ID |
Date |
Author |
Subject |
567
|
Fri May 31 21:03:21 2024 |
CB, EM | Run14 |
22:03
Before this run we checked quickly beam with the target on but it does not make sense with the current status of the beam.
We removed target and starte Run14
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable |
Attachment 1: Screenshot_from_2024-05-31_22-05-57.png
|
|
Attachment 2: Screenshot_from_2024-05-31_22-07-01.png
|
|
Attachment 3: Screenshot_from_2024-05-31_22-07-49.png
|
|
503
|
Fri Feb 9 17:04:39 2024 |
JM, AR, CB, NH | Run12 - Background |
|
502
|
Fri Feb 9 11:58:04 2024 |
TD + CB, JM, AR, NH, AB | Run10 contd, Run11 |
12.46 DAQ continues G22-88/R10_14
c. 20MeV 16O2+ + 4He
DSSSD bias OK - attachment 1
ambient temperature 17.7 deg C ( c. 1deg C increase cf. earlier this week )
FEE64 temperatures OK - attachment 2
All system wide checks OK
WR timestamp control OK - attachment 3
ADC data item stats - attachment 4
FEE64 rates c. 0Hz except 'hot' channels aida01 ( 71k ) and aida15 ( 8k )
per FEE64 Rate and Stat spectra - attachments 5-8
events ( beam? ) aida01 asic #4 and aida04 asic #4 ( c. 200k per ASIC channel )
per FEE64 1.8.L spectra - attachments 9-10
most spectra showing background and, possibly, a small amount of scattered beam?
per FEE64 1.8.W spectra - 20us FSR - attachments 11-12
aida04 3*L and 4*L spectra - attachments 13-15
distribution by strip # - uniform across asic# #4, continuing across asic #3 with abrupt fall off c. channel 9
not Rutherford - more consistent halo/beam scattering?
what are the current lab angles subtended by the DSSSDs?
Rutherford cross section for 20MeV 16O+4He
Reaction: 4He(16O,16O)4He
A: 4 16 16 4
Z: 2 8 8 2
Mass(amu): 4.002603 15.994915 15.994915 4.002603
Error(keV): 0.001 0.002 0.002 0.001
Origin:
Status:
Projectile lab energy: 20.000 MeV
CM energy: 4.003 MeV
Particle 3 Particle 4
( scattered ) ( recoil )
Lab Cross Section CM Lab/CM Lab Lab Lab
Angle Lab CM Angle Ratio Energy Angle Energy
(deg) (mb/sr) (deg) (MeV) (deg) (MeV)
0.00 NaN Inf 0.000 NaN 20.000 90.000 0.000
1.00 0.143028E+09 0.572301E+07 4.999 0.040 19.976 87.500 0.024
2.00 0.894079E+07 356425. 10.017 0.040 19.902 84.992 0.098
3.00 0.176669E+07 69982.6 15.072 0.040 19.780 82.464 0.220
4.00 559329. 21951.4 20.186 0.039 19.607 79.907 0.393
5.00 229335. 8887.05 25.382 0.039 19.382 77.309 0.618
6.00 110782. 4221.76 30.690 0.038 19.103 74.655 0.897
7.00 59959.1 2235.75 36.144 0.037 18.768 71.928 1.232
8.00 35299.6 1279.52 41.790 0.036 18.371 69.105 1.629
9.00 22192.2 775.052 47.692 0.035 17.907 66.154 2.093
10.00 14729.2 489.347 53.941 0.033 17.366 63.029 2.634
11.00 10261.1 317.919 60.685 0.031 16.732 59.658 3.268
12.00 7514.20 209.725 68.185 0.028 15.976 55.907 4.024
13.00 5900.47 137.756 77.018 0.023 15.035 51.491 4.965
14.00 5693.43 85.2271 89.184 0.015 13.688 45.408 6.312
Maximum lab scattering angle ( 14.49 deg ) exceeded
________________________________________________________________________________
aida01 4*L spectra - attachments 19-23
channels 0-8 - high stats structure(s) at low energy
channels 9-16 - high energy peak cf. aida04 asic #4 spectra
Merger, Merger stats, TapeServer - attachments 16-18
Started Run 11 with beam ON.
We observe the same pattern of counts in DSSD0 (Top left detector) but not in DSSD1,2,3 (attachment 25). Moving in the left arm with the pneumatic we see a large drop in
the rate on DSSD0 to essentially zero. The beam on the CRYRING monitor appears to disappear.
When we move the right pneumatic IN, we observe an increase in the number of counts in the bottom-left part of the detector. The position is likely to be a mistake in the
strip assignment.
No increase observed in the leakage current. Will try systematically tomorrow.
Beam position monitoring from the CRYRING are shown in attachments 25-27 |
Attachment 1: Screenshot_from_2024-02-09_12-45-57.png
|
|
Attachment 2: Screenshot_from_2024-02-09_12-46-32.png
|
|
Attachment 3: Screenshot_from_2024-02-09_12-46-59.png
|
|
Attachment 4: Screenshot_from_2024-02-09_12-47-37.png
|
|
Attachment 5: Screenshot_from_2024-02-09_12-47-59.png
|
|
Attachment 6: Screenshot_from_2024-02-09_12-48-59.png
|
|
Attachment 7: Screenshot_from_2024-02-09_12-49-17.png
|
|
Attachment 8: Screenshot_from_2024-02-09_12-49-39.png
|
|
Attachment 9: Screenshot_from_2024-02-09_12-50-18.png
|
|
Attachment 10: Screenshot_from_2024-02-09_12-51-04.png
|
|
Attachment 11: Screenshot_from_2024-02-09_12-51-48.png
|
|
Attachment 12: Screenshot_from_2024-02-09_12-52-25.png
|
|
Attachment 13: Screenshot_from_2024-02-09_12-54-05.png
|
|
Attachment 14: Screenshot_from_2024-02-09_12-56-04.png
|
|
Attachment 15: Screenshot_from_2024-02-09_12-56-20.png
|
|
Attachment 16: Screenshot_from_2024-02-09_12-56-52.png
|
|
Attachment 17: Screenshot_from_2024-02-09_12-57-06.png
|
|
Attachment 18: Screenshot_from_2024-02-09_12-57-36.png
|
|
Attachment 19: Screenshot_from_2024-02-09_13-03-20.png
|
|
Attachment 20: Screenshot_from_2024-02-09_13-03-38.png
|
|
Attachment 21: Screenshot_from_2024-02-09_13-03-54.png
|
|
Attachment 22: Screenshot_from_2024-02-09_13-04-22.png
|
|
Attachment 23: Screenshot_from_2024-02-09_13-04-59.png
|
|
Attachment 24: Screenshot_from_2024-02-10_10-07-57.png
|
|
Attachment 25: 20240209_174604.jpg
|
|
Attachment 26: 20240209_174705.jpg
|
|
Attachment 27: 20240209_174718.jpg
|
|
501
|
Fri Feb 9 07:57:38 2024 |
AR, CB, AB, JM | Run10 - First cool beam on target |
8:58
System wide checks okay, except
Base Current Difference
aida08 fault 0x0 : 0xc : 12
aida09 fault 0x0 : 0xb : 11
FPGA Timestamp error counter test result: Passed 14, Failed 2
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last
Temperatures okay
Attatchment 1: Layout 7 Waveforms
Attatchment 2: Temperatures
Attatchment 3: Layout 8 Waveforms
Attatchment 4: Layout 2 counter okay
Attatchment 5: Layout 4 spectrum okay
Attatchment 6: Leakage current okay
Switched to run 10 (run 9 is overnight background)
10:00 Target pressure on. Attachment 7
Beam on target
Observe low counts. Only aida4 showing anything (attachment 8)
See a peak around 19 MeV in LowEnergyExratedssd0 (can only see in the dssd0) that is assumed to be Rutherford scattering - Attachment 12
Temperatures ok
Wavefroms on attatchment 9 show only signals from aida04 - attachment 9
/////// UPDATE ////////
Peak observed is not from Rutherford, we think it is beam (or part of it) hitting the detector
Attachment 10 shows waveforms after the beam was turned off -> counts in aida 4 are gone.
Beam off = no peak at 19 MeV for LowEnergyExdssd0.
Leakage current ok (attachment 11)
//// BEAM ON AGAIN UPDATE ////
Still see peak at 20 MeV, see attachment 13
Peak is not in dssds 1,2,3 only in dssd 0 (Top left detector) (attatchments 14,15,16 respectively)
Leakage currents OK (attatchment 17)
Attachment 18: LowEnergyXYtotalDSSD0 plot showing all hits in one spot (this is with BEAM AND TARGET on)
TEST: Turning off target while beam is running, can still see hotspot in DSSD0 but no others, so definitely beam related. (attatchment 18)
Tom's estimate: 10^6 particles in 10 minutes - not as many as can be handled
/////// UPDATE ////////
No beam, no target (target off)
All checks ok (see attachment 20 and 21 for rates and temperatures)
Attachment 22: Layout 7 waveforms
Attatchment 23: Layout 8 waveforms
Attachment 24: Layout 2 waveforms
Attachment 25: Layout 1 waveforms (rate)
Atatchment 26: Leakage current
Attachment 27: LowEnergyExRatedssd 0,1,2,3 respectively
Note: Data similar to that seen in DSSD0 (19 MeV peak) is not observed in MIDAS for the other DSSDs -> Issue is not related to the monitoring code
|
Attachment 1: Screenshot_from_2024-02-09_09-03-01.png
|
|
Attachment 2: Screenshot_from_2024-02-09_09-03-28.png
|
|
Attachment 3: Screenshot_from_2024-02-09_09-04-12.png
|
|
Attachment 4: Screenshot_from_2024-02-09_09-08-45.png
|
|
Attachment 5: Screenshot_from_2024-02-09_09-11-08.png
|
|
Attachment 6: Screenshot_from_2024-02-09_09-13-29.png
|
|
Attachment 7: 20240209_100016.jpg
|
|
Attachment 8: Screenshot_from_2024-02-09_11-19-45.png
|
|
Attachment 9: Screenshot_from_2024-02-09_11-21-36.png
|
|
Attachment 10: Screenshot_from_2024-02-09_11-40-36.png
|
|
Attachment 11: Screenshot_from_2024-02-09_11-39-09.png
|
|
Attachment 12: dfcbe734-8730-4e0f-acfd-5e778e2b3f9f.JPG
|
|
Attachment 13: Screenshot_from_2024-02-09_13-17-12.png
|
|
Attachment 14: Screenshot_from_2024-02-09_13-17-27.png
|
|
Attachment 15: Screenshot_from_2024-02-09_13-17-31.png
|
|
Attachment 16: Screenshot_from_2024-02-09_13-17-40.png
|
|
Attachment 17: Screenshot_from_2024-02-09_13-18-46.png
|
|
Attachment 18: Screenshot_from_2024-02-09_13-41-38.png
|
|
Attachment 19: Screenshot_from_2024-02-09_13-55-21.png
|
|
Attachment 20: Screenshot_from_2024-02-09_16-25-22.png
|
|
Attachment 21: Screenshot_from_2024-02-09_16-25-38.png
|
|
Attachment 22: Screenshot_from_2024-02-09_16-27-06.png
|
|
Attachment 23: Screenshot_from_2024-02-09_16-27-40.png
|
|
Attachment 24: Screenshot_from_2024-02-09_16-28-11.png
|
|
Attachment 25: Screenshot_from_2024-02-09_16-28-39.png
|
|
Attachment 26: Screenshot_from_2024-02-09_16-29-55.png
|
|
Attachment 27: Screenshot_from_2024-02-09_16-30-11.png
|
|
486
|
Fri Feb 2 08:59:36 2024 |
JM, TD, CB | Run1 start - pulser widths spreadsheet |
Current pulser widths excel spreadsheet attached
09.58 DSSSD bias OK - attachment 1
Ambient temperature +16.1 deg C FEE64 temps OK - attachment 2
ADC data item stats OK - attachment 3
aida15 asic #2 high rate - asic u/s
system wide tests OK *except* aida09 - attachments 4-5
aida09 WR status 0x20 aida09 WR timestamp incorrect cf. all other FEE64s
per FEE64 Rate spectra - attachment 6
per FEE64 1.8.L spectra - attachments 7-8
aida14 pulse peak witdh 58 ch FWHM
aida11 pulse peak witdh 48 ch FWHM
per FEE64 1.8.W spetra - 20us FSR - attachments 9-10
12.00
aida09 WR timestamp issue resolved by FEE64 reboot
14.10
Safety pins removed from detector motion system. CB performs one, complete DSSSD movement cycle. Zero all histogram and stats per FEE64 Stat spectra ( log scale ) and
ADC
data item stats for the complete DSSSD movement cycle - attachments 11-12
No change observed in DSSSD leakage currents during DSSSD movement cycle.
15.35
Resolved issue from previous elog. Begin background run writing data to disk.
ASIC check load
Start DAQ G22-88/R1 background alpha
EG&G Ortec 448 Pulser OFF
ASIC settings 2024Jan20-12.02.08
slow comparator 0x64 ( all FEE64s *except aida15 asic #2 0xff )
All histograms set to zero
DSSSD bias OK - attachment 13
FEE64 temps OK - attachment 14
WR timestamps OK - attachment 15
ADC data item stats - attachment 16
aida15 asic #2 u/s - rate c. 250k
ADC data items/s Merger, Tape Server & Merger stats - attachments 17-19
data rate to disk c. 2Mb/s dominated by aida15 asic 2 |
Attachment 1: Screenshot_from_2024-02-02_09-59-10.png
|
|
Attachment 2: Screenshot_from_2024-02-02_10-10-46.png
|
|
Attachment 3: Screenshot_from_2024-02-02_10-10-59.png
|
|
Attachment 4: Screenshot_from_2024-02-02_10-11-18.png
|
|
Attachment 5: Screenshot_from_2024-02-02_10-11-37.png
|
|
Attachment 6: Screenshot_from_2024-02-02_10-12-32.png
|
|
Attachment 7: Screenshot_from_2024-02-02_10-13-17.png
|
|
Attachment 8: Screenshot_from_2024-02-02_10-15-11.png
|
|
Attachment 9: Screenshot_from_2024-02-02_10-16-33.png
|
|
Attachment 10: Screenshot_from_2024-02-02_10-18-01.png
|
|
Attachment 11: Screenshot_from_2024-02-02_14-09-42.png
|
|
Attachment 12: Screenshot_from_2024-02-02_14-09-46.png
|
|
Attachment 13: Screenshot_from_2024-02-02_15-41-14.png
|
|
Attachment 14: Screenshot_from_2024-02-02_15-41-35.png
|
|
Attachment 15: Screenshot_from_2024-02-02_15-42-14.png
|
|
Attachment 16: Screenshot_from_2024-02-02_15-42-31.png
|
|
Attachment 17: Screenshot_from_2024-02-02_15-43-02.png
|
|
Attachment 18: Screenshot_from_2024-02-02_15-43-11.png
|
|
Attachment 19: Screenshot_from_2024-02-02_15-43-25.png
|
|
Attachment 20: Pulser-widths-carme-2024.xlsx
|
488
|
Sat Feb 3 00:13:03 2024 |
CB TD | Run1 end - R3 start |
01.07 DAQ continues file G22-88/R1_46
background alpha
DSSSD bias OK - attachment 1
FEE64 temps OK - attachment 2
ADC data item stats - attachment 3
aida15 asic #2 u/s - rate c. 266k
aida11 1*L spectra - attachment 4
ADC data items/s Merger, Tape Server & Merger stats - attachments 5-7
data rate to disk c. 2Mb/s dominated by aida15 asic 2
10.05 DAQ continues file G22-88/R1_79
background alpha
DSSSD bias OK - attachment 8
ambient temperature 16.8 deg C
FEE64 temps OK - attachment 9
ADC data item stats - attachment 10
aida15 asic #2 u/s - rate c. 266k
aida11 1*L spectra - attachment 11
ADC data items/s Merger, Tape Server & Merger stats - attachments 12-14
data rate to disk c. 2Mb/s dominated by aida15 asic 2
10.12 per FEE64 Stat spectra - attachment 15
missing channels aida03, aida04, aida05 & aida15
Check ASIC load aida03, aida04, aida05 & aida15
13:25 Overnight leakage current plot - att 16
13.38 DAQ continues file G22-88/R1_92
background alpha
DSSSD bias OK - attachment 17
ambient temperature 16.8 deg C
FEE64 temps OK - attachment 18
ADC data item stats - attachment 19
aida15 asic #2 u/s - rate c. 266k
aida11 1*L spectra - attachment 20
ASIC check load at 10.12 should result in 1-2 hits per channel in the Stat spectra aida03, aida04, aida05 and aida15
Channels which continue to show zero counts likely indicate a FEE64-adaptor PCB connection issue
Channels which now show 1, or more, counts cf. attachment 15 likely correspond to channels now working following the check ASIC load
ADC data items/s Merger, Tape Server & Merger stats - attachments 21-23
data rate to disk c. 2Mb/s dominated by aida15 asic 2
13.48 Stopped DAQ file G22-88/R1_93
DSSSD bias OFF
FEE64 power OFF
16.00 Installed FEE64 power cable to aida16
Replaced aida15 with new FEE64 MAC 41:f6:5a
Update dhcpd.conf
Update MIDAS config to include aida16
FEE64 power ON
All system wide checks OK
DSSSD bias OK - attachment 24
FEE64 temperatures OK - attachment 25
ADC data item stats - attachment 26
aida15 rate changes from c. 260k to c. 25k ( 2x 'hot' channels, 1x channel missing, all other channels OK )
per FEE64 Rate spectra - attachment 27
per FEE64 1.8.L spectra - attachments 28-29
pulser peak width aida11 44 ch FWHM, aida14 57 ch FWHM
per FEE64 1.8.W spectra - 20us FSR - attachments 30-31
Update NewMerger startup script and Options for 16x FEE64s - see attachments 32-33
16.20 file G22-88/R2
Pulser 50Hz
16.28 DAQ stop
Pulser OFF
All histograms zero'd
16.33 DAQ starts file G22-88/R3
alpha background
ADC data item stats OK - attachment 34
NewMerger, TapeServer and Merger stats - attachments 35-38
16.50 Pulser peak width spreadsheet for file R2 - attachment 39
20.32 DAQ continues file G22-88/R3_9
DSSSD bias OK - attachment 40
FEE64 temperatures OK - attachment 41
ADC data item stats - attachment 42
aida15 rate c. 25k ( 2x 'hot' channels, 1x channel missing, all other channels OK )
per FEE64 Stat spectra - attachment 43
Merger, Merger stats and TapServer - attachments 44-46
Disc data items - attachment 47
aida15 disc rate c. 8k
Disable aida15 disc ( fast comparator ) outputs for asic#3 - attachment 48
Disc data items- attachment 49
all FEE64s rate << 1Hz
Update NewMerger startup script and Options for 16x FEE64s - see attachments 32-33 |
Attachment 1: Screenshot_from_2024-02-03_01-07-28.png
|
|
Attachment 2: Screenshot_from_2024-02-03_01-08-15.png
|
|
Attachment 3: Screenshot_from_2024-02-03_01-09-00.png
|
|
Attachment 4: Screenshot_from_2024-02-03_01-09-39.png
|
|
Attachment 5: Screenshot_from_2024-02-03_01-10-20.png
|
|
Attachment 6: Screenshot_from_2024-02-03_01-11-45.png
|
|
Attachment 7: Screenshot_from_2024-02-03_01-12-23.png
|
|
Attachment 8: Screenshot_from_2024-02-03_10-02-22.png
|
|
Attachment 9: Screenshot_from_2024-02-03_10-02-50.png
|
|
Attachment 10: Screenshot_from_2024-02-03_10-03-29.png
|
|
Attachment 11: Screenshot_from_2024-02-03_10-04-15.png
|
|
Attachment 12: Screenshot_from_2024-02-03_10-04-46.png
|
|
Attachment 13: Screenshot_from_2024-02-03_10-05-09.png
|
|
Attachment 14: Screenshot_from_2024-02-03_10-05-48.png
|
|
Attachment 15: Screenshot_from_2024-02-03_10-12-32.png
|
|
Attachment 16: Screenshot_from_2024-02-03_13-24-46.png
|
|
Attachment 17: Screenshot_from_2024-02-03_13-38-40.png
|
|
Attachment 18: Screenshot_from_2024-02-03_13-38-47.png
|
|
Attachment 19: Screenshot_from_2024-02-03_13-39-04.png
|
|
Attachment 20: Screenshot_from_2024-02-03_13-40-06.png
|
|
Attachment 21: Screenshot_from_2024-02-03_13-40-43.png
|
|
Attachment 22: Screenshot_from_2024-02-03_13-41-00.png
|
|
Attachment 23: Screenshot_from_2024-02-03_13-41-06.png
|
|
Attachment 24: Screenshot_from_2024-02-03_16-01-18.png
|
|
Attachment 25: Screenshot_from_2024-02-03_16-01-33.png
|
|
Attachment 26: Screenshot_from_2024-02-03_16-01-42.png
|
|
Attachment 27: Screenshot_from_2024-02-03_16-02-27.png
|
|
Attachment 28: Screenshot_from_2024-02-03_16-03-16.png
|
|
Attachment 29: Screenshot_from_2024-02-03_16-06-11.png
|
|
Attachment 30: Screenshot_from_2024-02-03_16-07-05.png
|
|
Attachment 31: Screenshot_from_2024-02-03_16-07-35.png
|
|
Attachment 32: CONTENTS
|
0x0006dead string 0x0000
MERGE.LinksAvailable string 16
Index string MERGE.LinksAvailable&&0x0006dead&&MERGE.RunOptions&&MERGE.LinksInUse
Merge.LinksInUse string 1%1%1%1%1%1%1%1%1%1%1%1%1%1%1%1%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%0%
MERGE.LinksInUse string 1%1%1%1%1%1%1%1%1%1%1%1%1%1%1%1%
MERGE.RunOptions string 3
|
Attachment 33: NewMerger
|
#!/bin/csh -f
echo "Tidy up"
killall -s9 MERGE
killall -s9 merge64.AD
killall -s9 master64
killall -s9 link64
killall -s9 merge64
killall -s9 statrate64
echo "Starting New Merger"
/bin/echo -e "\033];New Merger\007"
if (! $?MIDASBASE) setenv MIDASBASE /MIDAS
#
#
setenv PATH ${MIDASBASE}/Linux/bin64:${PATH}
setenv LD_LIBRARY_PATH ${MIDASBASE}/TclHttpd/Linux64:${MIDASBASE}/Linux/lib64:${MIDASBASE}/lib_Linux64
netint MERGE_SyncStep 0x0000000
cd ${MIDASBASE}/NewMerger/MergeServer/bin64
./master64 -i 16 -l 16 -p 11001 &
sleep 100000000000
#
echo startup complete
#
|
Attachment 34: Screenshot_from_2024-02-03_16-35-07.png
|
|
Attachment 35: Screenshot_from_2024-02-03_16-35-20.png
|
|
Attachment 36: Screenshot_from_2024-02-03_16-35-24.png
|
|
Attachment 37: Screenshot_from_2024-02-03_16-35-31.png
|
|
Attachment 38: Screenshot_from_2024-02-03_16-35-40.png
|
|
Attachment 39: Pulser-widths-carme-2024.xlsx
|
Attachment 40: Screenshot_from_2024-02-03_20-32-07.png
|
|
Attachment 41: Screenshot_from_2024-02-03_20-32-40.png
|
|
Attachment 42: Screenshot_from_2024-02-03_20-33-01.png
|
|
Attachment 43: Screenshot_from_2024-02-03_20-33-24.png
|
|
Attachment 44: Screenshot_from_2024-02-03_20-34-17.png
|
|
Attachment 45: Screenshot_from_2024-02-03_20-34-40.png
|
|
Attachment 46: Screenshot_from_2024-02-03_20-35-02.png
|
|
Attachment 47: Screenshot_from_2024-02-03_20-35-52.png
|
|
Attachment 48: Screenshot_from_2024-02-03_20-36-55.png
|
|
Attachment 49: Screenshot_from_2024-02-03_20-37-14.png
|
|
Attachment 50: Screenshot_from_2024-02-04_09-55-02.png
|
|
656
|
Wed Jun 19 12:06:50 2024 |
CB, OF, SS | Run 98/99/100 - higher energy |
10:21 June 17
New energy: 1.125 MeV/u
Target in. Run 98 started.
Run 99 overlap check.
Run 100 back to acquisition as run 98. |
655
|
Wed Jun 19 12:04:02 2024 |
CB + OF, SS | Run 96/97 |
Run 96 - as run 89, while we wait for beam energy change at higher energies
Run stopped around 08:56 June 17
Run 97 - target out, beam optimisation for higher energies.
|
757
|
Tue Jun 17 23:51:58 2025 |
TD, CB, JG | Run 96 - continued |
00.31 DAQ continues data file G22-205/R96_27
DSSSD bias & leakage currents - attachments 7-8
FEE64 temps OK - attachment 6
ADC data item stats - attachment 5
per FEE64 Rate spectra - attachment 4
Merger & TapeServer - attachments 2-3
Online analysis - attachment 1
Analysis data file R96_29 - attachment 9
no timewarps, all FEE64 deadtimes <<1%
04.04 FEE64 temps OK - attachment 10
ADC data item stats - attachment 11
DSSSD bias & leakage currents - attachments 12 & 15
per FEE64 Rate spectra - attachment 13
TapeServer - attachment 14
08.26 DAQ continues data file G22-205/R96_64
DSSSD bias & leakage currents - attachments 16-17
FEE64 temps OK - attachment 18
ADC data item stats - attachment 19
per FEE64 Rate spectra - attachment 20
TapeServer - attachment 21
11:03 All checks OK
EyTotal histogram attached.
12:25 Temps OK. Rates OK.
15:03 run stopped - beam time over
all checks ok
EyTotal hist attached |
Attachment 1: Screenshot_from_2025-06-18_00-51-35.png
|
|
Attachment 2: Screenshot_from_2025-06-18_00-50-35.png
|
|
Attachment 3: Screenshot_from_2025-06-18_00-50-18.png
|
|
Attachment 4: Screenshot_from_2025-06-18_00-50-00.png
|
|
Attachment 5: Screenshot_from_2025-06-18_00-48-42.png
|
|
Attachment 6: Screenshot_from_2025-06-18_00-31-31.png
|
|
Attachment 7: Screenshot_from_2025-06-18_00-31-06.png
|
|
Attachment 8: Screenshot_from_2025-06-18_00-30-49.png
|
|
Attachment 9: R96_29
|
*** TDR format 3.3.0 analyser - TD - May 2021
*** ERROR: READ I/O error: 5002
blocks: 32000
ADC data format: 261850116 ( 336634.1 Hz)
Other data format: 69888 ( 89.8 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 46 ( 0.1 Hz)
RESUME: 46 ( 0.1 Hz)
SYNC100: 34898 ( 44.9 Hz)
WR48-63: 34898 ( 44.9 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 71 ( 0.1 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
WR48-63: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
*** Timestamp elapsed time: 777.848 s
FEE elapsed dead time(s) elapsed idle time(s)
0 1.168 0.000
1 0.000 62.240
2 0.000 92.323
3 0.250 0.000
4 0.066 0.000
5 0.000 313.625
6 0.000 26.861
7 0.000 69.084
8 0.955 0.000
9 0.093 0.000
10 0.016 457.454
11 0.447 0.000
12 0.000 0.000
13 0.000 0.000
14 0.000 0.000
15 0.000 0.000
16 0.000 0.000
17 0.000 0.000
18 0.000 0.000
19 0.000 0.000
20 0.000 0.000
21 0.000 0.000
22 0.000 0.000
23 0.000 0.000
24 0.000 0.000
25 0.000 0.000
26 0.000 0.000
27 0.000 0.000
28 0.000 0.000
29 0.000 0.000
30 0.000 0.000
31 0.000 0.000
32 0.000 0.000
*** Statistics
FEE ADC Data Other Data Sample Undefined Pause Resume SYNC100 WR48-63 Disc MBS Other HEC Data
0 36906229 10016 0 0 10 10 4998 4998 0 0 0 41
1 10150880 2708 0 0 0 0 1354 1354 0 0 0 30
2 8283625 2156 0 0 0 0 1078 1078 0 0 0 0
3 37453715 10294 0 0 6 6 5141 5141 0 0 0 0
4 23358879 6376 0 0 4 4 3184 3184 0 0 0 0
5 4272841 1160 0 0 0 0 580 580 0 0 0 0
6 11147856 3068 0 0 0 0 1534 1534 0 0 0 0
7 9321327 2384 0 0 0 0 1192 1192 0 0 0 0
8 61453512 16278 0 0 14 14 8125 8125 0 0 0 0
9 19621565 5040 0 0 6 6 2514 2514 0 0 0 0
10 3914618 986 0 0 1 1 492 492 0 0 0 0
11 35964891 9422 0 0 5 5 4706 4706 0 0 0 0
12 36 0 0 0 0 0 0 0 0 0 0 0
13 5 0 0 0 0 0 0 0 0 0 0 0
14 137 0 0 0 0 0 0 0 0 0 0 0
15 0 0 0 0 0 0 0 0 0 0 0 0
16 0 0 0 0 0 0 0 0 0 0 0 0
17 0 0 0 0 0 0 0 0 0 0 0 0
18 0 0 0 0 0 0 0 0 0 0 0 0
19 0 0 0 0 0 0 0 0 0 0 0 0
20 0 0 0 0 0 0 0 0 0 0 0 0
21 0 0 0 0 0 0 0 0 0 0 0 0
22 0 0 0 0 0 0 0 0 0 0 0 0
23 0 0 0 0 0 0 0 0 0 0 0 0
24 0 0 0 0 0 0 0 0 0 0 0 0
25 0 0 0 0 0 0 0 0 0 0 0 0
26 0 0 0 0 0 0 0 0 0 0 0 0
27 0 0 0 0 0 0 0 0 0 0 0 0
28 0 0 0 0 0 0 0 0 0 0 0 0
29 0 0 0 0 0 0 0 0 0 0 0 0
30 0 0 0 0 0 0 0 0 0 0 0 0
31 0 0 0 0 0 0 0 0 0 0 0 0
32 0 0 0 0 0 0 0 0 0 0 0 0
*** Timewarps
FEE ADC Pause Resume SYNC100 WR48-63 Disc MBS Undefined Samples
0 0 0 0 0 0 0 0 0 0
1 0 0 0 0 0 0 0 0 0
2 0 0 0 0 0 0 0 0 0
3 0 0 0 0 0 0 0 0 0
4 0 0 0 0 0 0 0 0 0
5 0 0 0 0 0 0 0 0 0
6 0 0 0 0 0 0 0 0 0
7 0 0 0 0 0 0 0 0 0
8 0 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0 0
10 0 0 0 0 0 0 0 0 0
11 0 0 0 0 0 0 0 0 0
12 0 0 0 0 0 0 0 0 0
13 0 0 0 0 0 0 0 0 0
14 0 0 0 0 0 0 0 0 0
15 0 0 0 0 0 0 0 0 0
16 0 0 0 0 0 0 0 0 0
17 0 0 0 0 0 0 0 0 0
18 0 0 0 0 0 0 0 0 0
19 0 0 0 0 0 0 0 0 0
20 0 0 0 0 0 0 0 0 0
21 0 0 0 0 0 0 0 0 0
22 0 0 0 0 0 0 0 0 0
23 0 0 0 0 0 0 0 0 0
24 0 0 0 0 0 0 0 0 0
25 0 0 0 0 0 0 0 0 0
26 0 0 0 0 0 0 0 0 0
27 0 0 0 0 0 0 0 0 0
28 0 0 0 0 0 0 0 0 0
29 0 0 0 0 0 0 0 0 0
30 0 0 0 0 0 0 0 0 0
31 0 0 0 0 0 0 0 0 0
32 0 0 0 0 0 0 0 0 0
*** Program elapsed time: 26.919s ( 1188.754 blocks/s, 74.297 Mb/s)
|
Attachment 10: Screenshot_from_2025-06-18_04-04-16.png
|
|
Attachment 11: Screenshot_from_2025-06-18_04-04-35.png
|
|
Attachment 12: Screenshot_from_2025-06-18_04-04-51.png
|
|
Attachment 13: Screenshot_from_2025-06-18_04-22-08.png
|
|
Attachment 14: Screenshot_from_2025-06-18_04-22-29.png
|
|
Attachment 15: Screenshot_from_2025-06-18_04-22-47.png
|
|
Attachment 16: Screenshot_from_2025-06-18_08-25-28.png
|
|
Attachment 17: Screenshot_from_2025-06-18_08-25-46.png
|
|
Attachment 18: Screenshot_from_2025-06-18_08-49-30.png
|
|
Attachment 19: Screenshot_from_2025-06-18_08-49-45.png
|
|
Attachment 20: Screenshot_from_2025-06-18_08-50-00.png
|
|
Attachment 21: Screenshot_from_2025-06-18_08-50-17.png
|
|
Attachment 22: Screenshot_from_2025-06-18_12-02-48.png
|
|
Attachment 23: Screenshot_from_2025-06-18_15-00-16.png
|
|
756
|
Tue Jun 17 19:10:07 2025 |
NH, PMH | Run 96 |
20:00 - Go to YR cave to see what is going on
20:20 - Beam back
20:28 - Set Tape to Storage, rate ca 2500 kB/s
Temperature and rate OK, System wide checks OK
Target density low (4e12/cm2)
21:00 - target density still 4e12, Ey rate histograms attached (attach 4)
21:20 - Target density increased to 6e12, we see more indication of reactions
22:30 - All checks OK, tape rate still 2500 kB/s
Target density ~7e12/cm2, beam intensity ~1e9pps
23:00 - Target density 7.5e12/cm2, beam intensity: 1e9 pps. Particle detector rate: 8 kHz at injection
Gas bottle pressure still above first tick
Target position X=-1000, Y=1400
All running smoothly. CARME DAQ running OK. (Figs 5-7) |
Attachment 1: Screenshot_from_2025-06-17_20-28-22.png
|
|
Attachment 2: Screenshot_from_2025-06-17_20-28-43.png
|
|
Attachment 3: Screenshot_from_2025-06-17_20-58-26.png
|
|
Attachment 4: Screenshot_from_2025-06-17_21-02-19.png
|
|
Attachment 5: Screenshot_from_2025-06-17_22-59-58.png
|
|
Attachment 6: Screenshot_from_2025-06-17_22-59-48.png
|
|
Attachment 7: Screenshot_from_2025-06-17_22-59-38.png
|
|
755
|
Tue Jun 17 12:22:50 2025 |
CB EW JG AG SN | Run 94/95/96 |
Run 94
Background with beam at 150 keV and no target.
Note energy is 0.15% off vs. beam and target to have a more realistic orbit.
As prior run but thresholds changed to
DSSD1 - 0xc
DSSD2 - 0x8
DSSD3 - 0x8
DSSD4 - MEC
Run 95
DSSD 4 worked briefly. Tripped again a few minutes into this run. Ignore
Run 96 - as run 94
run stopped at 14:00
Zoran reset to with-target settings for later
The plotted readout for the G35 voltage divider over several injections for this run is attached. This kind of data can be obtained from Konstantin also after the experiment.
The average is something like 88.4808 V. |
Attachment 1: G35Voltage.pdf
|
|
653
|
Wed Jun 19 12:00:56 2024 |
CB | Run 93/94/95 - background |
04:00 17 June
Moved the detectors back to the usual configuration (as per run 89).
Moved target away - background run.
Some issues with run numbers - see attached run name R93, but run 94 is written. See attached. 93 and 95 contain data. 94 does *not* exist.
No obvious differences wrt. target run, except Rutherford. |
Attachment 1: Screenshot_from_2024-06-18_05-11-59.png
|
|
Attachment 2: Screenshot_from_2024-06-18_06-48-46.png
|
|
Attachment 3: Screenshot_from_2024-06-18_06-48-49.png
|
|
Attachment 4: Screenshot_from_2024-06-18_06-49-08.png
|
|
Attachment 5: Screenshot_from_2024-06-18_06-49-16.png
|
|
Attachment 6: Screenshot_from_2024-06-18_06-49-23.png
|
|
754
|
Tue Jun 17 12:14:30 2025 |
EW, CB, AG | Run 93- Background with beam 150keV |
Run 93- New run at 150keV, background with beam
13:15 All checks ok |
Attachment 1: Screenshot_from_2025-06-17_13-16-52.png
|
|
Attachment 2: Screenshot_from_2025-06-17_13-16-49.png
|
|
Attachment 3: Screenshot_from_2025-06-17_13-16-45.png
|
|
Attachment 4: Screenshot_from_2025-06-17_13-16-43.png
|
|
753
|
Tue Jun 17 08:04:09 2025 |
EW, AG, CB | Run 92- Background with beam 90keV |
Run 92- started at 09:00, background with beam
Same settings as run 91
See attachments for rates etc
All checks okay
09:58 Run stopped as now tuning 150keV beam
10:00 All checks ok
10:25 D2 bottle pack pressure 30 bar approximately
12:18 All checks ok |
Attachment 1: Screenshot_from_2025-06-17_09-03-43.png
|
|
Attachment 2: Screenshot_from_2025-06-17_09-03-38.png
|
|
Attachment 3: Screenshot_from_2025-06-17_09-03-34.png
|
|
Attachment 4: Screenshot_from_2025-06-17_09-03-29.png
|
|
652
|
Wed Jun 19 11:54:26 2024 |
CB | Run 91/92 - moving detectors |
~ 2:00 17 June
Run 91
Attempted to move the detectors closer to the beam, and move them in sync with the injection.
Found best position was
Right arm: s by 5 mm, r p in (not moving). No benefits from moving. 10 mm + r p in on sync results in beam clipping during injection. 69.55 mm not moving.
Left arm: s by 5 mm, l p in (in sync with motion). Due to the short beam preparation time, only the trigger out was used as per screenshots below. 23.58 -- 67.90 moving.
Saw spike on leakage current for both detectors on the right arm. Unsure if this because they are being moved, or because they were hit with beam. Leakage current recovered.
Run 92 - acquired data as above.
Rutherford locus more clear on left arm detectors. No clear sign of nuclear reactions signals. |
Attachment 1: Screenshot_from_2024-06-18_03-07-50.png
|
|
Attachment 2: Screenshot_from_2024-06-18_03-53-51.png
|
|
Attachment 3: Screenshot_from_2024-06-18_04-20-15.png
|
|
Attachment 4: Screenshot_from_2024-06-18_04-20-18.png
|
|
Attachment 5: Screenshot_from_2024-06-18_04-20-21.png
|
|
651
|
Mon Jun 17 23:10:08 2024 |
CB | Run 90 |
00:10
As run 89, detectors moved fully out.
01:20
TL detector noisy. Shook it. Fixed it.
Temps OK. Noise spike reported in Run 89 still there.
Target 6E11. Beam stable.
02:00
TL detector noisy. Shook it. Fixed it.
ELOG STOPPED WORKING
|
Attachment 1: Screenshot_from_2024-06-18_01-21-47.png
|
|
Attachment 2: Screenshot_from_2024-06-18_01-21-55.png
|
|
650
|
Mon Jun 17 16:26:00 2024 |
JM, RSS, CB | Run 89 |
17:20
(Run 87 Junk and Run 88 does not exist)
Beam charge state set to 5+. Energy is ~1.06 MeV/u
Cooler voltage setpoint 590.3 V.
FC20 read value 589.07 V.
Right arm at 64.60 mm
Left arm IN at 36.54 mm
FakeMotion script running
horizontal beam position 3.5 mm (new overlap position with new beam)
vertical beam position 0.0 mm
Beam intensity 7.0e+6 particles.
Target density ~5e11 at/cm2
Beam is stored in the ring for ~20 seconds. From beam overlap tests ~ 75% of the beam is destroyed in this time.
Beam halo appears slightly reduced. This may be due to reduced intensity and reduced injection rate (20 s vs 5s previously)
18:00
FEE64 temperatures OK
ADC data item stats OK
per FEE64 Rate spectra OK
Beam intensity 1.17e+7 particles
Target density ~5.25E11 atoms/cm2
20:28
FEE64 temperatures OK
ADC data item stats OK
per FEE64 Rate spectra OK
Beam intensity 1.08e+7 particles
Target density ~3.84E11 atoms/cm2
23:58
Temps OK. Stats OK. Merger OK. Write to disk OK.
Waveforms attached.
Beam keeps going away due to "solenoid" issues.
On-line analysis screenshots attached.
Without beam, we see a clear noise spike in all detectors every 50-200 seconds randomly. This is *not* correlated with beam injection or dump.
Checkloaded - no effect. Beam back - no changes. |
Attachment 1: Screenshot_from_2024-06-17_17-59-13.png
|
|
Attachment 2: Screenshot_from_2024-06-17_17-59-30.png
|
|
Attachment 3: Screenshot_from_2024-06-17_17-59-38.png
|
|
Attachment 4: Screenshot_from_2024-06-17_17-59-54.png
|
|
Attachment 5: Screenshot_from_2024-06-17_18-00-00.png
|
|
Attachment 6: Screenshot_from_2024-06-17_20-27-54.png
|
|
Attachment 7: Screenshot_from_2024-06-17_20-28-05.png
|
|
Attachment 8: Screenshot_from_2024-06-17_20-28-13.png
|
|
Attachment 9: Screenshot_from_2024-06-17_20-30-10.png
|
|
Attachment 10: Screenshot_from_2024-06-17_20-30-15.png
|
|
Attachment 11: Screenshot_from_2024-06-17_23-48-01.png
|
|
Attachment 12: Screenshot_from_2024-06-17_23-48-08.png
|
|
Attachment 13: Screenshot_from_2024-06-17_23-48-57.png
|
|
Attachment 14: Screenshot_from_2024-06-17_23-49-26.png
|
|
Attachment 15: Screenshot_from_2024-06-17_23-49-34.png
|
|
Attachment 16: Screenshot_from_2024-06-17_23-49-39.png
|
|
Attachment 17: Screenshot_from_2024-06-17_23-49-46.png
|
|
Attachment 18: Screenshot_from_2024-06-17_23-49-52.png
|
|
Attachment 19: Screenshot_from_2024-06-17_23-50-21.png
|
|
648
|
Mon Jun 17 05:47:35 2024 |
JM | Run 86 |
06:50
No sign of the resonance during night shift. New run started but when operators are avaiable we will likely move to higher energies to try and observe the (p,a) reaction.
Cooler voltage setpoint 247.8 V.
FC20 read value 247.25 V.
Right arm at 64.60 mm
Left arm IN at 36.54 mm
FakeMotion script running
horizontal beam position 1.0 mm
vertical beam position -5mm
Beam intensity 7.0e+7 particles.
Target density ~5e11 at/cm2 |
647
|
Mon Jun 17 04:01:16 2024 |
JM | Run 85 |
05:00
Cooler voltage setpoint 247.45 V.
FC20 read value 246.9 V.
Right arm at 64.60 mm
Left arm IN at 36.54 mm
FakeMotion script running
horizontal beam position 1.0 mm
vertical beam position -5mm
Beam intensity 7.0e+7 particles.
Target density ~5e11 at/cm2
Target temperature is 38.6 K. Values of E1 is 819 Hz, 2.7A, 1.85e-1 mbar
All checks OK |
646
|
Mon Jun 17 02:05:37 2024 |
JM | Run 84 |
03:00
Cooler voltage setpoint 247.7 V.
FC20 read value 247.15 V.
Right arm at 64.60 mm
Left arm IN at 36.54 mm
FakeMotion script running
horizontal beam position 1.0 mm
vertical beam position -5mm
Beam intensity 7.0e+7 particles.
Target density ~5e11 at/cm2
Target temperature is 38.6 K. Values of E1 is 820 Hz, 2.7A, 1.8e-1 mbar
All checks OK
03:34
Noise increase in top left dssd. Servo motor moved to resolve issue. |