Tue Nov 8 15:10:24 2016, Many, 717273Ni Run
|
F7 OUT F11 C-6mm
Start: 00:09
|
Tue Mar 28 09:10:09 2017, Many, 152Eu 
|
152Eu source inside BRIKEN for CLOVER calibration.
Also calibration of Si-dE with 60Co source
and AIDA_PL calibration with 60Co source.
|
Wed Mar 29 14:14:58 2017, Many, 40Mg RUN 000
|
Date: 29/03/2017
Start: 22:13
|
Wed Mar 29 15:16:17 2017, Many, 40Mg Run 001
|
Date: 29/03/2017
Start: 23:13
|
Mon Nov 5 09:59:36 2018, MW, KR, suspicious clover G7 temp sensor
|
We observed 6 pm LN2 filling of both clovers on Monday Nov 5th.
D4 went full within reasonable time of below 10 minutes.
Filling of G7 was stopped by "emergency sensor" after about 15 minutes of filling. |
Mon Nov 5 18:52:54 2018, MW, KR, suspicious clover G7 temp sensor
|
Quote:
We observed 6 pm LN2 filling of both clovers on Monday Nov 5th. |
Sat Apr 1 22:18:00 2017, Liu, YS, CG, GK, SN, JLT , 38Mg run 72
|
| DUMMY |
Thu Mar 30 22:19:22 2017, Liu DK KG JA ID, Run 28-29
|
Date: 31/03/2017
Start: 06:30
BRIKEN run: 28
|
Fri Jul 29 09:52:28 2016, Keishi Matsui, accuracy of Cf-252's intensity in RIKEN
|
According to the reply from Japan Radioisotope Association, the accuracy of the Cf-252's intensity is +0%/-15%.
( "+0%" is important, because 3.7 MBq is a standard limit for radio-active sources in Japan.) |
Sun Oct 22 08:52:17 2017, KR,JA,ATD,JLT, Clover resolution
|
Resolution check using 40K 1460 keV line
in the background measurement overnight
(File: 171022_0126_1101_bck.root)
|
Mon Nov 5 06:46:30 2018, KR, YR, MW, JLT, ..., Clover check 
|
With detectors still on the floor we check signals and found
that D4 gives no signal with PA power. Two power supplies/cables were checked.
|
Fri Oct 20 05:33:11 2017, KR, NB, SG, RY, JL, AT: , Calibration and Resolution of D4 (after Repair) 
|
Green and Red channels were repaired by crimping and securing the grounding of the cables and fixing the cables securely to the table.
All problems during the repair process were visible on the scope. See example attatched.
Calibration is as: |
Thu Nov 10 05:05:54 2016, KM, How to convert ridf-file to root-file
|
1. Login to briken@ribfana03.riken.jp
2. Check if the ridf-file you want to convert is in the directory; /home/briken16/rawdata/bigrips/
3. Login to anaroot; type "anarootlogin" |
Wed Jul 20 16:55:48 2016, K.Gabor, K. Matsui, A. Tarifeņo, Electronic
|
Drying of preamplifiers
PA8, PA9 and PA10 are removed and brought to a dryer box
|
Mon Apr 27 10:18:21 2015, K.Gabor S.Nishimura J.Agramunt, ORNL 3He tube at RIKEN
|
| |
Mon May 29 05:40:49 2017, K. Matsui, energy loss at logn
|
| |
Mon May 29 07:06:56 2017, K. Matsui, Energy loss in Long Plastic Scintillator
|
After Al degrader with the optimized width of 13 mm, the energy loss in Long Plastic Scintillator is 546(2) MeV at F11.
This value depends on |
Sat Nov 5 01:26:46 2016, Jose,Iris,Alvaro, Alejandro, Krzysztof, Background measurement 7x
|
Measurement overnight
Start: around 1:00
|
Sun Nov 20 12:28:27 2016, Jorge, Alvaro, Jose, DACQ
|
| Updated run protocol |
Sun Nov 20 13:02:06 2016, Jorge, Alavaro, Jose, DACQ
|
| Gasific-7.0 Short Manual (Preliminary) |