ID |
Date |
Author |
Subject |
|
784
|
Wed Nov 14 14:46:47 2018 |
CA | Thursday 15th November 00:00 - 08:00 | 23.58 DAQ continues ok (RIBF148/R5_505)
00.09 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 1)
good event statistics ok (attachment 2)
detector bias/leakage currents ok (attachment 3)
00.13 Merger and TapeServer run ok (attachment 4/5)
00.26 BRIKEN stop
AIDA file RIBF148/R5_509
00.29 BRIKEN start
AIDA file RIBF148/R5_510
BRIKEN run 70
BigRIPS run 1055
01.46 BRIKEN stop
AIDA file RIBF148/R5_523
01.49 BRIKEN start
AIDA file RIBF148/R5_523
BRIKEN run 71
BigRIPS run 1056
02.05 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 6)
good event statistics ok (attachment 7)
detector bias/leakage currents ok (attachment 8)
Merger and TapeServer still ok
02.34 XY hit patterns (attachment 9)
Ex vs Ey plots (attachment 10)
HEC spectra (attachment 11)
LEC spectra (attachment 12)
02.49 BRIKEN stop
AIDA file RIBF148/R5_534
02.50 BRIKEN start
AIDA file RIBF148/R5_534
BRIKEN run 72
BigRIPS run 1057
03.26 zeroed all spectra on online analysis and MIDASsort
04.04 BRIKEN stop
AIDA file RIBF148/R5_546
04.05 BRIKEN start
AIDA file RIBF148/R5_546
BRIKEN run 73
BigRIPS run 1058
04.10 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 13)
good event statistics ok (attachment 14)
detector bias/leakage currents ok (attachment 15)
Merger and TapeServer still ok
05.06 BRIKEN stop
AIDA file RIBF/R5_556
05.07 BRIKEN start
AIDA file RIBF148/R5_557
BRIKEN run 74
BigRIPS run 1059
05.08 First 200ms of decay correlations in all detectors
comparison between different Al degrader thickness values (3mm, 2.3mm, 2.8mm)
see attachments 16, 17 and 18 - graphs labelled with detector number and Al degrader thickness
06.09 BRIKEN stop
AIDA file RIBF148/R5_567
06.11 BRIKEN start
AIDA file RIBF148/R5_567
BRIKEN run 75
BigRIPS run 1060
06.16 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 19)
good event statistics ok (attachment 20)
detector bias/leakage currents ok (attachment 21)
Merger and TapeServer still ok
06.17 ASIC check load - ok
06.34 XY hit patterns (attachment 22)
Ex vs Ey plots (attachment 23)
HEC spectra (attachment 24)
LEC spectra (attachment 25)
07:12 BRIKEN stop
AIDA file RIBF148/R5_578
07:15 BRIKEN start
AIDA file RIBF148/R5_578
BRIKEN run 76
BigRIPS run 1061
|
|
783
|
Wed Nov 14 07:43:35 2018 |
TD | Wednesday 14 November 16:00 - 00:00 | 16.44 DAQ continues OK - AIDA file RIBF148/R5_432
165Pm BigRIPS setting + 2.8mm Al degrader
ASIC settings 2018Oct03-13.26.58
from c. 09.00 13.11.18
n+n FEE64s Ibias LF feedback 0xb -> 0xd
HEC fast comparator threshold 0x20 -> 0xff
BNC PB-4 pulser
Amplitude 90,000, x5 attenuator IN
t_r 50ns tau_d 1ms
Frequency min
Delay min
16.45 BigRIPS and BRIKEN stop
AIDA on File R5_432
16.48 BigRIPS starts run 1047
BRIKEN starts run 62
AIDA on file R5_433
16.57 online MIDASsort
m_p versus m_n (LEC) - attachment 1
x versus y (LEC) - attachment 2
E_p versus E_n (LEC) - attachment 3
m_p versus m_n (HEC) - attachment 4
x versus y (LEC) - attachment 5
E_p versus E_n (LEC) - attachment 6
all MIDASsort spectra zero'd
17.17 file RIBF128/R5_436 ADC data synchronised - attachment 7
analysis of RIBF148/R5_436 - attachment 8
max dead time nnaida19 ~0.25%
17.48 BigRIPS and BRIKEN stop
AIDA on File R5_443
17.48 BigRIPS starts run 1048
BRIKEN starts run 63
AIDA on file R5_443
18.45 BigRIPS and BRIKEN stop
AIDA on File R5_452
18.47 BigRIPS starts run 1049
BRIKEN starts run 64
AIDA on file R5_453
19.45 BigRIPS and BRIKEN stop
AIDA on File R5_463
19.46 BigRIPS starts run 1050
BRIKEN starts run 65
AIDA on file R5_463
20.49 BigRIPS and BRIKEN stop
AIDA on File R5_473
20.49 BigRIPS starts run 1051
BRIKEN starts run 66
AIDA on file R5_473
21.49 BigRIPS and BRIKEN stop
AIDA on File R5_483
ASIC check load
21.50 BigRIPS starts run 1052
BRIKEN starts run 67
AIDA on file R5_484
22.00 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 9 & 10
FEE64 temperatures OK - see attachment 11
good event statistics OK - see attachments 12
Merger OK - see attachment 13
TapeServer OK - see attachment 14
1x merger error message reported - trying to add items to Q index 12 (Link 12) but queue alreday full
per FEE64 Rate spectra - attachment 15
1.8.L spectra - attachments 16 & 17
nnaida1 pulser peak width 73 ch FWHM
1.8.H spectra - attachments 18 & 19
1.8.W spectra 20ms FSR - attachments 20 & 21
22.48 BigRIPS and BRIKEN stop
AIDA on File R5_493
22.50 BigRIPS starts run 1053
BRIKEN starts run 68
AIDA on file R5_493
23.24 BRIKEN DAQ found stopped - no sync check data - BigRIPS DAQ stopped and restarted
AIDA on File R5_499
23.26 BigRIPS starts run 1054
BRIKEN starts run 69
AIDA on file R5_499
|
|
782
|
Tue Nov 13 22:58:49 2018 |
OH | Wednesday 14 November 08:00 - 16:00 | 08:01 BigRips and BRIKEN stop
AIDA on file R5_346
08:02 BigRIPS starts run 1038
BRIKEN starts run 181114_0800_165Pm_053.dlt
AIDA on file R5_346
System wide checks ok * except nnaida 7 fails adc*
FEE Temps ok - attachment 1
Stats ok - attachment 2
Leakage currents ok - attachment 3
09:15 BigRIPS and BRIKEN stop
AIDA on file R5_358
09:16 BigRIPS starts run 1039
BRIKEN starts run 181114_0915_165Pm_054.dlt
AIDA on file R5_358
10:15 BigRIPS and BRIKEN stop
AIDA on file R5_368
10:15 BigRIPS starts run 1040
BRIKEN starts run 181114_1015_165Pm_055.dlt
AIDA on file R5_368
10:36 Analyser output of R5_368 - attachment 4
11:15 BugRIPS and BRIKEN stop
AIDA on File R5_378
11:16 BigRIPS starts run 1041
BRIKEN starts run
AIDA on file R5_378
08.03 DAQ continues - AIDA file RIBF148/R5_384
165Pm BigRIPS setting + 2.8mm Al degrader
ASIC settings 2018Oct03-13.26.58
from c. 09.00 13.11.18
n+n FEE64s Ibias LF feedback 0xb -> 0xd
HEC fast comparator threshold 0x20 -> 0xff
11.50 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 5 & 6
FEE64 temperatures OK - see attachment 7
good event & AIDA discriminator statistics OK - see attachments 8
Merger OK - see attachment 9
TapeServer OK - see attachment 10
no errors or warnings shown by server terminal sessions
per FEE64 Rate spectra - attachment 11
per FEE64 HitRate spectraq - attachment 12
1.8.L spectra - attachments 13-16
some evidence of pulser peak shifts, notably nnaida5, 11 & 15
nnaida1 pulser peak width 78 ch FWHM
1.8.H spectra - attachments 17 & 18
1.8.W spectra 20us FSR - attachments 19 & 23
1.8.W spectra 200us FSR - attachments 20 & 24
1.8.W spectra 2ms FSR - attachments 21 & 25
1.8.W spectra 20ms FSR - attachments 22 & 26
12.15 BugRIPS and BRIKEN stop
AIDA on File R5_388
12.16 BigRIPS starts run 1042
BRIKEN starts run 57
AIDA on file R5_388
12.22 *all* online and MIDASsort spectra zero'd
13:15 BigRIPS and BRIKEN stop
AIDA on file R5_398
13:16 BigRIPS starts run 1043
BRIKEN starts run58
AIDA on file R5_398
13:28 BigRIPS and BRIKEN stop
AIDA on file
beam halted to give access to F11 area
13.44 beam started
BigRIPS and BRIKEN start
AIDA file RIBF148/R5_402
BRIKEN run 59
BigRIPS run 1043
14.45 BRIKEN and BigRIPS stop
AIDA file RIBF148/R5_412
14.47 BRIKEN and BigRIPS start
AIDA file RIBF148/R5_413
BRIKEN run 60
BigRIPS run 1044
15.43 BRIKEN and BigRIPS stop
AIDA file RIBF148/R5_422
BRIKEN Resets the timestamp
15:44 BRIKEN and BigRIPS start
AIDA file RIBF148/R5_422
BRIKEN run 61
BigRIPS run 1045
16:02 System wide checks all ok *Except nnaida7 fails adc calibration*
Leakage currents ok - attachment 27
Stats ok - attachment 28
FEE Temps ok - attachment 29
16:27 First 200ms of decay correlations in DSSD 3 and 4 for the 3 degrader settings - attachment 30
DSSD 3 along the top row and DSSD 4 along the bottom.
Dgrader settings are 3mm, 2.3mm and 2.8mm
1 hour data in each |
|
781
|
Tue Nov 13 14:27:38 2018 |
LS | Wednesday 14 November 00:00 to 08:00 |
00:00 BRIKEN stop AIDA file RIBF148/R5_269
----------------------------------------------------------------
00.02 BRIKEN start
BRIKEN file 45
BigRIPS file 1030
AIDA file RIBF148/R5_270
00:07 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 1)
good event statistics ok (attachment 2)
detector bias/leakage currents ok (attachment 3)
01:01 BRIKEN stop
AIDA file RIBF148/R5_279
---------------------------------------------------------------
01.02 BRIKEN start
BRIKEN file 46
BigRIPS file 1031
AIDA file RIBF148/R5_279
02:00 BRIKEN stop
AIDA file RIBF148/R5_289
----------------------------------------------------------------
02.01 BRIKEN start
BRIKEN file 47
BigRIPS file 1032
AIDA file RIBF148/R5_289
03:00 BRIKEN stop
AIDA file RIBF148/R5_298
----------------------------------------------------------------
03.01 BRIKEN start
BRIKEN file 48
BigRIPS file 1033
AIDA file RIBF148/R5_298
03:04 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 4)
good event statistics ok (attachment 5)
detector bias/leakage currents ok (attachment 6)
04:00 BRIKEN stop
AIDA file RIBF148/R5_308
----------------------------------------------------------------
04.03 BRIKEN start
BRIKEN file 49
BigRIPS file 1034
AIDA file RIBF148/R5_308
05:00 BRIKEN stop
AIDA file RIBF148/R5_317
----------------------------------------------------------------
05.01 BRIKEN start
BRIKEN file 50
BigRIPS file 1035
AIDA file RIBF148/R5_318
06:00 BRIKEN stop
AIDA file RIBF148/R5_327
----------------------------------------------------------------
06.02 BRIKEN start
BRIKEN file 51
BigRIPS file 1036
AIDA file RIBF148/R5_327
06:07 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 7)
good event statistics ok (attachment 8)
detector bias/leakage currents ok (attachment 9)
07:00 BRIKEN stop
AIDA file RIBF148/R5_337
----------------------------------------------------------------
07.01 BRIKEN start
BRIKEN file 52
BigRIPS file 1037
AIDA file RIBF148/R5_327 |
|
780
|
Tue Nov 13 07:35:37 2018 |
CA, TD, OH | Tuesday 13 November 16:00 - 00:00 | 16.36 DAQ continues ok
AIDA file RIBF148/R5_198
BRIKEN file 37
BigRIPS file 1022
16.38 system wide checks ok, except nnaida7 ADC calibration failure
16.40 FEE64 temperatures ok (attachment 1)
16.42 good event statistics ok (attachment 2)
16.43 detector bias/leakage currents ok (attachment 3)
16.44 Merger and TapeServer ok, no error messages in terminals (attachments 4 and 5)
16.51 BRIKEN stop
AIDA file RIBF148/R5_200
17.00 Al Degrader thickness changed from 2.3mm to 2.8mm
The 2.3mm plate was previously assumed to be 2.5mm, but this has been corrected by measuring individual Al
plates with digital micrometer.
17.02 BRIKEN start
BRIKEN file 38
BigRIPS file 1023
AIDA file RIBF148/R5_202
18.01 BRIKEN stop
AIDA file RIBF148/R5_211
18.03 BRIKEN start
AIDA file RIBF148/R5_212
BRIKEN file 39
BigRIPS file 1024
18.08 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 6)
good event statistics ok (attachment 8)
detector bias/leakage currents ok (attachment 7)
18.11 Merger and TapeServer ok, no error messages in terminal
19.00 BRIKEN stop
AIDA file RIBF148/R5_221
19.02 BRIKEN start
AIDA file RIBF148/R5_221
BRIKEN file 40
BigRIPS file 1025
19.57 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 9)
good event statistics ok (attachment 10)
detector bias/leakage current ok (attachment 11)
20.01 BRIKEN stop
AIDA file RIBF148/R5_231
20.03 BRIKEN start
AIDA file RIBF148/R5_231
BRIKEN file 41
BigRIPS file 1026
21.01 BRIKEN stop
AIDA file RIBF148/R5_240
21.03 BRIKEN start
AIDA file RIBF148/R5_240
BRIKEN file 42
BigRIPS file 1027
21:18 system wide checks ok - except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 12)
good event statistics ok (attachment 13)
detector bias/leakage current ok (attachment 14)
22:01 BRIKEN stop
AIDA file RIBF148/R5_250
22:03 BRIKEN start
AIDA file RIBF148/R5_250
BRIKEN file 43
BigRIPS file 1028
22:10 system wide checks ok - except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 15)
good event statistics ok (attachment 16)
detector bias/leakage current ok (attachment 17)
Merger and TapeServer ok, no error messages in terminal
23:01 BRIKEN stop
AIDA file RIBF148/R5_260
23:03 BRIKEN start
AIDA file RIBF148/R5_260
BRIKEN file 44
BigRIPS file 1029
23.05 system wide checks ok - except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 18)
good event statistics ok (attachment 20)
detector bias/leakage current ok (attachment 19)
Merger and TapeServer ok, no error messages in terminal
|
|
779
|
Tue Nov 13 01:41:12 2018 |
TD | Offline analysis RIBF148/R5_130 to _139 |
10.41 FEE64 ADC data synchronised - attachment 1
LEC & HEC rates
*** DSSSD # 1 count: 5472516 old count: 5429034 dt: 1.05 s LEC rate: 41240.10 Hz
*** DSSSD # 2 count: 8599335 old count: 8530134 dt: 1.05 s LEC rate: 65633.05 Hz
*** DSSSD # 3 count: 1647116 old count: 1632316 dt: 1.05 s LEC rate: 14036.92 Hz
*** DSSSD # 4 count: 1825652 old count: 1810910 dt: 1.05 s LEC rate: 13981.91 Hz
*** DSSSD # 5 count: 384243 old count: 380911 dt: 1.05 s LEC rate: 3160.20 Hz
*** DSSSD # 6 count: 1246549 old count: 1236520 dt: 1.05 s LEC rate: 9511.91 Hz
*** DSSSD # 1 count: 31 old count: 30 dt: 1.05 s HEC rate: 0.95 Hz
*** DSSSD # 2 count: 258 old count: 255 dt: 1.05 s HEC rate: 2.85 Hz
*** DSSSD # 3 count: 1564 old count: 1548 dt: 1.05 s HEC rate: 15.18 Hz
*** DSSSD # 4 count: 3547 old count: 3512 dt: 1.05 s HEC rate: 33.20 Hz
*** DSSSD # 5 count: 680 old count: 676 dt: 1.05 s HEC rate: 3.79 Hz
*** DSSSD # 6 count: 10708 old count: 10629 dt: 1.05 s HEC rate: 74.93 Hz
Total heavy ion rate c. 130Hz of which c. 50% stop in DSSSDs # 1-6
Attachment 1 - *total* implant energy for AIDA DSSSD stack (5.6MeV/channel)
ch 1571 ~ 8.8GeV, ch 244 ~ 1.4GeV
Attachment 2 - implant energy for stopping DSSSD layer (20MeV/channel)
ch 390 ~ 7.8GeV |
|
778
|
Mon Nov 12 23:03:53 2018 |
TD | Tuesday 13 November 08:00 - 16:00 |
08.03 DAQ continues - AIDA file RIBF148/R5_117
165Pm BigRIPS setting + 3mm Al degrader
08.10 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event & AIDA discriminator statistics OK - see attachments 4 & 5
Merger OK - see attachment 6
TapeServer OK - see attachment 7
no errors or warnings shown by server terminal sessions
08.16 analysis of file RIBF148/R5_116 - see attachment 8
max deadtime nnaida19 ~ 1/420 ~ 0.25%, no timewarps
08.20 BRIKEN stop file
AIDA file RIBF148/R5_119
08.22 BRIKEN start file
AIDA file RIBF148/R5_120
BRIKEN file 28
BigRIPS file 1013
08.44 MIDASsort online spectra zero'd
08.56 1.8.W spectra 20ms FSR - see attachments 16 & 17
observe slow transient recovery per 33nF a.c. coupling but 'sawtooth' transients appear to be absent/rare
09.25 BRIKEN stop file
AIDA file RIBF148/R5_129
09.28 BRIKEN start file
AIDA file RIBF148/R5_130
BRIKEN file 29
BigRIPS file 1014
10.09 per FEE64 hit pattern (rate) - see attachment 9
per FEE64 1.8.L spectra - see attachments 10-13
nnaida1 pulser peak width 76 ch FWHM
per FEE64 1.8.H spectra - see attachments 14 & 15
10.25 BRIKEN stop file
AIDA file RIBF148/R5_139
10.27 Beam off - enter F11 to change BRIKEN amp gain
Correlation scaler pulser trigger frequency changed from 100Hz to 10Hz to minimise BigRIPS DAQ deadtime
Note - we will observe both edges of pulser trigger doubling the apparent frequency and the TDR data format
analyser reports each 16-bit word of the 48-bit correlation scaler so weill observe 6x frequency or 600Hz
(previously)
or 60Hz (now).
10.37 BRIKEN start file
AIDA file RIBF148/R5_141
BRIKEN file 30
BigRIPS file 1015
10.25 BRIKEN stop file
AIDA file RIBF148/R5_144
BRIKEN reloading DAQ configuration, will send correlation scaler reset
10.56 BRIKEN start file
AIDA file RIBF148/R5_145
BRIKEN file 31
BigRIPS file 1016
BRIKEN reload DAQ configuration again
11.00 BRIKEN stop file
AIDA file RIBF148/R5_145
11.06 BRIKEN start file
AIDA file RIBF148/R5_145
BRIKEN file 32
BigRIPS file 1017
11.50 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 18)
good event statistics ok (attachment 19)
detector bias/leakage currents ok (attachment 20)
12.05 BRIKEN stop file
AIDA file RIBF148/R5_155
12.07 BRIKEN start file
BRIKEN file 33
BigRIPS file 1018
AIDA file RIBF148/R5_155
13.06 BRIKEN stop file
AIDA file RIBF148/R5_165
13.08 BRIKEN start file
BRIKEN file 34
BigRIPS file 1019
AIDA file RIBF148/R5_165
13.35 BRIKEN stop file
AIDA file RIBF148/R5_169
BighRIPS DAQ reset required
13.38 BRIKEN start file
BRIKEN file 35
BigRIPS file 1020
AIDA file RIBF148/R5_170
14.39 BRIKEN stop file
AIDA file RIBF148/R5_179
14.40 BRIKEN start file
BRIKEN file 36
BigRIPS file 1021
AIDA file RIBF148/R5_179
15.38 BRIKEN stop file
AIDA file RIBF148/R5_189
15.49 System wide checks ok - except nnaida7 ADC calibration failure
FEE64 temperatures ok - attachment 21
good event statistics ok - attachment 22
detector bias/leakage currents ok - attachment 23
15.51 BRIKEN start file
BRIKEN file 37
BigRIPS file 1022
AIDA file RIBF148/R5_191
|
|
777
|
Mon Nov 12 14:51:44 2018 |
OH | Tuesday 13 November 00:00 - 08:00 | 00:00 AIDA currently on file R5_38
BRIKEN on file 181112_2321_165Pm_019.dlt (AIDA file at start R5_33)
BigRIPS run 1004
00:23 BRIKEN stop file
AIDA Currently on file R5_43
00:26 BRIKEN start 181113_0025_165Pm_020.dlt
BigRIIPS start 1005
AIDA Currently on file R5_43
System wide checks ok
01:13 Bias and leakage currents ok - attchment 1
Stats ok - attachment 2
FEE Temps ok - attachment 3
01:21 BigRIPS and BRIKEN stop
AIDA on file R5_52
01:25 BigRIPS starts run 1006
BRIKEN starts run 181113_0123_165Pm_021.dlt
AIDA on file R5_53
02:21 BigRIPS and BRIKEN stop
AIDA on file R5_62
02:23 BigRIPS starts run 1007
BRIKEN starts run 181113_0222_165Pm_022.dlt
AIDA on file R5_62
ASIC load check ok
03:15 Beam dropped out unsure for how long but is now back
03:21 BigRIPS and BRIKEN stopped
AiDA on file R5_71
03:23 BigRIPS starts run 1008
BRIKEN starts run 181113_0320_165Pm_023.dlt
AIDA on file R5_71
03:45 Output of analyser on R5_70 - attachment 4
03:59 System wide checks ok
FEE Temps ok - attachment 5
Statistics ok - attachment 6
04:21 BigRIPS and BRIKEN stop
AIDA on file R5_81
04:23 BigRIPS starts run 1009
BRIKEN starts run 181113_0421_165Pm_024.dlt
AIdA on file R5_81
05:19 BigRIPS and BRIKEN stop
AIDA on file R5_90
05:20 BigRIPS start run 1010
BRIKEN starts run 181113_0520_165Pm_025.dlt
AIDA on file R5_90
05:49 System wide checks all ok
Stats ok - attachment 7
FEE Temps ok - attchment 8
Bias and leakage currents ok - attachment 9
High energy hit pattern per dssd - attachment 10
Low energy Ex vs Ey plot per dssd - attachment 11
High energy Ex vs Ey plot per dssd - attachment 12
Also using the previously merged files produced from Phong's sort have confirmed that there is no to
little punch through on DSSD3
Did this through the use of a 2d plot of energy DSSD 3 vs the velocity
06:21 BigRIPS and BRIKEN stop
AIDA on file R5_100
06:23 BigRIPS starts run 1011
BRIKEN starts run 181113_0622_165Pm_026.dlt
AIDA on file R5_100
07:19 BigRIPS and BRIKEN stop
AIDA on file R5_110
07:20 BigRIPS starts run 1012
BRIKEN starts run
AIDA on file R5_110 |
|
776
|
Mon Nov 12 08:59:43 2018 |
LS, CA, TD | Monday 12 November 6pm onwards |
17:59 DAQ continues OK
Stats OK - attachment 1
FEE temperatures OK - attachment 2
Bias and leakage currents OK - attachment 3
System-wide checks OK *except* nnaida7 fails ADC calibration
19:07 DAQ continues OK
Stats OK - attachment 4
FEE temperatures OK - attachment 5
Bias and leakage currents OK - attachment 6
System-wide checks OK *except* nnaida7 fails ADC calibration
20.19 DAQ starts
AIDA file RIBF148/R5_4
BRIKEN file 16
BigRIPS 1001
165Pm setting + 3mm Al degrader
20.21 all online & MIDASsort histograms zero'd
20.23 FEE64 good event statistics - see attachment 7
Merger - see attachment 8
TapeServer - see attachment 9
HEC x versus y spectra - see attachment 10
SN thinks DSSSD #3 are the ions of interest, #4 are the ions of interest at different charge state
and #5 & #6 are lighter ions which do not stop in DSSSD #1-6
21.21 DAQ stops
DAQ starts
AIDA file RIBF148/R5_14
BRIKEN file 18
BigRIPS 1003
21.26 HEC x vs y multiplicity spectra - attachment 11
21.29 HEC Ex vs Ey spectra - attachment 12
FEE64 rate spectra - attachment 13
21.31 1.8.L front/rear energy spectra (attachments 14/15) LEC
21.33 1.8.H front/rear energy spectra (attachments 16/17) HEC
21.46 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 good event statistics ok (attachment 18)
FEE64 temperatures ok (attachment 19)
Leakage currents/detector bias ok (attachment 20)
21.49 Merger and TapeServer ok, no error messages in terminal
DAQ still running (RIBF148/R5_18)
22.10 attachment 21 - pulser spectra in all channels fired at same time
- FEE64 in sync
22.12 SYNC100 at 9153.8 Hz
FEE64 discriminator at ~15000 Hz - higher than expected
~1 second deadtime in FEE19 - due to nnaida12 running fast
no timewarps observed in FEE64
all of above in attachment 22
22:32 DAQ continues OK
Stats OK - attachment 24
FEE temperatures OK - attachment 23
Bias and leakage currents OK - attachment 25
System-wide checks OK *except* nnaida7 fails ADC calibration
23:30 DAQ continues OK
Stats OK - attachment 26
FEE temperatures OK - attachment 27
Bias and leakage currents OK - attachment 28
System-wide checks OK *except* nnaida7 fails ADC calibration
|
|
775
|
Sun Nov 11 21:35:42 2018 |
CB, OH, CA, LS | Monday 12 November | 06:35 DAQ Continues ok
Stats ok - attachment 1
FEE Temps ok - attachment 2
Bias and leakage currents ok - attachment 3
System wide checks all ok
Chiller refilled
07:57 DAQ continues OK
Stats OK - attachment 4
FEE temperatures OK - attachment 5
Bias and leakage currents OK - attachment 6
System-wide checks OK *except* nnaida7 fails ADC calibration
10:06 DAQ continues OK
Stats OK - attachment 7
FEE temperatures OK - attachment 8
Bias and leakage currents OK - attachment 9
System-wide checks OK *except* nnaida7 fails ADC calibration
Workstation occasionally feels sluggish. See iostat 5 (attach 10) & top -a (attach 11)
12:08 DAQ continues OK
Stats OK - attachment 12
FEE temperatures OK - attachment 13
Bias and leakage currents OK - attachment 14
System-wide checks OK *except* nnaida7 fails ADC calibration
15:40 DAQ continues OK
Stats OK - attachment 15
FEE temperatures OK - attachment 16
Bias and leakage currents OK - attachment 17
System-wide checks OK *except* nnaida7 fails ADC calibration
16:58 DAQ continues OK
Stats OK - attachment 18
FEE temperatures OK - attachment 19
Bias and leakage currents OK - attachment 20
System-wide checks OK *except* nnaida7 fails ADC calibration |
|
774
|
Sat Nov 10 21:53:29 2018 |
CA, CB, TD, OH | Sunday 11 November | 06.53 DAQ continues OK - data transfer disabled
06.55 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
Julabo FL11006 chiller set point +20 deg C actual +20 deg C, water level c. 75% (L-H)
ambient tempature + 26.9 deg C, RH 21.6%, d.p. +2.9 deg C
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachment 4
per FEE64 rate spectra - see attachment 5
ASIC check load
per FEE64 rate spectra - see attachment 6
good event statistics OK - see attachment 7
1.8.L spectra - see attachments 8 & 9
nnaida1 pulser peak width 72 ch FWHM
1.8.W spectra 20us FSR - see attachments 10 & 14
1.8.W spectra 200us FSR - see attachments 11 & 15
1.8.W spectra 2ms FSR - see attachments 12 & 16
1.8.W spectra 20ms FSR - see attachments 13 & 17
Note - LED threshold 9000 for *all* FEE64s
07.27 statistics zero'd
some 32bit counters (e.g. nnaida19 & nnaida20) had overflowed/wrapped around
18:19 Computer appears to be running slow though cannot see a reason why.
No programs using to much ram or cpu
FEE temp ok - attachment 18
bias and leakage currents ok - attachment 19
System wide checks ok
21:02 System wide checks ok
DAQ Continues ok - data transfer disabled.
Bias and leakage currents ok - attachment 20
FEE Temperatures ok - attachment 21 |
|
773
|
Sat Nov 10 04:06:20 2018 |
TD | Less commonly recorded AIDA control tabs |
Some less commonly recorded AIDA control tabs whilst DAQ going (data transfer disabled) |
|
772
|
Sat Nov 10 03:50:00 2018 |
CA, CB, TD, OH | Saturday 10 November |
12.35 DAQ continues OK - data transfer disabled
12.37 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
Julabo FL11006 chiller set point +20 deg C actual +20 deg C, water level topped up c. 1 litre - 80% (L-H)
ambient tempature + 27.3 deg C, RH 33.0%, d.p. +9.6 deg C
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachment 4
per FEE64 rate spectra - see attachment 5
1.8.L spectra - see attachments 6 & 7
nnaida1 pulser peak width 71 ch FWHM
1.8.W spectra 20us FSR - see attachments 8 & 9 |
|
771
|
Fri Nov 9 02:26:58 2018 |
TD | Offline analysis R18, R27, R29 - alpha background | Offline analysis of Oct18/R18_0 to 2, R27_0, 4 & 9, R29_0, 7 & 30
alpha background, BNC PB-4 pulser OFF, LEC/MEC slow comparator 0x64 (~1MeV)
*** ENTRY finish
*** data items: 2043072278 ( 246610.11 Hz)
*** ADC events: 10666744 ( 1287.53 Hz)
*** time warps: 2 ( 0.00 Hz)
*** DSSSD # 1 count: 1446 old count: 1446 dt: 231900.06 s LEC rate: 0.01 Hz
*** DSSSD # 2 count: 568 old count: 568 dt: 231900.06 s LEC rate: 0.00 Hz
*** DSSSD # 3 count: 644 old count: 644 dt: 231900.06 s LEC rate: 0.00 Hz
*** DSSSD # 4 count: 612 old count: 612 dt: 231900.06 s LEC rate: 0.00 Hz
*** DSSSD # 5 count: 591 old count: 591 dt: 231900.06 s LEC rate: 0.00 Hz
*** DSSSD # 6 count: 796 old count: 796 dt: 231900.06 s LEC rate: 0.00 Hz
*** DSSSD # 1 count: 6 old count: 6 dt: 231900.06 s HEC rate: 0.00 Hz
*** DSSSD # 2 count: 3 old count: 3 dt: 231900.06 s HEC rate: 0.00 Hz
*** DSSSD # 3 count: 0 old count: 0 dt: 231900.06 s HEC rate: 0.00 Hz
*** DSSSD # 4 count: 1 old count: 1 dt: 231900.06 s HEC rate: 0.00 Hz
*** DSSSD # 5 count: 0 old count: 0 dt: 231900.06 s HEC rate: 0.00 Hz
*** DSSSD # 6 count: 1 old count: 1 dt: 231900.06 s HEC rate: 0.00 Hz
*** ENTRY finish ends
attachment 1 - per DSSSD m_p versus m_n
2 - per DSSSD x versus y
3 - per DSSSD E_p versus E_n
4 - DSSSD channel versus E_p
5 - DSSSD channel versus E_p (expanded)
6 - hit pattern (ASIC & DSSSD order)
7 - titles
8 - variables
N.B. ADC offsets not set, 0 by default
| E_p - E_n | < 2000
DSSSD configuration looks OK. Conclude problems identified and addressed November 5 are confirmed
as fixed - see https://elog.ph.ed.ac.uk/AIDA/766 and https://elog.ph.ed.ac.uk/AIDA/761 |
|
770
|
Thu Nov 8 23:35:50 2018 |
OH, TD, CA, CB | Friday 9 November | 08:36 DAQ still running
Not writing data to disk
Bias and leakage currents ok - attachment 1
System wide checks all ok
FEE Temperatures ok - attachment 2
Statistics - attachment 3
Check ASIC Control done
08:56 Pulser peaks are ok and similar to what has been observed over the past two weeks
Layout 4 and 5 - attachments 4 and 5
11.55 Lost connection to FEE64 1-16 - fuse blown
stopped DAQ
switched all FEE64 off
Replaced 20A fuse
switched FEE64 back on individually
system wide checks ok, except nnaida7 failure in ADC calibration
DAQ start (not writing to file)
FEE64 temperatures ok (attachment 6)
Bias/Leakage currents ok (attachment 7)
good event statistics ok (attachment 8)
17.08 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 9 & 10
FEE64 temperatures OK - see attachment 11
good event statistics OK - see attachment 12
per FEE64 rate spectra - see attcahment 13
1.8.L spectra - see attachments 14 & 15
nnaida1 pulser peak width 72 ch FWHM
1.8.W spectra 20us FSR - see attachments 16 & 20
1.8.W spectra 200us FSR - see attachments 17 & 21
1.8.W spectra 2ms FSR - see attachments 18 & 22
1.8.W spectra 20ms FSR - see attachments 19 & 23 |
|
769
|
Thu Nov 8 00:55:18 2018 |
CA, CB, TD, OH | Thursday 8th November | 09.57 DAQ still running (R29_29)
System wide checks ok, except nnaida7 ADC calibration failure
09.59 FEE64 temperatures ok (attachment 1)
TapeServer ok
Link16 error reported in Merger (attachment 2)
10.04 Bias/Leakage currents ok (attachment 3)
10.07 Good event statistics ok, except nnaida17, 18, 22 which appears to intermittently run fast overnight, and
nnaida2 which runs fast due to hot channel (attachment 4)
10.11 1.8.W spectrum (attachment 5)
10.56 2.15.H spectra - show hot high energy channel in nnaida2 (attachment 6)
11.09 attachments 7,8,9 - High energy channels spectra, show hot channel in nnaida17
11.45 Stats - Attachment 10
15.07 analysis RIBF/R1_1 - see attachment 11
<1% deadtime for high rate FEE64s nnaida19 & nnaida20
15.14 AIDA shield (c. 3" Pb) in position between F11 and upstream BRIKEN (CH2)n shield - checked by TD & GK
Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C, water added to 80% level (L-H) or 2/3 (M-H)
16:41 DAQ stopped
slow comparator threshold at 10
DAQ restarted (\RIBF148\R2)
DAQ stopped (\RIBF148\R2_3)
slow comparator threshold raised back to 100
DAQ start (\RIBF148\R3_0)
17.10 DAQ stop (\RIBF148\R3_0)
slow comparator threshold lowered back to 10
DAQ start (not writing to file)
good event statistics ok (attachment 12)
17.15 system wide checks ok
FEE64 temperatures ok (attachment 13)
bias/leakage currents ok (attachment 14)
18:35 R4_0 started for background measurement for Phong.
R4_0 should be discarded as FEE1 was running fast at start. recovered with ASIC check
18:51 R1_2 stopped |
|
768
|
Wed Nov 7 01:00:58 2018 |
CA, CB, TD, OH, PJW | Wednesday 7 November | 10.02 DAQ still running, on file Oct18/R27_9
Merger and TapeServer ok, no error warnings in terminals
10.07 Good event statistics ok, except nnaida6 rate being high (attachment 1)
nnaida17 rate ok, but counts very high. May have run fast at some point overnight.
System wide checks ok, except nnaida7 failure in ADC calibration
FEE64 temperatures ok (attachment 2)
10.13 Bias/leakage currents slightly higher, but ok (attachment 3)
10:30 1.8.W spectra (attachment 4)
11.10 Stopped the DAQ (R27_9)
Checked ASIC control - ok
Started the DAQ again (Oct18/R28_0)
good event statistics ok, aside from nnaida6 running slightly fast (attachment 5)
15.07 DAQ stop - file Oct18/R28_2
Detector bias OFF
Move AIDA downstream to check position of Clover G7
16.05 Move back upstream into BRIKEN (CH2)n moderator assembly
Detector bias ON
16.12 all system wide checks OK *except* nnaida7 fails ADC calibration
sync all ASIC clocks using ReSYNC
detector biases & leakage currents OK - see attachments 6 & 7
FEE64 temperatures OK - see attachment 8
good event statistics OK *except* nnaida2 - see attachment 9
see attachment 10 for comparison to statistics 5.1.18
16.41 1.8.W spectra 20us FSR - see attachments 11 & 12
16.55 slow comparator thresholds raised from 10 to 100
DAQ start (Oct18/R29_0)
Merger and TapeServer ok - no error warnings in terminals
good event statistics ok, except nnaida6 running fast (attachment 13)
|
|
767
|
Tue Nov 6 00:13:00 2018 |
CB. PJW, TD, OH, CA | Tuesday 6 November | 09:13 DAQ still running R18_1
Merger and TapeServer OK
System wide checks ok, except for nnaida7 ADC calibration failure
Temperature OK (attachment 1)
Bias and leakage current OK (attachment 2)
Good event statistics OK (attachment 3)
11.20 1.8.W spectra - attachment 4 & 5
analysis file Oct18/R18_1 - attachment 6
15:35 Inserted AIDA snout in BRIKEN. Lowered threshold to 100 keV. Rate looks OK.
R19 pulser walkthrough 90 0000 ... 10 0000 (10 000 steps) - attach 7&8
R21 pulser walkthrough 12 000 ... 2 000 (2 000 steps)
Pulser back to 90 000. Frequency set to zero.
R22 background run starts.
Rates OK (attach 9)
16.07 Raised slow comparator threshold to 100 from 10
DAQ start (file R23)
Merger and TapeServer ok, no error messages in terminal
Good event statistics ok (attachment 10)
System wide checks ok, except nnaida7 failing ADC calibration
FEE64 temperatures ok (attachment 11)
bias/leakage currents ok (attachment 12)
16.58 Repeat pulser walkthroughs with additional ASIC check loads
R25 pulser walkthrough 90 0000 ... 10 0000 (10 000 steps) - attach 7&8
R26 pulser walkthrough 12 000 ... 2 000 (2 000 steps)
18.26 DAQ start - file Oct18/R27
alpha background
LEC/MEC slow comparator 0xa -> 0x64
BNC PB-4 pulser OFF @ 18.29!
18.34 1.8.W spectra 20us FSR - see attachments 13 & 14 |
|
766
|
Mon Nov 5 10:02:52 2018 |
TD | Offline analysis R16 - beam background | Offline analysis of Oct18/R16_0 to 11
BNC PB-4 pulser OFF, LEC/MEC slow comparator 0x64 (~1MeV)
background (presumably high energy, low Z particles) from beam stopping in F11 E detector upstream
of AIDA and BRIKEN - ~10-15cm Pb shielding installed in BRIKEN to stop heavy opns reaching AIDA
*** ENTRY finish
*** data items: ********** ( -13850.89 Hz)
*** ADC events: 2089104652 ( 24903.76 Hz)
*** time warps: 0 ( 0.00 Hz)
*** DSSSD # 1 count: 81173 old count: 81171 dt: 106837.14 s LEC rate: 0.76 Hz
*** DSSSD # 2 count: 73671 old count: 73668 dt: 106837.14 s LEC rate: 0.69 Hz
*** DSSSD # 3 count: 74114 old count: 74109 dt: 106837.14 s LEC rate: 0.69 Hz
*** DSSSD # 4 count: 73324 old count: 73320 dt: 106837.14 s LEC rate: 0.69 Hz
*** DSSSD # 5 count: 74312 old count: 74308 dt: 106837.14 s LEC rate: 0.70 Hz
*** DSSSD # 6 count: 74653 old count: 74649 dt: 106837.14 s LEC rate: 0.70 Hz
*** DSSSD # 1 count: 0 old count: 0 dt: 106837.14 s HEC rate: 0.00 Hz
*** DSSSD # 2 count: 0 old count: 0 dt: 106837.14 s HEC rate: 0.00 Hz
*** DSSSD # 3 count: 0 old count: 0 dt: 106837.14 s HEC rate: 0.00 Hz
*** DSSSD # 4 count: 0 old count: 0 dt: 106837.14 s HEC rate: 0.00 Hz
*** DSSSD # 5 count: 0 old count: 0 dt: 106837.14 s HEC rate: 0.00 Hz
*** DSSSD # 6 count: 0 old count: 0 dt: 106837.14 s HEC rate: 0.00 Hz
*** ENTRY finish ends
attachment 1 - per DSSSD m_p versus m_n
2 - per DSSSD x versus y
3 - per DSSSD E_p versus E_n
4 - DSSSD hit pattern (ASIC & DSSSD order) expanded view of LEC events
DSSSD configuration looks OK. With improved statistics (cf alpha background) some missing strips identifiable
- see also Elog https://elog.ph.ed.ac.uk/AIDA/764 |
|
765
|
Mon Nov 5 01:40:10 2018 |
CA, CB, TD, OH, PJW | Monday 5 November | 10.40 Lowered slow comparator threshold from 100 to 10
Good event statistics ok (attachment 1)
System wide checks ok, except for nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 2)
Bias/Leakage currents ok (attachment 3)
1.0.L spectra obtained, look ok (attachment 4)
DAQ stopped (R16)
10.49 Slow comparator threshold raised from 10 to 100
DAQ restarted (R17)
Merger and TapeServer ok, no error messages in terminals
Good event statistics ok (attachment 5)
nnaida2 is now at a good rate, after fixing kapton connection.
15:09 DAQ still running (Oct18/R17_0)
Merger and TapeServer ok
System wide checks ok, except for nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 6)
Bias/Leakage currents ok (attachment 7)
Good event statistics ok (attachment 8)
18.48 good event statistics OK *except* nnaida17 c. 400k
ASIC Control -> Check ASIC Control
*all* FEE64 good events statistics now OK
DAQ stop - current file Oct18/R17_20
18.50 DAQ start - file Oct18/R18
BNC PB-4 OFF
LEC/MEC slow comparator 0x64
alpha background run
18.54 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 9 & 10
FEE64 temperatures OK - see attachment 11
good event statistics OK - see attachment 12
Merger - Link16 errors reported - see attachment 13
TapeServer OK - see attachment 14
1.8.W spectra - attachment 15 & 16
|
|