Tue Mar 8 15:40:57 2016, TD, Tuesday 8 March
|
MSL type BB18(DS)-1000 serial 2998-22
Bias +200V, I_L +=6.705uA, ambient temperature +19.8 deg C
|
Tue Nov 6 00:13:00 2018, CB. PJW, TD, OH, CA, Tuesday 6 November 14x
|
09:13 DAQ still running R18_1
Merger and TapeServer OK
System wide checks ok, except for nnaida7 ADC calibration failure
|
Tue Jun 6 00:57:49 2017, PJW, TD, Tuesday 6 June 08.00-16.00 16x
|
08.57 Stop DAQ
AIDA file RIBF128/R6_24
|
Tue Dec 6 02:20:29 2016, TD, Tuesday 6 December   
|
11.20 DSSSD bias & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
Ambient temperature +24.8 deg C, d.p. +3.6 deg C, RH 25.4%
|
Mon May 4 18:09:42 2015, JA, TD, AE, CG & JLT, Tuesday 5 May 2015
|
02.09 R61 starts
nnaida1, 2, 6, 10, 13 & 14 fast discs disabled
|
Tue Jun 4 07:10:27 2019, TD, Tuesday 4 June 13x
|
15.01 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 70% M-H markers - topup to 100% with 1x jugful water
Temperature +25.8 deg C dew point +8.0 deg C RH 32.4%
|
Tue Oct 31 00:34:10 2017, TD, OH, Tuesday 31 October 8x
|
09.34 DAQ running OK - data transfer not enabled
BNC PB-4
|
Tue May 31 08:06:43 2016, TD, AE & CG, Tuesday 31 May 2016 7x
|
Attachment 3 - AIDA discriminator with pulser ~5Hz
standard ASIC settings (tau=8us, slow comparator 0x10, LEC/MEC comparator 0x9)
|
Mon May 29 16:02:21 2017, OH, TD, Tuesday 30th May 16x
|
00:02 Bias and leakage currents ok - attachment one
System wide check all ok
FEE64 Temperatures OK - see attachment 2
|
Mon Oct 29 23:40:41 2018, TD, OH, CA, Tuesday 30 October 16x
|
08.40 DAQ continues OK - current file Oct18/R9_15
08.42 all system wide checks OK *except* nnaida7 fails ADC calibration
|
Tue May 30 07:56:05 2017, CB, NN, OH, AE, Tuesday 30 May - evening shift 7x
|
16.00 Still no beam
Current run /RIBF127R1/R1_197
Found dropped channels on nnaida1, ASIC4 - see attach 1
|
Tue May 30 00:58:10 2017, TD, CB, Tuesday 30 May - day shift 27x
|
08.58 DAQ continues OK
File RIBF127R1/R1_99
BigRIPS run 1009
|
Tue Aug 30 08:26:33 2016, TD, Tuesday 30 August 28x
|
16.18 R10 stops
ASIC settings 2016Aug20-15.34.53
|
Tue Oct 3 09:14:59 2023, TD, Tuesday 3 October 
|
09.45 Power up
|
Tue May 3 07:51:04 2016, TD, Tuesday 3 May 2016  
|
32x FEE64 re-programmed from version 7 to version 8 per instructions
https://elog.ph.ed.ac.uk/AIDA/183
|
Tue Jul 3 02:24:45 2018, OH, Tuesday 3 July   
|
10:23 Bias and leakage currents ok - attachment 1
Fee temperatures ok - attachment 2
System wide checks ok
|
Mon Nov 28 18:07:14 2016, CG, TD, AE, DK, Tuesday 29 November 42x
|
01:39 : Attachment 26 (R30_70to72) shows a gradual change in the hit pattern for DSSD2 (at the lower left
corner) which may indicate the noise level was increasing.
|
Tue Apr 28 06:25:13 2015, JA, LHB, TD, AE & CG, Tuesday 28th April progress 6x
|
28/04/15
1420: Carried out ASIC optimisation tests to find best shaping time. Chose system wide shaping time of 0x6.
Waveforms of example FEEs/channels attachments 1-6. |
Tue Mar 28 01:28:12 2017, TD, CG, OH, DK, Tuesday 28 March 15x
|
09.13 DAQ run R3 ends OK (final file R3_9)
|
Tue Jun 26 05:44:44 2018, OH, Tuesday 26 June 
|
04:12 TD found the nnaida14 appeared to have become noisy at some point. Biases/;eakages/temps ok
ASIC check load appears to have fixed problem
Noise appears to affect R9_0 to R9_51
|