachment 3
      No warnings/errors reported by merger since DAQ GO - see attachment 4

      *all* system wide checks OK *except* ADC calibration of nnaida7

10.38 E17/F11 ambient temperature +24.7 deg C, RH 40.7%, d.p. +10.7 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)

10.39 DAQ stops OK
      Current file May2017/R19_253
      Deleted R19*

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate 250Hz range
      delay MIN, tail pulse

      Check ASIC Control x2
      Zero all spectra

10.47 DAQ starts
      File May2017/R20
      Ras Pi system console logs for USB0 attached - see attachment 5
      Analysed R20_0. All appears to be working fine. Highest dead time ~0.3%  - see attachment 6
      Off-line analysis of R20_0 confirms ADC data and discriminator are synchronised - see attachments 7,8,9

16.55 DAQ continues OK 
      Current file May2017/R20_71

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      Detector biases; leakage currents OK - see attachment 10
      FEE64 Temperatures OK - see attachment 12
      good events; statistics OK - see attachment 13
      No warnings/errors reported by merger since DAQ GO - see attachment 14

      *all* system wide checks OK *except* ADC calibration of nnaida7

17.02 E17/F11 ambient temperature +24.0 deg C, RH 40.3%, d.p. +9.7 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)

17.03 DAQ stops OK
      Current file May2017/R20_73
      Deleted R20*

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate 250Hz range
      delay MIN, tail pulse

      Check ASIC Control x2
      Zero all spectra

17.08 DAQ starts
      File May2017/R21
      Raspberry Pi system console logs for USB0 attached - see attachment 15

17.50 Changed permissions of options files /MIDAS/DB/EXPERIMENTS/Options/nnaida<n>/CONTENTS 
      from mode 644 to mode 777

18.09 DAQ stops OK
      Current file May2017/R21_11
      Deleted R21*

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x0a
      LEC/MEC fast comparator 0xff

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate 250Hz range
      delay MIN, tail pulse

      DAQ RESET
      DAQ SETUP

      *all* system wide checks OK *except* nnaida7, 9, 13 and 22 ADC calibration failed

      Check ASIC Control x2

17.08 DAQ starts
      File May2017/R22
      Raspberry Pi system console logs for USB0 attached - see attachment 16
  592   Wed May 24 03:20:35 2017 CB, TDWednesday 24 May
11.00 Merger found stopped at 03:27 - current file R24_7
      Restarted Merge. Stopped Run control. Run control stopped, got error message - see attachment 1
      nnaida 1-16 show error, nnaida 17-24 show stopped - see attachment 2
      No new panics in Raspberry Pi console monitor
      
13.10 Found FEE64 #1-16 power off - fuse failed in section #1 of USB-controlled ac mains relay - this fuse was
replaced yesterday

13.15 Closed file May2017/R24
      Closed DAQ, Merger & Tapeserver browsers & terminated terminal servers

13.22 Replaced 250V/16A/fast fuse *and* carrier in section #1 of USB-controlled ac mains relay - when fuse fails
there is some solder leaks from base of fuse into carrier 

13.25 Switch on relays # 1-8 at ~10s intervals
      Count 24 authenticated mount requests 13.26.02 to 13.27.30 - see attachment 5

13.31 E17/F11 ambient temperature +24.4 deg C, RH 40.0%, d.p. +9.9 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%)
      Detector biases & leakage currents OK - see attachments 3 & 4

13.39 Checked Pi-monitor - no further panics detected in FEE64 system logs

13.54 Turned on DAQ.
      FEE temperatures OK - see attach 6
      System wide checks ok *except* nnaida1 clock status (status 6) and nnaida1,7,8,9,13,21,24
      Manually recalibrated all *except* nnaida21

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x64
      LEC/MEC fast comparator 0x20

      BNC PB-4 settings 
      Amplitude 90,000
      x5 attenuator IN
      t_r 50ns, tau_d 50us
      polarity -
      rate 250Hz range
      delay MIN, tail pulse

      Current file R26

      ttyUSB0 attached - 7

14.04 good events statistics - see attachment 8